ACE Hemoglobin A1c (HbA1c) Reagent

K170623 · Alfa Wassermann Diagnostic Technologies, LLC · LCP · Feb 27, 2018 · Hematology

Device Facts

Record IDK170623
Device NameACE Hemoglobin A1c (HbA1c) Reagent
ApplicantAlfa Wassermann Diagnostic Technologies, LLC
Product CodeLCP · Hematology
Decision DateFeb 27, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7470
Device ClassClass 2
AttributesPediatric

Indications for Use

ACE Hemoglobin A1c (HbA1c) Reagent is intended for the quantitative determination of percent hemoglobin A1c in venous whole blood collected in K2-EDTA tubes, using the ACE Alera® and ACE Axcel® Clinical Chemistry Systems. This test is intended for use in clinical laboratories and physician office laboratories to monitor long term blood glucose control in individuals with diabetes mellitus. For in vitro diagnostic use only.

Device Story

The ACE Hemoglobin A1c (HbA1c) Reagent is an in vitro diagnostic assay for use on ACE Alera and ACE Axcel Clinical Chemistry Systems. It requires off-line pretreatment where whole blood samples are lysed and hemoglobin chains hydrolyzed. The assay uses latex agglutination inhibition: HbA1c in the sample competes with a synthetic polymer for antibody-coated microparticle binding sites; agglutination inhibition is inversely proportional to HbA1c concentration, measured monochromatically at 692 nm. Total hemoglobin is measured bichromatically (573/692 nm) after conversion to alkaline hematin. The system calculates the ratio of HbA1c to total hemoglobin to report percent HbA1c. Used in clinical and physician office laboratories by trained personnel to monitor long-term glycemic control in diabetic patients. Results assist clinicians in assessing patient glucose management; however, values cannot be used individually for diagnostic purposes. The device is sensitive to certain hemoglobinopathies (HbF, HbC, HbS), which are noted in labeling.

Clinical Evidence

Bench testing only. Precision studies (internal and external at POL sites) performed per CLSI EP05-A3. Linearity evaluated per CLSI EP06-A (range 2.7-13.0% HbA1c). Interference studies (endogenous substances, hemoglobin variants, labile A1c, carbamylated/acetylated Hb) performed per CLSI EP07-A2. Method comparison study against predicate (N=101-102 internal; N=50-52 per site external) showed high correlation (r > 0.98).

Technological Characteristics

Reagents: Porcine Pepsin, Sodium Hydroxide, HbA1c Hapten Polymer, mouse HbA1c antibody-coupled particles, BSA, surfactants. Principle: Latex agglutination inhibition; monochromatic (692 nm) and bichromatic (573/692 nm) absorbance. Connectivity: Integrated with ACE Alera/Axcel Clinical Chemistry Systems. Standardization: NGSP certified. Software: Delta data acquisition algorithm.

Indications for Use

Indicated for the quantitative determination of percent hemoglobin A1c in venous whole blood (K2-EDTA) to monitor long-term blood glucose control in individuals with diabetes mellitus. For use in clinical and physician office laboratories.

