K170722 · Moxxly, Inc. · HGX · Aug 28, 2017 · Obstetrics/Gynecology
Device Facts
Record ID
K170722
Device Name
Moxxly Flow Collection System
Applicant
Moxxly, Inc.
Product Code
HGX · Obstetrics/Gynecology
Decision Date
Aug 28, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5160
Device Class
Class 2
Indications for Use
The Moxxly Flow breast pump collection system is intended to be used in conjunction with a compatible powered breast pump for the purpose of expressing human milk.
Device Story
Moxxly Flow is a breast pump collection system used with compatible electric breast pumps; designed for lactating women to express milk. Device consists of two collection units featuring sloped flanges, flexible necks, internal valve assemblies, and low-profile bottles. Worn inside a nursing bra during operation; milk flows from breast through flange and valve into collection bottle. After pumping, user removes device and transfers milk to storage. Provides hands-free collection; aids comfort. Non-sterile, single-patient reusable device; cleaned by user via soap/water, dishwasher, or boiling. Benefits include discreet, convenient milk expression.
Clinical Evidence
Bench testing only. Verification and validation testing included flange design, vacuum performance, capacity, leakage, usability, and biocompatibility. Biocompatibility testing performed per ISO 10993-5:2009 (cytotoxicity) and ISO 10993-10:2010 (irritation/sensitization). Usability evaluated per IEC 62366-1:2015.
Technological Characteristics
Non-sterile, single-patient reusable breast pump collection system. Components: sloped flange, flexible neck, internal valve, collection bottle. Materials biocompatible per ISO 10993-5 and ISO 10993-10. Dimensions: 24mm/27mm flange sizes, 10 oz capacity. Connectivity: mechanical interface with compatible electric breast pumps. Cleaning: soap/water, dishwasher, or boiling.
Indications for Use
Indicated for lactating mothers requiring a collection system for use with compatible powered breast pumps to express human milk.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
Predicate Devices
DAO Health Freemie Breast Pump Collection System (K130349)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 28, 2017
Moxxly, Inc. Rory A. Carrillo Ouality/Regulatory 1777 Yosemite Ave. Suite 235 San Francisco. CA 94124
K170722 Re:
Trade/Device Name: Moxxly Flow Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: July 18, 2017 Received: July 24, 2017
Dear Rory A. Carrillo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170722
Device Name Moxxly Flow
Indications for Use (Describe)
The Moxxly Flow breast pump collection system is intended to be used in conjunction with a compatible powered breast pump for the purpose of expressing human milk
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### A. Submitter
Moxxly, Inc. 1777 Yosemite Avenue, Suite 235 San Francisco, CA 94124 Phone: (650) 416-6084 Email: cara@moxxly.com
#### B. Date Prepared
August 28, 2017
#### C. Contact Person
Rory A. Carrillo Quality/Regulatory Phone: Email:
| Device Proprietary<br>Name: | Moxxly Flow |
|-------------------------------------------------------------|------------------------------------|
| Device Common Name | Powered Breast Pump |
| Device Classification:<br>Regulatory Class:<br>Product Code | 21 CFR§884.5160<br>Class II<br>HGX |
#### D. Device Name and Classification
#### E. Predicate Device
The Moxxly Flow is substantially equivalent to the DAO Health Freemie Breast Pump Collection System, K130349.
The predicate devices have not been subject to a design-related recall.
#### F. Intended Use
The Moxxly Flow breast pump collection system is intended to be used in conjunction with a compatible powered breast pump for the purpose of expressing human milk.
#### G. Device Description
Moxxly, Inc. is the manufacturer of the Moxxly Flow, a device intended to be used in conjunction with a compatible breast pump for the purpose of expressing human milk.
The Moxxly Flow consists of two collection units designed to be compatible with select
{4}------------------------------------------------
compatible electric breast pumps and provide a more comfortable design to aid collection of breast milk. The Moxxly Flow is provided non-sterile and is a single patient reusable device. The system should be cleaned by the user prior to use.
