← Product Code [KHW](/productcode/KHW) · K170728

# E. HISTOLYTICA QUIK CHEK (K170728)

_Techlab, Inc. · KHW · Jun 7, 2017 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170728

## Device Facts

- **Applicant:** Techlab, Inc.
- **Product Code:** [KHW](/productcode/KHW.md)
- **Decision Date:** Jun 7, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3220
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K170728 · Jun 7, 2017 | E. HISTOLYTICA QUIK CHEK | Techlab, Inc. | Archived clinical specimen bank | Retrospective analysis of 96 archived, characterized clinical specimens from an endemic area to supplement prospective clinical performance data. | Retrospective study; Archived specimens; Clinical performance |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective analysis of archived specimens | Patients from an E. histolytica endemic area; Sample Size: 96 | Composite Reference Method (Microscopy and NAAT) | Sensitivity and Specificity |

## Indications for Use

The TECHLAB® E. HISTOLYTICA QUIK CHEK™ test is a rapid membrane enzyme immunoassay for the qualitative detection of adhesin from Entamoeba histolytica in a single use cassette. It is intended for use with human fecal specimens from patients with diarrhea or dysentery as an aid in the diagnosis of E. histolytica gastrointestinal infection. Test results should be considered in conjunction with patient history.

## Device Story

Rapid membrane enzyme immunoassay for qualitative detection of E. histolytica adhesin in human fecal specimens. Device uses single-use cassette containing immobilized monoclonal antibodies specific for adhesin in test line and anti-HRP antibodies in control line. Conjugate consists of polyclonal antibodies to adhesin coupled to horseradish peroxidase. User applies fecal sample to cassette; visual inspection of blue lines in reaction window indicates presence of adhesin (test line) and procedural validity (control line). Used in clinical settings; provides results in 30 minutes. Assists clinicians in diagnosing E. histolytica infection when combined with patient history. Benefits include rapid, point-of-care detection of pathogen in patients with diarrhea or dysentery.

## Clinical Evidence

Clinical performance evaluated via prospective (n=755) and retrospective (n=96) studies using a composite reference method (microscopy, FDA-cleared NAAT, and PCR sequencing). Prospective study showed 40% sensitivity (due to low positive count) and 100% specificity. Retrospective study of archived positive specimens showed 100% sensitivity and 100% specificity. Combined data supports diagnostic utility. No cross-reactivity observed with common enteric bacteria, viruses, or other parasites.

## Technological Characteristics

Rapid membrane enzyme immunoassay (EIA). Materials: monoclonal/polyclonal antibodies, horseradish peroxidase (HRP) conjugate, tetramethylbenzidine substrate. Form factor: single-use cassette. Analyte: E. histolytica adhesin. Storage: 2°C–8°C. No electronic components; manual visual interpretation.

## Regulatory Identification

Entamoeba histolytica serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Entamoeba histolytica in serum. Additionally, some of these reagents consist of antisera conjugated with a fluorescent dye (immunofluorescent reagents) used to identify Entamoeba histolytica directly from clinical specimens. The identification aids in the diagnosis of amebiasis caused by the microscopic protozoan parasite Entamoeba histolytica and provides epidemiological information on diseases caused by this parasite. The parasite may invade the skin, liver, intestines, lungs, and diaphragm, causing disease conditions such as indolent ulcers, an amebic hepatitis, amebic dysentery, and pulmonary lesions.

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.

## Predicate Devices

- E. histolytica II ([K994101](/device/K994101.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
K170728

B. Purpose for Submission:
To obtain a substantial equivalence determination for the detection of *Entamoeba histolytica* antigens in human stool

C. Measurand:
*Entamoeba histolytica* antigen

D. Type of Test:
Qualitative membrane enzyme immunoassay

E. Applicant:
TechLab Inc.

F. Proprietary and Established Names:
E. HISTOL YTICA QUIK CHEK

G. Regulatory Information:
1. Regulation section:
21 CFR Part 866.3220 *Entamoeba histolytica* Serological Reagents
2. Classification:
II
3. Product code:
KHW
4. Panel:

{1}

83 Microbiology

## H. Intended Use:

### 1. Intended use(s):

The TECHLAB E. HISTOLYTICA QUIK CHEK test is a rapid membrane enzyme immunoassay for the qualitative detection of adhesin from *Entamoeba histolytica* in a single use cassette. It is intended for use with human fecal specimens from patients with diarrhea or dysentery as an aid in the diagnosis of *E. histolytica* gastrointestinal infection. Test results should be considered in conjunction with patient history.

