The Universal Mesh is indicated for reconstruction, stabilization and/or rigid fixation subsequent to craniotomy, craniectomy, cranioplasty, and cranial fractures in adults and adolescents (age 12 and higher).
Device Story
Stryker Universal Mesh consists of 3D preformed titanium meshes designed for cranial reconstruction and stabilization. These devices are pre-bent versions of existing dynamic mesh designs, intended to reduce intraoperative contouring time. The meshes are available in symmetrical and anatomically preformed lateral variations, including a 1.0mm thickness option. The device is used by surgeons in clinical settings (e.g., OR) to bridge cranial defects. Fixation is achieved using screws. The mesh remains modifiable by the surgeon to ensure patient-specific fit. By providing a pre-contoured shape, the device aims to improve surgical efficiency while maintaining the structural integrity required for cranial stabilization. The device is MR conditional.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including static load/deformation testing over a 70mm diameter defect, sterilization validation, and MR compatibility testing per FDA guidance. Surgeon and non-trained user evaluations confirmed handling and usability requirements were met.
Technological Characteristics
Material: Titanium (ASTM F67). Form factor: 3D preformed mesh (symmetrical and anatomically preformed lateral). Thickness: Includes 1.0mm option. Fixation: Screw-based. Sterilization: Non-sterile, validated per ISO 14937, 17664, 17665. Connectivity: None. Software: None.
Indications for Use
Indicated for reconstruction, stabilization, and rigid fixation of cranial skeleton following craniotomy, craniectomy, cranioplasty, or cranial fractures in adults and adolescents aged 12 and older.
Regulatory Classification
Identification
A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 13, 2017
Stryker Jonathan Schell Staff Regulatory Affairs Specialist 750 Trade Centre Way-Suite 200 Portage, Michigan 49002
Re: K170725
Trade/Device Name: Stryker Universal Mesh Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed Alterable Cranioplasty Plate Regulatory Class: Class II Product Code: GWO Dated: October 10, 2017 Received: October 11, 2017
Dear Mr. Schell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Hoffmann -S
Carlos L. Peña. PhD. MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170725
Device Name Stryker Universal Mesh
Indications for Use (Describe)
The Universal Mesh is indicated for reconstruction, stabilization subsequent to craniotomy, craniectomy, cranioplasty, and cranial fractures in adults and adolescents (age 12 and higher).
Type of Use (Select one or both, as applicable)
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Stryker logo in black. The logo is a stylized wordmark with the word "stryker" in a bold, sans-serif font. A small circle with the letter 'c' inside is located to the upper right of the wordmark.
## 510(k) Summary
This section provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.
- SUBMITTER [§807.92(a)(1)] I.
| 510(k) Owner: | Stryker Leibinger GmbH & Co. KG<br>Boetzinger Strasse 41<br>79111 Freiburg, Germany |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter/ Contact<br>Person: | Jonathan Schell<br>Staff Regulatory Affairs Specialist<br>Stryker Craniomaxillofacial<br>750 Trade Centre Way<br>Portage, MI 49002<br>Phone: 269-389-5596<br>Fax: 877-648-7114 |
Date prepared: November 6, 2017
- DEVICE [§807.92(a)(2)] II.
Trade Name: Stryker Universal Mesh
Bone Plate Common or Usual name:
Classification Preformed Alterable Cranioplasty Plate name:
Regulatory Class: Class II
Product Code: GWO
## III. PREDICATE DEVICE [§807.92(a)(3)]
Universal Mesh System - K161821
### IV. DEVICE DESCRIPTION [§807.92(a)(4)]
The 3D Preformed meshes are added to the Universal Mesh system in order to offer a more comprehensive selection of mesh solutions for cranial reconstruction. Simply put, the 3D Preformed meshes are the 1.5/1.7mm Dynamic Mesh predicate design provided in a
{4}------------------------------------------------
preformed shape. The 3D Preformed meshes are provided non-sterile in 2 different variations; these meshes are offered in a Symmetrical Preformed mesh and an Anatomically preformed lateral mesh. The Anatomically preformed lateral 3D meshes add a 1.0mm thick mesh option to the portfolio, which was not available with the predicate device.
