The van Andel Dilatation Catheter is intended to facilitate percutaneous introduction of guide wires, catheters, and other devices into the femoral, popliteal, and infrapopliteal arteries during diagnostic and interventional procedures.
Device Story
Intravascular dilatation catheter; facilitates percutaneous introduction of guide wires, catheters, and other devices into femoral, popliteal, and infrapopliteal arteries. Used during diagnostic and interventional procedures. Operated by clinicians in clinical settings. Device consists of PTFE catheter; provides pathway for interventional tools. Benefits patient by enabling minimally invasive access to peripheral vasculature. Output is physical access/dilatation of vessel entry site; clinician uses fluoroscopic imaging to guide placement. Single-use, sterile device.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-1), tensile strength (BS EN ISO 10555-1), liquid leakage, dimensional verification, surface inspection, radiopacity, kink radius, and simulated use. All tests met predetermined acceptance criteria.
Technological Characteristics
Material: Polytetrafluoroethylene (PTFE). Dimensions: 5.0, 6.3, 7.0, 8.0, 9.0 French diameters; 60, 80, 100 cm lengths. Compatibility: 0.035 or 0.038 inch wire guides. Standards: ANSI AAMI ISO 10993-1:2009(R)2013 (biocompatibility), BS EN ISO 10555-1:2013 (tensile/leakage). Sterilization: Sterile, single-use.
Indications for Use
Indicated for patients requiring percutaneous introduction of guide wires, catheters, or other devices into femoral, popliteal, or infrapopliteal arteries during diagnostic or interventional procedures.
Regulatory Classification
Identification
A vessel dilator for percutaneous catheterization is a device which is placed over the guide wire to enlarge the opening in the vessel, and which is then removed before sliding the catheter over the guide wire.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, stacked on top of each other.
April 28, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Cook Incorporated Daniel Corbin Regulatory Affairs Specialist 750 Daniels Wav Bloomington, IN 47404
Re: K170616
Trade/Device Name: Van Andel Dilatation Catheter Regulation Number: 21 CFR 870.1310 Regulation Name: Vessel Dilator For Percutaneous Catheterization Regulatory Class: Class II Product Code: DRE Dated: February 28, 2017 Received: March 1, 2017
Dear Mr. Daniel Corbin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170616
Device Name van Andel Dilatation Catheter
Indications for Use (Describe)
The van Andel Dilatation Catheter is intended to facilitate percutaneous introduction of guide wires, and other devices into the femoral, popliteal, and infrapopliteal arteries during diagnostic and interventional procedures.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------|------------------------------------------------------|
| <div> <span> <input type="checkbox"/> </span> </div> | <div> <span> <input type="checkbox"/> </span> </div> |
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, block letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, block letters on a red background. The logo is simple and modern.
## 510(k) SUMMARY
# van Andel Dilatation Catheter Traditional 510(k) Summary 21 CFR §807.92
### Submitter Information
| Applicant: | Cook Incorporated | |
|-----------------------|---------------------------------------------------|--|
| Address: | 750 Daniels Way | |
| | Bloomington, IN 47404 | |
| Contact: | Daniel J. Corbin | |
| Email: | RegSubmissions@cookmedical.com | |
| Contact Phone Number: | 812-335-3575 ext. 104018 | |
| Contact Fax Number: | 812-332-0281 | |
| Date Prepared: | April 12, 2017 | |
| Device Information | | |
| Trade Name: | van Andel Dilatation Catheter | |
| Common Name: | Dilator, Vessel, For Percutaneous Catheterization | |
| Classification Name: | Vessel dilator for percutaneous catheterization | |
| | DRE (21 CFR §870.1310) | |
# Predicate Device
The van Andel Dilatation Catheter is substantially equivalent to the Lawmax Dilator, cleared under K123842, and the Fortress Introducer Sheath System, cleared under K153197.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is in white, sans-serif, uppercase letters. Below "COOK" is the word "MEDICAL" in white, sans-serif, uppercase letters.