FOR IN VITRO DIAGNOSTIC USE

### 2. Indication(s) for use:

The TECHLAB E. HISTOLYTICA QUIK CHEK test is a rapid membrane enzyme immunoassay for the qualitative detection of adhesin from *Entamoeba histolytica* in a single use cassette. It is intended for use with human fecal specimens from patients with diarrhea or dysentery as an aid in the diagnosis of *E. histolytica* gastrointestinal infection. Test results should be considered in conjunction with patient history.

FOR IN VITRO DIAGNOSTIC USE

### 3. Special conditions for use statement(s):

For prescription use only

### 4. Special instrument requirements:

None

## I. Device Description:

The *E. HISTOLYTICA QUIK CHEK* test uses antibodies to adhesin. The device contains a reaction window with two vertical lines of immobilized antibodies. The test line (“T”) contains monoclonal antibodies specific for *E. histolytica* adhesin. The control line (“C”) contains antibodies to horseradish peroxidase (HRP). The Conjugate consists of polyclonal antibodies to *E. histolytica* adhesin coupled to horseradish peroxidase. The “T” reaction is examined visually for the appearance of a vertical blue line on the “T” side of the Reaction Window. A blue line indicates a positive test. A positive “C” reaction, indicated by a vertical blue line on the “C” side of the reaction window, monitors/confirms that the sample and reagents were added correctly, the reagents were active at the time of performing the assay, and that the sample migrated properly through

{2}

the membrane device. It also confirms the reactivity of the other reagents associated with the assay.

# J. Substantial Equivalence Information:

1. Predicate device name(s):

E. histolytica II

2. Predicate  $510(\mathrm{k})$  number(s):

K994101

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device K170728 | Predicate K994101  |
|  Indications for use | The TECHLAB E. HISTOLYTICA QUIK CHEK™ test is a rapid membrane enzyme immunoassay for the qualitative detection of adhesin from Entamoeba histolytica in a single use cassette. It is intended for use with human fecal specimens from patients with diarrhea or dysentery as an aid in the diagnosis of E. histolytica gastrointestinal infection. Test results should be considered in conjunction with patient history. FOR IN VITRO DIAGNOSTIC USE. | The E. histolytica II is an enzyme immunoassay for the rapid detection of the adhesin of E. histolytica in human fecal specimens. It is indicated for use with fecal specimens from patients with diarrhea or dysentery to determine the presence of E. histolytica gastrointestinal infection. The test can be used for fecal specimens submitted for routine clinical testing from adults or children. Conventional microscopy is not a prerequisite for use of the test. FOR IN VITRO DIAGNOSTIC USE.  |
|  Measured analyte | E. histolytica specific antigen | Same  |
|  Target Population | Persons suspected of having E. histolytica infection | Same  |
|  Antibody Format | Monoclonal/Polyclonal | Same  |
|  Type of Test | Qualitative | Same  |
|  Sample Matrix | Human Fecal Specimen | Same  |

{3}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device K170728 | Predicate K994101  |
|  Controls | Positive and negative control included in the kit | Same  |
|  Storage | Refrigerated (2°C – 8°C) | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device K170728 | Predicate K994101  |
|  Format | Single use membrane cassette 25 tests | 96 well Mircoassay Plate  |
|  Specimen type | Fecal specimens in Cary-Blair and C&S Transport Media | Cannot use Transport Media  |
|  Time to result | 30 minutes | 2.5 hours  |
|  Controls | Internal “C” Control line | No internal control  |
|  Interpretation of result | Visual | Spectrophotometric and Visual  |

# K. Standard/Guidance Document Referenced (if applicable):

CLSI EP07-A2 Interference Testing In Clinical Chemistry; Approved Guideline - Second Edition

CLSI EP15-A3 User Verification of Precision And Estimation Of Bias; Approved Guideline - Third Edition

CLSI EP17-A2 Evaluation Of Detection Capability For Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition

# L. Test Principle:

Lateral flow immunochromatographic assay.

