K170714 · GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC · IYN · Jun 21, 2017 · Radiology
Device Facts
Record ID
K170714
Device Name
Venue
Applicant
GE Medical Systems Ultrasound & Primary Care Diagnostics, LLC
Product Code
IYN · Radiology
Decision Date
Jun 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Venue is a general purpose diagnostic ultrasound system for use by qualified healthcare professionals. The clinical environments where the Venue can be used include critical care and emergency room environments, as well as point-of care areas in offices, clinical and hospital settings for diagnosis of patients. The Venue is intended for ultrasound imaging, measurement and analysis of the human body and fluid for multiple clinical applications including: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), Peripheral vascular, neonatal and adult cephalic, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transvaginal, and imaging guidance of interventional procedures (e.g. Nerve block, vascular access). When Pinpoint GT Technology, from C.R. Bard, Inc., is included with the system, the Indications for Use include: Pinpoint GT Technology is intended to provide clinicians with visual tools for passive magnetic tracking of a needle with respect to ultrasound image data.
Device Story
Venue is a mobile, general-purpose diagnostic ultrasound system for point-of-care, critical care, and emergency settings. It acquires ultrasound signals via linear, convex, and phased array transducers; processes data for imaging, measurement, and analysis; and displays output on a 19-inch high-resolution touch-screen monitor. Operated by qualified healthcare professionals, the system supports standard acquisition modes (B, M, Doppler, etc.) and includes optional Pinpoint GT technology for passive magnetic needle tracking. The device features a battery for portable use, wireless connectivity, and DICOM-compliant storage/archiving. By providing real-time visualization and guidance for procedures like nerve blocks and vascular access, the system assists clinicians in diagnosis and interventional accuracy, potentially improving patient outcomes in acute care environments.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence consists of bench testing, including acoustic output measurement, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety testing in compliance with recognized standards (e.g., IEC 60601-1, ISO 10993-1).
Technological Characteristics
Mobile ultrasound system with 19-inch color LCD touch-screen. Utilizes linear, convex, and phased array transducers. Supports B, M, PW/CW Doppler, Color/Power Doppler, and harmonic imaging. Connectivity via wireless and DICOM. Biocompatible patient-contact materials. Software runs on Windows 10. Acoustic power levels meet FDA limits. Sterilization/disinfection via standard cleaning agents.
Indications for Use
Indicated for qualified healthcare professionals performing diagnostic ultrasound imaging, measurement, and analysis of the human body and fluid. Applications include abdominal, thoracic/pleural, ophthalmic, fetal/OB, small organ, peripheral vascular, neonatal/adult cephalic, musculoskeletal, cardiac, and transvaginal imaging, plus interventional guidance (nerve block, vascular access).
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 22, 2017
GE MEDICAL SYSTEMS ULTRASOUND & PRIMARY CARE DIAGNOSTICS LLC MS. TRACEY ORTIZ REGULATORY AFFAIRS DIRECTOR 9900 W. INNOVATION DRIVE WAUWATOSA, WI 53226
Re: K170714
Trade/Device Name: Venue Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, ITX, IYO Dated: May 25, 2017 Received: May 26, 2017
Dear Ms. Ortiz:
This letter corrects our substantially equivalent letter of June 21, 2017.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D. O'Hara For
Robert Ochs Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
K170714
Device Name Venue
Indications for Use (Describe)
The Venue is a general purpose diagnostic ultrasound system for use by qualified healthcare professionals. The clinical environments where the Venue can be used include critical care and emergency room environments, as well as point-ofcare areas in offices, clinical and hospital settings for diagnosis of patients.
The Venue is intended for ultrasound imaging, measurement and analysis of the human body and fluid for multiple clinical applications including: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), Peripheral vascular, neonatal and adult cephalic, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transvaginal, and imaging guidance of interventional procedures (e.g. Nerve block, vascular access).
When Pinpoint GT Technology, from C.R. Bard, Inc., is included with the system, the Indications for Use include: Pinpoint GT Technology is intended to provide clinicians with visual tools for passive magnetic tracking of a needle with respect to ultrasound image data.
