K170818 · Synthes USA Products, LLC · PBJ · Dec 7, 2017 · Neurology
Device Facts
Record ID
K170818
Device Name
Craniomaxillofacial Distraction System (CMFD)
Applicant
Synthes USA Products, LLC
Product Code
PBJ · Neurology
Decision Date
Dec 7, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5330
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K170818 · Dec 7, 2017
Craniomaxillofacial Distraction System (CMFD)
Synthes USA Products, LLC
Published clinical literature
A literature review of seven articles was conducted to demonstrate the effective use of the predicate CMF Distraction System in cranial distraction osteogenesis, supporting the substantial equivalence of the subject device.
Literature review; Cranial distraction osteogenesis; Clinical performance
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Literature review of CMF Distraction System (K060138) in the cranium; Literature review
Adult and pediatric patients requiring cranial distraction osteogenesis
Not applicable for this study
Effective use of the device in cranial distraction osteogenesis
Indications for Use
The DePuy Synthes CMF Distraction System is intended for use as a bone stabilizer and lengthening (and/or transport) device. The DePuy Synthes CMF Distraction System is indicated for correction of congenital deficiencies or post-traumatic defects of the cranium, where gradual bone distraction is required in adults and pediatric patients. Cranium The 1.5 mm and 2.0 mm mesh and cloverleaf footplates and screws are intended for infants, children, and adults.
Device Story
Modular distraction system for cranial reconstruction; consists of distractor body, footplates, extension arms, and bone screws. Surgeon secures distractor body to bone via footplates/screws; percutaneous activation performed by rotating advancement/lead screw using activation instrument. System features reversing prevention mechanism. Used in hospital or home settings. Provides gradual bone distraction to correct congenital or post-traumatic defects; benefits patient through bone lengthening/stabilization.
Clinical Evidence
No new clinical trials conducted. Literature review of seven articles identified regarding use of previous design (K060138) in cranial distraction osteogenesis; evidence provided via narrative descriptions, x-rays, and images.
Technological Characteristics
Modular metallic construct. Materials: TAN (ASTM F1295), L605 (ASTM F90), CoCrMo (ASTM F1537), MP35N (ASTM F562), CP Titanium (ASTM F67), Silicone (ASTM F2042). Sensing/Actuation: Mechanical lead screw advancement. Sterilization: Steam (moist-heat) to SAL 10^-6 per ANSI/AAMI/ISO 17665-1.
Indications for Use
Indicated for correction of congenital deficiencies or post-traumatic defects of the cranium requiring gradual bone distraction in adults, adolescents, children, and infants.
Regulatory Classification
Identification
A preformed nonalterable cranioplasty plate is a device that is implanted in a patient to repair a skull defect and is constructed of a material, e.g., stainless steel or vitallium, that cannot be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
Synthes Craniomaxillofacial Distraction System (K060138)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized eagle. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with "ADMINISTRATION" underneath.
December 7, 2017
Synthes USA Products, LLC Nicholas Fountoulakis Regulatory Affairs Specialist 1301 Goshen Parkway West Chester, Pennsylvania 19380
Re: K170818
Trade/Device Name: Craniomaxillofacial Distraction System (CMFD) Regulation Number: 21 CFR 882.5330 Regulation Name: Preformed Non-alterable Cranioplasty Plate Regulatory Class: Class II Product Code: PBJ Dated: November 8, 2017 Received: November 9, 2017
Dear Mr. Fountoulakis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170818
Device Name
Craniomaxillofacial Distraction System (CMFD)
Indications for Use (Describe)
The DePuy Synthes CMF Distraction System is intended for use as a bone stabilizer and lengthening (and/or transport) device.
The DePuy Synthes CMF Distraction System is indicated for correction of congenital deficiencies or post-traumatic defects of the cranium, where gradual bone distraction is required in adults and pediatric patients.
Cranium
The 1.5 mm and 2.0 mm mesh and cloverleaf footplates and screws are intended for infants, children, and adults.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 1. 510(k) Summary
Date Prepared: November 7, 2017
#### 1.1.Submitter
Primary Contact: Nicholas Fountoulakis Senior Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-6553 nfounto@its.jnj.com
#### Alternate Contact:
Susan Lewandowski Associate Director, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5852 slewand2@its.jnj.com
## 1.2.Device
Name of Device: Craniomaxillofacial Distraction System (CMFD) Common or Usual Name(s): Cranial Distraction System Classification Name(s): Preformed nonalterable cranioplasty plate Regulatory Class: Class II - 882.5330 Product Code: PBJ Review Panel: Neurology
#### 1.3.Predicate Devices
Primary Predicate: Zurich Distraction System's Arnaud Distractor (K010139) Additional Predicate: Synthes Craniomaxillofacial Distraction System (K060138)
#### 1.4.Indications for Use
The DePuy Synthes CMF Distraction System is intended for use as a bone stabilizer and lengthening (and/or transport) device.
