# HydroMARK Breast Biopsy Site Markers (K170803)

_Devicor Medical Products, Inc. · NEU · Jun 15, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170803

## Device Facts

- **Applicant:** Devicor Medical Products, Inc.
- **Product Code:** [NEU](/productcode/NEU.md)
- **Decision Date:** Jun 15, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The HydroMARK® Breast Biopsy Site Marker is intended to mark tissue during a percutaneous breast biopsy procedure, be visible under ultrasound for at least 6 weeks, and be permanently visible by x-ray and MRI.

## Device Story

HydroMARK® Breast Biopsy Site Marker consists of a resorbable hydrogel component and a permanent metallic (titanium/stainless steel) coil. Device is pre-loaded in a sterile, disposable applicator. During percutaneous breast biopsy, clinician deploys marker into biopsy tract. Hydrogel expands upon fluid contact to anchor marker at biopsy site; provides ultrasound visibility for at least 6 weeks. Hydrogel degrades via hydrolysis over time, leaving metallic coil for permanent x-ray and MRI visibility. Device aids clinicians in identifying biopsy sites for follow-up or surgical planning. Benefits include accurate site localization for patient management.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including material characterization (GPC, NMR) and finished device performance testing (deployment, hydration, imaging visibility, sterility, and biocompatibility per ISO 10993).

## Technological Characteristics

Implantable marker with resorbable hydrogel and permanent titanium (grade 2) or stainless steel coil. Dimensions: 0.070" OD, 0.20" length. Sensing principle: hydrogel expansion for ultrasound contrast; metallic coil for x-ray/MRI. Sterilization: Ethylene Oxide (ETO). Complies with ISO 11135, ISO 10993 (1, 5, 11, 18), and ASTM standards (F2182, F2119, F2213, F2503, F2052).

## Regulatory Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- HydroMARK® Breast Biopsy Site Markers ([K161021](/device/K161021.md))
- HydroMARK® Breast Biopsy Site Markers ([K121113](/device/K121113.md))
- HydroMARK® Breast Biopsy Site Markers ([K130537](/device/K130537.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 15, 2017

Devicor Medical Products, Inc. Ms. Shawna Rose Sr. Director, ORA 300 E-Business Wav. Fifth Floor Cincinnati. Ohio 45241

Re: K170803

Trade/Device Name: HydroMARK® Breast Biopsy Site Markers Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: NEU Dated: March 13, 2017 Received: March 17, 2017

Dear Ms. Rose:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K170803

Device Name HydroMARK® Breast Biopsy Site Marker

Indications for Use (Describe)

The HydroMARK® Breast Biopsy Site Marker is indicated to mark tissue during a percutaneous breast biopsy procedure, be visible under ultrasound for at least 6 weeks, and be permanently visible by x-ray and MRI.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Page 1 of 1

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

# 510(k) Summary

The following information is provided as required by 21 CFR § 807.92 for the HydroMARK® Breast Biopsy Site Marker 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990 the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.

# Company:

Devicor® Medical Products, Inc. 300 E-Business Way, Fifth Floor Cincinnati, OH 45241 Establishment Registration Number: 3008492462

## Contact:

Shawna Rose Sr. Director, Quality and Regulatory Affairs Devicor Medical Products, Inc. 300 E-Business Way, Fifth Floor Cincinnati, OH 45241 Ph: 513-864-9178 Fax: 513-864-9011 E-mail: shawna.rose@leicabiosystems.com

Date of Submission: March 13, 2017

Proprietary Name: HydroMARK® Breast Biopsy Site Marker

Common Name: Breast Biopsy Site Marker

Regulation: 21 CFR 878.4300

Regulatory Class: II

Product Codes: NEU

Classification Name: General and Plastic Surgery Panel

- HydroMARK® Breast Biopsy Site Markers (K161021) (Primary Predicate) Predicate Devices: HydroMARK® Breast Biopsy Site Markers (K121113) (Reference Predicate) HydroMARK® Breast Biopsy Site Markers (K130537) (Reference Predicate)

{4}------------------------------------------------

#### Device Description:

The HydroMARK® Breast Biopsy Site Marker contains a resorbable hydrogel component and a metallic component for permanent marking. The hydrogel has features that are unique and highly desirable for breast tissue marking.

The HydroMARK® Breast Biopsy Site Marker is provided pre-loaded in a sterile, disposable applicator that is compatible with specified commercially available biopsy devices. The marker is deployed by the delivery system and is left in the tract created during the biopsy procedure.

