K170700 · Merit Medical Systems, Inc. · DQX · Apr 6, 2017 · Cardiovascular
Device Facts
Record ID
K170700
Device Name
Inqwire Diagnostic Guide Wire
Applicant
Merit Medical Systems, Inc.
Product Code
DQX · Cardiovascular
Decision Date
Apr 6, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
Merit Medical guide wires are used to facilitate the placement of devices during diagnostic and interventional procedures.
Device Story
InQwire Amplatz Super Stiff Guide Wire facilitates device placement during diagnostic and interventional procedures. Device consists of blue PTFE-coated flat wire coil with internal core wire for stiffness. Core wire welded to coil at distal end, proximal end, and one intermediate point. Available in 0.035-inch and 0.038-inch diameters, lengths 75cm to 260cm, with various straight and J-tip configurations. Provided in spiral hoop dispenser with flush port luer adapter and J-straightener. Used by clinicians in clinical settings to guide catheters or other interventional devices through vasculature or body lumens. Enhances procedural efficiency by providing trackability and support for device delivery.
Clinical Evidence
Bench testing only. No clinical data. Performance testing conducted per risk analysis and industry standards including ISO 11070:2014, ISO 11135:2014, ASTM F1980-07, and ISO 10993-1:2009. Evaluated attributes: size, radiopacity, tensile strength, torque, tip flexibility, coating integrity, lubricity, catheter/needle compatibility, and corrosion resistance. All results met predetermined acceptance criteria.
Technological Characteristics
Materials: Blue PTFE-coated flat wire coil, internal core wire. Dimensions: 0.035"/0.038" OD, 75-260cm length. Configuration: Straight/J-tip. Sterilization: Ethylene oxide (ISO 11135:2014). Packaging: ISO 11607-1:2006. Biocompatibility: ISO 10993-1:2009. Mechanical: Welded core-to-coil assembly.
Indications for Use
Indicated for patients undergoing diagnostic and interventional procedures requiring device placement.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
April 6, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Merit Medical Systems, Inc. Mr. Michael O'Sullivan Senior Regulatory Affairs Specialist Parkmore Business Park West Galway, Ireland
Re: K170700
Trade/Device Name: InOwire Amplatz Super Stiff Guide Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: March 6, 2017 Received: March 7, 2017
Dear Mr. O'Sullivan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Fernando Aguel
Fernando Aguel
-S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170700
#### Device Name
InQwire Amplatz Super Stiff Guide Wire
Indications for Use (Describe)
Merit Medical guide wires are used to facilitate the placement of devices during diagnostic and interventional procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(+353) 91 703700 (3061)<br>(+353) 91 680 104<br>Mark Mullaney<br>1721504 |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Correspondent Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date of Preparation:<br>Registration Number: | Merit Medical Ireland Ltd.<br>Parkmore Business Park West<br>Galway, Ireland<br>(+353) 91 703700 (3223)<br>(+353) 91 680 104<br>Michael O'Sullivan<br>5/Apr//2017<br>9616662 |
| Subject<br>Device | Trade Name:<br>Common/Usual Name:<br>Classification Name: | InQwire® Amplatz<br>Merit Medical Guide Wire<br>21 CFR 870.1330 Catheter guide wire |
| Predicate<br>Device | Premarket Notification Predicate #1: (Primary Predicate)<br>Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer:<br><br>Premarket Notification Reference Device<br>Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | InQwire®<br>21 CFR 870.1330 Catheter guide wire<br>K163575<br>Merit Medical Systems, Inc.<br><br>Amplatz Super Stiff Guidewire<br>21 CFR 870.1330 Catheter guide wire<br>K843012<br>Boston Scientific Corporation |
| Classification | Class II<br>21 CFR 870.1330 Catheter guide wire<br>FDA Product Code: DQX<br>Review Panel: Division of Cardiovascular Devices | |
| Intended Use | Merit Medical guide wires are used to facilitate the placement of devices during<br>diagnostic and interventional procedures. | |
| Device<br>Description | Merit InQwire® Amplatz Super Stiff Guide Wires are intended to facilitate the<br>placement of devices during diagnostic and interventional procedures. The guide<br>wires consist of a Blue PTFE (Polytetrafluoroethylene) coated flat wire coil, with<br>an inside core wire. The core wire enhances the "Super Stiff" characteristic of<br>the guide wire.<br>The core wire extends the full length of the coil and is welded to the coil at three<br>(3) points; a) the distal end, b) the proximal end and c) a spot weld approx. 22cm<br>from the distal end. These welds are designed to provide integrity and ensure<br>that the guide wire components remain together The outside coil is PTFE coated | |
