← Product Code [LOD](/productcode/LOD) · K170802

# TRAUMACEM™ V+ Injectable Bone Cement (K170802)

_Synthes (USA) Products, LLC · LOD · Sep 28, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170802

## Device Facts

- **Applicant:** Synthes (USA) Products, LLC
- **Product Code:** [LOD](/productcode/LOD.md)
- **Decision Date:** Sep 28, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3027
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K170802 · Sep 28, 2017 | TRAUMACEM™ V+ Injectable Bone Cement | Synthes (USA) Products, LLC | Real-world market experience (outside the US); Peer-reviewed clinical literature (case series/outcomes) | The sponsor utilized real-world market experience and published clinical literature on PMMA bone cement augmentation to support the safety profile and clinical performance of the device, specifically regarding cut-out resistance and cement leakage rates. | Postmarket experience; Clinical literature; PMMA bone cement; Trochanteric fractures |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Outside the US market experience; Real-world market surveillance | Patients with trochanteric fractures | Not applicable for this study | Safety (cement leakage), clinical performance |
| Peer-reviewed publications (case series); Retrospective/Observational case series; Follow-up/Duration: Short-term and long-term outcomes | Patients with trochanteric fractures treated with cephalomedullary nail augmentation with PMMA bone cements | Not applicable for this study | Clinical outcomes, cut-out incidence, cement leakage |

## Indications for Use

The TRAUMACEM™ V+ Injectable Bone Cement is indicated for augmentation of the TFN-ADVANCED® Proximal Femoral Nailing System through cannulated implants and instruments for patients with poor bone quality (e.g., osteoporosis).

## Device Story

TRAUMACEM™ V+ is a two-component PMMA bone cement kit consisting of sterile powder and liquid. Components are mixed to initiate an exothermic polymerization reaction, creating a radiopaque, injectable cement. Used in clinical settings by physicians to augment TFN-ADVANCED® Proximal Femoral Nailing systems via cannulated implants. Increased radiopacity and extended working time facilitate delivery through specialized instruments. Provides mechanical reinforcement for implants in patients with poor bone quality; reduces risk of implant cut-out.

## Clinical Evidence

Evidence includes one prospective, multicenter, randomized postmarket study and two peer-reviewed publications on cephalomedullary nail augmentation with PMMA cement. Studies evaluated outcomes in patients with unstable trochanteric fractures. Results showed no cases of cut-out and low rates of cement leakage. Data aligns with real-world international market experience, supporting that benefits outweigh risks.

## Technological Characteristics

Two-component self-polymerizing PMMA system (powder/liquid). Meets ASTM F451-08/ISO 5833:2002 standards (excluding setting/dough time). Features increased radiopacity and extended working time compared to predicate. Sterile kit includes mixer and transfer lid.

## Regulatory Identification

Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”

## Predicate Devices

- Simplex P ([K062553](/device/K062553.md))

## Reference Devices

- BonOs Inject ([K090460](/device/K090460.md))
- Vertecem ([K090435](/device/K090435.md))
- Vertecem II ([K121876](/device/K121876.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, arranged in a way that they resemble a bird in flight. The profiles are connected and flow into each other, creating a sense of unity and movement. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Synthes (USA) Products, LLC Christopher J. Medberry, Ph.D. Project Leader, Regulatory Affairs 1301 Goshen Parkway West Chester, Pennsylvania 19380 September 28, 2017

Re: K170802

Trade/Device Name: TRAUMACEM™ V+ Injectable Bone Cement Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: LOD Dated: August 25, 2017 Received: August 25, 2017

Dear Dr. Medberry:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K170802

Device Name

TRAUMACEM™ V+ Injectable Bone Cement

Indications for Use (Describe)

The TRAUMACEM™ V+ Injectable Bone Cement is indicated for augmentation of the TFN-ADVANCED® Proximal Femoral Nailing System through cannulated implants and instruments for patients with poor bone quality (e.g., osteoporosis).

Type of Use (Select one or both, as applicable)

| <span style="font-size:100%;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:100%;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|

escription OSC (Part 21 CFR 801 Subpart D)

Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| K170802 510(k) Summary      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor:                    | Synthes (USA) Products, LLC                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                             | Primary Contact:                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                             | Christopher J. Medberry, Ph.D.                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                             | Project Leader, Regulatory Affairs                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                             | Synthes (USA) Products, LLC                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                             | 1301 Goshen Parkway                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                             | West Chester, PA 19380 Phone:                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                             | (610) 719- 6806                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                             | cmedberr@its.jnj.com                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                             | Alternate Contact:                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                             | Stacey Bonnell                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                             | Associate Director, Regulatory Affairs                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                             | Synthes (USA) Products, LLC                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                             | 1301 Goshen Parkway                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                             | West Chester, PA 19380                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                             | Phone: (610) 719-5895                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                             | sbonnell@its.jnj.com                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Date<br>Prepared:           | September 28th, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Proprietary<br>Name:        | TRAUMACEM™ V+ Injectable Bone Cement                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Classification:             | Classification: 888.3027<br>Product Code: LOD                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Predicate<br>Device:        | Simplex P (K062553) - primary                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Reference<br>Devices:       | BonOs Inject (K090460), Vertecem (K090435), Vertecem II<br>(K121876)                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device<br>Description:      | TRAUMACEM™ V+ Injectable Bone Cement consists of sterile<br>packed powder and liquid components to create radiopaque<br>TRAUMACEM™ V+ Injectable Bone Cement. The powder component<br>is pre-packed in the mixer. The liquid component is stored in a glass<br>ampoule. A transfer lid for the mixing and transfer of the bone cement<br>in the application system is also contained in the kit. Mixing the two<br>sterile components produces the injectable bone cement. |
| Indications<br>for Use:     | The TRAUMACEM™ V+ Injectable Bone Cement is indicated for<br>augmentation of the TFN-ADVANCED® Proximal Femoral Nailing<br>System through cannulated implants and instruments for patients with<br>poor bone quality (e.g., osteoporosis).                                                                                                                                                                                                                                 |
| Substantial<br>Equivalence: | The proposed TRAUMACEM™ V+ Injectable Bone Cement has<br>comparable indications, design characteristics, materials, and<br>performance characteristics in comparison to the predicate device.                                                                                                                                                                                                                                                                              |

{4}------------------------------------------------

Summary Comparison of Technological Characteristics In specific, the subject, predicate, and reference devices are selfpolymerizing two-component systems comprising a powder and a liquid that polymerize at room temperature immediately after they are mixed together in an exothermic reaction. The major powder component of the subject and predicate devices is acrylic polymer. The indications for the TRAUMACEM™ V+ Injectable Bone Cement are highly similar to a subset of the indications for predicate device Simplex P (K062553) and utilized in the same patient population (e.g., patients with poor bone quality or loss of bone substance). Both TRAUMACEM™ V+ Injectable Bone Cement and Simplex P (K062553) meet the requirements of ASTM F451-08/ISO 5833:2002 (apart from setting and dough time). The major differences between the subiect and predicate cement are radiopacity and handling properties. TRAUMACEM™ V+ Injectable Bone Cement was designed to have increased radiopacity and working time to facilitate delivery through cannulated implants and instruments. As these differences influence physical and chemical properties, the subject device was then compared to reference devices BonOs Inject (K090460) and Vertecem (K090435) to demonstrate compliance with the Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement; Guidance for Industry and FDA (dated July 17th, 2002). Reference device Vertecem II (K121876) was also included as a material reference in support of the biological risk evaluation. The following performance tests have also been provided in support of this 510(k):

- Biomechanical evaluation of cut-out resistance ●
- Cadaveric evaluation of cut-out resistance ●
- Cannulated implant removal evaluation ●
- Radiopacity comparison ●
- Cytotoxicity ●
- Irritation ●
- 0 Sensitization
- Material mediated pyrogenicity ●
- Genotoxicity (in vitro and in vivo) ●
- Implantation ●
- Chemical characterization. ●

### Summary of Clinical Evidence

Three clinical studies conducted outside the US were presented in support of the subject device. One was a prospective, multicenter, randomized, postmarket study comparing outcomes in patients with unstable trochanteric fractures that were treated with or without Traumacem V+ Injectable Bone Cement Augmentation. Two were

{5}------------------------------------------------

| summaries of peer-reviewed publications summarizing the short-term<br>and long-term outcomes of patients with trochanteric fractures who<br>were treated with cephalomedullary nail augmentation with PMMA<br>bone cements. The three studies were evaluated with respect to<br>clinical indications, method of use and patient populations, which<br>were confirmed to align with the intended use of the subject device.<br>No cases of cut-out were observed. Potential risks associated with<br>augmentation (e.g., cement leakage) had a low rate of occurrence.<br>These data correlated well with presented real-world market<br>experience on the use of Traumacem V+ Injectable Bone Cement<br>outside the United States. |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusions<br>Overall, the performance testing and clinical evidence from the<br>postmarket randomized study, postmarket case series, and outside<br>the US market experience support that the benefits of using<br>Traumacem V+ Injectable Bone Cement outweigh the potential risks<br>when utilized in accordance with its indications for use and supports<br>the substantial equivalence of the subject device.                                                                                                                                                                                                                                                                                                               |

---

**Source:** [https://fda-staging.innolitics.com/device/K170802](https://fda-staging.innolitics.com/device/K170802)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).