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| <div> <span style="text-decoration: overline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo, which is a blue circle with the letters 'GE' in a stylized script in the center. The letters are also blue, matching the color of the circle. There are three white swirls evenly spaced around the inside of the circle, adding a dynamic element to the design. The logo is simple, recognizable, and represents the multinational conglomerate corporation.
## Indications for Use Forms
The following forms represent indications with clinical applications and exam types along with the modes of operation for the Venue. Combinations identified "P" represents those previously cleared with another GE Ultrasound system. Combinations identified as "N" are new.
The following Indication for Use forms are appended:
System: Venue Transducer: 3Sc-RS Transducer: 9L-RS Transducer: C1-5-RS Transducer: 8C-RS Transducer: E8C-RS Transducer: 12L-RS Transducer: L12n-RS
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The frame has decorative, swirling elements around the letters. The logo is colored in a light blue hue, and the background is white.
GE Healthcare 510(k) Premarket Notification Submission
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-----------------------|---------------------|-----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*1] | Harmonic<br>Imaging | Coded<br>Pulse• | Other |
| Ophthalmic | N | N | N | N | N | N | N | N | N | N | N |
| Fetal / Obstetrics | N | N | N | N | N | N | N | N | N | N | N |
| Abdominal [1] | N | N | N | N | N | N | N | N | N | N | N |
| Pediatric | N | N | N | N | N | N | N | N | N | N | N |
| Small Organ [2] | N | N | N | N | N | N | N | N | N | N | N |
| Neonatal Cephalic | N | N | N | N | N | N | N | N | N | N | N |
| Adult Cephalic | N | N | N | N | N | N | N | N | N | N | N |
| Cardiac [3] | N | N | N | N | N | N | N | N | N | N | 8 |
| Peripheral Vascular | N | N | N | N | N | N | N | N | N | N | N |
| Musculo-skeletal Conventional | N | N | N | N | N | N | N | N | N | N | N |
| Musculo-skeletal Superficial | N | N | N | N | N | N | N | N | N | N | N |
| Thoracic/Pleural [4] | N | N | N | N | N | N | N | N | N | N | N |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | N | N | N | N | N | N | N | N | N | N | N |
| Transvaginal | N | N | N | N | N | N | N | N | N | N | N |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Vascular Access (IV, PICC) | N | N | N | N | N | N | N | N | N | N | 6,7,9 |
| Nonvascular [5] | N | N | N | N | N | N | N | N | N | N | 6,7,9 |
#### GE Venue Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA K151028; P1 = previously cleared by FDA K163596;
Notes: [1] Abdominal includes GYN and Urology (includes prostate);
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac is Adult and Pediatric;
[4] Including detection of fluid and pleural motion/sliding;
[5] Nonvascular includes nerve block or biopsy;
[6] Needle guidance with Pinpoint™ GT needle technology;
[7] Biopsy bracket available;
[8] Combined modes also include: B/CWD, B/Color/CWD;
[9] Image guidance supports freehand needle/catheter placement;
[*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
[↓] Coded pulse is for digitally encoded harmonics.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The frame has decorative, swirling elements around the perimeter. The logo is presented in a blue color.
GE Healthcare 510(k) Premarket Notification Submission
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse• | Other |
| Ophthalmic | P | | | | P | | P | | P | | |
| Fetal / Obstetrics | P | P | P | | P | P | P | P | P | | |
| Abdominal [1] | P | P | P | | P | P | P | P | P | | |
| Pediatric | | | | | | | | | | | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | P | P | P | | P | P | P | P | P | | |
| Cardiac [3] | P | P | P | P | P | P | P | P | P | | 8 |
| Peripheral Vascular | N | N | N | N | N | N | N | N | N | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural [4] | P | P | P | | P | P | P | P | P | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Vascular Access (IV, PICC) | P | P | P | | P | P | P | P | P | | 9 |
| Nonvascular [5] | | | | | | | | | | | |
#### GE Venue with 3Sc-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA K151028; P1 = previously cleared by FDA K163596;
Notes: [1] Abdominal includes GYN and Urology (includes prostate);
- [2] Small Organ includes breast, testes, thyroid;
[3] Cardiac is Adult and Pediatric;
- [4] Including detection of fluid and pleural motion/sliding;
- [5] Nonvascular includes nerve block or biopsy;
- [6] Needle guidance with Pinpoint™ GT needle technology;
[7] Biopsy bracket available;
- [8] Combined modes also include: B/CWD, B/Color/CWD;
[9] Image guidance supports freehand needle/catheter placement;
- [*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
- [↓] Coded pulse is for digitally encoded harmonics.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular border. The color of the logo is a light blue.