The DePuy Synthes CMF Distraction System is indicated for correction of congenital deficiencies or post-traumatic defects of the cranium, where gradual bone distraction is required in adults and pediatric patients.
Cranium
- . The 1.5 mm and 2.0 mm mesh and cloverleaf footplates and screws are intended for infants, children, adolescents, and adults.
{4}------------------------------------------------
# 1.5.Device Description
DePuy Synthes Craniomaxillofacial (CMF) Distraction System is a modular distractor system intended for correction and reconstruction of the cranium in adults, adolescents, children, and infants. The distractor construct consists of the distractor body, footplates, extension arms, and bone screws, all of which are available in multiple configurations to meet patient and surgeon needs as detailed in the table below.
| | Distractor Body | Footplates | Extension Arms | Bone Screws |
|------------|------------------------------------------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Dimensions | 10, 15, 20, 25,<br>30, 35, and 40<br>mm lengths | Cloverleaf and<br>mesh designs<br>Accept 1.5, 2.0,<br>and 2.4 mm<br>diameter screws | Rigid lengths are<br>20, 40, and 60<br>mm<br>Flexible lengths<br>are 30, 40, and<br>60 mm | 1.5, 2.0, and 2.4<br>mm diameter<br>4 – 12 mm<br>lengths |
| Materials | TAN (ASTM F1295) L605 (ASTM F90) CoCrMo (ASTM F1537) MP35N (ASTM F562) | CP2 and CP4 Titanium (ASTM F67) | TAN (ASTM F1295) L605 (ASTM F90) MP35N (ASTM F562) Silicone (ASTM F2042) | CP Titanium (ASTM F67) TAN (ASTM F1295) |
The distractor body is first secured to bone using footplates and screws, and then activation of the distractor is accomplished through the rotation of an advancement/lead screw with an activation instrument percutaneously.
{5}------------------------------------------------
# 1.6.Comparison to Predicate Devices
| System/Device Name | Synthes Craniomaxillofacial<br>Distraction System | KLS Zurich Distraction System's<br>Arnaud Distractor | Synthes Craniomaxillofacial<br>Distraction System (previous<br>design) |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications | The DePuy Synthes CMF<br>Distraction System is intended for<br>use as a bone stabilizer and<br>lengthening (and/or transport)<br>device.<br><br>The DePuy Synthes CMF<br>Distraction System is indicated for<br>correction of congenital<br>deficiencies or post-traumatic<br>defects of the cranium, where<br>gradual bone distraction is<br>required in adults and pediatric<br>patients. DePuy Synthes CMF<br>Distraction System is intended for<br>single use only.<br><br>Cranium<br>The 1.5 mm and 2.0 mm<br>mesh and cloverleaf<br>footplates and screws are<br>intended for infants,<br>children, adolescents, and<br>adults. | The Zurich Distraction System<br>includes devices intended as a<br>bone stabilizer and lengthening<br>(and/or transport) device when<br>correction of congenital<br>deficiencies or post traumatic<br>defects of the mandible (including<br>ramus, body, alveolar ridge, palate,<br>symphysis), midface, and cranial<br>bones require gradual distraction. | The Synthes Craniomaxillofacial<br>Distraction System (CMF<br>Distraction System) is intended for<br>use as a bone stabilizer and<br>lengthening (and/or transport)<br>device for correction of congenital<br>deficiencies or post-traumatic<br>defects of the mandibular body<br>and ramus where gradual bone<br>distraction is required.<br><br>The Synthes CMF Distraction<br>System is intended for single use<br>only. The Synthes Pediatric CMF<br>Distraction System is intended for<br>use as a bone stabilizer and<br>lengthening (and/or transport)<br>device for correction of congenital<br>deficiencies or post-traumatic<br>defects of the mandibular body<br>and ramus where gradual bone<br>distraction is required in children<br>under the age of 12 months. The<br>Synthes Pediatric CMF Distraction<br>System is intended for single use<br>only. |
| Contraindications | The Synthes CMF Distraction<br>System is contraindicated for | | Use of the Synthes CMF Distraction<br>System is contraindicated in |
| System/Device Name | Synthes Craniomaxillofacial<br>Distraction System | KLS Zurich Distraction System's<br>Arnaud Distractor | Synthes Craniomaxillofacial<br>Distraction System (previous<br>design) |
| | patients sensitized to nickel, cobalt<br>chromium, silicone or<br>molybdenum. | | patients previously sensitized to<br>nickel |
| Where Used | Hospital, home | Hospital, home | Hospital, home |
| System Features | Modular design includes body,<br>plates, and removable activation<br>arms | Pre-assembled distractor body and<br>plates with removable and non-<br>removable activation arm | Modular design includes body,<br>plates, and removable activation<br>arms |
