LifeWatch Mobile Cardiac Telemetry 3 Lead LifeWatch MCT 3L
K170565 · Lifewatch Services, Inc. · QYX · Aug 1, 2017 · Cardiovascular
Device Facts
Record ID
K170565
Device Name
LifeWatch Mobile Cardiac Telemetry 3 Lead LifeWatch MCT 3L
Applicant
Lifewatch Services, Inc.
Product Code
QYX · Cardiovascular
Decision Date
Aug 1, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The MCT 3L (CG-6108 ACT-3L) Continuous ECG Monitor and Arrhythmia Detector is intended for use by patients who experience transient symptoms that may suggest cardiac arthythmia. The device continuously monitors patient ECG, automatically generates an alert triggered by an arrhythmia detection algorithm, or generates an alert manually triggered by the patient, and transmits the recorded data transtelephonically to a monitoring center provides the ECG data to the medical practitioner for evaluation. The MCT 3L (CG-6108 ACT-3L) Continuous ECG Monitor and Arrhythmia Detector is intended to be prescribed for patients who have demonstrated a need for cardiac monitoring and are at low risk of developing life-threatening arrhythmias. Conditions where the system should not be used include patients likely to experience primary Ventricular Fibrillation or Ventricular Tachycardia and patients who have other co-morbid cardiovascular conditions where an arrhythmia could be potentially life threatening. The device has been validated for use on patients 6 years of age and older.
Device Story
MCT 3L is a chest-worn sensor with 4 electrodes on a harness; acquires, records, and transmits ECG signals. Device includes 3.6V AA battery, Bluetooth transceiver, and patient alert buzzer. ECG signals transmit via Bluetooth to a handheld mobile computing device (e.g., cell phone) running proprietary application. System processes and transmits ECG recordings via cellular link to a remote Monitoring Center for professional analysis; landline transmission available if cellular service is unavailable. Used by patients at home; operated by patients or caregivers. Monitoring center medical practitioners evaluate transmitted ECG data to inform clinical decision-making. Device provides continuous monitoring and automated arrhythmia detection to identify transient cardiac symptoms, benefiting patients requiring long-term monitoring.
Clinical Evidence
No clinical data was necessary to support substantial equivalence. Bench testing only, including human factors validation for pediatric use (ages 6-21), electromagnetic compatibility (ANSI/AAMI/IEC 60601-1-2), and wireless coexistence testing.
Technological Characteristics
Chest-worn sensor with 4-electrode harness; 3.6V AA battery power; Bluetooth connectivity; cellular/landline data transmission. Standards: IEC 62366-1:2015 (human factors), ISO 14971:2012 (risk management), ANSI/AAMI/IEC 60601-1-2:2014 (EMC). Software-based arrhythmia detection algorithm. Non-sterile, multi-use device.
Indications for Use
Indicated for patients 6 years and older experiencing transient symptoms suggesting cardiac arrhythmia; prescribed for those requiring cardiac monitoring at low risk for life-threatening arrhythmias. Contraindicated for patients likely to experience primary Ventricular Fibrillation or Ventricular Tachycardia, or those with co-morbid cardiovascular conditions where arrhythmia could be life-threatening.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
CG-6108 ACT 3L Continuous ECG Monitor and Arrhythmia Detector (K143359)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the rest of the name. The text reads "U.S. FOOD & DRUG ADMINISTRATION".
November 15, 2023
LifeWatch Services, Inc Stefanie Martinez-Koenig Sr. Director Regulatory and Clinical Trials 10255 W Higgins Road Suite 100 Rosemont, Illinois 60018
Re: K170565
Trade/Device Name: LifeWatch Mobile Cardiac Telemetry 3 Lead LifeWatch MCT 3L Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II Product Code: QYX, DSI
Dear Stefanie Martinez-Koenig:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated August 1, 2017. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Kozen, OHT2: Office of Cardiovascular Devices, 301-796-5813, jennifer.kozen@fda.hhs.gov.