510(k) Premarket Notification Submission
#### GE Venue with 9L-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse♦ | Other | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | | |
| Abdominal [1] | P | P | P | P | P | P1 | P | P | P | P | | |
| Pediatric | P | P | P | P | P | P1 | P | P | P | P | | |
| Small Organ [2] | P | P | P | P | P | P1 | P | P | P | P | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac [3] | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | P | P | P1 | P | P | P | P | | |
| Musculo-skeletal Conventional | P | P | P | P | P | N | P | P | P | P | | |
| Musculo-skeletal Superficial | P | P | P | P | P | N | P | P | P | P | | |
| Thoracic/Pleural [4] | P | P | P | P | P | N | P | P | P | P | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | | |
| Vascular Access (IV, PICC) | P | P | P | P | P | N | P | P | P | P | 9 | |
| Nonvascular [5] | P | P | P | P | P | N | P | P | P | P | 9 | |
N = new indication; P = previously cleared by FDA K151028; P1 = previously cleared by FDA K163596;
Notes: [1] Abdominal includes GYN and Urology (includes prostate);
- [2] Small Organ includes breast, testes, thyroid;
[3] Cardiac is Adult and Pediatric;
- [4] Including detection of fluid and pleural motion/sliding;
- [5] Nonvascular includes nerve block or biopsy;
- [6] Needle quidance with Pinpoint™ GT needle technology:
[7] Biopsy bracket available;
- [8] Combined modes also include: B/CWD, B/Color/CWD;
[9] Image guidance supports freehand needle/catheter placement;
[*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
- [ +] Coded pulse is for digitally encoded harmonics.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a circular frame. The logo is colored in a shade of blue.
510(k) Premarket Notification Submission
### GE Venue with C1-5-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|----|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|------------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse * | Other | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | P1 | P | P | P | P | | |
| Abdominal [1] | P | P | P | | P | P1 | P | P | P | P | | |
| Pediatric | P | P | P | | P | P1 | P | P | P | P | | |
| Small Organ [2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac [3] | | | | | | | | | | | | |
| Peripheral Vascular | P1 | P1 | P1 | | P1 | P1 | P1 | P1 | P1 | P1 | | |
| Musculo-skeletal Conventional | P | P | P | | P | N | P | P | P | P | | |
| Musculo-skeletal Superficial | P | P | P | | P | N | P | P | P | P | | |
| Thoracic/Pleural [4] | N | N | N | | N | N | N | N | N | N | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | | |
| Vascular Access (IV, PICC) | N | N | N | | N | N | N | N | N | N | 9 | |
| Nonvascular [5] | P | P | P | | P | N | P | P | P | P | 9 | |
N = new indication; P = previously cleared by FDA K151028; P1 = previously cleared by FDA K163596;
Notes: [1] Abdominal includes GYN and Urology (includes prostate);
- [2] Small Organ includes breast, testes, thyroid;
[3] Cardiac is Adult and Pediatric;
- [4] Including detection of fluid and pleural motion/sliding;
- [5] Nonvascular includes nerve block or biopsy;
- [6] Needle quidance with Pinpoint™ GT needle technology:
[7] Biopsy bracket available;
- [8] Combined modes also include: B/CWD, B/Color/CWD;
[9] Image guidance supports freehand needle/catheter placement;
[*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
- [ +] Coded pulse is for digitally encoded harmonics.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle is surrounded by decorative flourishes, giving it a classic and recognizable appearance.