| | Surgeon and Patient Facing<br>Instrumentation | Surgeon and Patient Facing<br>Instrumentation | Surgeon and Patient Facing<br>Instrumentation |
| Distractor Body<br>Specifications | End and center translating 10 - 40 mm lengths | End and center translating 20 mm and 30 mm lengths | End and center translating 10 - 40 mm lengths |
| | Accommodates left and right<br>placement Available with or without<br>universal joint Spring clip to prevent<br>inadvertent reversing | Accommodates left and right<br>placement Available with or without<br>universal joint (2.0 mm size) | Accommodates left and right<br>placement Available with or without<br>universal joint |
| Plate Specifications | Cloverleaf and mesh designs Symmetric design May be cut and bent Fixed with 1.5, and 2.0 mm | Mesh design Symmetric design May be cut and bent Fixed with 1.5, and 2.0 mm | Cloverleaf, mesh, and elevated<br>mesh designs Symmetric design May be cut and bent Fixed with 1.0, 1.3, 1.5, and 2.0 mm |
| System/Device Name | Synthes Craniomaxillofacial<br>Distraction System | KLS Zurich Distraction System's<br>Arnaud Distractor | Synthes Craniomaxillofacial<br>Distraction System (previous<br>design) |
| | Fixed with 2.0 and 2.4 mm Emergency Screws Fixed with 2.0 mm Locking Screws | Fixed with 1.8, and 2.3 mm Emergency Screws | Fixed with 1.2, 1.7, 2.0, 2.4 mm PlusDrive Emergency Screws Fixed with 2.0 mm Locking Screws |
| Extension Arm<br>Specifications | Point of activation moved away from distractor for patient access Rigid and flexible options May be removed without surgical procedure Rigid lengths are 20, 40, and 60 mm Flexible lengths are 30, 40, and 60 mm May be combined to increase length | Point of activation moved away from distractor for patient access Rigid and flexible options (2.0 mm size) May be removed without surgical procedure (2.0 mm size) Rigid lengths are 25, 33, 35, 43, 45, 50, and 53 mm Flexible lengths are 30, 33, 40, and 50 mm (2.0 mm size) May be combined to increase length (2.0 mm size) | Point of activation moved away from distractor for patient access Rigid and flexible options May be removed without surgical procedure Rigid lengths are 20, 40, and 60 mm Flexible lengths are 30, 40, and 60 mm May be combined to increase length |
| Operating Principle(s) | Advancement/lead screw<br>adjustment of metallic plate and<br>screw construct to achieve<br>distraction osteogenesis | Advancement/lead screw<br>adjustment of metallic plate and<br>screw construct to achieve<br>distraction osteogenesis | Advancement/lead screw<br>adjustment of metallic plate and<br>screw construct to achieve<br>distraction osteogenesis |
| Materials | Distractor Bodies:<br>Titanium Alloy TAN (ASTM F1295) Chromium Cobalt L605 (ASTM F90) CoCrMo (ASTM F1537) MP35N (ASTM F562) | Distractor Body:<br>Titanium Alloy TAV (ASTM F136) | Distractor Bodies:<br>Titanium Alloy TAN (ASTM F1295) Chromium Cobalt L605 (ASTM F90) |
| System/Device Name | Synthes Craniomaxillofacial<br>Distraction System | KLS Zurich Distraction System's<br>Arnaud Distractor | Synthes Craniomaxillofacial<br>Distraction System (previous<br>design) |
| | Plates:<br>CP2 or CP4 Titanium (ASTM F67)<br><br>Extension Arms:<br>Titanium Alloy<br>• TAN (ASTM F1295)<br>Chromium Cobalt<br>• L605 (ASTM F90)<br>• MP35N (ASTM F562)<br>Silicone (ASTM F2042) | Plates:<br>CP Titanium (ASTM F67)<br><br>Extension Arms:<br>Titanium Alloy<br>• TAV (ASTM F136) | Plates:<br>CP2 or CP4 Titanium (ASTM F67)<br><br>Extension Arms:<br>Titanium Alloy<br>• TAN (ASTM F1295)<br>Chromium Cobalt<br>• L605 (ASTM F90)<br>• MP35N (ASTM F562)<br>Silicone (ASTM F2042) |
| | Screws:<br>• TAN (ASTM F1295) | Screws:<br>• CP Titanium (ASTM F67) | Screws:<br>• CP Titanium (ASTM F67)<br>• TAN (ASTM F1295) |
| Sterility | Nonsterile<br>Sterilized to SAL of 10-6 via steam<br>(moist-heat) prior to use | Nonsterile<br>Sterilized via steam (moist-heat)<br>prior to use | Nonsterile<br>Sterilized to SAL of 10-6 via steam<br>(moist-heat) prior to use |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
#### 1.6.1. Summary Comparison of Intended Use
Both the subject DePuy Synthes CMFD System and the KLS Zurich Distraction System's Arnaud Distractor (K010139) are indicated for correction of congenital deficiencies or post-traumatic defects of the cranium where gradual bone distraction is required. Additionally, clinical literature is provided in this submission discussing the effective use of the previous design of Synthes CMF Distraction System (K060138) in cranial distraction osteogenesis. The proposed use of the subject CMFD in the cranium does not constitute a new intended use not already addressed by the predicate device.
