K170738 · Inmode MD , Ltd. · GEX · Aug 7, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170738
Device Name
InMode Diolaze XL
Applicant
Inmode MD , Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 7, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The InMode Diolaze XL System is intended for hair removal and permanent hair reduction defined as stable, long-term reduction in hair counts at 6,9 or 12 months following a treatment regime.
Device Story
InMode Diolaze XL is a diode laser system for hair removal and permanent hair reduction. Device delivers 810nm optical energy to skin via handpiece with integrated sapphire cooling block; requires water-based gel for optical contact. System comprises AC/DC power supply, diode driver, water cooling, controller, and touch screen interface. Operator (clinician) adjusts light fluence (5-40 J/cm²) and pulse duration (5-200 ms) via interface; foot switch activates energy delivery. Integrated cooling enhances patient comfort and safety. Device used in clinical settings. Output is optical energy; clinical benefit is stable, long-term reduction in hair counts.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing and performance evaluation against recognized consensus standards (AAMI ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1).
Technological Characteristics
Diode laser system; 810nm wavelength; 5-40 J/cm² fluence; 5-200 ms pulse width; 11mm x 27.5mm spot size. Features integrated sapphire block with water-based cooling (7℃ or 12℃). Console includes AC/DC power supply, diode driver, and touch screen interface. Complies with IEC 60601-1, 60601-1-2, 60601-2-22, and 60825-1 standards. Standalone unit.
Indications for Use
Indicated for hair removal and permanent hair reduction in subjects seeking hair removal.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows a partial view of a logo or emblem. The visible portion includes the text "DEPARTMENT OF HEALTH & HUMAN" arranged in a circular fashion, suggesting it's part of a larger circular design. To the right of the text, there's a stylized graphic featuring three abstract profiles facing the same direction, stacked one above the other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 7, 2017
InMode MD Ltd. % Amit Goren Regulatory Manager A. Stein- Regulatory Affairs Consulting Ltd. 20 Hata'as Str. Suite 102 Kfar Saba. 44425 Israel
Re: K170738 Trade/Device Name: InMode Diolaze XL Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: July 4, 2017 Received: July 10, 2017
Dear Amit Goren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S.
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170738
Device Name InMode Diolaze XL
Indications for Use (Describe)
The InMode Diolaze XL System is intended for hair removal and permanent hair reduction defined as stable, long-term reduction in hair counts at 6,9 or 12 months following a treatment regime.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY INMODE DIOLAZE XL DEVICE
# 510(k) Number K170738
### Applicant Name:
| Company Name: | InMode MD Ltd. |
|---------------|------------------------------------------------------------|
| Address: | Tabor Building, Shaar Yokneam<br>Yokneam 2069200<br>Israel |
| Tel: | +972-4-9097470 |
| Fax: | +972-4-9097471 |
| E-mail: | amit@asteinrac.com |
#### Contact Person:
| Official Correspondent: Amit Goren |
|------------------------------------|
|------------------------------------|
| Company Name: | A. Stein - Regulatory Affairs Consulting Ltd. |
|----------------------|-----------------------------------------------------------|
| Address: | 20 Hata'as Str., Suite 102<br>Kfar Saba 4442500<br>Israel |
| Tel: | +972-9-7670002 |
| Fax: | +972-9-7668534 |
| E-mail: | amit@asteinrac.com |
| Date Prepared: | July 04, 2017 |
| Trade Name: | InMode Diolaze XL Device |
| Classification Name: | CFR Classification section 878.4810; (Product code GEX) |
| Classification: | Class II Medical Device |
### Predicate Device:
The InMode Diolaze XL device is substantially equivalent to the following predicate devices.
| Manufacturer | Device | 510(k) No. |
|--------------|------------------------|------------|
| Lumenis Ltd. | Lightsheer Desire (ET) | K151947 |
{4}------------------------------------------------
#### Device Description:
The InMode Diolaze XL device is designed to deliver optical energy to the skin via a pre-cooled sapphire block. The good optical contact between the sapphire block and skin is achieved by using water based gel. The device provides individual adjustment of light fluence and pulse duration to achieve maximum efficiency and safety for each patient. The hand piece has integrated skin cooling to enhance safety and comfort of the treatment.
The InMode Diolaze XL device consists of an AC/DC power supply unit, a diode driver, water cooling system, controller and a touch screen user interface. The 810nm diode laser hand piece is connected to the console via a cable and a foot switch activates the energy delivery to the hand piece comprises the InMode Diolaze XL with 20 diode laser bars stacked vertically. The diode is enclosed to the sapphire light guide. The diode laser produces up to 3000W peak optical power.
