K170740 · W. L. Gore & Associates, Inc. · FGE · May 8, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K170740
Device Name
GORE VIABIL Short Wire Biliary Endoprosthesis
Applicant
W. L. Gore & Associates, Inc.
Product Code
FGE · Gastroenterology, Urology
Decision Date
May 8, 2017
Decision
SESU
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The GORE® VIABIL® Short Wire Biliary Endoprosthesis is intended for palliation of malignant strictures in the biliary tree.
Device Story
The GORE® VIABIL® Short Wire Biliary Endoprosthesis is a flexible, radially compressed stent mounted on a delivery catheter for endoscopic implantation in the biliary tract. The device is designed to palliate malignant biliary strictures. The subject modification introduces a guidewire exit port near the distal end of the catheter, allowing for the use of shorter (minimum 260cm) guidewires, which enables single-user control during the procedure. The stent itself remains unchanged from the predicate, consisting of an ePTFE/FEP film tube supported by an FEP-coated Nitinol stent. The device is used by clinicians in an endoscopic setting. The output is the physical placement of the stent to maintain biliary patency, providing symptomatic relief for patients with malignant obstructions.
Clinical Evidence
No clinical evaluations or animal studies were required to support this submission. Substantial equivalence was demonstrated through non-clinical bench testing, including simulated use deployment, dimensional analysis, tensile/bond strength, and biocompatibility testing per ISO 10993-1.
Technological Characteristics
Stent: ePTFE/FEP film tube with FEP-coated Nitinol support. Delivery system: Dual lumen catheter with distal guidewire exit port. Sterilization: Ethylene Oxide (SAL 10^-6). Biocompatibility: ISO 10993-1 compliant. MR safety: Passive implant evaluated per FDA guidance.
Indications for Use
Indicated for the palliation of malignant strictures in the biliary tree in patients requiring biliary stenting.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the text "Public Health Service". The text is in a simple, sans-serif font and is left-aligned. The text is likely part of a document or sign.
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the most complete and the bottom profile being the least complete.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W O66-G609 Silver Spring, MD 20993-0002
May 8, 2017
W.L. Gore & Associates, Inc. Barbara L. Smith, RAC Regulatory Associate 301 Airport Road Elkton, MD 21921
Re: K170740
Trade/Device Name: GORE® VIABIL® Short Wire Biliary Endoprosthesis Regulation Number: 21 CFR§ 876.5010 Regulation Name: Biliary Catheter and Accessories Regulatory Class: II Product Code: FGE Dated: March 9, 2017 Received: March 10, 2017
Dear Barbara L. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:
The safety and effectiveness of this device for use in the vascular system have not been established
{1}------------------------------------------------
Furthermore, the indication for biliary use must be prominently displayed in all labeline. including pouch box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.
Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.
The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
{2}------------------------------------------------
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# William H. Maisel -S
William H. Maisel, MD, MPH Director, Office of Device Evaluation (Acting) Deputy Center Director for Science Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K170740
Device Name
GORE® VIABIL® Short Wire Biliary Endoprosthesis
Indications for Use (Describe)
The GORE® VIABIL® Short Wire Biliary Endoprosthesis is intended for palliation of malignant strictures in the biliary tree.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 7 510(K) SUMMARY [21 CFR 807.92]
#### 510(k) Submitter
W. L. Gore & Associates, Inc. 301 Airport Road Elkton, Maryland 21921 Regulatory contact: Barbara L. Smith, RAC
#### Date Prepared
May 1, 2017
#### Device Names/Classification
| Device Name: | GORE® VIABIL® Short Wire Biliary Endoprosthesis |
|----------------------|-------------------------------------------------|
| Common Name: | Biliary Stent |
| Classification Name: | Catheter, biliary, diagnostic |
| Classification: | 21CFR876.5010 – Class II |
| Product Code: | FGE |
#### Predicate Devices
- . K081184 GORE® VIABIL® Biliary Endoprosthesis
#### Device Description
The GORE® VIABIL® Short Wire Biliary Endoprosthesis is a flexible endoprosthesis (stent) that is radially compressed and secured onto the distal end of a delivery catheter. The catheter provides a means for implanting the GORE® VIABIL® Short Wire Biliary Endoprosthesis at the target site in the biliary tract. The subject of this 510(k) pre-market notification is a modification to the catheter component of the endoscopic (GI) delivery system to include a guidewire exit port located near the distal end of the catheter which will allow use of a shorter length (minimum 260cm) guidewire enabling single user control of the guidewire during use. No changes have been made to the implanted endoprosthesis of the predicate GORE® VIABIL® Biliary Endoprosthesis device.