GE Healthcare 510(k) Premarket Notification Submission
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse• | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal [1] | P | P | P | | P | P1 | P | P | P | P1 | |
| Pediatric | P | P | P | | P | P1 | P | P | P | P1 | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | P | P | P | | P | P1 | P | P | P | P1 | |
| Adult Cephalic | P | P | P | | P | N | P | P | P | N | |
| Cardiac [3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | P | P | P | | P | N | P | P | P | N | |
| Musculo-skeletal Superficial | P | P | P | | P | N | P | P | P | N | |
| Thoracic/Pleural [4] | P | P | P | | P | N | P | P | P | N | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nonvascular [5] | | | | | | | | | | | |
#### GE Venue with 8C-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA K151028; P1 = previously cleared by FDA K163596;
- [2] Small Organ includes breast, testes, thyroid;
- [3] Cardiac is Adult and Pediatric;
- [4] Including detection of fluid and pleural motion/sliding;
- [5] Nonvascular includes nerve block or biopsy;
- [6] Needle guidance with Pinpoint™ GT needle technology;
- [7] Biopsy bracket available;
- [8] Combined modes also include: B/CWD, B/Color/CWD;
- [9] Image guidance supports freehand needle/catheter placement;
- *1 Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
- + ] Coded pulse is for digitally encoded harmonics.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. The circle is surrounded by decorative flourishes, giving it a classic and recognizable appearance.
GE Healthcare 510(k) Premarket Notification Submission
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|-----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[1] | Harmonic<br>Imaging | Coded<br>Pulse• | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | P1 | P | P | P | P1 | |
| Abdominal [1] | P | P | P | | P | P1 | P | P | P | P1 | |
| Pediatric | | | | | | | | | | | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac [3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural [4] | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | P | P | P | | P | P1 | P | P | P | P1 | |
| Transvaginal | P | P | P | | P | P1 | P | P | P | P1 | |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nonvascular [5] | | | | | | | | | | | |
#### GE Venue with E8C-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA K151028; P1 = previously cleared by FDA K163596;
- [2] Small Organ includes breast, testes, thyroid;
- [3] Cardiac is Adult and Pediatric;
- [4] Including detection of fluid and pleural motion/sliding;
- [5] Nonvascular includes nerve block or biopsy;
- [6] Needle guidance with Pinpoint™ GT needle technology;
- [7] Biopsy bracket available;
- [8] Combined modes also include: B/CWD, B/Color/CWD;
- [9] Image guidance supports freehand needle/catheter placement;
- [*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
- · ] Coded pulse is for digitally encoded harmonics.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a circular frame. The logo is colored in a shade of blue, and the background is white.
GE Healthcare 510(k) Premarket Notification Submission
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|------------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse • | Other |
| Ophthalmic | P | P | P | P | P | | P | P | P | P | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal [1] | P | P | P | P | P | P1 | P | P | P | P | |
| Pediatric | P | P | P | P | P | P1 | P | P | P | P | |
| Small Organ [2] | P | P | P | P | P | P1 | P | P | P | P | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac [3] | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | P | P | P1 | P | P | P | P | |
| Musculo-skeletal Conventional | P | P | P | P | P | P1 | P | P | P | P | |
| Musculo-skeletal Superficial | P | P | P | P | P | P1 | P | P | P | P | |
| Thoracic/Pleural [4] | P | P | P | P | P | N | P | P | P | P | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Vascular Access (IV, PICC) | P | P | P | P | P | N | P | P | P | P | 7,9 |
| Nonvascular [5] | P | P | P | P | P | N | P | P | P | P | 7,9 |
#### GE Venue with 12L-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA K151028; P1 = previously cleared by FDA K163596;
- [2] Small Organ includes breast, testes, thyroid;
- [3] Cardiac is Adult and Pediatric;
- [4] Including detection of fluid and pleural motion/sliding;
- [5] Nonvascular includes nerve block or biopsy;
- [6] Needle guidance with Pinpoint™ GT needle technology;
- [7] Biopsy bracket available;
- [8] Combined modes also include: B/CWD, B/Color/CWD;
- [9] Image guidance supports freehand needle/catheter placement;
- [*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
- · ] Coded pulse is for digitally encoded harmonics.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a blue circle. The circle is surrounded by decorative swirls or flourishes, giving it a classic and recognizable appearance.