Following are the InMode Diolaze XL device specifications:
Laser Output Parameters:
| Wavelength | 810 nm |
|-----------------------------------|----------------------------------------------------------|
| Fluence | 5-40 J/cm² |
| Pulse width (duration) | 5-200 ms (pulse type : Short/Long) |
| Light guide cooling | Strong: 7℃, Normal: 12℃ |
| Spot size | 11mm x27.5 mm |
| Dimensions: | 46cm W x 46cm D x 100cm H (18.2'' W x 18.2'' D x 40'' H) |
| Weight: | 32 Kg (70.548 lbs) |
| Main Line Frequency<br>(nominal): | 50-60 Hz |
| Input Voltage (nominal): | 100-240 VAC |
{5}------------------------------------------------
#### Intended Use/Indication for Use:
The InMode Diolaze XL System is intended for hair removal and permanent hair reduction defined as the stable, long-term reduction in hair counts at 6. 9, or 12 months following a treatment regime.
#### Performance Standards:
The InMode Diolaze XL Device has been tested and complies with the following voluntary recognized standards:
- AAMI ANSI: ES 60601-1:2005/(R)2012 and A1:2012. C1:2009/(R)2012 . and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment -Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-2:2007-03(Modified), Medical Electrical Equipment Part 1-● 2: General requirements for basic safety and essential performance -Collateral standard: Electromagnetic compatibility - Requirements and tests
- . IEC 60601-2-22 Edition 3.1 2012-10, Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- . IEC 60825-1: Edition 2.0 2007-03, Safety of laser products - Part 1: Equipment classification and requirements [Including: Technical Corrigendum 1 (2008), Interpretation Sheet 1 (2007), Interpretation Sheet 2 (2007)].
#### Non-Clinical Performance Data:
The InMode Diolaze XL device laser output parameters were evaluated as part of the test performance evaluation conducted in adherence with the FDA recognized consensus standards 60601-1, 60601-1-2, 60601-2-22 and 60825. The results of these performance tests demonstrated that the InMode Diolaze XL device operates in accordance with the device design requirements and performs within the range of laser specifications as to that of the predicate device and therefore, is substantially equivalent to the predicate device.
#### Pre-Clinical Performance Data:
Not Applicable
#### Clinical Performance Data:
Not Applicable
{6}------------------------------------------------
#### Substantial Equivalence:
The following table presents a comparison between the subject device and its predicate device:
| Technological<br>Characteristic | InMode Diolaze XL<br>InMode MD Ltd. | LightSheer<br>Desire, ET (Lumenis<br>Ltd.)<br>K151947 |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code, Class | GEX,<br>Class II | GEX,<br>Class II |
| Indications for Use | Hair removal and<br>permanent hair<br>reduction defined as the<br>stable, long-term<br>reduction in hair counts<br>at 6, 9, or 12 months<br>following a treatment<br>regime. | Hair removal and<br>permanent hair reduction<br>defined as the stable, long-<br>term reduction in hair<br>regrowing when measured<br>at 6, 9, or 12 months after<br>the completion of<br>treatment regime. |
| Target Population | Subjects seeking hair<br>removal | Subjects seeking hair<br>removal |
| Wavelength | 808 nm | 805 nm, ET (two<br>handpiece options are<br>available 790-830) |
| Fluence | 5-40 J/cm² | 10-100 J/cm² |
| Pulse duration | 5-200 msec | 5-400 msec |
| Cooling<br>temperature | Skin sapphire cooling | Chilled sapphire tip water-<br>cooled |
The indications for use of the InMode Diolaze XL device are substantially equivalent to the indications for use of predicate device and are identical to the indications for use of the InMode Diolaze device (cleared as 510 (k) 142952). The design and components in the InMode Diolaze XL device, including the console (with power supply, software, cooling system and touch screen user interface), the water-cooled hand piece connected to the console via a cable and the foot switch are identical to the design and components found in the InMode Diolaze device (cleared as 510 (k) 142952) and very similar to those of the predicate device. The performance specifications (including wavelength, fluence, pulse width, pulse repetition rate, spot size and cooling) in the InMode Diolaze XL device are very similar to those at the predicate device. The safety features in the InMode Diolaze XL device are substantially equivalent to the safety features found in the predicate device. Consequently, the InMode Diolaze XL device is substantially equivalent to the Lightsheer Desire (ET) predicate device, cleared in 510(k) K151947, and therefore, may be legally marketed in the USA.
{7}------------------------------------------------
## Conclusions:
Based on the performance testing and comparison to predicate device, the InMode Diolaze XL device is as safe and effective as the predicate device and consequently is substantially equivalent to the predicate device and can be sold in the US market.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.