{5}------------------------------------------------
The sizes of the GORE® VIABIL® Short Wire Biliary Endoprosthesis device are the same as the predicate GORE® VIABIL® Biliary Endoprosthesis as noted in the following table.
| Endoprosthesis -<br>Diameter (mm) x Length<br>(cm) | Drainage Holes<br>Located at the<br>Hilar Region |
|----------------------------------------------------|--------------------------------------------------|
| 8 × 4 | No holes |
| 8 x 6 | No holes |
| 8 × 8 | No holes |
| 8 x 10 | No holes |
| 10 x 4 | No holes |
| 10 x 6 | No holes |
| 10 x 8 | No holes |
| 10 x 10 | No holes |
| 8 × 6 | Holes |
| 8 × 8 | Holes |
| 8 x 10 | Holes |
| 10 x 6 | Holes |
| 10 x 8 | Holes |
| 10 x 10 | Holes |
#### Indications for Use
The GORE® VIABIL® Short Wire Biliary Endoprosthesis is intended for palliation of malignant strictures in the biliary tree.
{6}------------------------------------------------
### Comparison of Technological Characteristics
The table below describes the similarities/differences between the catheter component of the subiect GORE® VIABIL® Short Wire Biliary Endoprosthesis and the catheter component of the predicate GORE® VIABIL® Biliary Endoprosthesis (K081184).
| | Modified Device | Predicate Device |
|-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| | GORE® VIABIL® Short Wire Biliary | GORE® VIABIL® Biliary |
| | Endoprosthesis | Endoprosthesis<br>K081184 |
| Indications for use | No change - Same as predicate | Intended for palliation of malignant<br>strictures in the biliary tree. |
| Endoprosthesis<br>(Stent device) | No change - Same as predicate | ePTFE/FEP film tube supported by<br>an FEP coated Nitinol stent. |
| Deployment<br>System | No change - Same as predicate | Pull-line style deployment using an<br>ePTFE zipper |
| | Single port hub | Y-arm (dual port) catheter hub |
| GI Delivery<br>Catheter<br>Component -<br>Modified for use<br>with shorter<br>guidewire | Dual lumen catheter tubing<br>-Added guidewire exit port on catheter<br>shaft; 2nd catheter hub port eliminated.<br>Catheter material formulation change<br>to different durometer of same<br>material type | Dual lumen catheter tubing<br>-Guidewire exits from 2nd catheter<br>hub port |
| Biocompatibility | Same as predicate | Biocompatible per ISO10993-1. |
| Device Packaging | • HDPE coil with modified connector<br>• No change to sterile barrier | • HDPE coil and connector for<br>physical protection of device<br>within sterile barrier |
| Sterilization<br>method/SAL | Same as predicate | Ethylene Oxide/SAL of 10-6 |
| Labeling | Same as predicate<br>MR safety information<br>Update to warning statements | Complies with applicable regulations &<br>FDA guidance on MR safety |
#### Summary of Performance Testing
#### Non-Clinical
The device specific guidance document, FDA's Guidance for the Content of Premarket Notifications for Metal Expandable Biliary Stents, was consulted in determining the appropriate testing for the subject catheter modification. Testing was conducted on representative sterile, aged samples. The test samples were subjected to simulated use deployment testing, dimensional testing, tensile/bond strength testing, deployment force, accessory component compatibility, deployment accuracy, and deployment reliability. Testing demonstrated the subject Viabil device met the established acceptance criteria for up to the intended device shelf life, consistent with the predicate Viabil device. Biocompatibility testing was conducted per ISO10993-1 and confirmed the subject device, like the predicate device, is biocompatible for its intended use. Packaging testing of the subject Viabil device in the predicate Viabil device packaging system confirmed that the packaging system
{7}------------------------------------------------
was acceptable for the GORE® VIABIL® Short Wire Biliary Endoprosthesis device. MR Safety testing was conducted in recognition of FDA's guidance, Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment and the results included in the device labeling.
Animal study: No animal studies were required to support this submission.
Clinical
No clinical evaluations were required to support this submission.
#### Conclusion
Based on the information contained within this 510(k) premarket notification, W. L. Gore & Associates concludes that the subject GORE® VIABIL® Short Wire Biliary Endoprosthesis device is substantially equivalent to the predicate GORE® VIABIL® Biliary Endoprosthesis device in terms of indications for use, materials, biocompatibility, sterilization, packaging, labeling and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.