K170664 · Heraeus Medical Components, LLC · DQY · Nov 9, 2017 · Cardiovascular
Device Facts
Record ID
K170664
Device Name
Odyssey Micro Catheter
Applicant
Heraeus Medical Components, LLC
Product Code
DQY · Cardiovascular
Decision Date
Nov 9, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Odyssey Micro Catheter is intended to provide support to facilitate the placement of guidewires in the coronary and peripheral vasculatures and can be used to exchange one guide wire for another. The Odyssey Micro Catheter is also intended to assist in the infusion of contrast media. The micro catheter is not intended to be used in the neurovasculature.
Device Story
Single-lumen percutaneous catheter; used to support guidewire placement/exchange and contrast media infusion in coronary and peripheral vasculatures. Manually inserted by physician through guiding catheter over guidewire; advanced to target region. Construction: PTFE liner, metallic coil, polymer outer layer, radiopaque polymer tip; hydrophilic coating on distal 30-60 cm. Provides mechanical support for navigation; facilitates contrast delivery. Benefits patient by enabling access to complex vascular anatomy. Used in clinical settings (e.g., cath lab).
Clinical Evidence
Bench testing only. No clinical data. Performance demonstrated via in vitro testing per ISO 10555-1:2013. Biocompatibility established per ISO 10993-1:2009. Tests included: cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogenicity, hemolysis, complement activation, in-vivo thrombogenicity, chemical characterization, particulate testing, corrosion resistance, dimensional inspection, package integrity, tensile/torque strength, kink resistance, radiopacity, flow rate, and burst pressure/leakage. All acceptance criteria met.
Technological Characteristics
Single-lumen catheter; materials: stainless steel, Pebax, Pellethane, PTFE, adhesive, SBC, polycarbonate. Features radiopaque polymer tip and hydrophilic coating. Dimensions: 90-225 cm length; 0.032-0.042 in. OD. Sterilization: Ethylene Oxide. Mechanical sensing/actuation; no software/electronics.
Indications for Use
Indicated for support to facilitate guidewire placement and exchange in coronary and peripheral vasculatures, and for contrast media infusion. Not for use in neurovasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 9, 2017
Heraeus Medical Components, LLC Margaret Batchelder Principal Regulatory Specialist 2605 Fernbrook Lane North, Suite J Plymouth, Minnesota 55447
Re: K170664
Trade/Device Name: Odyssey Micro Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: October 11, 2017 Received: October 12, 2017
## Dear Margaret Batchelder:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
## Page 2 - Margaret Batchelder
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K170664
Device Name Odyssey Micro Catheter
Indications for Use (Describe)
The Odyssey Micro Catheter is intended to provide support to facilitate the placement of guidewires in the coronary and peripheral vasculatures and can be used to exchange one guide wire for another. The Odyssey Micro Catheter is also intended to assist in the infusion of contrast media. The micro catheter is not intended to be used in the neurovasculature.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 100%;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size: 100%;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Summary | K170664 |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Heraeus Medical Components, LLC<br>2605 Fernbrook Lane North, Suite J<br>Plymouth, MN 55447 USA |
| Contact Person: | Margaret Batchelder,<br>Principal Regulatory Specialist<br>2605 Fernbrook Lane North, Suite J<br>Plymouth, MN 55447 USA<br>763-225-6675 (voice)<br>763-559-7676 (fax) |
| Date Prepared: | October 25, 2017 |
| Trade Name: | Odyssey Micro Catheter |
| Classification: | Class II<br>Regulation Number: 21 CFR 870.1250<br>Percutaneous Catheter |
| Product Code: | DQY |
| Predicate Device: | The subject device is substantially equivalent to K151103;<br>ASAHI Corsair Microcatheter manufactured by Asahi Intecc<br>Co., LTD. |
| Device Description: | The Odyssey Micro Catheter is a single lumen catheter<br>constructed using a PTFE liner inside of metallic coil and a<br>polymer outer layer. The distal tip is a radiopaque polymer<br>material and the distal 30 cm to 60 cm of the micro catheter has<br>a hydrophilic coating. The key components of the catheter are<br>the core, coil(s), tip, and coatings. |
| Indication for Use: | The Odyssey Micro Catheter is intended to provide support to<br>facilitate the placement of guidewires in the coronary and<br>peripheral vasculatures and can be used to exchange one guide<br>wire for another. The Odyssey Micro Catheter is also intended<br>to assist in the infusion of contrast media. This micro catheter is<br>not intended to be used in the neurovasculature. |
| Contraindications: | None known. |
| Principle of<br>Operation: | The Odyssey Micro Catheter is manually inserted into<br>vasculature through a compatible guiding catheter over a<br>compatible guidewire and advance with the guidewire to the<br>target region. |