| The Merit InQwire® Amplatz Super Stiff Guide Wires are offered in 0.035-inch<br>and 0.038-inch outer diameter with a 1cm, 3cm, 3.5cm ,4cm, 6cm and 7cm<br>straight and J tip configuration and are available in lengths from 75cm to 260cm.<br>The wires will be provided in a spiral hoop dispenser sized appropriately for the<br>wires diameter and length.<br>The dispenser has a standard flush port luer adapter that accepts any standard<br>luer lock or slip tip syringe to facilitate flushing of the guide wire prior to use. A J-<br>straightener is provided on the dispenser to facilitate the advancement of the<br>wire tip into other devices. | | |
| Comparison to<br>Predicate | The changes to the device are as follows; a) A new variant of the core wire OD<br>(0.38") has been added to the range, b) 5 new Tip lengths have been added<br>(1cm, 3cm, 3.5cm, 4cm and 6cm), spread across the two ODs and c) A J Tip<br>shape variant is also added to the range, spread across the two ODs. None of<br>the above changes would be deemed sufficient to affect the technological<br>characteristics of the device and thus the subject Merit InQwire® Amplatz Super<br>Stiff Guide Wire is substantially equivalent to Predicate Device #1, the Merit<br>InQwire® Amplatz Super Stiff Guide Wire [K163575 and the Reference Device,<br>the Amplatz Super Stiff Guidewire [K843012].]. | |
| | The proposed new wires add in a new OD of 0.038" as well as Tip length (1cm,<br>3cm, 3.5cm, 4cm and 6cm) and Tip shape (J tip) variants to the range. The<br>lengths, of 75cm to 260cm, are within the bracketed range of the predicate<br>(75cm to 260cm). The Indications for use of the subject wire are identical to<br>Predicate Device #1. | |
| Safety &<br>Performance<br>Tests | No performance standards have been established under section 514 of the<br>Food, Drug and Cosmetic Act for these devices. Performance testing of the<br>subject Merit InQwire® Amplatz Super Stiff Guide Wire was conducted based on<br>risk analysis. A battery of testing was conducted in accordance with protocols<br>based on requirements outlined in guidance's and industry standards and these<br>were shown to meet the acceptance criteria that were determined to<br>demonstrate substantial equivalence. | |
| | Where appropriate, the tests were based on the requirements of the following<br>documents:<br>FDA guidance Coronary and Cerebrovascular Guide Wire Guidance<br>•<br>January 1995.<br>ISO 11070:2014, Sterile Single-Use Intravascular Catheter Introducers.<br>•<br>ISO 11135:2014 Sterilization of health care products-Ethylene oxide-<br>•<br>Part 1: Requirements for the development, validation and routine control<br>of a sterilization process for medical devices.<br>ASTM F1980-07 Standard Guide for Accelerated Aging of Sterile Barrier<br>•<br>Systems for Medical Devices<br>ISO 10993-1:2009, Biological Evaluation of Medical Devices Part 1:<br>•<br>Evaluation and Testing within a risk management process, and the FDA<br>Modified ISO 10993 Test Profile FDA Memo G95-1.<br>ISO 11607-1:2006, Packaging for terminally sterilized medical devices -<br>•<br>Part 1: Requirements for materials, sterile barrier systems and<br>packaging systems | |
| | The Merit InQwire® Amplatz Super Stiff Guide Wire was compared to the<br>predicate device for various performance attributes that support substantial<br>equivalence of the device. The difference in assembly between the modified<br>device and the cleared device, Merit InQwire® Amplatz Super Stiff Guide Wire<br>[K163575] has raised no new issues. In some instances, performance | |
## Section 5 510(k) Summary
{4}------------------------------------------------
and this Blue PTFE coating extends from the distal tip of the wire to within 7cm of the proximal tip, which remains uncoated.
characteristic testing was based on the Reference Device, the Amplatz Super
{5}------------------------------------------------
Stiff Guidewire [K843012], when it was deemed to be relevant.
| | Size Designation | Radiopacity | Tensile | Torque | | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|-----------------|-----------|--|--|
| | | | Strength | Strength | | |
| | Tip Flexibility | Coating | Catheter<br>and | Lubricity | | |
| | | Adherence/Integrity | Needle | | | |
| | | Fracture<br>● | Compatibility | | | |
| | | Flex<br>● | | | | |
| | | Particulate<br>● | | | | |
| | | Evaluation | | | | |
| | Surface | Biocompatibility | Corrosion | Tip Shape | | |
| | All test results were comparable to the predicate devices and the subject Merit<br>InQwire® Amplatz Super Stiff Guide Wire met the predeterminded acceptance<br>criteria. This has demonstrated that the subject device is substantially equivalent<br>to the predicate devices. | | | | | |
| | Based on the Indications for Use, design, safety and performance testing, the | | | | | |
| Summary of<br>Substantial | subject Merit InQwire® Amplatz Super Stiff Guide Wire is substantially | | | | | |
| Equivalence | equivalent to the predicate devices, the cleared Predicate Device #1, the Merit<br>InQwire® Amplatz Super Stiff Guide Wire [K163575] and the Reference Device, | | | | | |
| | the Amplatz Super Stiff Guidewire [K843012] | | | | | |
The following is a list of all testing that was successfully completed:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.