This Traditional 510(k) addresses the qualification and validation of a new source of raw material polymer of the same original formulation, produced in a new facility and in smaller batch sizes, to be used by the manufacturer of the hydrogel component of the HydroMARK® Breast Biopsy Site Marker. The hydrogel component is manufactured by Coldstream Laboratories, Inc. The new source of polymer for the hydrogel is Corden Pharma, replacing Genzyme polymer.

#### Intended Use:

The HydroMARK® Breast Biopsy Site Marker is intended to mark tissue during a percutaneous breast biopsy procedure, be visible under ultrasound for at least 6 weeks, and be permanently visible by x-ray and MRI.

The indications are identical to those of the predicate device, the HydroMARK® Breast Biopsy Site Marker, indicated "to mark tissue during a percutaneous breast biopsy procedure, be visible under ultrasound for at least 6 weeks, and be permanently visible by x-ray and MRI."

## Technological Characteristics:

The hydrogel component expands on fluid contact to fill the track of the biopsy needle anchoring the HydroMARK® Breast Biopsy Site Marker at the exact location of biopsy. Because the hydrogel is hydrophilic, it is clearly distinct from normal breast structure under ultrasound imaging. The hydrogel material degrades via hydrolysis over time leaving the internal stainless steel or titanium coil which provides permanent visibility under x-ray and MRI. The hydrogel material uses polymer raw material from a new suppler which has been shown to be chemically equivalent to the original raw material using molecular weight distribution, nuclear magnetic resonance spectra, and degradation profiles from gel permeation chromatography and high performance liguid chromatography.

Non-clinical testing of the modified finished HydroMARK® Breast Biopsy Site Markers is identical to the tests used in the original 510(k) and includes the following:

{5}------------------------------------------------

- . Visual inspection
- Critical dimensions ●
- % Moisture
- Functional Deployment
- Hydration Rate
- Visibility under ultrasound, x-ray and MRI
- Sterility, bioburden, pyrogens, and EO residuals

The finished devices performed as intended according to the specifications established for the original device. The change to the polymer vendor had no impact to the finished products as demonstrated by repeating the original test protocols and meeting the pre-established acceptance criteria. The modified devices are substantially equivalent to the predicate devices.

{6}------------------------------------------------

The modified HydroMARK® Breast Biopsy Site Markers are compared to those cleared in the prior 510(k) submissions in the table below.

| Feature / Technological<br>Characteristics  | Modified HydroMARK®<br>Breast Biopsy Site Marker                     | HydroMARK® Breast Biopsy<br>Site Markers (predicates)<br>K161021 (Primary)<br>K121113 (Reference)<br>K130537 (Reference) |
|---------------------------------------------|----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Device                                      | Breast biopsy marker                                                 | Breast biopsy marker                                                                                                     |
| Resorbable component                        | Polymerized and Desiccated<br>Hydrogel with Corden<br>Pharma polymer | Polymerized and Desiccated<br>Hydrogel with Genzyme<br>polymer                                                           |
| Permanent component                         | Titanium (grade 2)<br>Stainless steel                                | Titanium (grade 2)<br>Stainless steel                                                                                    |
| Marker outer diameter                       | 0.070 inches                                                         | 0.070 inches                                                                                                             |
| Marker length                               | 0.20inches                                                           | 0.20inches                                                                                                               |
| Coil (permanent component)<br>length        | 0.035-0.0625inches                                                   | 0.035-0.0625inches                                                                                                       |
| Coil (permanent component)<br>width         | 0.039 inches                                                         | 0.039 inches                                                                                                             |
| Detection                                   | Ultrasound, x-ray and MRI                                            | Ultrasound, x-ray and MRI                                                                                                |
| Sterilization                               | ETO                                                                  | ETO                                                                                                                      |
| Delivery system (non-patient<br>contacting) | Flexible, Rigid , and Rigid<br>Sharp styles                          | Flexible, Rigid , and Rigid<br>Sharp styles                                                                              |

# PREDICATE DEVICE COMPARISON TABLE

{7}------------------------------------------------

# Finished Product Performance Testing

The polymer used in the hydrogel was historically manufactured by Genzyme in their Lexington, Massachusetts facility. Genzyme transferred the process to their Swiss facility. This facility was later purchased by Corden Pharma. This polymer will now be made in the Corden Pharma Liestal, Switzerland facility in smaller batches than Genzyme originally used, but to the same specifications. The equipment, batch size and processes were qualified by testing 3 consecutive test batches from Liestal, Switzerland to the original specifications. The three manufacturing batches were combined and used to prepare HydroMARK® Breast Biopsy Site Markers which met all original finished product specifications to demonstrate that the changes in polymer manufacturing have had no effect on the finished product.

