K170655 · Mynosys Cellular Devices, Inc. · PUL · Jun 2, 2017 · Ophthalmic
Device Facts
Record ID
K170655
Device Name
Zepto
Applicant
Mynosys Cellular Devices, Inc.
Product Code
PUL · Ophthalmic
Decision Date
Jun 2, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Zepto™ is a surgical instrument to be used by ophthalmic surgeons for performing anterior capsulotomy during cataract surgery. The device includes durable hardware for multi-use, a disposable handpiece and fluid isolator assembly for single-patient use . The Zepto™ is a prescription (Rx) device to be used by, or on the order of a licensed physician.
Device Story
Zepto is an electrosurgical system for anterior capsulotomy during cataract surgery. System components: console, disposable handpiece, fluid isolator assembly. Input: pulsed DC energy from console storage capacitor; suction pressure to engage silicone suction cup with capsule. Operation: surgeon places suction cup on anterior capsule; suction creates tensile stress; device delivers pulsed DC energy to circular element to create 5mm diameter capsulotomy. Used in clinical/surgical settings by ophthalmic surgeons. Output: circular anterior capsulotomy. Benefits: standardized capsulotomy diameter, reduced procedure time compared to manual techniques, improved edge strength/extensibility. Healthcare providers use visual confirmation via surgical microscope to ensure proper apposition before energy delivery.
Clinical Evidence
Clinical study of 100 eyes. Primary safety endpoint: no posterior capsule rupture or vitreous loss. Primary effectiveness: 98/100 successful 360-degree capsulotomies; 2 cases required manual completion. Mean capsulotomy diameter 5.14 mm (SD 0.14). 97/100 cases reported cortex removal ease similar/easier than manual CCC. Adverse events included anterior capsule tears (2/100), epithelial erosion, macular edema, and IOP elevation; only one case (AC/PC tear) deemed device-related. Mean 1-month BCVA 20/20. Bench testing included paired human cadaver eye study showing superior edge extensibility compared to CCC and femtosecond laser.
Technological Characteristics
Electrosurgical system; 110 VAC 50-60 Hz power; Type BF electrical classification. Disposable handpiece with silicone suction cup and circular metallic cutting element. Biocompatible materials (ISO 10993-5, -10, -11). Sterilization: EtO. Connectivity: standalone console. Software: microprocessor-controlled firmware with button-based activation. Dimensions: 13.5" x 10" x 6.5" console.
Indications for Use
Indicated for use in performing anterior capsulotomy during cataract surgery in patients undergoing cataract extraction.
Regulatory Classification
Identification
A radiofrequency electrosurgical cautery apparatus is an AC-powered or battery-powered device intended for use during ocular surgery to coagulate tissue or arrest bleeding by a high frequency electric current.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird-like shape.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 2, 2017
Mynosys Cellular Devices, Inc. Dan Marinsik VP, Quality, Clinical & Regulatory Affairs 46710 Fremont Blvd. Fremont, CA 94538-6538
Re: K170655
Trade/Device Name: Zepto Regulation Number: 21 CFR 886.4100 Regulation Name: Radiofrequency Electrosurgical Cautery Apparatus Regulatory Class: Class II Product Code: PUL Dated: April 25, 2017 Received: April 28, 2017
Dear Dan Marinsik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation
(21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170655
Device Name Zepto™M
Indications for Use (Describe)
Zepto™ is indicated for use in performing anterior capsulotomy during cataract surgery.
