K170690 · Medacta International S.A. · LZO · Nov 29, 2017 · Orthopedic
Device Facts
Record ID
K170690
Device Name
M-Vizion Femoral Revision System
Applicant
Medacta International S.A.
Product Code
LZO · Orthopedic
Decision Date
Nov 29, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The hip prosthesis M-Vizion is designed for cementless use in total or partial hip arthroplasty in primary or revision surgery. Hip Replacement is indicated in the following cases: - Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia. - Avascular necrosis of the femoral head. - Acute traumatic fracture of the femoral head or neck. - Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.
Device Story
Modular cementless femoral stem system for hip arthroplasty; consists of proximal body, distal stem, and locking screw. Components assembled via conical coupling and secured with locking screw. Proximal body features titanium alloy with TiGrowth-C (Mectagrip) coating; distal stem features titanium alloy with shard fins for rotational stability; locking screw features titanium alloy with TiNbN coating. Used by orthopedic surgeons in clinical settings for primary or revision hip replacement. Device provides structural support for femoral component of hip joint; restores joint function and stability; alleviates pain associated with degenerative or traumatic hip conditions.
Clinical Evidence
No clinical data. Bench testing only: range of motion (EN ISO 21535), shaft fatigue (ISO 7206-4), neck fatigue (ISO 7206-6), pull-off force (ASTM F2009-00), and coating characterization. Bacterial endotoxin testing met limits <20 EU/device.
Technological Characteristics
Modular cementless femoral stem. Materials: Titanium alloy. Coatings: TiGrowth-C (proximal), TiNbN (locking screw). Standards: ISO 7206-4 (shaft fatigue), ISO 7206-6 (neck fatigue), ASTM F2009-00 (pull-off force), EN ISO 21535 (ROM). Conical coupling assembly. Sterile.
Indications for Use
Indicated for patients requiring total or partial hip arthroplasty (primary or revision) due to severe joint pain/disability from arthritis, traumatic arthritis, rheumatoid polyarthritis, congenital hip dysplasia, avascular necrosis, acute traumatic fracture of femoral head/neck, or failure of previous hip surgeries.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
November 29, 2017
Medacta International SA % Ms. Elizabeth Rose Manager, Regulatory Affairs Mapi USA, Inc. 2343 Alexandria Drive, Suite 100 Lexington, Kentucky 40504
Re: K170690
Trade/Device Name: M-Vizion Femoral Revision System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, KWY Dated: October 30, 2017 Received: October 31, 2017
Dear Ms. Rose:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170690
### Device Name
M-Vizion Femoral Revision System
## Indications for Use (Describe)
The hip prosthesis M-Vizion is designed for cementless use in total or partial hip arthroplasty in primary or revision surgery.
Hip Replacement is indicated in the following cases:
- · Severely painful and/or disabled joint as a result of arthritis, traumatoid polyarthritis, or congenital hip dysplasia.
- · Avascular necrosis of the femoral head.
- · Acute traumatic fracture of the femoral head or neck.
- · Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div style="display:inline-block;">☒</div> Presentation Use (Part 21 CFR 301 Subject to Reporting) |
|----------------------------------------------------------------------------------------------------|
| <div style="display:inline-block;">☐</div> Own-Firm Use (Part 21 CFR 301 Subject to Reporting) |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 3.0 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory Affairs Manager Date Prepared: March 6, 2017 Date Revised: November 21, 2017
#### II. Device
| Device Proprietary Name: | M-Vizion Femoral Revision System |
|--------------------------|-----------------------------------------------------------------------------------------|
| Common or Usual Name: | Total Hip Prosthesis |
| Classification Name: | Hip joint metal/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis |
| Primary Product Code: | LZO |
| Secondary Product Code: | KWY |
| Regulation Number: | 21 CFR 888.3353, 21 CFR 888.3390 |
| Device Classification | 2 |
#### III. Predicate Device
Substantial equivalence is claimed to the following devices: Primary Predicate:
- . Revision Femoral Stem, K151739, Limacorporate S.p.A.
