← Product Code [KCT](/productcode/KCT) · K170640

# Endoscope Sterilization Tray (K170640)

_Intuitive Surgical, Inc. · KCT · Sep 17, 2017 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170640

## Device Facts

- **Applicant:** Intuitive Surgical, Inc.
- **Product Code:** [KCT](/productcode/KCT.md)
- **Decision Date:** Sep 17, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6850
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Intuitive Surgical Endoscope Sterilization Tray is intended for use to encase and protect da Vinci Xi endoscopes (Model #'s 470026 and 470027) for sterilization in any of the following sterilization machines/cycles: - STERRAD 100NX sterilization system using the Express cycle - STERRAD 100S sterilization system using the Standard cycle - Steris V-PRO maX using the Non Lumen, Flexible, or Lumen cycles - Steris V-PRO 1 Plus using the Non Lumen or Lumen cycles - Steris V-PRO 1 using the V-PRO/Lumen cycle The sterilization cycle parameters of the sterilizers are preset by the manufacturers and are not adjustable. The maximum product load per tray is one da Vinci Xi Endoscope. The length of the da Vinci Xi Endoscope is approximately 600 mm and the diameter of the shaft is 8.8 mm. The maximum weight of the tray and endoscope is 8.9 lbs. The Endoscope Sterilization Tray is not intended to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared STERRAD and Steris compatible sterilization wrap in order to maintain sterility of the enclosed medical instrument.

## Device Story

Thermoformed plastic tray with silicone inserts and clear lid; designed to encase and protect da Vinci Xi endoscopes during H2O2 chemical sterilization. Tray and lid feature perforations to allow sterilization gas penetration. Used in clinical settings to prepare endoscopes for sterilization in specific STERRAD and Steris systems. Device does not maintain sterility independently; must be used with validated, FDA-cleared sterilization wrap. Protects instruments during handling and processing; ensures compatibility with preset sterilization cycles.

## Clinical Evidence

Bench testing only. Cleaning validation and human factors validation were performed. Design validation studies were not repeated due to identical design, materials, and manufacturing processes compared to the predicate.

## Technological Characteristics

Thermoformed plastic tray with silicone inserts and clear lid. Features perforations for H2O2 gas sterilization. Reusable. Identical to predicate device.

## Regulatory Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

## Predicate Devices

- Endoscope Sterilization Tray ([K151450](/device/K151450.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 17, 2017

Intuitive Surgical, Inc. Kunal Gunjal Regulatory Affairs Specialist 1266 Kifer Road Building 101 Sunnyvale, California 94086

Re: K170640

Trade/Device Name: Endoscope Sterilization Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: August 24, 2017 Received: August 25, 2017

Dear Kunal Gunjal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply 

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Tara A. Ryan -S 2017.09.17 18:52:36 -04'00'

for

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K170640

Device Name Endoscope Sterilization Tray

Indications for Use (Describe)

The Intuitive Surgical Endoscope Sterilization Tray is intended for use to encase and protect da Vinci Xi endoscopes (Model #'s 470026 and 470027) for sterilization in any of the following sterilization machines/cycles:

- · STERRAD 100NX sterilization system using the Express cycle
- · STERRAD 100S sterilization system using the Standard cycle
- · Steris V-PRO maX using the Non Lumen, Flexible, or Lumen cycles
- · Steris V-PRO 1 Plus using the Non Lumen or Lumen cycles
- · Steris V-PRO 1 using the V-PRO/Lumen cycle

The sterilization cycle parameters of the sterilizers are preset by the manufacturers and are not adjustable. The maximum product load per tray is one da Vinci Xi Endoscope. The length of the da Vinci Xi Endoscope is approximately 600 mm and the diameter of the shaft is 8.8 mm. The maximum weight of the tray and endoscope is 8.9 lbs. The Endoscope Sterilization Tray is not intended to maintain sterility; it is intended to be used in conjunction with a legally marketed, validated, FDA-cleared STERRAD and Steris compatible sterilization wrap in order to maintain sterility of the enclosed medical instrument.

| Type of Use (Select one or both, as applicable)            |                                                           |  |  |  |
|------------------------------------------------------------|-----------------------------------------------------------|--|--|--|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |  |  |  |
| CONTINUE ON A SEPARATE PAGE IF NEEDED                      |                                                           |  |  |  |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary (K170640)