# M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

The precision of the E. HISTOLYTICA QUIK CHEK test was determined using an eight member masked fecal specimen panel. The panel consisted of 2 negative, 2 high negative (just below C5), 2 low positive (just above LoD), and 2 moderate positive (2-3x higher than the C95) specimens. Each fecal specimen was spiked using whole

{4}

organism to achieve the desired level. The Specimens were tested twice a day over a 12 day period by multiple technicians using two different kit lots. Positive and negative controls were run with each sample panel of masked specimens. The positive specimens tested positive 100% of the time and the negative specimens tested negative 100% of the time. The samples produced the expected result 100% of the time.

The reproducibility of the E. HISTOLYTICA QUIK CHEK test was determined using eight masked fecal specimen panel prepared the same way as described for the precision study. Testing was performed at 2 independent laboratories and on-site at TECHLAB, Inc. The Specimens were tested twice a day over a five day period by multiple technicians at each site using two different kit lots. Positive and negative controls were run with each sample panel of masked specimens. The result from each laboratory were submitted to TECHLAB Inc. and compared. The results were consistent among all three locations and exhibited a correlation of 100%. The samples produced the expected result 100% of the time.

b. Linearity/assay reportable range:

Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Storage stability study:

The effect of specimen storage on antigen stability was evaluated for both fresh samples and samples in transport media. For the analysis, a total of 15 fecal specimens were tested with the E. HISTOLYTICA QUIK CHEK test. The fecal specimens consisted 3 negative fecal samples, 3 E. histolytica low positive fecal samples (1-2 times C95 for whole organisms), 3 E. histolytica moderate positive fecal samples (2-3 times C95 for whole organism), and 3 E. histolytica high positive fecal samples (4-5 times C95 for whole organism). Both fresh samples and samples stored in transport media were tested.

For fresh samples, samples were stored between 2°C and 8°C and tested at 24 hour intervals for 120 hours. At the same time positive and negative controls were tested. The following transport media were used for the study: Thermo Scientific Protocol Cary Blair media and Meridian Bioscience Inc., Para-Pak C&amp;S. As recommended in the Cary Blair package insert, Cary Blair samples were stored between 20°C and 25°C. C&amp;S samples were stored at both refrigerated (2°C to 8°C) and prevailing room temperature (15°C to 30°C) conditions as indicated in the package insert. Samples stored in transport media were tested at 24 hour intervals for 120 hours. At the same time positive and negative controls were tested.

Positive samples remained positive and negative samples remained negative throughout the study. Storage in transport media did not affect the stability of the samples. Based on these results, the recommended storage time for fresh samples is

5

{5}

up to 120 hours between 2°C and 8°C, Cary Blair transport media is up to 96 hours between 20°C and 25°C and C&amp;S transport media is up to 120 hours between 2°C and 30°C based on the data and the package insert recommendation.

Frozen stability:

The frozen stability tests were determined for raw stool specimens using 15 masked fecal specimens panel prepared the same way as was described for the storage stability study. Samples were stores at ≤ -10°C for 28 weeks. Specimens were tested at 0, 1, 4, 6, 12, 16, 20, 24 and 28 weeks. Positive samples remained positive and negative samples remained negative throughout the study.