- Wagner SL Revision Stem Lateral, Wagner Cone Prosthesis® System, K161192, Zimmer o GmbH
Reference Devices
- MasterLoc Stem, K151531, Medacta International SA
- MasterLoc Stem, K160289, Medacta International SA ●
{4}------------------------------------------------
#### IV. Device Description
The purpose of this submission is to gain clearance for the M-Vizion Femoral Revision System. The M-Vizion Femoral Revision System is a modular cementless stem intended to be used for hip arthroplasty in primary or revision surgery. The system is composed of the proximal body, the distal stem, and the locking screw. The proximal body and the distal stem are intended to be assembled together on a conical coupling and tightened by the locking screw.
The proximal body is made of titanium alloy and coated with a titanium coating, TiGrowth -C (Medacta commercial name: Mectagrip). The distal stem is a straight stem made of titanium alloy and the principal feature consists of shard fins that potentially increase the rotation stability. The locking screw is made of titanium alloy and coated with TiNbN.
The M-Vizion Femoral Revision System is similar to competitor predicate devices Limacorporate S.p.A.'s Revision Femoral Stem (K151739), Zimmer GmbH's Wagner SL Revision Stem Lateral, Wagner Cone Prosthesis® System (K161192) and reference device Medacta's MasterLoc Stem (K151531 and K160289).
#### V. Indications for Use
The hip prosthesis M-Vizion is designed for cementless use in total or partial hip arthroplasty in primary or revision surgery.
Hip Replacement is indicated in the following cases:
- Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, ● rheumatoid polyarthritis, or congenital hip dysplasia.
- Avascular necrosis of the femoral head. ●
- Acute traumatic fracture of the femoral head or neck.
- Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.
#### VI. Comparison of Technological Characteristics
The M-Vizion Femoral Revision System and the predicate devices share the following characteristics:
- indications for use:
- materials: ●
- design;
- sterile: ●
- coating: and ●
- device usage.
{5}------------------------------------------------
The M-Vizion Femoral Revision System is technologically different from the predicate devices as follows:
- . sizes:
- lengths: and
- diameters.
#### VII. Performance Data
The following performance tests are being provided in support of a substantial equivalence determination. Testing was conducted to written protocols with acceptance criteria that were based on standards.
Non-Clinical Studies
- 0 Performance Tests
- range of motion (ROM): EN ISO 21535:2009 Non-Active Surgical Implants o Joint Replacement Implants - Specific Requirements For Hip-Joint Replacement Implants;
- shaft fatigue testing: ISO 7206-4 Third Edition 2010-06-15 Implants for surgery o Partial and total hip joint prostheses - Part 4: Determination of endurance properties and performance of stemmed femoral components [Including AMENDMENT 1 (2016)];
- modular conical connection fatigue test, and
- post fatigue fretting corrosion analysis.
- neck fatigue testing: ISO 7206-6 Second edition 2013-11-15 Implants For Surgery o - Partial And Total Hip Joint Prostheses - Part 6: Determination Of Endurance Properties Of Head And Neck Region Of Stemmed Femoral Components;
- pull off force testing: ASTM F2009-00 (Reapproved 2011) Standard Test Method o for Determining the Axial Disassembly Force of Taper Connections of Modular Prostheses; and
- coating characterization tests: o
- characterization report of titanium coating on Ti6A17Nb substrates for ■ orthopedic; and
- characterization report of TiNbN coating on cobalt chrome, titanium alloy, and stainless steel substrates for orthopedic.
- Pyrogenicity ●
- о Bacterial Endotoxin Test (LAL test) was conducted and met the endotoxin limit of <20 EU/device; and
- Medacta has no intentions of labeling the subject devices as non-pyrogenic or o pyrogen free.
Clinical Studies
- No clinical studies were conducted. 0
{6}------------------------------------------------
## VIII. Conclusion
The information provided above supports that the M-Vizion Femoral Revision System is as safe and effective as the predicate devices. Therefore, it is concluded that the M-Vizion Femoral Revision System is substantially equivalent to the predicate devices.
-----------------------------------This space intentionally left blank--------------------------------------------------------------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.