The Acculoc Total Knee System is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, primary and secondary traumatic arthritis, avascular necrosis of the femoral condyle, posttraumatic loss of joint configuration. Moderate valgus, varus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. This device may also be indicated in correction of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure. The Acculoc Total Knee System is designed for cemented use only.
Device Story
Fixed-bearing, posterior-stabilized, and cruciate-retained total knee prosthesis; consists of femoral component, tibial insert, tibial tray, and patellar component. Femoral component articulates with tibial insert; tibial insert snaps into tibial baseplate. Designed to restore natural femoral anatomy, rotation, extension, and flexion. Used by orthopedic surgeons in clinical settings for total knee arthroplasty. Provides mechanical joint replacement to reduce pain and restore function in patients with severe knee disability.
Clinical Evidence
Bench testing only. Performance testing included fatigue testing of tibial tray, interlock mechanism strength, shear fatigue strength of tibial insert post, contact pressures/areas, lateral subluxation, knee constraint, jump distance, and range of motion. Biocompatibility risk assessment and pyrogenicity testing conducted. No clinical data presented.
Technological Characteristics
Fixed-bearing, semi-constrained, cemented knee prosthesis. Materials: Femoral component (CoCrMo, ASTM F75), Tibial Tray (Ti-6Al-4V ELI, ASTM F136), Tibial Insert/Patella (UHMWPE GUR 1050, ASTM F648). Mechanical sensing/actuation via anatomical articulation. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients requiring knee arthroplasty for pain relief or improved function due to rheumatoid arthritis, primary/secondary traumatic arthritis, avascular necrosis of femoral condyle, or posttraumatic joint loss. Also indicated for moderate valgus, varus, or flexion deformity with stable ligamentous structures, and for correction of failed osteotomy, arthrodesis, or previous arthroplasty. Designed for cemented use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 23, 2018 RootLoc Co., Ltd. Jungsun Ha RA Specialist #1210, 68, Digital-ro 9-gil Geumcheon-gu, Seoul, 08512 Korea Re: K170753 Trade/Device Name: Acculoc Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: January 31, 2018 Received: February 2, 2018
Dear Jungsun Ha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Katherine D. Kavlock -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K170753
Device Name Acculoc Total Knee System
### Indications for Use (Describe)
The Acculoc Total Knee System is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, primary and secondary traumatic arthritis, avascular necrosis of the femoral condyle, posttraumatic loss of joint configuration. Moderate valgus, varus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. This device may also be indicated in correction of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure. The Acculoc Total Knee System is designed for cemented use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
The following 510(k) summary is being submitted as required by 21 CFR 807.92(a):
| Submitter: | RootLoc Co., Ltd.<br>#1210, 68, Digital-ro 9-gil<br>Geumcheon-gu, Seoul, 08512, Korea<br>Phone. 82-2-6941-4341<br>e-mail: jsha0314@gmail.com |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jungsun Ha |
| Date prepared: | March 15, 2018 |
#### 2. Device Identification
1.
| Trade Name | Acculoc Total Knee System |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL,<br>SEMI-CONSTRAINED, CEMENTED,<br>POLYMER/METAL/POLYMER |
| Product Code | JWH |
| Regulatory Class | 2 |
| Classification Name | Knee joint patellofemorotibial polymer/metal/polymer<br>semi-constrained cemented prosthesis per 21CFR<br>888.3560. This falls under the Orthopedics panel/87<br>as a ClassⅡ device. |
#### 3. Purpose of 510(k)
The RootLoc Co., Ltd. here by submits this traditional 510(k)
- 4. Predicate or legally marketed devices which are substantially equivalent : United Orthopedic U2 Total Knee System (K051640) Progressive Orthopaedic Total Knee System (K142649, K150783) Corentec EAUM/LOSPA TKR System (K110404) NovoKnee Total Knee System (K123339)
#### 5. Description of the Device
The Acculoc Total Knee System is of the fixed bearing type with a posterior stabilized design and cruciate retained design. It is a Patellofemorotibia, polymer.metal/polymer, semi-constrained, cemented knee prosthesis that consists of a femoral component, tibial insert, tibial tray and patellar component. The femoral
{4}------------------------------------------------
component articulates with the tibial insert component. The underside of the tibial insert component is flat and "snaps" into the tibial baseplate component. The design and sizing of the femoral components correspond to the natural femoral anatomy, enhancing stress distribution and restoring original femoral dimensions and normal rotation, extension and flexion. Each femoral component has the same intercondylar distance and radius of curvature. Each tibial insert component is complimentarily shaped to conform to the femoral component. The dome shape of each UHMWPE patellar component provides excellent contact with the femoral component and evenly distributes stresses. The dome shape of each patellar component also simplifies implantation by eliminating the need for rotational orientation.
| Product | Material | Standard | |
|---------------|-------------------|-----------|--|
| Femoral | CoCrMo Alloy | ASTM F75 | |
| component | | | |
| Tibial Tray | Ti-6Al-4V ELI | ASTM F136 | |
| Tibial Insert | UHMWPE (GUR 1050) | ASTM F648 | |
| Patella | UHMWPE (GUR 1050) | ASTM F648 | |
Materials:
#### Indication for Use 6.
The Acculoc Total Knee System is indicated in knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, avascular necrosis of the femoral condyle, posttraumatic loss of joint configuration. Moderate valgus, varus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. This device may also be indicated in correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure. The Acculoc Total Knee System is designed for cemented use only.
### Comparison of the technological characteristics of the subject and predicate 7. devices
The Acculoc Total Knee System has the same intended use and indications as the U2 Total Knee system, Progressive Orthopaedic Total Knee System, LOSPA TKR System and NovoKnee Total Knee System. The Acculoc Total Knee System is manufactured from the same, or better materials as the U2 Total Knee System, Progressive Orthopaedic Total Knee System, LOSPA TKR System and NovoKnee
{5}------------------------------------------------
Total Knee System. The range of sizes available for the Acculoc Total Knee System is the same as the range of sizes cleared for the U2 Total Knee system, Progressive Orthopaedic Total Knee System, LOSPA TKR System and NovoKnee Total Knee System. The Acculoc Total Knee System design is substantially similar to the U2 Total Knee system design, Progressive Orthopaedic Total Knee System, LOSPA TKR System and NovoKnee Total Knee System. Based on these similarities, the Acculoc Total Knee System believes that the Acculoc Total Knee System is substantially equivalent to the U2 Total Knee system, Progressive Orthopaedic Total Knee System, LOSPA TKR System and NovoKnee Total Knee System.
#### 8. Performance Testing
The Acculoc Total Knee System was tested for fatigue performance of the tibial tray, interlock mechanism strength (between the tibial tray and tibial insert), shear fatigue strength of the tibial insert post, contact pressures and areas, lateral subluxation of patellar component, Knee constraint tests ,Jump distance test and range of motion performance. Test results indicate that the Acculoc Total Knee System performs as well, or better than, the U2 Total Knee system , Progressive Orthopaedic Total Knee System ,LOSPA TKR System and NovoKnee Total Knee System and is capable of withstanding expected in vivo loading without failure. Biocompatibility risk assessment and pyrogenicity testing has been conducted.
#### 9. Conclusion
The Acculoc Total Knee System is substantially equivalent to legally marketed predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.