Stainless Steel Suture is indicated for use in abdominal wound closure, hernia repair, sternal closure and orthopaedic procedures including cerclage and tendon repair.
Device Story
Stainless Steel Suture is a sterile, non-absorbable surgical suture composed of 316L stainless steel, available as monofilament or multifilament, with or without needles. Used by surgeons in clinical settings for tissue approximation in abdominal, hernia, sternal, and orthopaedic procedures. The device provides mechanical wound support; it elicits minimal acute inflammatory reaction and maintains tensile strength over time. The device is MR Conditional, allowing safe use in 1.5T or 3.0T MRI environments under specified conditions (gradient, SAR, scan time). Healthcare providers use the suture to secure tissues; the MR Conditional labeling informs clinical decision-making regarding patient eligibility for MRI scans post-implantation.
Clinical Evidence
Bench testing only. No clinical data presented. Verification activities included USP testing for needle attachment, tensile strength, and diameter. Biocompatibility and sterilization testing (EO residuals, pyrogenicity) were performed. MRI safety testing (deflection, torque, heating, and artifact assessment) was conducted per ASTM F2503, F2052, F2119, F2182, and F2213 to establish MR Conditional status.
Technological Characteristics
Material: 316L stainless steel. Form: Monofilament or multifilament, non-absorbable. Sterilization: Gamma irradiation or Ethylene Oxide. Compliance: USP monographs 861, 871, 881; European Pharmacopoeia for Sterile Non-Absorbable Strands. MR Conditional per ASTM F2503, F2052, F2119, F2182, F2213.
Indications for Use
Indicated for abdominal wound closure, hernia repair, sternal closure, and orthopaedic procedures (cerclage, tendon repair). Contraindicated in patients with known sensitivities or allergies to 316L stainless steel or constituent metals (chromium, nickel).
Regulatory Classification
Identification
A stainless steel suture is a needled or unneedled nonabsorbable surgical suture composed of 316L stainless steel, in USP sizes 12-0 through 10, or a substantially equivalent stainless steel suture, intended for use in abdominal wound closure, intestinal anastomosis, hernia repair, and sternal closure.
Special Controls
*Classification.* Class II (special controls). The device, when it is a steel monofilament suture that is uncoated and does not incorporate barbs, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9. The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 1, 2017
Ethicon Incorporated Mr. Hrishikesh Khatav, RAC Senior Regulatory Affairs Specialist U.S. Route 22 West, P.O. Box 151 Somerville, New Jersey 08876-0151
Re: K170767
Trade/Device Name: Stainless Steel Suture Regulation Number: 21 CFR 878.4495 Regulation Name: Stainless Steel Suture Regulatory Class: Class II Product Code: GAQ Dated: March 13, 2017 Received: March 14, 2017
Dear Mr. Khatav:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017 See PRA Statement below.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) к170767
Device Name Stainless Steel Suture
Indications for Use (Describe)
mainless Steel Suture is indicated for use in abdominal wound closure, hernia repair, sternal closure and orthopaedic procedures including cerclage and tendon repair.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
| Sponsor: | ETHICON, Inc.<br>P.O. Box 151<br>Route 22 West<br>Somerville, New Jersey 08876 |
|-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Hrishikesh Khatav, RAC<br>ETHICON, Inc.<br>P.O. Box 151<br>Route 22 West<br>Somerville, New Jersey 08876<br>Phone: 908-218-2348<br>Fax: 908-218-2595<br>Email: hkhatav@its.jnj.com |
| Date of Submission: | March 13, 2017 |
| Trade/ Proprietary<br>Name: | Stainless Steel Suture |
| Common Name: | Surgical Stainless Steel Suture<br>Suture, Non-absorbable, Steel, Monofilament and Multifilament,<br>Sterile |
| Regulation Number: | 21 CFR 878.4495 |
| Regulatory Class: | II |
| Product Code: | GAQ |
| Predicate Devices: | Surgical Stainless Steel Suture; K946173, cleared January 9, 1995; Labeling Changes for Generic Suture Types, by Ethicon, Inc., Surgical Stainless Steel Suture; Pre-amendment Device by Ethicon, Inc.; marketed prior to May 28, 1976 |