The following tests were performed to demonstrate substantial equivalence:

| Reference No.  | Title                                                                                                                                                                           |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 11135      | Sterilization of health care products-Ethylene Oxide, Part I, Requirements<br>for development, validation, and routine control of sterilization process<br>for medical devices. |
| ISO 10993-1    | Biological evaluation of medical devices, Part 1: Evaluating and testing<br>within a risk management process.                                                                   |
| ISO 10993-5    | Biological evaluation of medical devices, Part 5: Tests for in vitro<br>Cytoxicity                                                                                              |
| ISO 10993-11   | Biological evaluation of medical devices, Part 11: Tests for systemic<br>toxicity                                                                                               |
| ISO 10993-18   | Biological evaluation of medical devices, Part 18: Chemical<br>characterization of materials                                                                                    |
| ASTM F2182-11a | Standard test method for measurement of radio frequency induced<br>heating on or near passive implants during magnetic resonance imaging.                                       |
| ASTM F2119-07  | Standard test method for evaluation of magnetic resonance image<br>artifacts from passive implants                                                                              |
| ASTM F2213-06  | Standard test method for measurement of magnetically induced torque<br>on medical devices in the magnetic resonance environment                                                 |
| ASTM F2503-13  | Marking medical devices and other items for safety in the magnetic<br>resonance environment                                                                                     |
| ASTM F2052-15  | Standard test method for measurement for magnetically induced<br>displacement force on medical devices in the magnetic resonance<br>environment.                                |

{8}------------------------------------------------

- . Visual Inspection – free of visual defects
- Critical Dimensions – outer diameter and length
- % Moisture - pre- and post-sterilization
- Functional Deployment
- Hydration Rate
- . Visibility under ultrasound, x-Ray and MRI imaging modalities
- Sterilization method
- Bioburden
- . Pyrogens
- . EO residuals

## Polymer Equivalence Testing

Gel permeation chromatography, NMR analyses and degradation profiles were used to establish the equivalency of the raw material polymers

Three samples each of the predicate polymer and the new polymer, both of identical formulations, were tested by gel permeation chromatography with differential refractive index detection. Results show the number (Mn) and weight averaged (Mw) molecular weights obtained from both sources are comparable. The polymer molecular weight distributions for both manufacturers are equivalent. There is no evidence of changed polymer composition that would have ramifications in further processing of the polymer. Therefore no unanticipated reactions are expected to be encountered during subsequent hydrogel production.

The possibility of new chemical entities being created at the new polymer manufacturing facility was studied using nuclear magnetic resonance (NMR). Ratios of chemicals from the first reaction step demonstrate equivalency. NMR was also used to determine molar ratios of the secondary reaction step and also showed equivalency.

Polymer from both synthesis sites was compared in every way. The results of the processes for the predicate and the new material are identical. No different or new chemical entities are identified.

## Substantial Equivalence Discussion

The raw material polymer used to make the hydrogel component of the modified HydroMARK® Breast Biopsy Site Markers was demonstrated to be equivalent to the original polymer. Material characterization testing with gel permeation chromatography established equivalent molecular weight distribution and polydispersity between the two sources of the material. Nuclear magnetic resonance established the chemical ratios from each of the two reaction steps were equivalent. No new or different chemical entities were identified. Finally, hydrolysis of both

{9}------------------------------------------------

predicate and new polymers contained the same components, in the same proportions. Both raw material polymers are chemically equivalent.

#### Conclusion

HydroMARK® Breast Biopsy Site Markers manufactured with hydrogel made from Corden Pharma raw material polymer met all original finished product specifications including visual inspection, dimensions, % moisture, functional deployment, hydration rate, visibility by ultrasound, x-ray and MRI, sterility, bioburden, pyrogens and ethylene oxide residuals. The finished product HydroMARK® Breast Biopsy Site Markers made with Corden Pharma raw material polymer are substantially equivalent to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K170803](https://fda-staging.innolitics.com/device/K170803)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).