#### INDICATIONS FOR USE [§807.92(a)(5 )] V.
| | Subject Device | Predicate – K161821 |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Universal Mesh is indicated<br>for reconstruction, stabilization<br>and/or rigid fixation subsequent<br>to craniotomy, craniectomy,<br>cranioplasty, and cranial fractures<br>in adults and adolescents (age 12<br>and higher). | The Universal Mesh is indicated<br>for reconstruction, stabilization<br>and/or rigid fixation subsequent to<br>craniotomy, craniectomy,<br>cranioplasty, and cranial fractures<br>in adults and adolescents (age 12<br>and higher). |
#### TABLE 1: COMPARISON OF INDICATIONS FOR USE
The predicate device is indicated for use in the cranial skeleton. The Indications for Use statement of the subject Universal 3D Preformed Mesh is unchanged and remains identical to the predicate device.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE [§807.92(a)(6)]
The Stryker Universal Mesh is compared to its predicate devices for substantial equivalence based on the following criteria:
- A. Principle of Operation
- B. Technological Characteristics
## A. Principle of Operation
The Stryker Universal 3D Preformed Mesh implants are designed for the reconstruction, stabilization and rigid fixation of the bony areas of the cranial skeleton. The necessary instrumentation are included with the system.
The Principle of Operation of the subject Universal 3D Preformed Mesh is unchanged and remains identical to the predicate device.
## B. Technological and Operational Characteristics
At a high level, the Subject devices and Predicate devices are based on the following technological elements:
{5}------------------------------------------------
- Same materials: Both the subject device and predicate device are made of titanium according to ASTM F67
- -Same mode of fixation: implant fixation with screws
- -Same mesh pattern: The Subject device has the same Dynamic Mesh pattern as the predicate device, however is provided in 1.0mm thickness.
- -Similar Design/Sizes: The Subject device 3D Preformed mesh is designed to offer the surgeon a pre-bent version of an existing mesh, to minimize the time the surgeon needs for bending the implant into the desired shape. The 3D Preformed mesh are designed to be modified by the surgeon as needed to fit the patient needs, the same as the predicate mesh.
## VII. PERFORMANCE DATA [§807.92(b)(1)]
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility Testing
The biocompatibility evaluation for the subject device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The tests supported the biocompatibility of the device.
The implant is made from titanium which conforms to ASTM F67 for chemical composition, which is the same as the predicate. Cytotoxicity testing was performed using DIN EN 10993-5:2009, Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity. The tests supported the biocompatibility of the device.
## Performance Bench Testing
| Test | Test Method Summary | Results |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Sterilization<br>Validation | Validate the sterilization instructions for the Subject device according<br>to internal standard. Referenced standards: ISO 14937, ISO 17664,<br>ISO 17665 - 1&2, AAMI TIR 12, AAMI ST 81, AAMI ST 79. | Method<br>validated. |
| Static Load Max.<br>Defect | Determination of defined load and deformation over maximum defect<br>according to labeling of fixated meshes with bone simulation material | Acceptance<br>criterion<br>were met, tests<br>were passed. |
The following performance bench tests were completed.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase, and the letters are closely spaced together. A small registered trademark symbol is located to the right of the letter "r".
| Static Load | Determination of defined load and deformation over 70mm diameter defect according to labeling of fixated meshes with bone simulation material. Comparison to predicate device tested over same defect. | Acceptance criterion were met, tests were passed. |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| End User Validation Testing | Surgeon evaluation | Acceptance criterion was met, test was passed |
| Internal Handling Test | Non-trained user evaluation | Acceptance criterion was met, test was passed |
| MR Compatibility | Magnetic Resonance (MR) testing was performed on the subject device mesh implants in accordance with FDA guidance for industry "Assessment of Radiofrequency - Induced Heating in the Magnetic Resonance (MR) Environment for Multi-Configuration Passive Medical Devices,"March 22, 2016. MR heat testing has been performed on the Subject device according to FDA final guidance "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment," December 11, 2014. | Based on the results of these evaluations, the Stryker Universal mesh will be labeled as MR conditional. |
The Subject device met all pre-defined acceptance criteria and standards and was found to not represent a new worst case. Overall, the results of the performance bench tests support the substantial equivalence of the Subject device to the predicate.
## Animal Testing
Animal testing was not required as a basis for substantial equivalence.
## Clinical Testing [§807.92(b)(2)]
Clinical testing was not required as a basis for substantial equivalence.
## VIII. CONCLUSIONS [§807.92(b)(3)]
The results of the non-clinical data demonstrate the Stryker Universal 3D Preformed Mesh will perform as intended in the specified use conditions. According to the comparison based on the requirements of 21 CFR 807.92 and the information provided herein, it is concluded that the information included in this submission supports substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.