| Comparison of<br>Technological<br>Characteristics: | The key technological and performance similarities examined<br>between the approved devices and the proposed device are as<br>follows: |
| | Indications for use - The Indications for use for the proposed<br>device is a subset of indications for use of the predicate device<br>as proposed device, |
| | Fundamental scientific technology, including design are<br>equivalent to the predicate devices |
| | Operating principle - equivalent to the predicate devices |
| | Packaging materials - equivalent to the predicate device |
| | Sterility assurance level and method of sterilization -<br>equivalent to the predicate devices |
| | The length and diameter of the device are similar to the<br>dimensions of the predicate devices. |
| | The proposed device and that of the predicate device are<br>equivalent in that they are constructed with equivalent materials<br>to provide equivalent performance characteristics and coating<br>properties. |
| Substantial<br>Equivalence: | The Heraeus Odyssey Micro Catheter is substantially equivalent<br>to the ASAHI Corsair Microcatheter (K151103). Substantial<br>equivalence, which is summarized in the following table, is<br>based on indications for use, physical and technological<br>characteristics, and comparative device testing. |
{4}------------------------------------------------
| | Odyssey Micro Catheter | Predicate:<br>Asahi Corsair Microcatheter<br>(K151103) |
|--------------------------------|------------------------|--------------------------------------------------------|
| Device<br>Common/Usual<br>Name | Catheter, Percutaneous | Catheter, Percutaneous |
| Device Class | Class II | Class II |
| Product Code /<br>Regulation | DQY / 21 CFR 870.1250 | DQY / 21 CFR 870.1250 |
| Regulation Name | Percutaneous catheter | Percutaneous catheter |
{5}------------------------------------------------
| | Odyssey Micro Catheter | Predicate:<br>Asahi Corsair Microcatheter<br>(K151103) |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Odyssey Micro Catheter is<br>intended to provide support to<br>facilitate the placement of<br>guidewires in the coronary and<br>peripheral vasculatures and can be<br>used to exchange one guide wire<br>for another. The Odyssey Micro<br>Catheter is also intended to assist<br>in the infusion of contrast media.<br>This micro catheter is not intended<br>to be used in the neurovasculature. | The ASAHI Corsair Microcatheter<br>is intended to provide support to<br>facilitate the placement of Guide<br>wires in the coronary and<br>peripheral vasculatures and can be<br>used to exchange one guide wire<br>for another. The Corsair<br>Microcatheter is also intended to<br>assist in the delivery of contrast<br>media into the coronary,<br>peripheral, and abdominal<br>vasculatures. |
| Catheter OD | 0.032 in. / 0.042 in. | 0.037 in. |
| Catheter Length | 90 - 225 cm | 90 - 150 cm |
| Catheter Materials | Stainless Steel, Pebax, Pellethane,<br>PTFE, Adhesive, SBC,<br>Polycarbonate | Stainless Steel, Tungsten, Polymer<br>Jacket, PTFE, Adhesive, SBC,<br>Polycarbonate |
| Catheter Coating | Hydrophilic coating | Hydrophilic coating |
| Sterile Device | Yes | Yes |
| Sterilization Type | Ethylene Oxide | Ethylene Oxide |
| Disposable /<br>Reusable | Disposable | Disposable |
Performance In vitro bench tests were utilized to demonstrate equivalence with reference to ISO 10555-1:2013, Intravascular catheters - sterile and single-use catheters -Testing: Part 1: General requirements.
> The performance testing assessment supports that the biocompatibility, shelf life, and functional specifications of the proposed micro catheter device were met.
> The Odyssey Micro Catheter device test data supports the claims of substantial equivalence to the predicate devices. Biological Safety of the predicate device has been established through biocompatibility testing carried out in compliance with ISO10993-1:2009 and G95-1, FDA General Program Memorandum: Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing.
The following bench tests were conducted or evaluated to support the proposed device:
{6}------------------------------------------------
- Biocompatibility testing: .
- Cytotoxicity o
- Sensitization o
- Intracutaneous Reactivity O
- Acute Systemic Toxicity o
- Pyrogenicity Testing o
- Hemolysis o
- Complement Activation o
- In-Vivo Thrombogenicity o
- o Chemical Characterization
- Particulate testing
- Corrosion resistance .
- Dimensional inspection ●
- Sterile package integrity testing ●
- Tensile strength
- Torque strength ●
- Kink resistance ●
- Radiopacity .
- Flow rate ●
- Burst pressure/freedom from leakage .
The Odyssey Micro Catheter met all predetermined acceptance criteria and compared favorably with the predicate device.
- Heraeus considers the Odyssey Micro Catheter to be equivalent to the Conclusion: predicate device. This conclusion is based upon the fact that device has an equivalent intended use, and there are no differences that raise new types of questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.