Type of Use (Select one or both, as applicable)
|--|--|
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT | Mynosys Cellular Devices, Inc.<br>46710 Fremont Blvd.<br>Fremont, CA. 94538 |
|--------------------------|-----------------------------------------------------------------------------|
| Contact | Dan Marinsik |
| Phone Number | 510-857-6296 |
| FAX Number | 510-689-2349 |
| Date Prepared | 5-26-17 |
| TRADE NAME: | ZEPTO™ |
| COMMON NAME: | Capsulotomy Device |
| DEVICE | Device Class - II |
| CLASSIFICATION:<br>NAME: | Radiofrequency electrosurgical cautery apparatus |
| PRODUCT CODE | NCR |
| REGULATION<br>NUMBER: | 21 CFR § 886.4100 |
| PREDICATE DEVICE | The Fugo Blade<br>(K001498) |
## SUBSTANTIALLY EQUIVALENT TO:
| The Fugo Blade | (K001498) | |
|----------------|-----------|--|
| | | |
| | | |
| | | |
{4}------------------------------------------------
## Description of the Device Subject to Premarket Notification:
The Zepto™ is an electrosurgical device indicated for use in performing anterior capsulotomies during cataract surgery. The Zepto™ Capsulotomy System consists of a console, disposable handpiece, and a fluid isolator assembly (consisting of the fluid isolator, roller dispenser, and clamp). The handpiece power cord connector is connected to the front panel of the console to provide power for the capsulotomy procedure. The medical grade PVC suction tubing from the handpiece is connected to the fluid isolator. The fluid isolator is then connected to the console's front panel to provide suction during the treatment to a silicone suction cup containing the capsulotomy element. The Zepto 110 console delivers pulsed DC energy from a storage capacitor to a circular capsulotomy element on the tip of the disposable handpiece for creation of a nominal 5mm diameter capsulotomy.
### Statement of Intended Use
The intended use for Zepto™ is as a surgical instrument to be used by ophthalmic surgeons for performing anterior capsulotomy during cataract surgery. The device includes durable hardware for multi-use, a disposable handpiece and fluid isolator assembly for single-patient use . The Zepto™ is a prescription (Rx) device to be used by, or on the order of a licensed physician.
### Indication for Use
Zepto™ is indicated for use in performing anterior capsulotomy during cataract surgery.
### Technological Characteristics Comparison
The Zepto™ shares many technological characteristics including indications and technological features to those of the predicate device, The Fugo Blade. The Zepto "" also utilizes the same mechanism of action to perform the capsulotomy and is therefore substantially equivalent to the legally marketed predicate device. Differences between the Zepto and the predicate device are well-defined and well-characterized and do not raise any new concerns of safety. Key characteristics are provided in the table below.
{5}------------------------------------------------
# 510(K) SUMMARY
| Attributes | Zepto™<br>(K170655) | The Fugo Blade<br>(K001498) | Comment |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | INTENDED USE | | |
| Manufacturer | Mynosys Cellular<br>Devices | Medisurg Ltd. | |
| Clearance Date | - | 08/10/2000 | |
| Model Number | 12684, 12430, 12686 | M100 | |
| FDA Regulation | 21 CFR 886.4100 | 21 CFR 886.4100 | Same |
| Regulation Name | Radiofrequency<br>electrosurgical<br>cautery apparatus | Radiofrequency<br>electrosurgical<br>cautery apparatus | Same |
| FDA Classification<br>Product Code | NCR | NCR | Same |
| FDA<br>Medical Device<br>Classification | Class - II | Class - II | Same |
| Single Use<br>Disposable<br>Component(s) | Zepto™ Handpiece<br>and fluid isolator<br>assembly are single<br>use disposables | The Fugo Blade tip<br>is a single use<br>disposable | Same |
| Prescription<br>(Rx)/Over-the-<br>Counter (OTC) | Rx | Rx | Same |
| | TECHNOLOGICAL CHARACTERISTICS | | |
| Energy Type | Rectified RF Pulsed-<br>DC | RF | Fugo Blade uses full<br>wave RF, Zepto™<br>uses rectified RF |
| Induction of capsule<br>tensile stress | Yes, by suction<br>pressure | Yes, by manipulation<br>of probe tip against<br>capsule tissue to be<br>cut | SE, induced by<br>different mechanical<br>means |
| Electrical | 110 VAC 50-60 Hz | 4 size "C"<br>rechargable batteries<br>provide DC current<br>for the Fugo Blade<br>Battery Charger:<br>100-240 VAC, 50/60<br>Hz, 1.0A | Same, both<br>electrically powered |
| Electrical<br>Classification | Type BF | Type BF | Same |
| Attributes | Zepto™<br>(K170655) | The Fugo Blade<br>(K001498) | Comment |