The device has a sloped flange (cup insert), flexible neck, internal valve assembly and low-profile bottle. Variations of the device incorporate minor differences to replicate the functional and performance characteristics necessary for use with additional pump brands.
The Moxxly device is designed to fit within a lactating woman's ordinary or nursing bra and held in place there while breast pumping. When the pump extracts milk, the milk flows out through the end of the flange, where it passes through the neck and valve system at the bottom of the bra and into the attached collection bottle. When the lactating woman is done pumping, she turns off the pump, removes the Moxxly device from under her clothes and transfers the milk to a storage container for later use.
| MFR | DAO Health | Moxxly, Inc. |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model<br>Name | Freemie Breast Pump<br>Collection System (K130349) | Moxxly Flow (K170722) |
| Indications<br>for Use | The Freemie breast pump<br>collection system is intended to<br>be used in conjunction with an<br>approved powered breast pump<br>for the purpose of expressing<br>human milk. | The Moxxly Flow breast pump<br>collection system is intended to<br>be used in conjunction with a<br>compatible powered breast<br>pump for the purpose of<br>expressing human milk. |
| MFR | DAO Health | Moxxly, Inc. |
| Model Name | Freemie Breast Pump<br>Collection System | Moxxly Flow |
| 510(k)<br>Number | K130349 | K170722 |
| Classification | 21 CFR§884.5160 | 21 CFR§884.5160 |
| Device Type | Pump, Breast, Powered | Pump, Breast, Powered |
| Product Code | HGX | HGX |
| User Population | Lactating Mothers | Lactating Mothers |
| Target Area | Breast | Breast |
| How Supplied | Non-sterile | Non-sterile |
| Single User | Yes | Yes |
| Cleaning Methods | Soap and warm water<br>Dishwasher<br>Boiling | Soap and warm water<br>Dishwasher<br>Boiling |
| Flange Size(s) | 25mm<br>28mm<br>32mm | 24mm (Medium)<br>27mm (Large) |
| Overflow Protection | Yes | Yes |
| Volume Capacity | 16 oz | 10 oz |
| Pump Compatibility | Yes | Yes |
| | • Pump in Style Advance<br>• Medela Symphony<br>• Lactina<br>• Ameda Purely Yours<br>• Twin Electric Breast Pump<br>• Spectra Dew 300<br>• Spectra Dew 350 | • Ameda Purely Yours<br>• Evenflo Advanced<br>• Freemie Freedom<br>• Hygeia Enjoye<br>• Medela Pump In Style<br>Advanced<br>• Medela Symphony |
#### H. Comparison of Indications for Use with Predicate Device
The intended use is similar to that of the predicate device.
{5}------------------------------------------------
#### I. Comparison of Technical Characteristics with Predicate Device
The subject device and the predicate device do not have the same technological characteristics. The flange size, collection bottle size, material, and compatible pumps are different between the subject device and the predicate device.
The different technological characteristics of the subject device do not raise different safety or
{6}------------------------------------------------
effectiveness questions because all devices used for collection of breast milk in conjunction with breast must demonstrate that they are biocompatible and work in conjunction with the identified compatible breast pump.
#### J. Non-Clinical Performance Test Summary
Verification and Validation was conducted to demonstrate safety and effectiveness and for comparison to the predicate device. Testing included: flange design, compatible pumps, vacuum performance, capacity and from leakage, and ability to be supported, usability, and biocompatibility. Test results established that the device meets its design requirements and intended use, that it is as safe, as effective, and performs as well as the predicate device. Testing data demonstrate the device meets all of its specifications including compliance with the following:
Biocompatibility:
- ISO 10993-5:2009 Biological Evaluation of Medical Devices Part 5: For In . vitro Cytotoxicity
- . ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Test for Irritation and Skin Sensitization
Usability:
- IEC 62366-1:2015 Medical Devices Part 1: Application of Usability . Engineering for Medical Devices
#### K. Conclusion
The Moxxly Flow is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.