← Product Code [HWC](/productcode/HWC) · K170571

# Tyber Medical BioTy(TM) Nanotopography Trauma Screw, TyFix Screw (K170571)

_Tyber Medical, LLC · HWC · Jun 20, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170571

## Device Facts

- **Applicant:** Tyber Medical, LLC
- **Product Code:** [HWC](/productcode/HWC.md)
- **Decision Date:** Jun 20, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

A Tyber Medical BioTy™ Nanotopography Trauma Screw is designed to apply fixation between two adjacent segments of cortical and/or cancellous bone.

## Device Story

Bone fixation fastener for compression and fixation of cortical and/or cancellous bone. Features BioTy™ nanotopography (microscopic roughened surface with nano-scaled features) to potentially enhance bone-implant interface. Provided in solid or cannulated forms; manufactured from titanium and stainless steel. Used by surgeons in orthopedic procedures. Device provides mechanical fixation to stabilize bone segments during healing.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing, including pull-out testing (ASTM F543), static torsion testing (ASTM F543), and pyrogenicity testing (ST72:2011).

## Technological Characteristics

Metallic bone fixation fastener (titanium and stainless steel). Features nanotopography surface treatment. Provided sterile. Mechanical testing per ASTM F543. No software or electronic components.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Tyber Medical BioTy™ Nanotopography Trauma Screw ([K161597](/device/K161597.md))
- Tyber Medical Trauma Screw ([K133842](/device/K133842.md))

## Reference Devices

- Wright Medical PRO-TOE® ([K120645](/device/K120645.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 20, 2017

Tyber Medical, LLC Mark Schenk Vice President Of Regulatory And Quality 83 S Commerce Way. Suite 310 Bethlehem. Pennsylvania 18017

Re: K170571

Trade/Device Name: Tyber Medical BioTy™ Nanotopography Trauma Screw, TyFix Screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: May 15, 2017 Received: May 16, 2017

Dear Mark Schenk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K170571

Device Name

Tyber Medical BioTy" Nanotopography Trauma Screw, TyFix Screw

#### Indications for Use (Describe)

The Tyber Medical BioTy" Nanotopography Trauma Screw, TyFix Screw is indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion appropriate for the device. The screw is intended for single use only.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

### Traditional 510(k) Summary

as required by section 807.92(c).

### Tyber Medical

## BioTy™ Nanotopography Trauma Screw, TyFix Screw

### K170571

| Submitted                                                                               | 6/5/17                                                                                                                                                                                                                                                                                                                                                                                                                        |
|-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                                                                              | Tyber Medical, LLC<br>83 South Commerce Way, Ste 310<br>Bethlehem, PA 18017                                                                                                                                                                                                                                                                                                                                                   |
| Contact Person                                                                          | Mark F Schenk<br>VP of QA/RA<br>Phone: (610) 849-0645 Fax: (866) 889-9914<br>Email: mschenk@tybermed.com                                                                                                                                                                                                                                                                                                                      |
| Trade Name                                                                              | Tyber Medical BioTy™ Nanotopography Trauma Screw, TyFix Screw                                                                                                                                                                                                                                                                                                                                                                 |
| Common Name                                                                             | Bone Compression Screw                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Class                                                                            | Class II                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Classification Name<br>and Number                                                       | Smooth or threaded metallic bone fixation fastener<br>21 CFR 888.3040                                                                                                                                                                                                                                                                                                                                                         |
| Classification Panel:                                                                   | Orthopedic                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Product Code                                                                            | HWC                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Predicate Devices                                                                       | Tyber Medical BioTy™ Nanotopography Trauma Screw - K161597<br>Tyber Medical Trauma Screw - K133842                                                                                                                                                                                                                                                                                                                            |
| Reference Predicate                                                                     | Wright Medical PRO-TOE® - K120645                                                                                                                                                                                                                                                                                                                                                                                             |
| Device Description                                                                      | General trauma screw for compression and fixation of bone. The<br>BioTy™ Nanotopography provides a microscopic roughened surface<br>with nano-scaled features. The screws will be provided sterile in both<br>solid and cannulated form, made of from titanium and stainless steel.                                                                                                                                           |
| Intended Use                                                                            | A Tyber Medical BioTy™ Nanotopography Trauma Screw is designed<br>to apply fixation between two adjacent segments of cortical and/or<br>cancellous bone.                                                                                                                                                                                                                                                                      |
| Indications for Use                                                                     | The Tyber Medical BioTy™ Nanotopography Trauma Screw, TyFix<br>Screw is indicated for use in bone reconstruction, osteotomy,<br>arthrodesis, joint fusion appropriate for the size of the device. The<br>screw is intended for single use only.                                                                                                                                                                               |
| Statement of<br>Technological<br>Comparison and<br>Fundamental Scientific<br>Technology | Tyber Medical BioTy™ Nanotopography Trauma Screw, TyFix Screw<br>and its predicate device have the same indications for use and the<br>same design on the threaded end of the device, and the same<br>materials, technology and principles of operation and test results.<br><br>This submission adds a barbed end that is a similar design, materials,<br>technology and principles of operation to the reference predicate. |

{4}------------------------------------------------

| Nonclinical Testing<br>Summary | The following tests were performed to demonstrate that the<br>Tyber Medical BioTy™ Nanotopography Trauma Screw is substantially<br>equivalent to other predicate device.                                                                                  |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                | Pull out Test per ASTM F543 Static Torsion Test per ASTM F543 Pyrogenicity testing was performed per ST72:2011              The results of these studies showed that the Tyber Medical BioTy™<br>Nanotopography Trauma Screw met the acceptance criteria. |
| Clinical Test Summary          | n/a                                                                                                                                                                                                                                                       |

| Conclusion | Tyber Medical BioTy™ Nanotopography Trauma Screw, TyFix Screw and its predicate devices have the same indications for use, similar design, and test results. Both devices are manufactured using materials with a long history of use in orthopedic implants. |
|------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

---

**Source:** [https://fda-staging.innolitics.com/device/K170571](https://fda-staging.innolitics.com/device/K170571)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com