← Product Code [HRS](/productcode/HRS) · K170771

# VariAx 2 Foot System (K170771)

_Stryker GmbH · HRS · May 8, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170771

## Device Facts

- **Applicant:** Stryker GmbH
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** May 8, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

VariAx 2 Foot System is indicated for internal fixation of bones in the foot, ankle, hand, and wrist in adult and adolescent (12 - 21 years) patients, which includes replantation, joint fusions, corrective osteotomies, and osteopenic bone.

## Device Story

VariAx 2 Foot System is an orthopedic internal fixation system consisting of contoured titanium plates, screws, and instruments. Used by surgeons in clinical settings for bone stabilization; the system facilitates fixation via locking and non-locking screws and k-wires for temporary placement. The device provides mechanical support for bone healing in procedures such as joint fusions and corrective osteotomies. It functions as a mechanical implant system; no software or algorithmic processing is involved.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including dynamic bending and static stiffness mechanical performance tests, and LAL testing for endotoxin limits.

## Technological Characteristics

Constructed from Grade II commercially pure titanium with Type II anodization. System includes contoured plates, locking/non-locking screws, and instruments. Mechanical fixation principle. No software, energy source, or connectivity features.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- VariAx 2 System ([K141992](/device/K141992.md))
- Ortholoc 2.0/2.4 Plate & Ortholoc 2.0/2.4 Screw ([K090692](/device/K090692.md))
- Modular Mini Fragment LCP System ([K063049](/device/K063049.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

DEPARTMENT OF HEALTH

May 8, 2017

Stryker GmbH Mr. Paul Nelson Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430

Re: K170771

Trade/Device Name: VariAx 2 Foot System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: March 10, 2017 Received: March 14, 2017

Dear Mr. Nelson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K170771

Device Name VariAx 2 Foot System

#### Indications for Use (Describe)

VariAx 2 Foot System is indicated for internal fixation of bones in the foot, ankle, hand, and adolescent (12 - 21 years) patients, which includes replantation, joint fusions, corrective osteotomies, and osteopenic bone.

| Type of Use (Select one or both, as applicable)            |                                                           |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

| Submitter:             | Stryker GmbH<br>Bohnackerweg 1<br>CH-2545 Selzach<br>Switzerland                                      |
|------------------------|-------------------------------------------------------------------------------------------------------|
| Contact Person:        | Paul Nelson<br>Regulatory Affairs Specialist<br>Phone: (201) 831-5691<br>Fax: (201) 831-6500          |
| Date Prepared:         | February 24, 2017                                                                                     |
| Name of Device:        | VariAx 2 Foot System                                                                                  |
| Common or Usual Name:  | Bone Plates                                                                                           |
| Classification Name:   | Single/multiple component metallic bone fixation appliances and<br>accessories<br>(21 CFR § 888.3030) |
| Regulatory Class:      | Class II                                                                                              |
| Product Codes:         | HRS                                                                                                   |
| Primary Predicate:     | VariAx 2 System, K141992                                                                              |
| Additional Predicates: | Ortholoc 2.0/2.4 Plate & Ortholoc 2.0/2.4 Screw, K090692                                              |
| Additional Predicates: | Modular Mini Fragment LCP System, K063049                                                             |

#### Description:

VariAx 2 Foot System is a system used for internal fixation and is composed of sterile and nonsterile plates, and instruments. This system also uses compatible instruments and compatible screws from other systems. These contoured plates are made of Grade II, commercially pure titanium, with Type II anodization. They are available in a variety of sizes and shapes, and are designed to accept locking and non-locking screws, as well as k-wires for temporary fixation during initial placement.

This submission extends K141992 by introducing additional plates, additional instruments, and expands the indications.

## Indications for Use:

VariAx 2 Foot System is indicated for internal fixation of bones in the foot, ankle, hand, and wrist in adult and adolescent (12 - 21 years) patients, which includes replantation, joint fusions, corrective osteotomies, and osteopenic bone.

{4}------------------------------------------------

## Summary of Technologies:

A comparison of the systems demonstrated that the subject VariAx 2 Foot System is substantially equivalent to the previously cleared VariAx 2 system, K141992, to the Ortholoc 2.0/2.4 Plate & Ortholoc 2.0/2.4 Screw system, K090692, and to the Modular Mini Fragment LCP System, K063049, in regards to intended use, material, design, and operating principles.

#### Performance Data:

#### Non-Clinical Testing

Comparative mechanical testing to predicate systems demonstrated substantial equivalence.

The following tests were performed:

- . Dynamic Bending
- o Static Stiffness

Testing demonstrated that the VariAx 2 Foot System is equivalent in mechanical performance to the predicate devices of the ORTHOLOC 2.0/2.4 Plate & Ortholoc 2.0/2.4 Screw system.

LAL testing was performed to establish that the subject devices meet the specified 20 EU/device limit.

#### Clinical Testing

Clinical testing was not a requirement of this submission.

#### Conclusion:

With the exception of the additional components being introduced, the VariAx 2 Foot System described in this submission is identical to the previously cleared, and primary predicate, K141992. Compared to the Ortholoc 2.0/2.4 Plate & Ortholoc 2.0/2.4 Screw system, and to the Modular Mini Fragment LCP system reference predicates, the subject system shares similar intended use, patient population, and principles of operation, as well as similar technological characteristics and materials. Mechanical testing demonstrated that subject system performs as intended and at least as well as the predicates. Based on these attributes, the subject system is deemed substantially equivalent.

---

**Source:** [https://fda-staging.innolitics.com/device/K170771](https://fda-staging.innolitics.com/device/K170771)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).