← Product Code [FLL](/productcode/FLL) · K170769

# wireless thermometer (K170769)

_Ech (Changzhou) Medical Instrument Co., Ltd. · FLL · Jul 26, 2017 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170769

## Device Facts

- **Applicant:** Ech (Changzhou) Medical Instrument Co., Ltd.
- **Product Code:** [FLL](/productcode/FLL.md)
- **Decision Date:** Jul 26, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2910
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Software as a Medical Device, Pediatric

## Indications for Use

The Wireless Thermometer is a battery-operated electronic device with intended use of measuring and monitoring human armpit temperature continuously via wireless signal transmission of the measuring result. The system is intended for armpit temperature monitoring for persons with all ages. The products can be used in hospital and at home.

## Device Story

Device consists of a wearable patch and a mobile application; patch contains NTC thermal resistor, MCU, and Bluetooth transmitter; patch adheres to patient's armpit via medical adhesive; MCU collects voltage from NTC resistor; Bluetooth chip performs table lookup to convert resistance to Celsius temperature; data transmitted wirelessly to compatible smartphone; mobile app displays and manages temperature data; no alarming or interpretation features; used by patients or clinicians in home or hospital; provides continuous monitoring to assist in temperature tracking.

## Clinical Evidence

No clinical data; bench testing only. Compliance verified via ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization/irritation), IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 80601-2-56 (clinical thermometer performance), and IEC 60601-1-11 (home healthcare environment).

## Technological Characteristics

Patch includes NTC thermal resistor, MCU, Bluetooth 4.0 transmitter, and button cell battery. Materials: medical adhesive and foam. Connectivity: Bluetooth to iOS/Android smartphones. Standards: ISO 10993-5, ISO 10993-10, IEC 60601-1, IEC 60601-1-2, ISO 80601-2-56, IEC 60601-1-11. Single-use patch. No display on device.

## Regulatory Identification

A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.

## Special Controls

(1) Device is not a clinical thermometer with telethermographic functions; 
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and 
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.

## Predicate Devices

- TempTraq, Model TT-100 ([K143267](/device/K143267.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Tab #2 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: K170769

- 1. Date of Preparation: 07/24/2017
- 2. Sponsor Identification

## ECH (ChangZhou) Medical Instrument Co., Ltd.

No.65 Huilin west road, zhouqu town, zhong lou District Changzhou city , Jiangsu Province ,213144

Establishment Registration Number: Not yet registered

Contact Person: Mr. Yuan Hu Position: deputy general manager Tel: +86-0519-88161606 Fax: +86-0519-83635058 Email: huyuan@ech-med.com

- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Cindy Wang (Alternative Contact Person)

#### Mid-Link Consulting Co., Ltd

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net

{1}------------------------------------------------

- 4. Identification of Proposed Device
Trade Name: Wireless Thermometer, Model(s): ECH-b1-L, ECH-b1-M, and ECH-b1-S

Common Name: Clinical electronic thermometer Model(s): ECH-b1-L, ECH-b1-M, ECH-b1-S

Regulatory Information

Classification Name: Clinical electronic thermometer Classification: Class II Product Code: FLL Regulation Number: 21 CFR 880.2910 Review Panel: General Hospital

- 5. Intended Use Statement:
The Wireless Thermometer is a battery-operated electronic device with intended use of measuring and monitoring human armpit temperature continuously via wireless signal transmission of the measuring result. The system is intended for armpit temperature monitoring for persons with all ages.

The products can be used in hospital and at home.

- 6. Device Description
The MCU chip in the wireless electronic thermometer collects the voltage value of the NTC thermal resistor, a thermo-sensitivity resistance, whose resistance will change according to temperature. After collecting and filtering algorithm, a resistance value data is obtained. After that, the Bluetooth chip performs a table lookup operation to find the Celsius temperature value corresponding to the resistance value. The temperature data is transmitted through the Bluetooth transmission to the compatible smart phone, in order to achieve the function of temperature measurement.

The wireless thermometer is composed of a Patch and a Mobile App.

- > Patch: One side of Patch is medical adhesive which is used to fix the patch on the armpit of patients. The other side is foam. There is a circuit board inside the patch. The patch is used for detecting of human body temperature and then transferred to Mobile App for display and further management. There is no display function on the patch, except for a indicator to indicating power on/off status.
Patches are single use only, and available in different configuration. The models ECH-b1-S, ECH-b1-M and ECH-b1-L all share the same design principles, materials, mechanism of actions. However they are different in size and intended patient.

- > Mobile APP: The APP is intended for display and management of obtained human body temperature data.

{2}------------------------------------------------

The mobile APP is intended to be run on compatible smart phones with iOS operating system or Android operating system, including:

- iPhone 5, iPhone 5s, iphone 5C, iPhone 6, iPhone 6 Plus, iPhone 6s, iPhone 6s Plus, Iphone SE, iPhone 7, iPhone 7 Plus smart phone run on IOS 7.1 or advanced operation system or
- Nexus 5, Sumsang Galaxy S6, Huawei Mate 8 smart phone run on Android 4.4 or advanced operation system.

Its function includes:

- Patient management;
- Communication with Patch;
- Simultaneous display of temperature;
- Indication of low battery content of patch.

There is no alarming and interpretation feature.