| Device Description: | Stainless Steel Suture is a sterile monofilament or multifilament, non-<br>absorbable sterile surgical suture composed of 316L stainless steel.<br>The suture is also available coated with polyethylene. Stainless Steel<br>Suture is available in a range of gauge sizes and lengths, non-needled<br>or attached to needles of various types and sizes. Stainless Steel Suture<br>complies with the requirements of the European Pharmacopoeia (Ph.<br>Eur.) for Sterile Non-Absorbable Strands and the United States<br>Pharmacopoeia (USP) for Non Absorbable Surgical Sutures. The<br>European Pharmacopoeia recognizes units of measure Metric and Ph.<br>Eur. sizes as equivalent which is reflected on the labeling. |
| | Stainless Steel Suture elicits a minimal acute inflammatory reaction in<br>tissue and is not absorbed. Implantation studies in animals show that no<br>significant change in the retention of tensile strength of the suture<br>occurs during the lifetime of the implantation. |
| Indications for Use: | Stainless Steel Suture is indicated for use in abdominal wound closure,<br>hernia repair, sternal closure and orthopaedic procedures including<br>cerclage and tendon repair. |
| Contraindications: | The use of this suture is contraindicated in patients with known<br>sensitivities or allergies to 316L stainless steel, or constituent metals<br>such as chromium and nickel |
| Summary of<br>Technological<br>Characteristics of<br>Proposed Device to<br>Predicate- Labeling<br>changes only: | The principle of operation and fundamental scientific technology of the<br>proposed device are equivalent to the predicate device. The<br>performance modification was accomplished via testing in MRI<br>environment and update labeling per FDA guidance "Establishing<br>Safety and compatibility of Passive implants in the Magnetic<br>Resonance (MR) environment" dated December 11, 2014. |
{4}------------------------------------------------
A comparison between the proposed and the predicate device is given below
| | Predicate Device | Proposed Device |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | K946173 | K170767 |
| Product<br>Code | GAQ | Same |
| Regulation | 21 CFR 878.4495 | Same |
| Absorbable | No | Same |
| Trade<br>Name | Surgical Stainless Steel<br>Suture | Stainless Steel Suture |
| Intended<br>Use | Surgical Stainless Steel<br>Suture is indicated for use<br>in abdominal wound<br>closure, hernia repair,<br>sternal closure and<br>orthopaedic procedures<br>including cerclage and<br>tendon repair. | Same |
| Contraindic<br>ation | The use of this suture is<br>contraindicated in<br>patients with known<br>sensitivities or allergies to<br>316L stainless steel, or<br>constituent metals such as<br>chromium and nickel | Same |
| MRI Safety<br>Information<br>in<br>Warnings<br>Section | N/A | Image: MR Conditional symbol<br>MR Conditional<br>Non-clinical testing has<br>demonstrated that<br>Stainless Steel Suture is<br>MR Conditional. A patient<br>with this device can be<br>safely scanned in an MR<br>system meeting the<br>following conditions:<br>• Static magnetic field of<br>1.5 T or 3.0 T<br>• Maximum spatial field<br>gradient of 2000 gauss/cm<br>(20 T/m)<br>• Maximum MR system<br>reported, whole body<br>averaged specific<br>absorption rate (SAR) of 1<br>W/kg<br>• Under the scan<br>conditions defined, the<br>Stainless Steel Suture is<br>expected to produce a<br>maximum temperature<br>rise of less than 4.5°C<br>after 15 minutes of<br>continuous scanning,<br>using RF coils running in<br>quadrature mode.<br>In non-clinical testing, the<br>image artifact caused by<br>the Stainless Steel Suture<br>extends approximately 20<br>mm from the device when<br>imaged with a gradient<br>echo pulse sequence and a<br>2.0 T MRI system. |
| How<br>Supplied | Surgical Stainless steel<br>sutures are available in<br>sizes 10-0 through 7<br>(metric sizes 0.2 -9.0) in a<br>variety of lengths with<br>and without permanently<br>attached needles in one,<br>two and three dozen<br>boxes. | Stainless steel sutures are<br>available in sizes 5-0<br>through 7 (metric sizes 1.0<br>- 9.0) in a variety of<br>lengths with and without<br>permanently attached<br>needles in one, two and<br>three dozen boxes. |