Sincerely.
Jennifer W. Shih -S
Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a family.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 1. 2017
LifeWatch Services, Inc. Stefanie Martinez-Koenig Sr. Director Regulatory and Clinical Trials 10255 W Higgins Road Suite 100 Rosemont, Illinois 60018
Re: K170565
Trade/Device Name: LifeWatch Mobile Cardiac Telemetry 3 Lead LifeWatch MCT 3L (CG-6108 ACT-3L) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: DSI Dated: June 30, 2017 Received: July 3, 2017
Dear Stefanie Martinez-Koenig:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{2}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
M.A. Wilhelmsson
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170565
Device Name
LifeWatch Mobile Cardiac Telemetry 3 Lead LifeWatch MCT 3L (CG-6108 ACT-3L)
#### Indications for Use (Describe)
The MCT 3L (CG-6108 ACT-3L) Continuous ECG Monitor and Arrhythmia Detector is intended for use by patients who experience transient symptoms that may suggest cardiac arthythmia. The device continuously monitors patient ECG, automatically generates an alert triggered by an arrhythmia detection algorithm, or generates an alert manually triggered by the patient, and transmits the recorded data transtelephonically to a monitoring center provides the ECG data to the medical practitioner for evaluation.
The MCT 3L (CG-6108 ACT-3L) Continuous ECG Monitor and Arrhythmia Detector is intended to be prescribed for patients who have demonstrated a need for cardiac monitoring and are at low risk of developing life-threatening arrhythmias. Conditions where the system should not be used include patients likely to experience primary Ventricular Fibrillation or Ventricular Tachycardia and patients who have other co-morbid cardiovascular conditions where an arrhythmia could be potentially life threatening.
The device has been validated for use on patients 6 years of age and older.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------|--------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the LifeWatch logo, which features a series of curved lines in shades of green and blue, resembling a waveform. To the right of the waveform is the word "LifeWatch" in a stylized font, with a trademark symbol next to it. The text "K170565" is located in the upper right corner of the image.
### 510(k) Summary
This document contains the 510(k) summary for the LifeWatch Mobile Cardiac Telemetry 3 Lead Continuous ECG Monitor and Arrhythmia Detector. The content of this summary is based on the requirements of 21 CFR Section 807.92(c)
Submission Date: February 24, 2017
#### Submitter Information:
| Submitted By: | Stefanie Martinez-Koenig, Senior Director, Regulatory and Clinical<br>Trials<br>10255 W Higgins Rd Suite 100<br>Rosemont, IL 60018<br>Email: SMartinezKoenig@LifeWatch.com |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Robin Martin, Regulatory Consultant<br>10255 W Higgins Rd Suite 100<br>Rosemont, IL 60018<br>Email: CT-RMartin@LifeWatch.com |
| Device Information: | |
| Trade Name: | LifeWatch Mobile Cardiac Telemetry 3 Lead LifeWatch MCT3L<br>(CG-6108 ACT-3L) |
| Common Name: | MCT 3L |
| Classification Name: | Detector and Alarm, Arrhythmia<br>Transmitters and Receivers, Electrocardiograph, Telephone |
| Device Classification: | 870.1025<br>870.2920 |
| Predicate Device(s): | K143359 CG-6108 ACT 3L Continuous ECG Monitor and<br>Arrhythmia Detector<br>LifeWatch Technologies, Ltd. (DSI and DXH) |
| Device Description: | The MCT 3L Continuous ECG Monitor and Arrhythmia Detector is<br>designed for self-testing by patients at home and for analysis by<br>medical professionals at a remote Monitoring Center. |
| 510(k) Premarket Notification | |
| | The chest-worn sensor is used for the acquisition, recording, and<br>transmission of the ECG signal. The device is equipped with 4<br>electrodes on a harness and it houses a 3.6V AA battery, a<br>Bluetooth transceiver and a patient alert buzzer.<br>The ECG signals are transmitted via Bluetooth to a handheld<br>device with a proprietary interactive application, configured to<br>process and transmit the ECG recordings. The handheld device is a<br>mobile computing device with a display and a touch input such as<br>a cell phone. It has sufficient memory and processing capability to<br>run the proprietary application.<br>When an arrhythmia event is detected, the handheld device<br>transmits the recorded ECG information automatically via cellular<br>link, to the Monitoring Center for professional analysis. When<br>cellular service is unavailable the patient has an option to transmit<br>via a landline telephone. |