COOK INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A.
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
#### Comparison to Predicate Device
It has been demonstrated that the van Andel Dilatation Catheter and the predicate devices are substantially equivalent in terms of intended use, duration of use, principles of operation, fundamental technological characteristics, and insertion method. The design, dimensions, indications for use, duration of use, and materials of the subject device is similar to the attributes of the predicate devices. The predicate devices are available in 4.0 - 24.0 French diameters, 26 - 100 centimeter lengths, and are compatible with a 0.035 or 0.038 inch wire guide; the subject device is available in 5.0, 6.3, 7.0, 8.0, and 9.0 French diameters, 60 - 100 centimeter lengths, and is compatible with a 0.035 or 0.038 inch wire guide. Therefore, the specifications of the subject device fall within the range of specifications of the predicate device in terms of French size, catheter length, and wire guide compatibility. This, combined with the provided testing, supports a determination of substantial equivalence between the predicate and subject devices.
### Device Description
The van Andel Dilatation Catheter are intravascular dilation catheters comprised of polytetrafluoroethylene, designed for use in diagnostic and interventional procedures. The van Andel Dilatation Catheters are provided sterile and are meant for single use only. The van Andel Dilatation Catheter is available in 5.0, 6.3, 7.0, 8.0, and 9.0 French diameters and is manufactured in lengths of 60, 80, and 100 centimeters.
#### Intended Use
The van Andel Dilatation Catheter is intended to facilitate percutaneous introduction of guide wires, catheters, and other devices into the femoral, popliteal, and infrapopliteal arteries during diagnostic and interventional procedures.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Cook Medical logo. The logo consists of the word "COOK" in white, block letters on a red background. Below the word "COOK" is the word "MEDICAL" in smaller, white block letters. The red background is a rectangle shape.
COOK INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A.
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
# Test Data
The van Andel Dilatation Catheter, subject of this submission, was subjected to applicable testing to assure reliable design and performance under the testing parameters. The following tests were conducted to ensure reliable design and performance:
- . Biocompatibility testing - Testing (i.e., cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, pyrogenicity, hemocompatibility, complement activation, and partial thromboplastin time) demonstrated that the device is biocompatible. In conformance with the applicable sections of ANSI AAMI ISO 10993-1:2009(R)2013, the predetermined acceptance criteria were met.
- Tensile Testing of the Hub to Shaft Bond Testing verified that under proper ● clinical use of the catheter, the hub-to-shaft connection is expected to withstand peak load values specified in BS EN ISO 10555-1:2013 – Section 4.6. The predetermined acceptance criterion was met.
- Liquid Leakage Testing Testing in accordance with BS EN ISO 10555-1:2013 -. Section 4.7.1.verified that under proper clinical use of the catheter, there shall be no liquid leakage. The predetermined acceptance criterion was met.
- Dimensional Verification Testing Testing verified that the outer diameter, inner . diameter, taper length and overall length of the device are within a specified tolerance. The predetermined acceptance criteria were met.
- Exterior Surface Condition Testing and Distal Tip Inspection – Testing verified that exterior surface of the device was smooth and free from any process defects. The predetermined acceptance criteria were met.
- . Radiopacity Testing – Testing verified that the device would be visible under fluoroscopic imaging in simulated clinical conditions. The predetermined acceptance criterion was met.
- Kink Radius Testing – Testing verified that the device would not kink under clinical condition in which the test is expected to be used. The predetermined acceptance criteria were met.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in large, white, sans-serif font, stacked on top of the word "MEDICAL" in a smaller, white, sans-serif font. Both words are set against a solid red background. The "COOK" has a registered trademark symbol after it.
- Simulated Use Testing verified the performance parameters were acceptable for . clinical use. The predetermined acceptance criteria were met.
In conclusion, the results of these tests support a determination of substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.