| Color | Undyed | Same |
| Material | 316L Stainless Steel | Same |
| Sterilization | Sterilized by Gamma<br>Irradiation OR Ethylene<br>Oxide | Same |
| Packaging | Surgical Stainless Steel<br>Suture is packaged in<br>paper folders. The folders<br>are packaged in<br>poly/paper and/or<br>poly/Tyvek overwraps.<br>The overwraps are then<br>packed into cartons with<br>IFU's. The cartons are<br>then wrapped with a film<br>or closed with a security<br>label. | Same |
| U.S.P.<br>continued<br>Compliance<br>Additional<br>Compliance<br>to<br>European<br>Pharmacop<br>eia | Surgical Stainless steel<br>suture complies with all<br>the requirements of the<br>United States<br>Pharmacopoeia U.S.P.<br>Monograph 861, 871 and<br>881 for Non Absorbable<br>Surgical Sutures | Stainless Steel Suture<br>complies with all the<br>requirements of the<br>European Pharmacopoeia<br>(Ph. Eur.) for Sterile Non-<br>Absorbable Strands and<br>the United States<br>Pharmacopoeia (U.S.P.)<br>Monograph 861, 871 and<br>881 for non-absorbable<br>surgical sutures. |
Table 1: Device Comparison Table
{5}------------------------------------------------
{6}------------------------------------------------
Performance Data: The proposed device is identical with the predicate device in respect to technological characteristics. No changes to material, construction, design or specifications. Only device labeling is updated to reflect MRI safety information, updated "How Supplied" section and additional compliance to European Pharmacopeia requirement for non
{7}------------------------------------------------
absorbable sutures per "Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Surgical Sutures".
The following Bench/Design Verification activities were performed on Stainless Steel Suture in compliance with design control requirements per 820.30 to demonstrate the continued conformance to respective FDA recognized standards for the proposed device:
Sutures - Needle Attachment per U.S.P. < 871> Tensile Strength per U.S.P. <881> Sutures - Diameter per U.S.P. <861>
## Sterilization and Biocompatibility testing
EO residual testing per AAMI/ANSI/ISO 10993-7 Biological Evaluation of Medical devices. Pyrogenicity testing - Bacterial Endotoxin Test (BET) per ANSI/AAMI ST72:2011- Bacterial Endotoxins - Test Methods, routine monitoring and alternatives to batch testing USP <161> Transfusion and Infusion Assemblies and Medical Devices Bacterial Endotoxin Testing (BET) as defined in USP <85>
## MRI specific Testing:
Magnetic field interactions at 3-Tesla considering worst case for magnetic field interactions: Deflection Angle test for Translational Attraction and Qualitative Assessment of Torque MRI-related Heating, 1.5-Tesla and 3.0 Tesla MR Image Artifacts at 3-Tesla considering worst case for artifacts
Above MRI tests are in compliance with below ASTM Standards: F2503 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment F2052 Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment F2119 Test Method for Evaluation of MR Image Artifacts from Passive Implants F2182 Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants during Magnetic Resonance Imaging F2213 Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment
Conclusions: For the purposes of determining substantial equivalence per FD&C Act Section 510(k), the proposed device is compared with the same device and design that is currently being marketed. The proposed Stainless
{8}------------------------------------------------
Steel Suture is identical to the legally marketed predicate device. Surgical Stainless Steel Suture except for labeling and trade name modifications. The modification does not change the intended use or the fundamental scientific technology. The proposed device uses the same operating principle, incorporates the same basic design and is packaged, manufactured and sterilized using the same materials and processes.
In summary, the Stainless Steel Suture described in this submission is, in our opinion, substantially equivalent to the predicate device which is legally marketed, by Ethicon, Inc. class II device with FDA product code (GAQ), regulatory definition (21 CFR 878.4495) and intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.