Gentrix Surgical Matrix Thick and Gentrix Surgical Matrix Extend are intended for implantation to reinforce soft tissue where weakness exists in patients requiring gastroenterological or plastic & reconstructive surgery. Reinforcement of soft tissue within gastroenterological and plastic & reconstructive surgery includes, but is not limited to, the following procedures: hernia and body wall repair, colon and rectal prolapse repair, tissue repair, and esophageal repair.
Device Story
Porcine-derived extracellular matrix (urinary bladder matrix) scaffold; supplied as 8-layer sheets up to 30 cm x 40 cm. Implanted by surgeons during gastroenterological or plastic/reconstructive procedures to reinforce soft tissue. Device acts as resorbable scaffold; integrates into host tissue via cellular infiltration and capillary growth; resorbs in approximately 240 days. Provides mechanical support for soft tissue defects.
Clinical Evidence
No clinical data provided. Evidence based on bench testing (biocompatibility per ISO 10993, mechanical strength, material characterization) and pre-clinical porcine model of incisional ventral hernia repair demonstrating host tissue integration.
Technological Characteristics
Porcine-derived extracellular matrix (urinary bladder matrix); collagen-based. 8-layer sheet configuration. Resorbable. Sterilized via electron beam irradiation. Single-use. Dimensions up to 30 cm x 40 cm. No electronic or software components.
Indications for Use
Indicated for patients requiring soft tissue reinforcement during gastroenterological or plastic & reconstructive surgery, including hernia, body wall, colon/rectal prolapse, and esophageal repair.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 25, 2017
ACell, Incorporated % John Smith, M.D., J.D. Hogan Lovells US LLP 555 Thirteenth Street. NW Washington DC, District of Columbia 20004
Re: K170763
Trade/Device Name: Gentrix™ Surgical Matrix Thick; Gentrix™ Surgical Matrix Extend Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTM, OXH, OXK Dated: March 13, 2017 Received: March 13, 2017
Dear Dr. Smith:
This letter corrects our substantially equivalent letter of June 8, 2017. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170763
#### Device Name
Gentrix Surgical Matrix Thick and Gentrix Surgical Matrix Extend
Indications for Use (Describe)
Gentrix Surgical Matrix Thick and Gentrix Surgical Matrix Extend are intended for implantation to reinforce soft tissue where weakness exists in patients requiring gastroenterological or plastic & reconstructive surgery. Reinforcement of soft tissue within gastroenterological and plastic & reconstructive surgery includes, but is not limited to, the following procedures: hernia and body wall repair, colon and rectal prolapse repair, and esophageal repair.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 100%;"> <span style="font-size: 100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
| <span style="font-size: 100%;"> <span style="font-size: 100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY ACell, Inc.'s Gentrix™ Surgical Matrix Thick and Gentrix™ Surgical Matrix Extend
March 13, 2017 Date Prepared:
# Manufacturer Information
| Submitted By: | ACell, Inc.<br>6640 Eli Whitney Drive<br>Columbia, MD 21046 |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Salman Elmi<br>VP and Deputy General Counsel; Head of Regulatory Affairs<br>ACell, Inc.<br>Phone: (410) 953-8500<br>Facsimile: (240) 465-8187 |
# Device Name and Classification
| | Trade/Proprietary Name: Gentrix™ Surgical Matrix Thick and Gentrix™ Surgical Matrix Extend |
|--------------------|--------------------------------------------------------------------------------------------|
| | Common or Usual Name: Surgical Mesh, ECM, Surgical Scaffold |
| Regulation Name: | 21 CFR 878.3300; Surgical Mesh |
| Regulatory Class: | Class II |
| Product Code: | FTM, OXH, OXK |
| Predicate Devices: | ACell, Inc. Gentrix™ Surgical Matrix 8 Layer (K162554) |
# Device Description
The Gentrix™ Surgical Matrix Thick and Gentrix™ Surgical Matrix Extend devices are composed of porcine-derived extracellular matrix scaffolds, specifically known as urinary bladder matrix. The devices are supplied in an eight-layer sheet configuration in sizes up to 30 cm x 40 cm, and packaged in double peel-open pouches. The devices are terminally sterilized using electron beam irradiation. The implantable biomaterial is a resorbable extracellular matrix scaffold that will incorporate (remodel) into the body through cellular infiltration, capillary growth, and integration by the surrounding host tissue. Animal studies have shown device resorption in approximately 240 days.
#### Intended Use / Indications for Use
Gentrix Surgical Matrix Thick and Gentrix Surgical Matrix Extend are intended for implantation to reinforce soft tissue where weakness exists in patients requiring gastroenterological or plastic & reconstructive surgery. Reinforcement of soft tissue within gastroenterological and plastic & reconstructive surgery includes, but is not limited to, the following procedures: hernia and body wall repair, colon and rectal prolapse repair, tissue repair, and esophageal repair.