Regulatory Classification

Identification

A glycosylated hemoglobin assay is a device used to measure the glycosylated hemoglobins (A1a , A1b , and A1c ) in a patient's blood by a column chromatographic procedure. Measurement of glycosylated hemoglobin is used to assess the level of control of a patient's diabetes and to determine the proper insulin dosage for a patient. Elevated levels of glycosylated hemoglobin indicate uncontrolled diabetes in a patient.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k170623 B. Purpose for Submission: Modification of previously cleared systems with a new data acquisition method C. Measurand: Whole Blood Glycosylated Hemoglobin (HbA1c) D. Type of Test: Quantitative, Antibody-coated latex agglutination inhibition E. Applicant: Alfa Wassermann Diagnostic Technologies, LLC. F. Proprietary and Established Names: ACE Hemoglobin A1c (HbA1c) Reagent G. Regulatory Information: | Classification Name | Regulation Section | Device Class | Product Code | Panel | | --- | --- | --- | --- | --- | | Glycosylated Hemoglobin Assay | 21 CFR 864.7470 | II | LCP | Hematology (81) | H. Intended Use: 1. Intended use(s): See Indication(s) for use below. {1} 2. Indication(s) for use: ACE Hemoglobin A1c (HbA1c) Reagent is intended for the quantitative determination of percent hemoglobin A1c in venous whole blood collected in K2-EDTA tubes using the ACE Alera and ACE Axcel Clinical Chemistry Systems. This test is intended for use in clinical laboratories and physician office laboratories to monitor long term blood glucose control in individuals with diabetes mellitus. For in vitro diagnostic use only. 3. Special conditions for use statement(s): For prescription use only. For in vitro diagnostic use only. This test is not intended for screening or diagnosis of diabetes mellitus. This test is not intended for use in neonates. This test is not useful in judging day-to-day glucose control and should not be used to replace daily home testing of urine and blood glucose. Diseases and other conditions that affect red blood cell survival (e.g., hemolytic anemia or other hemolytic diseases, pregnancy, significant or chronic blood loss), will cause a decrease in results for % Hemoglobin A1c and are not reliable. Hemoglobinopathies may interfere with glycated hemoglobin analysis. The results from both ACE Alera and ACE Axcel Clinical Chemistry Systems show that there is no significant interference for Hemoglobin D (≤ 36.3%) or Hemoglobin E (≤ 22.5%). High Hemoglobin F (&gt; 10.1%) will result in lower than expected HbA1c values. High Hemoglobin C (&gt; 14.0%) and high Hemoglobin S (&gt; 17.1%) will result in higher than expected HbA1c values. 4. Special instrument requirements: ACE Alera and ACE Axcel Clinical Chemistry Systems I. Device Description: The ACE Hemoglobin A1c (HbA1c) Reagent kit contains the following: - 2 x 25 mL Hemoglobin Denaturant: Porcine Pepsin, Buffer and Preservative - 2 x 12 mL Total Hemoglobin Reagent: Sodium Hydroxide and Surfactant - 1 x 11 mL Hemoglobin A1c Agglutinator Reagent: HbA1c Hapten Polymer, Bovine Serum Albumin, Buffer, Preservative and Surfactant - 1 x 11 mL Hemoglobin A1c Antibody Reagent: HbA1c Antibody (mouse) Coupled Particles, Bovine Serum Albumin, Buffer, Preservative and Surfactant {2} J. Substantial Equivalence Information: 1. Predicate device name(s): Hemoglobin A1c Reagent on DCA2000+ System 2. Predicate 510(k) number(s): k951361 3. Comparison with predicate: | Similarities/Differences Assay | | | | --- | --- | --- | | Item | New Device ACE Hemoglobin A1c (HbA1c) Reagent (k170623) | Predicate Hemoglobin A1c Reagent on DCA2000+ System (k951361) | | Intended Use | Quantitative determination of percent hemoglobin A1c in venous whole blood. This test is intended to monitor long term blood glucose control in individuals with diabetes mellitus. | Same | | Methodology | Antibody-coated latex agglutination inhibition | Same | | Instrument Platforms | ACE Alera and ACE Axcel Clinical Chemistry Systems | DCA 2000+ Analyzer | | Method Traceability or Standardization | National Glycohemoglobin Standardization Program (NGSP) | Same | | Calibration | 6 points for HbA1c, 1 point for total hemoglobin | Calibration card | | Sample Type | Venous whole blood (K2-EDTA) | Venous whole blood (EDTA, heparin, citrate and fluoride/oxalate), capillary whole blood | | Sample Pretreatment | Manual sample pretreatment with hemoglobin denaturant | Automatic sample pretreatment | | Testing Environment | Clinical laboratories and physician office laboratories | Same | | Measuring range | 2.7% - 13.0% HbA1c | 2.5% - 14.0% HbA1c | {3} K. Standard/Guidance Document Referenced (if applicable): CLSI EP05-A3, Evaluation of Precision Performance of Quantitative Measurement Methods, Approved Guideline, Third Edition CLSI EP06-A, Evaluation of the Linearity of Quantitative Analytical Measurement Procedure: A Statistical Approach, Approved Guideline CLSI EP07-A2, Interference Testing in Clinical Chemistry, Approved Guideline, Second Edition CLSI: EP09-A3, Measurement Procedure Comparison and Bias Estimation Using Patient Samples, Approved Guideline, Third Edition CLSI EP15-A3, User Verification of Performance for Precision and Estimation of Bias; Approved Guideline, Third Edition CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures, Approved Guideline, Second Edition L. Test Principle: The ACE Hemoglobin A1c (HbA1c) Reagent assay requires a pretreatment step of the whole blood samples, which is done off-line. The red blood cells in the sample are lysed by the Hemoglobin Denaturant and the hemoglobin chains are hydrolyzed. For determination of HbA1c, a latex agglutination inhibition assay is used. HbA1c in the sample inhibits agglutination of the polymer agglutinator and the antibody-coated microparticles in the HbA1c Antibody Reagent. The increase in absorbance, monitored monochromatically at 692 nm, is inversely proportional to the HbA1c present in the sample. The HbA1c concentration is calculated using delta data acquisition method, instead of the previously cleared quadratic method, for the same ACE HbA1c Reagent on the ACE Alera and ACE Axcel Clinical Chemistry Systems. All hemoglobin derivatives in the sample are converted to alkaline hematin in order to determine total hemoglobin. The reaction produces a green colored solution, which is measured bichromatically at 573 nm/692 nm. The intensity of color produced is directly proportional to the total hemoglobin concentration in the sample. The concentrations of both HbA1c and total hemoglobin are measured, the ratio is calculated and the result reported as percent HbA1c. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: {4} # Internal Precision Study Precision studies were performed according to CLSI EP05-A3 guideline. The study included four whole blood samples prepared from pooled $\mathrm{K}_2$ -EDTA venous specimens. Samples were analyzed in duplicate per run, two runs per day for 20 days on both ACE Alera and ACE Axcel Clinical Chemistry Systems. Results are shown below: | ACE Alera Clinical Chemistry Systems | | | | | | | --- | --- | --- | --- | --- | --- | | Sample | Mean (N=80) | Within-Run | | Total | | | | | SD | %CV | SD | %CV | | 1 | 5.1 | 0.07 | 1.4 | 0.13 | 2.5 | | 2 | 6.8 | 0.14 | 2.1 | 0.18 | 2.6 | | 3 | 8.2 | 0.10 | 1.3 | 0.20 | 2.4 | | 4 | 11.2 | 0.32 | 2.8 | 0.34 | 3.1 | | ACE Axcel Clinical Chemistry Systems | | | | | | | Sample | Mean (N=80) | Within-Run | | Total | | | | | SD | %CV | SD | %CV | | 1 | 5.0 | 0.08 | 1.6 | 0.12 | 2.4 | | 2 | 6.8 | 0.14 | 2.0 | 0.17 | 2.6 | | 3 | 8.2 | 0.13 | 1.6 | 0.21 | 2.6 | | 4 | 11.1 | 0.27 | 2.4 | 0.32 | 2.9 | # External Precision Study An external precision study using four $\mathrm{K}_2$ -EDTA venous whole blood samples was performed at three Physician's Office Laboratories for ACE Alera and ACE Axcel Clinical Chemistry Systems using one lot of reagent, respectively. Samples were tested in triplicates per day for 5 days on one instrument by one