- 7. Identification of Predicate Device(s)
510(k) Number: K143267 Product Name: TempTraq, Model TT-100

- Non-Clinical Test Conclusion 8.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- ♪ ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Tests for in vitro cvtotoxicitv
- > ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Tests for irritation and skin sensitization
- > IEC 60601-1:2012, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.
- > IEC 60601-1-2:2014, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests.
- > ISO 80601-2-56 First Edition 2009-10-01, Medical Electrical Equipment - Part 2-56: Particular Requirements For Basic Safety And Essential Performance Of Clinical Thermometers For Body Temperature Measurement.
- > IEC 60601-1-11:2010 Medical electrical equipment- Part 1-11:General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical system used in the home healthcare environment

{3}------------------------------------------------

## 9. Clinical Test Conclusion

No clinical study is included in this submission.

#### 10. Substantially Equivalent (SE) Comparison

| Item                   | Proposed Device                                                                                                                                                                                                                                                                                                                                       | Predicate Device                                                                                                                                                                                                                                    | Remark                                                     |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Product Code           | FLL                                                                                                                                                                                                                                                                                                                                                   | FLL                                                                                                                                                                                                                                                 | Similar                                                    |
| Regulation Number      | 21 CFR 880.2910                                                                                                                                                                                                                                                                                                                                       | 21 CFR 880.2910                                                                                                                                                                                                                                     | Similar                                                    |
| Class                  | Clinical electronic thermometers                                                                                                                                                                                                                                                                                                                      | Clinical electronic thermometers                                                                                                                                                                                                                    | Similar                                                    |
| Intended Use           | The Wireless Thermometer is a battery-operated electronic device with intended use of measuring and monitoring human armpit temperature continuously via wireless signal transmission of the measuring result. This system is intended for armpit temperature monitoring for persons with all ages. The products can be used in hospital and at home. | The wireless thermometer, model TT-100, is a battery-operated electronic device with intended use of measuring human body temperature precisely. This device is a single-use and intended for armpit temperature measurement of persons of all age. | Similar                                                    |
| Measurement Site       | Under the armpit                                                                                                                                                                                                                                                                                                                                      | Under the armpit                                                                                                                                                                                                                                    | Similar                                                    |
| Features               | Compatible with iOS or Android Mobile Device via Bluetooth connection<br>Display in mobile application, no display on device                                                                                                                                                                                                                          | Compatible with iOS or Android Mobile Device via wireless connection<br>Display in mobile application, no display on device                                                                                                                         | Similar                                                    |
| Component              | Temperature Sensitive Sensor<br>Circuit Board<br>No Cable<br>Battery (button cell)                                                                                                                                                                                                                                                                    | Temperature Sensitive Sensor<br>Circuit Board<br>No cable<br>Battery(button cell)                                                                                                                                                                   | Similar                                                    |
| Working voltage        | 3.0Vdc                                                                                                                                                                                                                                                                                                                                                | 3.0 Vdc                                                                                                                                                                                                                                             | Similar                                                    |
| Measurement range      | 25.0~45.0 ℃                                                                                                                                                                                                                                                                                                                                           | 30~42.4 ℃                                                                                                                                                                                                                                           | Similar                                                    |
| Accuracy               | $\pm$ 0.2 °C for 25-45 °C,                                                                                                                                                                                                                                                                                                                            | $\pm$ 0.1 °C for 30-42.4 °C,                                                                                                                                                                                                                        | Similar                                                    |
| Signal<br>transmission | Bluetooth Version 4.0                                                                                                                                                                                                                                                                                                                                 | Bluetooth Version 4.0                                                                                                                                                                                                                               | Similar                                                    |
| Receiver               | IOS 7.1 or later/ Android 4.4<br>or later based smart phones<br>with Bluetooth® 4.0interface.                                                                                                                                                                                                                                                         | IOS 7.1 through 8.1 or<br>Android operating 4.3<br>through 4.4 operating<br>system supportable smart<br>phones                                                                                                                                      | Similar                                                    |
| Unit                   | °C or °F                                                                                                                                                                                                                                                                                                                                              | °C or °F                                                                                                                                                                                                                                            | Similar                                                    |
| Biocompatibility       | Cytotoxicity<br><br>Sensitization<br><br>Irritation                                                                                                                                                                                                                                                                                                   | No Cytotoxicity<br><br>No sensitization<br><br>Negligible                                                                                                                                                                                           | Comply with ISO 10993-5<br>and ISO 10993-10<br><br>Similar |
| Electrical Safety      | Complied with IEC 60601-1                                                                                                                                                                                                                                                                                                                             | Complied with IEC<br>60601-1                                                                                                                                                                                                                        | Similar                                                    |
| EMC                    | Complied with IEC<br>60601-1-2                                                                                                                                                                                                                                                                                                                        | Complied with IEC<br>60601-1-2                                                                                                                                                                                                                      | Similar                                                    |

Table 1 Comparison of Technology Characteristics

{4}------------------------------------------------

The intended use, measurement range and accuracy of proposed device and predicate device are different. Intended use of the proposed device and predicate device are the same. The difference is only expression difference. The measurement range and accuracy of proposed device is different from that of the predicated device, but both the proposed device and predicated device comply with IEC 80601-2-56. Therefore, these differences will not affect the substantially equivalency concerning safety and effectiveness.

- 11. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 26, 2017

ECH (Changzhou) Medical Instrument Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd Room 405, Ziguang Mansion No.341 Shangcheng Road Shanghai. 200120 CHINA

Re: K170769

Trade/Device Name: Wireless Thermometer, Models: ECH-b1-L, ECH-b1-M, and ECH-b1-S Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: June 26, 2017 Received: June 27, 2017

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely,

for Michael Ryan Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

# Please wait...

If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.

You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.

For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.

Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.

---

**Source:** [https://fda-staging.innolitics.com/device/K170769](https://fda-staging.innolitics.com/device/K170769)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