GE Healthcare 510(k) Premarket Notification Submission
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|----|---------------|---------------|------------------|--------------------|------------------|----------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes[*] | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic | P1 | | | | P1 | | P1 | P1 | P1 | N | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal [1] | P1 | P1 | N | | P1 | N | P1 | P1 | P1 | N | |
| Pediatric | P1 | P1 | N | | P1 | N | P1 | P1 | P1 | N | |
| Small Organ [2] | P1 | P1 | N | | P1 | N | P1 | P1 | P1 | N | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac [3] | | | | | | | | | | | |
| Peripheral Vascular | P1 | P1 | N | | P1 | N | P1 | P1 | P1 | N | |
| Musculo-skeletal Conventional | P1 | P1 | N | | P1 | N | P1 | P1 | P1 | N | |
| Musculo-skeletal Superficial | P1 | P1 | N | | P1 | N | P1 | P1 | P1 | N | |
| Thoracic/Pleural [4] | N | N | N | | N | N | N | N | N | N | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Vascular Access (IV, PICC) | N | N | N | | N | N | N | N | N | N | 6, 9 |
| Nonvascular [5] | N | N | N | | N | N | N | N | N | N | 6, 9 |
#### GE Venue with L12n-RS Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA K151028; P1 = previously cleared by FDA K163596;
- [2] Small Organ includes breast, testes, thyroid;
- [3] Cardiac is Adult and Pediatric;
- [4] Including detection of fluid and pleural motion/sliding;
- [5] Nonvascular includes nerve block or biopsy;
- [6] Needle guidance with Pinpoint™ GT needle technology;
- [7] Biopsy bracket available;
- [8] Combined modes also include: B/CWD, B/Color/CWD;
- [9] Image guidance supports freehand needle/catheter placement;
- [*] Combined modes are: B/M, B/Color M, B/PWD, B/Color/PWD, B/Power/PWD;
- · ] Coded pulse is for digitally encoded harmonics.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The frame has decorative swirls or flourishes around the perimeter. The logo is presented in a blue color scheme.
510(k) Premarket Notification Submission
## 510(k) Summary
| | In accordance with 21 CFR 807.92 the following summary of information is provided: |
|---------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date:<br>Submitter: | March 8, 2017<br>GE Medical Systems Ultrasound and Primary Care Diagnostics<br>9900 Innovation Drive<br>Wauwatosa, WI 53226 |
| Primary Contact Person: | Tracey Ortiz<br>Regulatory Affairs Director<br>GE Healthcare<br>T:(262)676-6120<br>F:(414)918-8275 |
| Secondary Contact Person: | Karin Shimoni<br>Regulatory Affairs Leader<br>GE Medical Systems Israel LTD. |
| Device Trade Name:<br>Common/Usual Name:<br>Classification Names: | Venue<br>Diagnostic Ultrasound System<br>Class II<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90- |
| Product Code: | IYN; Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560,<br>90-IYO; Diagnostic Ultrasound Transducer, 21 CFR 892.1570,<br>90-ITX |
| Primary Predicate Device:<br>Secondary Predicate<br>Device(s): | Venue 50 (K152758)<br>LOGIQe (K151028)<br>Vivid S60/S70 (K142323)<br>LOGIQ P9 and LOGIQ P7 (K163596) |
| Device Description: | The proposed Venue system is a general-purpose, Track 3,<br>diagnostic ultrasound device, intended for ultrasound imaging,<br>measurement and analysis of the human body and fluid that<br>provides digital acquisition, processing and display capabilities.<br>Venue can be used in offices, clinical areas and hospitals, with a<br>focus in critical care and emergency rooms settings.<br>The Venue is a mobile system with a small footprint that easily<br>fits into tight spaces and positioned to accommodate the<br>sometimes-awkward work settings of the point of care user.<br>The Venue has a 19" high resolution color LCD monitor, with a<br>simple, multi-touch user interface that makes the system intuitive.<br>The single surface screen can be cleaned with disinfectants. |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a blue circle. There are decorative swirls around the circle, giving it a classic and recognizable appearance.
Articulated monitor arm enables flexible display positions in order to be accessible and clearly visible in both user-standing and sitting positions.
The Venue has a battery that allows scanning continuously for hours without the need to plug in to an electrical outlet. The system is capable of wireless communication and a barcode reader is optionally available to be used as an input device. System meets DICOM requirements to support users image storage and archiving needs and allows for output to printing devices.
The Venue utilizes a variety of linear, convex, and phased array transducers which provide high imaging capability, supporting all standard acquisition modes. Some biopsy kits are available for needle-guidance procedures and Pinpoint GT option is also available.