## 1.6.2. Summary Comparison of Technological Characteristics
The major technological differences between the subject CMFD in comparison to predicate KLS Zurich Distraction System's Arnaud Distractor (K010139) are:
- The subject device features a reversing prevention mechanism while the predicate does ● not. Non-clinical performance data of the reversing prevention mechanism demonstrates its efficacy.
- . The subject device is made of commercially pure titanium alloy (TAN), cobalt chromium alloys (L605, MP25N, and CoCrMo), and silicone while the predicate is made of titanium and titanium alloy (TAV). However, these materials were used in other implantable components of the previous design of CMFD and are materials with established biocompatibility.
# 1.7.Non-clinical performance data
Non-clinical performance data demonstrates that the technological differences of the CMFD System do not raise new questions of safety and efficacy with respect to the predicate devices, and therefore support substantial equivalence:
| Test Name | Test Method Summary | Results |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Reversing Prevention | Anatomically simulated<br>spinning and vibration of the<br>subject distractor and the<br>previous design of CMFD. | The subject device does not<br>reverse unintentionally under<br>the same conditions where<br>the previous design of CMFD<br>reversed. |
| Construct bending strength | Four-point bend test setup of<br>a subject device construct in<br>comparison to the predicate<br>KLS Zurich Distraction<br>System's Arnaud Distractor,<br>and analysis comparing the<br>subject design to the<br>previous design of CMFD. | The CMFD construct is<br>stronger than the predicate<br>KLS Arnaud distractor, and<br>the design changes relative<br>to the previous design in<br>K060138 do not affect the<br>strength of the distractor<br>construct. |
| Torque Input/Force Output<br>and Construct Torsional<br>Strength | This testing challenged the<br>construct in torsion by<br>applying an increasing torque<br>(torque input) to the<br>activation end of the device<br>through an extension arm<br>until failure of the device<br>occurred. | The device is capable of<br>outputting the force required<br>to distract bone in the<br>cranium without failing. |
| Extension arm interface<br>tensile strength | This testing challenged the<br>interface between the<br>distractor and extension arm<br>to ensure the interface can<br>withstand a tensile force<br>based on a human factors<br>benchmark. | The redesigned CMFD meets<br>the same human factors<br>benchmark as the previous<br>design of CMFD. |
| Biocompatibility | All subject devices with the<br>exception of the modified<br>distractor bodies utilize the<br>same materials and<br>manufacturing processes as<br>the predicate devices cleared<br>in K060138, and therefore<br>biocompatibility testing was<br>not provided.<br>For the modified distractor<br>bodies, a risk based<br>assessment per ISO 10993-1<br>was conducted, and<br>supporting biocompatibility<br>testing was provided. | The results of<br>biocompatibility risk<br>assessment and testing<br>demonstrate that the<br>material change to the<br>distractor body does not<br>raise new questions of safety<br>or efficacy with respect to<br>biocompatibility. |
| Steam Sterilization Validation | The sterilization parameters<br>were validated using the<br>"overkill" method as<br>referenced in Annex D of<br>ANSI/AAMI/ISO 17665-1 in<br>order to assure a Sterility<br>Assurance Level (SAL) of 10-6. | The devices may be<br>effectively steam sterilized by<br>the proposed parameters,<br>which are the same<br>parameters as the previous<br>design of CMFD. |
| Endotoxin Testing | Endotoxin levels for finished<br>subject devices were<br>quantified using a Kinetic<br>Turbidimetric method. | The subject devices meet the<br>maximum endotoxin testing<br>limit of 2.15 EU/device. |
{10}------------------------------------------------
{11}------------------------------------------------
# 1.8.Clinical performance data
Clinical literature discussing the application of the previous design of CMFD (K060138) in the cranium is provided in this submission. A literature review was conducted and seven articles were identified discussing the use of the previous design of CMFD (K060138) in the cranium. The previous design of CMFD (K060138) was identified through narrative descriptions, x-rays, and/or images within the articles.
# 1.9.Substantial Equivalence
Results of non-clinical performance testing, discussions of clinical literature, and comparison to predicate devices included in this submission demonstrate that:
- Technological differences of the CMFD System in comparison to the predicate devices does not raise new questions of safety and efficacy
- The proposed CMFD System has the same intended use as the predicate devices
It is concluded that the information provided in this submission supports substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.