Freeze/Thaw:

Freeze/Thaw stability for Cary Blair and C&amp;S media compared to raw specimens was demonstrated for one Freeze/Thaw cycle. 60 masked samples were tested, ten true negative, ten high negative (just below C5 (whole organisms)), thirty low positives (1-2 times C95 (whole organisms)), and ten high positives (4-5 x LoD (whole organisms)) were tested for each transport media and raw stool. Based on the data submitted, the freeze-thaw cycle neither enhances nor impairs the performance of the E. HISTOLYTICA QUIK CHEK test.

d. Detection limit:

The cutoff point (limit of detection, LOD) for the E. HISTOLYTICA QUIK CHEK test was determined using whole organism spiked into unpreserved (raw stool) and preserved (Cary Blair and C&amp;S media). Additional testing of spiked purified native antigen into raw stool was also performed. The cutoff is the concentrations of organisms or antigen that yields a positive result 95% of the time and a negative result 5% of the time. The cutoff point was determined as the concentration that provided the correct result 95% of the time based on testing dilutions of whole organism and purified antigen in a negative fecal sample matrix, in replicates of 20.

The LoD for the E. HISTOLYTICA QUIK CHEK test was determined to be 320 organisms/mL (equivalent to 15 organisms detected per test) or 0.2 ng/mL of adhesin in raw stool. For specimens in Cary Blair media, the LoD was established at 275 organisms/mL (equivalent to 14 organisms detected per test). For specimens in C&amp;S media, the LoD was established at 245 organisms/mL (equivalent to 12 organism detected per test). No testing with adhesin was conducted in Cary Blair media or C&amp;S.

6

{6}

e. Analytical specificity:

Cross reactivity.

The E. HISTOLYTICA QUIK CHEK test was evaluated for cross-reactivity with the bacteria and viruses listed below. None of the strains were shown to interfere with the performance of the E. HISTOLYTICA QUIK CHEK test. Bacteria was spiked at concentration of &gt;10⁸ CFU/mL and viruses at a range from 10³·³ to 10⁸·⁷⁵ TCID₅₀ units per 0.2 mL.

|  Aeromonas hydrophila | Bacillus cereus  |
| --- | --- |
|  Bacillus subtilis | Bacteroides fragilis  |
|  Campylobacter coli | Campylobacter fetus  |
|  Campylobacter jejuni | Candida albicans  |
|  Clostridium bifermentans | Clostridium difficile  |
|  Enterococcus faecalis | Escherichia coli  |
|  Escherichia coli O157:H7 | Escherichia coli EIEC  |
|  Escherichia coli EPEC | Escherichia coli ETEC  |
|  Klebsiella pneumoniae | Salmonella typhimurium  |
|  Shigella dysenteriae | Shigella flexneri  |
|  Shigella sonnei | Staphylococcus aureus  |
|  Staphylococcus aureus (Cowan’s) | Staphylococcus epidermidis  |
|  Vibrio parahaemolyticus | Yersinia enterocolitica  |
|  Adenovirus, 2, 5, 40, 41 | Coxsackievirus B5  |
| --- | --- |
|  Echovirus 11, 18, 22, 33 | Enterovirus 68, 69  |
|  Human Adenovirus 1, 3 | Human Coronavirus  |
|  Human Coxsackievirus B2, B3, B4 | Human Echovirus 9  |
|  Human Enterovirus 69, 70, 71 | Human parechovirus 1 [Echovirus 22]  |
|  Human Rotavirus |   |

Cross reactivity with Norovirus is unknown because it was not tested in analytical studies. However, Norovirus GI/GII was identified in 50 clinical specimens using an FDA cleared multiplex NAAT assay during clinical testing and no cross reactivity was found using the E. HISTOLYTICA QUIK CHEK in those samples

The following parasites were found during clinical testing and found not to be cross reactant. The fecal specimens were documented to be positive for parasites by microscopy. The number in parenthesis is the number of individual clinical samples that were microscopy positive for the parasites listed below. No cross-reactivity was seen with any of the following organisms.