## IS4000 Endoscope Sterilization Tray

| 510(k)<br>Owner                          | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086                                                                                              |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact                                  | Kunal Gunjal<br>Regulatory Affairs Specialist, Regulatory Affairs<br>Phone Number: 408-523-8017<br>Fax Number: 408-523-8907<br>Email: Kunal.Gunjal@intusurg.com |
| Date<br>Summary<br>Prepared              | September 13, 2017                                                                                                                                              |
| Trade Name                               | Endoscope Sterilization Tray                                                                                                                                    |
| Common<br>Name                           | Sterilization Tray                                                                                                                                              |
| Classification<br>and<br>Regulation      | Class II,<br>21 CFR 880.6850                                                                                                                                    |
| Product<br>Codes                         | KCT                                                                                                                                                             |
| Classification<br>Advisory<br>Committee: | General Hospital                                                                                                                                                |
| Predicate<br>Device                      | K151450                                                                                                                                                         |

{4}------------------------------------------------

## Device Description

## IS4000 Endoscope Sterilization Tray

| Trade<br>Name         | Endoscope Sterilization Tray                                                                                                                                                                                                                                                                                                                                  |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Description | The <b>Endoscope Sterilization Tray</b> is intended for use to encase and protect <i>da Vinci Xi</i><br>Endoscopes during the sterilization process. It is a thermoformed plastic tray with<br>silicone inserts and a clear lid. The tray and the lid contain perforations to allow<br>sterilization gases to penetrate the tray and sterilize the endoscope. |

## Technological Comparison Table (Endoscope Sterilization Tray)

| Characteristic                   | Subject Device<br>Endoscope Sterilization Tray<br>K170640                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Predicate Device<br>Endoscope Sterilization Tray<br>K151450 |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| Manufacturer                     | Intuitive Surgical, Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Intuitive Surgical, Inc.                                    |
| Trade Name                       | Endoscope Sterilization Tray                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | SAME as subject device                                      |
| Common Name                      | Sterilization Tray                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | SAME as subject device                                      |
| Regulation No.                   | 21 CFR 880.6850                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | SAME as subject device                                      |
| Product Code                     | KCT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | SAME as subject device                                      |
| Device Class/<br>Regulation Name | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | SAME as subject device                                      |
|                                  | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Predicate Device                                            |
| Characteristic                   | Endoscope Sterilization Tray<br>K170640                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Endoscope Sterilization Tray<br>K151450                     |
| Indications for Use              | The Intuitive Surgical Endoscope Sterilization<br>Tray is intended for use to encase and protect<br>da Vinci Xi endoscopes (Model #'s 470026<br>and 470027) for sterilization in any of the<br>following sterilization machines/cycles:<br>STERRAD 100NX sterilization system using the Express cycle STERRAD 100S sterilization system using the Standard cycle Steris V-PRO maX using the Non Lumen, Flexible, or Lumen cycles Steris V-PRO 1 Plus using the Non Lumen or Lumen cycles Steris V-PRO 1 using the V-PRO/Lumen cycle  The sterilization cycle parameters of the<br>sterilizers are preset by the manufacturers and<br>are not adjustable. The maximum product<br>load per tray is one da Vinci Xi Endoscope.<br>The length of the da Vinci Xi Endoscope is<br> | SAME as subject device                                      |
| Sterilization Method             | H2O2 chemical sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | SAME as subject device                                      |
| Reusable                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | SAME as subject device                                      |
| Reprocessing<br>Instructions     | The following changes have been made to the Reprocessing instructions for the Endoscope<br>Sterilization Tray, since the most recently cleared legally marketed device (K151450):  Re-organization of content and formatting Remove or add warning and caution statements Revise warning and caution statements Reference to specific washer/disinfector models was removed (Removed the automated cleaning process) Addition of parameter based thermal disinfection process                                                                                                                                                                                                                                                                                                 |                                                             |

{5}------------------------------------------------

{6}------------------------------------------------

## Technological Characteristics:

The technological characteristics of the subject device are identical to the predicate device.

#### Performance Data:

Performance test data demonstrates that the subject device is substantially equivalent to the predicate device and that the design output meets the design input requirements. The testing conducted consisted of Cleaning Validation and Human Factors Validation. Design validation studies were not repeated as the subject device design, materials, and manufacturing processes are identical to the predicate device.

#### Summary:

Based on the intended use, indications for use, technological characteristics, performance data, and nonclinical tests performed, the subject device is substantially equivalent and is as safe, as effective, and performs as well as the legally marketed predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K170640](https://fda-staging.innolitics.com/device/K170640)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).