| | The MCT 3L does contain software. The device is not supplied<br>sterile or intended to be sterilized by the operator prior to use,<br>and is not intended for single use. |
| Indications for Use: | The MCT 3L (CG-6108 ACT-3L) Continuous ECG Monitor and<br>Arrhythmia Detector is intended for use by patients who<br>experience transient symptoms that may suggest cardiac<br>arrhythmia. The device continuously monitors patient ECG,<br>automatically generates an alert triggered by an arrhythmia<br>detection algorithm, or generates an alert manually triggered by<br>the patient, and transmits the recorded data transtelephonically<br>to a monitoring center. The monitoring center provides the ECG<br>data to the medical practitioner for evaluation. |
| | The MCT 3L (CG-6108 ACT-3L) Continuous ECG Monitor and<br>Arrhythmia Detector is intended to be prescribed for patients who<br>have demonstrated a need for cardiac monitoring and are at low<br>risk of developing life-threatening arrhythmias. Conditions where<br>the system should not be used include patients likely to<br>experience primary Ventricular Fibrillation or Ventricular<br>Tachycardia and patients who have other co-morbid<br>cardiovascular conditions where an arrhythmia could be<br>potentially life threatening.<br>The device has been validated for use on patients 6 years of age<br>and older. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for LifeWatch, which includes a series of curved lines in shades of green and blue, followed by the company name in a stylized font. To the right of the logo, the text "K170565" is visible.
# It C />
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the LifeWatch logo. The logo features a series of curved lines in shades of blue and green, resembling a wave pattern. To the right of the wave pattern is the word "LifeWatch" in a stylized font, with a trademark symbol next to it.
K170565
## 510(k) Premarket Notification
| Comparison to Predicate: | The Indications for Use for the proposed MCT 3L device are identical to those of the predicate device with the exception that use for pediatrics age 6 and older is not excluded in the MCT 3L labeling. Bench testing of the proposed MCT 3L was conducted to demonstrate substantial equivalence to the predicate device. Insignificant historical changes are listed within the device description. No changes have been made to the user interface, device algorithms, or applications software compared to the predicate device. |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The proposed device has the same intended use and employs the same fundamental scientific technology. |
| Performance Testing: | |
| Non-Clinical: | The design and development of the MCT 3L device conforms to LifeWatch Service's Quality System. Because the subject of this premarket notification is a labeling change, and certain predicate standards test reports remain applicable to the device, testing to only the following voluntary standards are submitted to support substantial equivalence to the predicate device: <ul><li>IEC 62366-1:2015, including human factors validation testing related to the use of the device in the pediatric patient population (age 6 to 21) with adult caregivers, where appropriate.</li><li>ISO 14971:2012, including evaluation of use-related risks</li><li>ANSI/AAMI/IEC 60601-1-2 Edition 4:2014-02, Electromagnetic Compatibility</li><li>Wireless Coexistence Testing in accordance with FDA's guidance entitled, "Radio Frequency Wireless Technology in Medical Devices", dated August 14, 2013</li></ul> |
| Clinical: | No clinical data was necessary to support substantial equivalence of the proposed device compared to the predicate. |
| Conclusion: | The above testing supports that the proposed device is as safe, effective and performs as well as the legally marketed predicate in its intended use. Therefore, the proposed device is substantially equivalent to the predicate. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.