- Intended Use: The Venue is a general purpose diagnostic ultrasound system for use by qualified healthcare professionals. The clinical environments where the Venue can be used include critical care and emergency room environments, as well as point-of care areas in offices, clinical and hospital settings for diagnosis of patients. The Venue is intended for ultrasound imaging, measurement and analysis of the human body and fluid for multiple clinical applications including: abdominal (GYN and Urology), thoracic/pleural, ophthalmic, Fetal/OB, Small Organ (including breast, testes, thyroid), Peripheral vascular, neonatal and adult cephalic, pediatric, musculoskeletal (conventional and superficial), cardiac (adults and pediatric), Transrectal, Transvaginal, and imaging guidance of interventional procedures (e.g. Nerve block, vascular access). When PinpointTM GT Technology, from C.R. Bard, Inc., is included with the system, the Indications for Use include: Pinpoint™ GT Technology is intended to provide clinicians with visual tools for passive magnetic tracking of a needle with respect to ultrasound image data.
- Technology: The Venue employs the same fundamental scientific technology as its predicate devices.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. The circle has a decorative, wave-like border, also in blue, adding a touch of elegance to the design.
510(k) Premarket Notification Submission
Determination of Comparison to Predicate Devices The Venue system is substantially equivalent to the predicate Substantial Equivalence: devices with regards to intended use, imaging capabilities, technological characteristics and safety and effectiveness. ● The systems are all intended for ultrasound imaging.
- measurement and analysis of the human body and fluid for multiple clinical applications.
- . The Venue and predicate Venue 50 (K152758) have similar clinical indications for use. Proposed Venue has Transrectal indication which has been cleared on predicate LOGIQe (K151028).
- The Venue and predicate LOGIQ e (K151028) have similar imaging modes. Venue has additional Combined Modes (B/Color M, B/CWD, B/Color/CWD).
- The Venue and predicate LOGIQ e (K151028) systems transducers are similar however there are new applications added to the 3Sc-RS and C1-5-RS probes. The Venue includes the L12n-RS transducer which was cleared on LOGIQ P9 and LOGIQ P7 (K163596). Two new applications are being added to this probe.
- New automated features are being added to assist the user workflow.
- The embedded operating system used is Windows 10, which is latest version introduced by Microsoft.
- The Venue and predicates Venue 50 (K152758) and LOGIQ e (K151028) have similar capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
- The system is manufactured with materials which have ● been evaluated and found to be safe for the intended use of the device.
- The systems have acoustic power levels which are below the applicable FDA limits.
- o The Venue and predicate Venue 50 (K152758) have been designed in compliance with approved electrical and physical safety standards.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a blue circle. The letters are stylized and appear to be handwritten or cursive. The blue circle has a white outline, and the letters 'GE' are also white.
### Summary of Non-Clinical Tests:
Venue has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to comply with applicable medical device safety standards. The Venue complies with voluntary standards:
- o AAMI/ANSI ES60601-1, Medical Electrical Equipment -Part 1: General Requirements for Safety, 2005/ A2:2012
- IEC60601-1-2, Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests, 2014
- IEC60601-2-37, Medical Electrical Equipment - Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment, 2007
- NEMA UD 3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment, 2004
- ISO10993-1. Biological Evaluation of Medical Devices-. Part 1: Evaluation and Testing- Third Edition, 2009
- NEMA UD 2, Acoustic Output Measurement Standard for ● Diagnostic Ultrasound Equipment, 2004
- ISO14971, Application of risk management to medical ● devices, 2007
- NEMA, Digital Imaging and Communications in Medicine (DICOM) Set. (Radiology), 2011
The following quality assurance measures are applied to the development of the system:
- Risk Analysis
- Requirements Reviews ●
- Design Reviews
- Testing on unit level (Module verification)
- Integration testing (System verification) 0
- Performance testing (Verification) o
- . Safety testing (Verification)
Transducer material and other patient contact materials are biocompatible.
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a circular border. The logo is colored in a light blue hue, and the border has decorative swirls around it.
Summary of Clinical Tests:
The subject of this premarket submission, Venue, did not require clinical studies to support substantial equivalence.
Conclusion: GE Healthcare considers the Venue to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.