| Control | Firmware and button<br>control | Firmware, footswitch<br>and variable power<br>and intensity control | Both are<br>microprocessor<br>controlled. Fugo<br>Blade has power and<br>intensity controls on<br>the console based on<br>the individual<br>Physician's speed in<br>performing<br>Capsulotomy.<br>Zepto™ has a factory<br>preset power output<br>that is slightly lower<br>due to the cutting<br>element being circular<br>in design. |
| System Components | Handpiece, control<br>console, fluid isolator<br>assembly<br>(see Fig. 12.1) | Handpiece, control<br>console footswitch<br>(see Fig. 12.1) | Equivalent |
| Dimensions (Control<br>Box) | $13.5" X 10" X 6.5"$ | $10.00" X 7.05" X6.12"$ | Both consoles are<br>portable |
| Emergency Off | Yes | Yes | Same |
| Foot Switch<br>Controlled | No | Yes | The Fugo Blade<br>utilizes a footswitch to<br>provide power to the<br>treatment tip. The<br>power of the Zepto is<br>delivered via the<br>Cut/Release button |
| Procedure Halt Using<br>Foot Switch | No | Yes | The Fugo Blade<br>utilizes a footswitch,<br>Zepto utilizes an<br>emergency Stop<br>Button on console |
| On Switch | Yes | Yes | Same |
| Display Panel | Yes | Yes | Same |
| Mains Switch | Yes | Yes | Same |
| Disposable Cutting<br>Element | Yes | Yes | Same |
| Attributes | Zepto™<br>(K170655) | The Fugo Blade<br>(K001498) | Comment |
| Cutting Element<br>Shape | Circular | Straight | SE - Fugo Blade<br>straight cutting<br>element requires 7-15<br>seconds to perform<br>capsulotomy, Zepto™<br>circular cutting<br>element completes<br>capsulotomy in 4<br>msec |
| Capsulotomy Size | 5.0 mm (nominal) | Variable Dependent<br>on Surgeon | SE - Fugo Blade<br>capsulotomy varies in<br>diameter due to<br>manual circular<br>manipulation of the<br>straight cutting tip.<br>Zepto™ has uniform<br>capsulotomy diameter<br>based on a circular<br>cutting element |
| MANUFACTURING | | | |
| Biocompatibility | The Zepto™<br>Handpiece passed<br>testing performed<br>according to ISO<br>10993-5<br>(Cytotoxicity), 10993-<br>10 (Acute Systemic<br>Toxicity) and 10993-<br>11(Sensitization and<br>Irritation). | There is no mention<br>in the company's<br>literature, website or<br>510(k) summary as<br>to whether the Fugo<br>Blade was tested for<br>biocompatibility. | The Zepto™ is made<br>from Biocompatible<br>materials, The Fugo<br>Blade is presumed to<br>be the same. |
| Sterilization | EtO | Heat Autoclave or<br>EtO | Same - Both use<br>validated sterile<br>methods presumed to<br>provide SAL of 1 X 10-6 |
| Packaging | Zepto™ Handpiece:<br>sterile single<br>pouched Tyvek® bag | Fugo Blade tip:<br>single pouched<br>Tyvek® bag in box of<br>12 tips | Same |
| Shelf Life | 12 months | Unknown | The Zepto™ has a<br>validated shelf life.<br>The Fugo Blade I<br>presumed to be the<br>same. |
{6}------------------------------------------------
# 510(K) SUMMARY
{7}------------------------------------------------
# 510(K) SUMMARY
{8}------------------------------------------------
# Brief Summary of Nonclinical Tests and Results
Mynosys has developed the Zepto™ for use in anterior capsulotomy. Nonclinical testing included pre-clinical animal studies and bench testing.
# Summary of Pre-Clinical Testing
A GLP study was conducted that demonstrated the effectiveness of the device in 20 out of 20 rabbit eyes. This study also verified that there was negligible temperature change in the anterior chamber during Zepto "" use. In addition, there were no differences in the endothelial cell condition and ocular histopathology of eyes or the overall condition of eyes that received a Zepto"10 capsulotomy compared to fellow eyes that received a CCC. The study also confirmed that there was no observable zonular stress during a Zepto "10 capsulotomy.
# Summary of Bench Testing
A paired human cadaver eves study compared the strength and extensibility of the Zepto "" capsulotomy edge with that of femtosecond laser and CCC. (The predicate device was not available for this study.) This study demonstrated overall general safety of the Zepto™. The biomechanical testing of capsulotomy edge properties resulting from different methods of capsulotomy in paired human cadaver eves provides direct quantitative assessment of edge strength and extensibility. Pairwise comparison of Zepto 100 capsulotomies in one eve and either CCC or femtosecond laser capsulotomy in the fellow eye of the same donor provide the most scientifically rigorous evaluation of their relative performance and rule out confounding donor related factors.
