K170826 · Xvivo Perfusion AB · KDN · Jun 22, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K170826
Device Name
Perfadex Plus
Applicant
Xvivo Perfusion AB
Product Code
KDN · Gastroenterology, Urology
Decision Date
Jun 22, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5880
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K170826 · Jun 22, 2018
Perfadex Plus
Xvivo Perfusion AB
UNOS (United Network for Organ Sharing) registry
The sponsor utilized data from the UNOS registry to compare patient survival outcomes between lung transplants using the proposed device formulation (supplemented with THAM and calcium) and the predicate device (not supplemented with calcium) to demonstrate safety and substantial equivalence.
Registry data; Lung transplantation; Survival outcomes; UNOS
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
UNOS Registry Analysis; Retrospective registry analysis
Lung transplant recipients
Perfadex not supplemented with Calcium
Patient survival
Indications for Use
Perfadex Plus is indicated for the flushing, static cold storage and transportation of isolated lungs after removal from the donor in preparation for eventual transplantation into a recipient.
Device Story
Perfadex Plus is a pre-buffered, extracellular solution containing dextran 40 and calcium; used for rapid cooling, perfusion, and static cold storage of isolated donor lungs. It replaces the predicate Perfadex, which required manual buffering with THAM and manual calcium supplementation at the clinic. The device is supplied in an 'Ecobag' featuring a 'Click System' connector for sterile connection, replacing the traditional spike port. It is used by transplant clinicians in a clinical or surgical setting to preserve donor lungs during transport. By providing a ready-to-use, pre-buffered, and calcium-supplemented solution, it simplifies preparation, reduces potential for manual errors, and maintains physiological electrolyte balance, ultimately supporting lung viability for transplantation.
Clinical Evidence
Clinical evidence derived from a federally mandated registry (UNOS) and a clinical trial. Comparison of Perfadex supplemented with THAM and calcium ions (PTHCa) versus Perfadex without calcium. Results demonstrate substantial equivalence in patient survival outcomes, indicating that supplementation with THAM and calcium does not increase patient risk.
Technological Characteristics
Pre-buffered, extracellular solution containing dextran 40 and calcium. Packaged in an 'Ecobag' with a 'Click System' connector. Sterile barrier maintenance tested per ISO 11607. Mechanical testing includes pressure, leakage, drop resistance, penetration, tensile strength, and flow testing. Stability testing covers pH, calcium, osmolality, dextran, glucose, and particulate analysis.
Indications for Use
Indicated for flushing, static cold storage, and transport of isolated donor lungs prior to transplantation.
Regulatory Classification
Identification
An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 22, 2018
XVIVO Perfusion AB % Kathleen Johnson President Medical Device Approvals, Inc. P.O. Box 421 Fairfield, Iowa 52556
Re: K170826
Trade/Device Name: Perfadex Plus Regulation Number: 21 CFR 876.5880 Regulation Name: Isolated kidney perfusion and transport system and accessories Regulatory Class: Class II Product Code: KDN Dated: May 24, 2018 Received: May 25, 2018
Dear Kathleen Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
### Page 2 - Kathleen Johnson
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Carolyn Y. Neuland -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170826
Device Name Perfadex Plus
Indications for Use (Describe)
Perfadex Plus is indicated for the flushing, static cold storage and transportation of isolated lungs after removal from the donor in preparation for eventual transplantation into a recipient.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 7. 510(k) SUMMARY
| DATE: | June 22, 2018 |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER: | XVIVO Perfusion AB<br>Mässans gata 10<br>SE-41251 Göteborg<br>Sweden<br><br>Contact: Kathleen Johnson<br>Regulatory Affairs Consultant<br>610-527-0585<br>kathleen@mdapprovals.com |
| REGULATION NAME: | Isolated kidney perfusion and transport system and accessories |
| COMMON OR<br>USUAL NAME: | Perfadex Plus® |
| REGULATION<br>NUMBER | 21 CFR §876.5880 |
| CLASSIFICATION: | Class II |
| PRODUCT CODE | KDN |
| PREDICATE DEVICE: | Perfadex® with THAM K091989 |
| DESCRIPTION: | Perfadex® Plus is a pre-buffered, extracellular solution of dextran 40<br>containing calcium used for rapid cooling, perfusion and storage of lungs in<br>connection with transplantation. |
| INDICATIONS FOR USE: | Perfadex Plus is indicated for the flushing, static cold storage and<br>transportation of isolated lungs after removal from the donor in<br>preparation for eventual transplantation into a recipient. |
| SUBSTANTIAL<br>EQUIVALENCE: | Perfadex® Plus is substantially equivalent to the predicate Perfadex®<br>with THAM cleared under K091989. The devices have the same<br>intended use.<br><br>The differences between Perfadex® Plus and the predicate device<br>Perfadex® with THAM (K091989) are the following:<br><ul><li>The subject device Perfadex® Plus, is buffered with<br/>Tromethamine (THAM) and does not require buffering prior<br/>use, while the predicate device Perfadex with THAM<br/>(K091989) has a slightly acidic pH and requires buffering prior<br/>use. The predicate device is supplied with a sterile 50 ml glass<br/>bottle containing THAM.</li><li>The subject device Perfadex® Plus contains calcium to further<br/>mimic extracellular plasma concentration of electrolytes<br/>while the predicate device (K091989) does not contain<br/>calcium which is therefore often added at the clinics prior to<br/>use.</li><li>Perfadex and Perfadex Plus both are packaged in the Ecobag<br/>however the Perfadex Plus Ecobag utilizes the "Click System"</li></ul> |
{4}------------------------------------------------
instead of a spike port. The connection to the bag is made with a connector that "clicks" onto the port rather than by inserting a spike into the rubber stopper as in the Perfadex Ecobag. Complete information on the "Click System" can be found in the Device Description.
- Package testing to ensure maintenance of the sterile barrier has been performed according to ISO 11607. The bags have been tested by the manufacturer for bag welding, port system welding, tube welding, and fill volume.
#### NONCLINICAL TESTING:
Nonclinical testing including biological safety including:
- · Cytotoxicity
- Intracutaneous Irritation
- · Sensitization
- · Hemolysis
- · Material Mediated Pyrogenicity
- · Leachables
As well as Stability Studies including: •pH •Calcium
-
- •Osmolality
- •Dextran
- •Glucose
- ·Particulate analysis
Additionally, mechanical testing on the new bag and port system includes:
- · Pressure and Leakage testing
- · Resistance to dropping
- · Penetration testing
- · Tensile strength testing
- · Flow testing
has shown the changes to the product do not produce any new safety or effectiveness issues.
- CLINICAL DATA: Using a statistically powered sample size with p-values obtained from a federally mandated registry [UNOS] and from a clinical trial, Clinical data on Perfadex® supplemented at point of use with THAM and calcium ions (PTHCa) shows it is substantially equivalent in survival to Perfadex not supplemented with Calcium. Therefore, we can conclude that supplementation with THAM and Calcium does not add any additional risk to patients receiving a lung transplant.
- CONCLUSION: The non-clinical testing conducted as well as clinical data and realworld evidence with the proposed device demonstrate that the proposed device does not raise any new issues of safety and effectiveness and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.