operator at each site. Results from each site are shown below: | ACE Alera Clinical Chemistry Systems | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | Lab # | Sample | Mean (N=15) | Within-Run | | Total | | | | | | SD | %CV | SD | %CV | | 1 | 1 | 5.3 | 0.09 | 1.6 | 0.09 | 1.6 | | 2 | | 5.2 | 0.11 | 2.1 | 0.12 | 2.4 | | 3 | | 4.6 | 0.08 | 1.8 | 0.13 | 2.9 | | 1 | 2 | 6.8 | 0.12 | 1.7 | 0.12 | 1.7 | | 2 | | 6.7 | 0.08 | 1.2 | 0.14 | 2.1 | | 3 | | 6.0 | 0.07 | 1.1 | 0.14 | 2.3 | | 1 | 3 | 8.8 | 0.24 | 2.7 | 0.27 | 3.1 | | 2 | | 8.5 | 0.11 | 1.3 | 0.18 | 2.1 | | 3 | | 7.9 | 0.07 | 0.9 | 0.25 | 3.1 | | 1 | 4 | 12.5 | 0.29 | 2.4 | 0.32 | 2.5 | | 2 | | 12.5 | 0.36 | 2.9 | 0.36 | 2.9 | | 3 | | 11.5 | 0.13 | 1.1 | 0.27 | 2.3 | {5} | ACE Axcel Clinical Chemistry Systems | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | Lab # | Sample | Mean (N=15) | Within-Run | | Total | | | | | | SD | %CV | SD | %CV | | 1 | 1 | 5.4 | 0.12 | 2.2 | 0.24 | 4.3 | | 2 | | 4.6 | 0.09 | 2.0 | 0.11 | 2.3 | | 3 | | 4.8 | 0.10 | 2.1 | 0.10 | 2.2 | | 1 | 2 | 6.8 | 0.08 | 1.1 | 0.13 | 1.9 | | 2 | | 6.1 | 0.13 | 2.2 | 0.16 | 2.7 | | 3 | | 6.3 | 0.11 | 1.7 | 0.12 | 1.8 | | 1 | 3 | 8.8 | 0.12 | 1.3 | 0.20 | 2.2 | | 2 | | 8.8 | 0.16 | 1.8 | 0.24 | 2.8 | | 3 | | 9.1 | 0.15 | 1.7 | 0.27 | 2.9 | | 1 | 4 | 11.9 | 0.19 | 1.6 | 0.29 | 2.5 | | 2 | | 11.8 | 0.34 | 2.9 | 0.42 | 3.5 | | 3 | | 12.2 | 0.15 | 1.2 | 0.35 | 2.9 | # b. Linearity/assay reportable range: Linearity was evaluated according to CLSI EP06-A. Specific dilutions of a high sample were prepared to make a total of 11 samples with concentrations covering the assay range (2.7, 3.7, 4.8, 5.8, 6.8, 7.8, 8.9, 9.9, 10.9, 11.9 and $13.0\%$ ). All samples were analyzed in triplicates on one ACE Alera and one ACE Axcel Clinical Chemistry Systems. The mean of these replicates was compared to the expected values and the results of linear regressions are summarized below: ACE Alera Clinical Chemistry System: $y = 0.987x + 0.3$ , $r^2 = 0.995$ ACE Axcel Clinical Chemistry System: $y = 0.954x + 0.3$ , $r^2 = 0.994$ The study supports the sponsor's claimed linearity range of 2.7 - $13.0\%$ HbA1c on both ACE Alera and ACE Axcel Clinical Chemistry Systems. Acceptable total hemoglobin/hematocrit ranges for the HbA1c measurement was determined following CLSI EP06-A. Specific dilutions of a high hemoglobin sample were prepared to make a total of 11 samples. All samples were analyzed in four replicates on one ACE Alera and one ACE Axcel Clinical Chemistry Systems. The mean of these replicates was compared to the expected values and the results of linear regressions are summarized below: ACE Alera Clinical Chemistry System: $y = 1.006x + 0.10$ , $r^2 = 0.998$ ACE Axcel Clinical Chemistry System: $y = 0.997x + 0.20$ , $r^2 = 0.996$ The study supports the sponsor's claimed reportable range of 1.4 to $21.8\mathrm{g / dL}$ for total hemoglobin on both ACE Alera and ACE Axcel Clinical Chemistry Systems. {6} c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: The ACE Hemoglobin A1c (HbA1c) Reagent is certified by the National Glycohemoglobin Standardization Program (NGSP). The NGSP certification expires in one year. See NGSP website for current certification at http://www.ngsp.org. d. Detection limit: A Limit of Quantitation (LoQ) study was performed according to the CLSI EP17-A2 guideline. Five whole blood samples with low analyte concentrations were tested over 5 days with 8 replicates per day using one lot of the ACE Hemoglobin A1c (HbA1c) Reagent on both ACE Alera and ACE Axcel Clinical Chemistry Systems. LoQ was defined as the lowest concentration at which $\% \mathrm{CV}$ was $&lt; 20\%$ and was determined to be $2.5\%$. e. Analytical specificity: Endogenous substances: Interference studies were performed according to recommendations in the CLSI EP07-A2 guideline. Various concentrations of potential endogenous interferents were spiked into two levels of whole blood samples (4.9-5.5% HbA1c and 6.9-7.8% HbA1c). All samples were tested