The results of the Capsule Edge Strength Study (COCESS) using paired cadaver eyes show that the extensibility of the Zepto "" capsule edges as quantified by stretch ratios were significantly greater than that of capsule edges produced by CCC or by femtosecond laser. The data show that Zepto™ capsulotomies should also exhibit greater extensibility when compared to the predicate device.
# Microcalorimetry Validation Test Report
Microcalorimetry testing demonstrated that the Zepto™ is equivalent to the predicate device in energy used during performance of capsulotomies in cataract surgery.
# Brief Summary of Clinical Tests and Results
A clinical trial of the Zepto™ was conducted to evaluate its performance in the creation of anterior capsulotomies during cataract surgery. This clinical study was undertaken to address specific questions of safety and effectiveness related to demonstrating substantial
{9}------------------------------------------------
equivalence.
Primary Safety Endpoint: Safety of the ZeptoTM was evaluated by comparing the observed event rate against the historically derived rate for the "gold standard" surgical technique of continuous curvilinear capsulorrhexis (CCC). No posterior capsule rupture and vitreous loss in 100 eyes treated with Zepto.
Primary Effectiveness Endpoint: Capsulotomy effectiveness was determined by evaluating the completeness of the Zepto capsulotomy. The effectiveness analysis revealed 2 cases of incomplete capsulotomy out of 100 cases performed. Successful Zepto 360 degree capsulotomy in 98 of 100 subjects treated with Zepto. Two eyes required manual capsulotomy, both with good visual outcome and IOL capsular fixation, and without adverse outcomes:
- 1 case was attributed to user error (simultaneous energy applied and suction release):
- . 1 case with small tissue bridge observed after Zepto procedure was completed manually, attributed to suboptimal surgical microscope visualization, leading to application of Zepto energy before complete apposition of Zepto capsulotomy ring with the capsule was observed.
# Additional Safety Parameters:
- · AC Tear: 2 of 100 eyes treated with Zepto had anterior capsule tear, neither with vitreous loss:
- o 1 case noted at end of surgery with secondary PC tear extending during exchange of damaged IOL;
- o 1 case attributed to cataract chopping technique, did not extend to posterior capsule.
- . Corneal Touch: No cases reported.
- . Adverse Events: evithelial erosion and macular edema, anterior capsule tear. anterior and posterior capsule tear without vitreous loss, IOP elevation, and ache in treated eye. Only 1 case (ac/pc tear) was noted to be device-related.
# Additional Effectiveness Parameters:
- Diameter and Circularity of Capsulotomy: Mean anterior capsule diameter ● 5.14 mm +/- S.D. 0.14 mm (range 4.9-5.5 mm). 99 cases recorded as circular without zonular damage.
- . Pre- and Post Zepto Corneal Incision Size: Mean increase in incision size post-Zepto treatment 0.0305 mm (range 0-0.2 mm).
- . Ease of cortex removal: 97/100 cases reported as similar or easier ease of cortex aspiration as compared to manual capsulorhexis.
- Capsulotomy centration: 96/100 reported as centered.
- IOL centration: 100% reported with IOL intracapsular fixation and centered.
{10}------------------------------------------------
# Clinical Findings:
- Mean BCVA at 1 month was 20/20 (Snellen equivalent to ETDRS assessment).
- . No capsular abnormalities reported.
- . No clinically significant slit lamp exam findings reported.
## Diameter of Anterior Capsulotomy:
- Mean diameter of ZeptoTM capsulotomies = 5.14 mm +/- S.D. 0.14 mm (median = 5.2 mm; range 4.9-5.5 mm).
# Minimum Incision Size
The minimum incision size required to accommodate the Zepto Handpiece tip is 2.2 mm. The minimum incision size used during the clinical study was 2.4 mm.
# Conclusion of Nonclinical and Clinical Tests and Results
In conclusion, the data from this study, combined with the rigorous lab, preclinical testing and OUS clinical experience demonstrate that the Mynosys Zepto™ Capsulotomy System is as safe and effective as the legally marketed predicate with which it is compared for regulatory purposes and is therefore substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.