{4}------------------------------------------------
# Summary of Technological Characteristics
Gentrix™ Surgical Matrix Thick and Gentrix™ Surgical Matrix Extend configurations have identical indications for use, intended use, materials, manufacturing processes, packaging materials, and sterilization requirements as the predicate device, and are identical or similar in performance characteristics when compared to the predicate Gentrix™ Surgical Matrix 8-Layer (K162554) device. The Performance evaluations described below were completed to demonstrate equivalence to the predicate device.
| | ACell, Inc.<br>Gentrix™ Surgical Matrix Thick and<br>Gentrix™ Surgical Matrix Extend | ACell, Inc.<br>Gentrix™ Surgical Matrix 8-Layer |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) No. | TBD | K162554 |
| Device Class | Class II | Class II |
| Product Code | FTM, OXH | FTM, OXH |
| Classification | ECM, Surgical Mesh | ECM, Surgical Mesh |
| Intended Use /<br>Indications for Use | The Gentrix Surgical Matrix Thick and<br>Gentrix Surgical Matrix Extend are<br>intended for implantation to reinforce soft<br>tissue where weakness exists in patients<br>requiring gastroenterological or plastic &<br>reconstructive surgery. Reinforcement of<br>soft tissue within gastroenterological and<br>plastic & reconstructive surgery includes,<br>but is not limited to, the following<br>procedures: hernia and body wall repair,<br>colon and rectal prolapse repair, tissue<br>repair, and esophageal repair. | Gentrix™ Surgical Matrix 6-layer and 8-<br>Layer are intended for implantation to<br>reinforce soft tissue where weakness<br>exists in patients requiring<br>gastroenterological or plastic &<br>reconstructive surgery. Reinforcement of<br>soft tissue within gastroenterological and<br>plastic & reconstructive surgery includes,<br>but is not limited to, the following<br>procedures: hernia and body wall repair,<br>colon and rectal prolapse repair, tissue<br>repair, and esophageal repair. |
| Material Source | Porcine Urinary Bladder | Porcine Urinary Bladder |
| Material Type | Collagen, Extracellular Matrix | Collagen, Extracellular Matrix |
| Resorbable | Yes | Yes |
| Configuration | Sheets | Sheets |
| Nominal Sizes (cm) | Up to 30 x 40 | Up to 10 x 15 |
| Reusable | Single Use Device | Single Use Device |
| Packaging | Dual Foil:PET Pouch System | Dual Foil:PET Pouch System |
| Sterilization | electron beam irradiation | electron beam irradiation |
A table comparing the key features of the subject and predicate device is provided below.
# Performance Data
#### Bench - Biocompatibility Testing
The subject devices were evaluated for biocompatibility per ISO-10993 including the following: cytotoxicity, sensitization, irritation/intracutaneous reactivity, acute systemic toxicity, subacute and subchronic toxicity and implantation, genotoxicity, hemocompatibility, and LAL endotoxin. The evaluation demonstrates that the subject devices meet the biocompatibility requirements of the ISO standard.
{5}------------------------------------------------
#### Bench - Mechanical Testing
Gentrix™ Surgical Matrix Thick and Gentrix™ Surgical Matrix Extend configurations were tested for the following: tensile strength, suture retention strength, and tear strength, and tear strength. The results of the mechanical testing provided evidence that the subject devices are substantially equivalent to the predicate device, and provide adequate mechanical strength for its application throughout its labeled shelf life.
# Bench - Material Characterization
Gentrix™ Surgical Matrix Thick and Gentrix™ Surgical Matrix Extend configurations were tested for the following: stiffness, moisture content, hydration uptake, and hydrated onset temperature. The results of the material characterization testing provided evidence that the subject devices are substantially equivalent to the predicate device and provide adequate performance for its application throughout its labeled shelf life.
# Animal Testing
Gentrix™ Surgical Matrix Thick and Gentrix™ Surgical Matrix Extend configurations were evaluated in a pre-clinical porcine model of incisional ventral hernia repair through full device degradation and remodeling. These animal studies provide evidence of full host tissue integration and demonstrate substantially equivalent performance between the subject and predicate devices. In addition, these results support the biocompatibility of the subject devices.
# Clinical
No Clinical data was provided in support of this clearance.
# Conclusions
The Gentrix™ Surgical Matrix Thick and Gentrix™ Surgical Matrix Extend are as safe and effective as the Gentrix™ Surgical Matrix 8-Layer. The Gentrix™ Surgical Matrix Thick and Gentrix™ Surgical Matrix Extend have identical intended uses, indications, and principles of operation; additionally the subject and predicate devices have identical or similar technological characteristics. The minor technological differences between the Gentrix™ Surgical Matrix Thick and Gentrix™ Surgical Matrix Extend and the predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the Gentrix™ Surgical Matrix Thick and Gentrix™ Surgical Matrix Extend are as safe and effective as the Gentrix™ Surgical Matrix 8-Layer. Thus, the Gentrix™ Surgical Matrix Thick and Gentrix™ Surgical Matrix Extend are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.