in triplicate on both ACE Alera and ACE Axcel Clinical Chemistry Systems. The sponsor defined non-significant interference as $\leq 10\%$ bias in recovery for spiked samples when compared to control samples. The highest endogenous substance concentrations tested that demonstrated no interference are shown below: | Substance | Highest Concentration tested at which no interference was observed | | --- | --- | | Bilirubin | 53 mg/dL | | Triglycerides | 1100 mg/dL | | Ascorbic Acid | 6 mg/dL | | Acetaldehyde | 100 mg/dL | | Sodium Fluoride | 1200 mg/dL | Hemoglobin variant interference: A hemoglobin variant study was performed using 15 K2-EDTA venous whole blood samples (4.7 – 12.0% HbA1c) containing known levels of hemoglobin variants C, D, E, F and S. Three to five samples per hemoglobin variant were tested in four replicates on both ACE Alera and ACE Axcel Clinical Chemistry Systems for $\% \mathrm{HbA1c}$ results. Percent recovery was calculated by comparing to the results obtained on comparator methods free of interference with the respective hemoglobin variant (Trinity Biotech Premier Hb9210, k112015, for HbC, HbD, HbS, Primus {7} Ultra 2 Boronate Affinity HPLC, k891235, for HbE and Tosoh G8, k071132 for HbF). Non-significant interference was defined as recovery within $100 \pm 10\%$. The testing results indicate that there is no significant interference from Hemoglobin D ($\leq 36.3\%$) or Hemoglobin E ($\leq 22.5\%$). Hemoglobin F ($&gt;10.1\%$) will result in lower than expected HbA1c results, while Hemoglobin C ($&gt;14.0\%$) and Hemoglobin S ($&gt;17.1\%$) will result in higher than expected HbA1c results. The labeling contains the following limitation statements for Hb variant interference: “Hemoglobinopathies may interfere with glycated hemoglobin analysis. The results from both ACE Alera and ACE Axcel Clinical Chemistry Systems show that there is no significant interference for Hemoglobin D ($\leq 36.3\%$) or Hemoglobin E ($\leq 22.5\%$). High Hemoglobin F ($&gt;10.1\%$) will result in lower than expected HbA1c values. High Hemoglobin C ($&gt;14.0\%$) and high Hemoglobin S ($&gt;17.1\%$) will result in higher than expected HbA1c values.” ## Labile A1c interference: Two HbA1c levels of whole blood samples were spiked with glucose to contain 360, 720, 1080 and $1440\mathrm{mg/dL}$ glucose and then incubated for four hours at $37^{\circ}\mathrm{C}$ to facilitate the formation of labile A1c. The samples were tested in four replicates on both ACE Alera and ACE Axcel Clinical Chemistry Systems. The sponsor defined non-significant interference as within $10\%$ difference between samples containing glucose and the control sample. The results support the sponsor’s claim that labile A1c showed no significant interference at $1440\mathrm{mg/dL}$ glucose level. ## Carbamylated hemoglobin interference: Two HbA1c levels of whole blood samples were spiked with Sodium Cyanate to contain 500, 1000, 1500 and $2000\mathrm{mg/dL}$ of sodium cyanate. The samples were tested in four replicates on both ACE Alera and ACE Axcel Clinical Chemistry Systems. The sponsor defined non-significant interference as within $10\%$ difference between samples containing Sodium Cyanate and the control sample. The results support the sponsor’s claim that Carbamylated Hb showed no significant interference at $2000\mathrm{mg/dL}$ Sodium Cyanate level. ## Acetylated hemoglobin interference: Two HbA1c levels of whole blood samples were spiked with Acetylsalicylic Acid to contain 500, 1000, 1500 and $2000\mathrm{mg/dL}$ of Acetylsalicylic Acid. The samples were tested in four replicates on both ACE Alera and ACE Axcel Clinical Chemistry Systems. The sponsor defined non-significant interference as within $10\%$ difference between samples containing Acetylsalicylic Acid and the control sample. The results support the sponsor’s claim that acetylated hemoglobin showed no significant interference at $2000\mathrm{mg/dL}$ Acetylsalicylic Acid level. ## HbA0 and HbA1a+b interference: Two HbA1c levels of whole blood samples were spiked with $\mathrm{HbA1a + b}$ to contain 25, 8 {8} 50, 75 and 100 mg/dL of HbA1a+b. The samples were tested in four replicates on both ACE Alera and ACE Axcel Clinical Chemistry Systems. The sponsor defined non-significant interference as within 10% difference between samples containing HbA1a+b and the control sample. The results support the sponsor's claim that HbA1a+b fraction showed no significant interference at 100 mg/dL. Two HbA1c levels of whole blood samples were spiked with HbA0 to contain 575, 1150, 1725 and 2300 mg/dL of HbA0. The samples were tested in four replicates on both ACE Alera and ACE Axcel Clinical Chemistry Systems. The sponsor defined non-significant interference as within 10% difference between samples containing HbA0 and the control sample. The results support the sponsor's claim that HbA0 fraction showed no significant interference at 1725 mg/dL. f. Assay cut-off: Not applicable. 2. Comparison studies: a. Method comparison with predicate device: A method comparison study was conducted internally by testing 101 and 102 K₂-EDTA venous whole blood patient samples in singlicate on one ACE Alera and one ACE Axcel Clinical Chemistry System, respectively, against the predicate device. Results of the linear regression are shown below: | Device | ACE Alera vs. DCA 2000+ | ACE Axcel vs. DCA 2000+ | | --- | --- | --- | | N | 101 | 102 | | Range (x-method) | 3.2% - 12.8% HbA1c | 2.5% - 12.8% HbA1c | | Regression Equation | y=0.979x+0.05 | y=0.983x-0.03 | | Correlation Coefficient | 0.9839 | 0.9832 | | Confidence Interval – slope | 0.944 to 1.015 | 0.948 to 1.019 | | Confidence Interval - intercept | -0.21 to 0.31 | -0.29 to 0.24 | External method comparison studies were conducted at 3 Physician Office Laboratory (POL) sites for ACE Alera Clinical Chemistry System and 3 other POL sites for ACE Axcel Clinical Chemistry System. A total of 50-52 K₂-EDTA venous whole blood samples with ≤ 10% altered samples were tested on ACE Alera or ACE Axcel Clinical Chemistry System in singleton at each POL site with one lot of the ACE HbA1c reagent and compared to the predicate method. Deming regression analysis results are shown below: {9} 10 ACE Alera | Site # | N | Range (x-method) %HbA1c | Regression Equation | Correlation Coefficient | Confidence Interval – slope | Confidence Interval - intercept | | --- | --- | --- | --- | --- | --- | --- | | 1 | 50 | 3.2 – 13.4 | y=0.967x+0.34 | 0.9892 | 0.925 to 1.008 | 0.04 to 0.65 | | 2 | 50 | 3.1 – 12.9 | y=0.984x-0.02 | 0.9945 | 0.955 to 1.014 | -0.23 to 0.18 | | 3 | 50 | 3.2 – 12.4 | y=0.981x-0.09 | 0.9939 | 0.950 to 1.013 | -0.30 to 0.12 | ACE Axcel | Site # | N | Range (x-method) %HbA1c | Regression Equation | Correlation Coefficient | Confidence Interval – slope | Confidence Interval - intercept | | --- | --- | --- | --- | --- | --- | --- | | 4 | 52 | 3.2 – 12.9 | y=1.000x-0.28 | 0.9885 | 0.957 to 1.043 | -0.59 to 0.03 | | 5 | 50 | 3.4 – 12.8 | y=0.993x-0.12 | 0.9932 | 0.960 to 1.027 | -0.37 to 0.13 | | 6 | 50 | 3.5 – 12.9 | y=0.980x+0.02 | 0.9960 | 0.955 to 1.006 | -0.16 to 0.20 | b. Matrix comparison: Not applicable. The ACE Hemoglobin A1c (HbA1c) Reagent is for use with $\mathrm{K}_2$-EDTA venous whole blood samples only. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: Not applicable. {10} 5. Expected values/Reference range: | Glycemic recommendation for many non-pregnant adults with diabetes: | | | --- | --- | | A1c | < 7.0%* | *More or less stringent glycemic goals may be appropriate for individual patients. Goals should be individualized based on duration of diabetes, age/life expectancy, comorbid conditions, known CVD or advanced microvascular complications, hypoglycemia unawareness, and individual patient considerations. Source: American Diabetes Association. Glycemic Targets: Standards of Medical Care in Diabetes - 2018. Diabetes Care 2018 Jan; 41 (Supplement 1): S55-S64. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...