Diamondback 360 Peripheral Orbital Atherectomy System
K170792 · Cardiovascular Systems, Inc. · MCW · Jun 9, 2017 · Cardiovascular
Device Facts
Record ID
K170792
Device Name
Diamondback 360 Peripheral Orbital Atherectomy System
Applicant
Cardiovascular Systems, Inc.
Product Code
MCW · Cardiovascular
Decision Date
Jun 9, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4875
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Diamondback 360 Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy. The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.
Device Story
Diamondback 360 Peripheral Orbital Atherectomy System (OAS) is a percutaneous device for removing stenotic material from peripheral arteries and AV dialysis fistulae. System components include an Orbital Atherectomy Device (OAD), guide wire (e.g., ViperWire Advance), saline pump, and lubricant. Operation involves an eccentrically orbiting diamond-coated crown driven by a brushless DC electric motor; the crown sands stenotic material into particles small enough for bodily absorption. Used in clinical settings by physicians; device provides tactile feedback during procedure. Proposed modifications include increased drive shaft lengths (180 cm, 200 cm) and compatible 475 cm guide wires to facilitate alternative access methods like radial access. Device benefits include effective removal of arterial/graft occlusions, restoring vessel patency.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified via system life, stall, introducer compatibility, switch logic, dimensional analysis, tactile feedback, temperature, tensile, flexibility, torque, orbit, brake, saline flow, track, and particulate analysis testing.
Technological Characteristics
Electrically driven orbital atherectomy catheter. Materials: 304V stainless steel (drive shaft, guide wire core/coil), nylon/Pebax (saline sheath). Energy source: Brushless DC electric motor. Dimensions: 180 cm/200 cm drive shaft lengths; 475 cm guide wire length. Sterilization: Ethylene oxide. Connectivity: None (standalone).
Indications for Use
Indicated for patients with occlusive atherosclerotic disease in peripheral arteries or occluded hemodialysis grafts (AV shunts) who are candidates for percutaneous transluminal atherectomy or angioplasty.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
Predicate Devices
Diamondback 360 Peripheral Orbital Atherectomy System (K152694)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
June 9, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Cardiovascular Systems, Inc. Ms. Kara Cutshall Regulatory Affairs Specialist 1225 Old Highway 8 NW Saint Paul, Minnesota 55112
Re: K170792
Trade/Device Name: Diamondback 360 Peripheral Orbital Atherectomy System Regulation Number: 21 CFR 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW Dated: May 10, 2017 Received: May 12, 2017
Dear Ms. Cutshall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Fernando
Aguel-S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170792
Device Name
Diamondback 360 Peripheral Orbital Atherectomy System
Indications for Use (Describe)
The Diamondback 360 Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.
The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| <span></span> | <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <span></span> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Company Name: | Cardiovascular Systems, Inc.<br>1225 Old Highway 8 NW<br>Saint Paul, MN 55112 | | |
|----------------------------|-------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|--|
| Contact: | Kara Cutshall | | |
| Phone: | (651) 202-4886 | | |
| Summary Date: | March 15, 2017 | | |
| Trade Name: | Diamondback 360® Peripheral Orbital Atherectomy System | | |
| Product Code: | MCW-Catheter, Peripheral, Atherectomy | | |
| Classification Regulation: | | 21 CFR 870.4875—Intraluminal Artery Stripper | |
| Classification: | II | | |
| Predicate Devices: | 510(k) Number:<br>Manufacturer:<br>Trade Name: | K152694<br>Cardiovascular Systems, Inc.<br>Diamondback 360® Peripheral Orbital<br>Atherectomy System | |
#### 6.1 Description of Device
The Diamondback 360 Peripheral Orbital Atherectomy System (OAS) is intended for use in the treatment of peripheral arteries and A-V graft (shunt) stenosis.
The OAS provides a method of removing stenotic material from peripheral arteries and A-V grafts. The Diamondback 360 uses an eccentrically orbiting sanding surface (crown) to remove stenotic material on the vessel wall. The stenotic particles that are removed are small enough to be absorbed by the body.
{4}------------------------------------------------
The Diamondback 360 Peripheral Orbital Atherectomy System consists of the following components:
- 1) Orbital Atherectomy Device (OAD)
- 2) Atherectomy Guide Wire (e.g., ViperWire Advance)
- 3) Saline Pump (OAS Pump)
- 4) Atherectomy Lubricant (e.g., ViperSlide)
#### 6.2 Description of Proposed
Additional 1.25 Solid, 1.50 Solid, and 1.75 Solid Diamondback 360 Peripheral Orbital Atherectomy Devices (OADs) with increased drive shaft lengths of 180 cm and 200 cm are proposed as additional length options to accommodate patient vessel anatomy and physicians preferred access methodology, for example radial access. The proposed OADs are designed to maintain 5Fr compatibility. Additional accessories to accommodate the increased drive shaft lengths were necessitated such as additional saline sheaths with increased lengths and additional lengths of ViperWire Advance Peripheral guide wires to accommodate the working length of the proposed OADs. The proposed 0.014 in core guide wires are 475 cm in length and compatible with the 180 cm and 200 cm length drive shafts. The proposed 0.014 in core guide wires are designed with a 0.014 in or 0.018 in distal spring tip options. The proposed dimensions of the 1.75 Solid crown are configured to accommodate the smaller diameter drive shaft while maintaining 5Fr compatibility. Packaging dimensions were increased in comparison to that of the predicate packaging to accommodate for the increased OAD and guide wire lengths.
#### 6.3 Indications for Use
The Diamondback 360 Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.
The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.
The Intended Use has not changed from the predicate/current Diamondback 360 Peripheral Orbital Atherectomy System.