Ascaris lumbricoides and with eggs (21)
Entamoeba bangladeshi (3)
Giardia spp. (45)
Blastocystis hominis (12)

{7}

Entamoeba coli (13)
Iodamoeba bütschlii (10)
Cryptosporidium spp. (30)
Entamoeba moshkovskii (3)
Trichuris trichiura eggs (11)
Cryptosporidium spp. (30)
Entamoeba moshkovskii (3)
Trichuris trichiura eggs (11)

f. Assay cut-off:
Not applicable

g. Prozone:
The purpose of this study was to demonstrate that a high concentration of *E. histolytica* adhesion (antigen) does not interfere with a positive reaction in the *E. HISTOLYTICA QUIK CHEK* test. High samples were prepared by spiking a negative fecal pool with 20,000 x the LoD described in section M.1.c above down to a concentration possibly observed in clinical specimens. A total of 5 different dilutions of the analyte, including the clinically observed high concentration, were prepared and tested. Testing was performed in triplicate according to the Package Insert instructions. The results below demonstrated that there was no overall hook affect, that elevated levels of analyte did not affect the detection of the analyte.

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable

3. Clinical studies:
a. Clinical Sensitivity:
Prospective Study

The performance of the *E. HISTOLYTICA QUIK CHEK* test was evaluated at three independent sites. The sensitivity and specificity for the *E. HISTOLYTICA QUIK CHEK* device was determined using the following composite reference method (CRM) table 1 at all three sites:

{8}

Table 1. CRM algorithm

|  Microscopy | FDA cleared multiplex NAAT test | Alternate PCR with sequencing | CRM Result  |
| --- | --- | --- | --- |
|  Pos | Pos | Pos | Pos  |
|  Pos | Pos | Neg | Neg  |
|  Pos | Neg | Pos | Pos  |
|  Pos | Neg | Neg | Neg  |
|  Neg | Pos | Pos | Pos  |
|  Neg | Pos | Neg | Neg  |
|  Neg | Neg | N/A | Neg  |

Prospective testing consisted of 755 specimens from two non-endemic areas in the US and one endemic area in Bangladesh. The endemic area had 5 CRM positive specimens. The remaining 750 samples were all negative. Table 2 below summarizes the performance observed, which mainly focused on evaluation of specificity due to the low number of positive specimens (See retrospective study).

Table 2. Performance...

|  N = 755 | Composite Reference Method Positive | Composite Reference Method Negative  |
| --- | --- | --- |
|  E. HISTOLYTICA QUIK CHEK™ Positive | 2 | 0  |
|  E. HISTOLYTICA QUIK CHEK™ Negative | 3 | 750  |
|   |   | 95% Confidence Limits  |
| --- | --- | --- |
|  Sensitivity | 40.0% | 7.3% - 83.0%  |
|  Specificity | 100% | 99.4% - 100%  |
|  Predictive Positive Value | 100% | 19.8% - 100%  |
|  Predictive Negative Value | 99.6% | 98.7% - 99.9%  |
|  Correlation | 99.6% | 99.6% - 99.6%  |

It should be noted additional discrepant analysis was done on the three false positive results because all three were microscopy negative, FDA cleared multiplex NAAT test positive and alternate PCR positive. Additional testing was performed and all three samples were determined to be negative by an alternative antigen test.

{9}

Retrospective Study

Retrospective testing consisted of 96 archived specimens previously collected and frozen. The Frozen samples are characterized as microscopy and NAAT positive for inclusion into the specimen bank. The specimens were collected form an *E. histolytica* endemic area. Table 3 below summarizes the performance observed

Table 3...
|  N = 96 | Composite Reference Method Positive | Composite Reference Method Negative  |
| --- | --- | --- |
|  E. HISTOLYTICA QUIK CHEK™ Positive | 30 | 0  |
|  E. HISTOLYTICA QUIK CHEK™ Negative | 0 | 66  |
|   |   | 95% Confidence Limits  |
| --- | --- | --- |
|  Sensitivity | 100% | 85.9% - 100%  |
|  Specificity | 100% | 93.1% - 100%  |

The age range for the 851 samples is one year to 100 years old. 161 of the sample tested were from children less than 21 years old. 57.3 percent of the specimens were from males and 42.7% of the specimens were from females.

b. Clinical specificity:

See section M3a. above.

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable.

4. Clinical cut-off:

Not applicable.

5. Expected values/Reference range:

Not applicable.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

{10}

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

11

---

**Source:** [https://fda-staging.innolitics.com/device/K170728](https://fda-staging.innolitics.com/device/K170728)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