#### 6.4 Technology
The Diamondback 360 Peripheral Orbital Atherectomy System provides a method of removing or reducing occlusive atherosclerotic or stenotic material. The OAS applies a diamond coated, eccentrically rotating sanding surface to reduce or remove stenotic
{5}------------------------------------------------
material on the vessel wall. The stenotic particles are small enough to be absorbed by the body. The proposed OADs and guide wires have the same mechanism of operation, materials, chemical composition, and energy source characteristics that were cleared to market for use in the peripheral arteries for Diamondback 360 Peripheral OAS in 510(k) K152694. The proposed OADs and guide wires have increased lengths as options to accommodate patient vessel anatomy and physicians preferred access methodology.
| Characteristic | Predicate<br>OADs and Guide Wires | Proposed<br>OADs and Guide Wires | |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | OAD | CSI | Same |
| Guide Wires | CSI | Same | |
| Indication for Use | The Diamondback 360<br>Peripheral Orbital<br>Atherectomy System is a<br>percutaneous orbital<br>atherectomy system<br>indicated for use as a<br>therapy in patients with<br>occlusive atherosclerotic<br>disease in peripheral<br>arteries and who are<br>acceptable candidates for<br>percutaneous<br>transluminal<br>Atherectomy.<br>The OAS supports<br>removal of stenotic<br>material from artificial<br>arteriovenous dialysis<br>fistulae (AV shunt). The<br>system is a percutaneous<br>orbital atherectomy<br>system indicated as a<br>therapy in patients with<br>occluded hemodialysis<br>grafts who are acceptable<br>candidates for<br>percutaneous<br>transluminal angioplasty. | Same | |
| Characteristic | | Predicate<br>OADs and Guide Wires | Proposed<br>OADs and Guide Wires |
| Orbital Atherectomy Device Platform | | Electrically driven, diamond crown, atherectomy catheter | Same |
| Design/Construction | Crown<br>Assembly | Eccentric crown assembly bonded to a concentric drive shaft | Same |
| | Drive<br>Mechanism | Brushless DC Electric Motor | Same |
| | Guide Wire | 0.014 in atherectomy guide wire designs compatible with 145 cm drive shaft lengths | 0.014 in atherectomy guide wire designs compatible with 180 cm and 200 cm drive shaft lengths |
| Materials | OAD | Drive Shaft: 304V Stainless Steel Saline Sheath: Nylon body and Pebax distal end | Same |
| | Guide Wire | Core Material: 304V Stainless Steel Support Coil: 304V Stainless Steel | Same |
| Packaging | OAD | 1 unit per package | Same Increase in length of drive shaft dispenser coil to accommodate the 180 cm and 200 cm length drive shafts |
| | Guide Wire | 5 units packaged per shelf carton | Same Increase in the dimensions of the guide wire dispenser coil, guide wire pouch, shelf carton, and shipper box to accommodate the 475 cm guide wires Increase in the density of the shelf carton |
| Characteristic | | Predicate | Proposed |
| | | | OADs and Guide Wires OADs and Guide Wires |
| | Drive Shaft | · Length: 145 cm | Length: 180 cm and |
| | | • OD: 0.029 in; 0.031 in | 200 cm |
| | | ID: 0.0156 in; 0.019<br>● | OD: 0.029 in |
| | | in | ID: 0.017 in<br>● |
| Specifications | Saline<br>Sheath | Introducer Size<br>● | Introducer Size<br>● |
| | | Compatibility: 4 Fr; 5 | Compatibility: 5Fr |
| | | Fr; 6 Fr | · Length: 70.92 in; |
| | | Length: 57.26 in;<br>● | 78.75 in |
| | | 57.35 in | • OD: 0.063 in |
| | | · OD: 0.052 in; | ID: 0.049 in |
| | | 0.063 in | |
| | | ID: 0.041 in; 0.049 in | |
| | Guide Wire | Core: 0.014 in | Core: Same |
| | | Spring Tip: 0.014 in; | Spring Tip: 0.014 in;<br>● |
| | | 0.017 in | 0.018 in |
| | | Length: 335 cm | Length: 475 cm |
| Sterilization | | Ethylene oxide | Same |
{6}------------------------------------------------
{7}------------------------------------------------
#### 6.5 Performance Data
The Diamondback 360 Peripheral Orbital Atherectomy System was evaluated using the following performance bench testing to confirm the performance characteristics as compared to the predicate device.
- System Life Testing including guide wire ●
- . Stall Testing
- Introducer Compatibility Testing .
- . Switch Logic Testing
- Dimensional Analysis .
- Tactile Feedback ●
- Temperature Testing .
- Tensile Testing ●
- Flexibility Testing .
- Delivered Torque Testing .
- Orbit Testing .
- Guide Wire Brake Testing .
- Saline Flow Testing ●
- Track Testing ●
- Packaging/Simulated Distribution Testing for Guide Wire ●
{8}------------------------------------------------
- Guide Wire Tip Flexibility
- Guide Wire Trackability
- Guide Wire Torque Transfer
- Guide Wire Torque Strength .
- . Particulate Inspection
- Effluent Particulate Analysis .
All test results demonstrate that the materials, manufacturing process, and design utilized for the Diamondback 360 Peripheral Orbital Atherectomy System met the established specifications necessary for consistent performance during its intended use.
#### Conclusions 6.6
The Diamondback 360 Peripheral Orbital Atherectomy System met all predetermined acceptance criteria of design verification and validation testing as specified by applicable standards, test protocols, and/or customer inputs. Testing results demonstrate that the Diamondback 360 Peripheral Orbital Atherectomy System is substantially equivalent to the legally marketed predicate devices and does not raise any new safety or effectiveness questions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.