K170557 · Genepoc, Inc. · NJR · May 25, 2017 · Microbiology
Device Facts
Record ID
K170557
Device Name
GenePOC GBS LB
Applicant
Genepoc, Inc.
Product Code
NJR · Microbiology
Decision Date
May 25, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3740
Device Class
Class 1
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K170557 · May 25, 2017
GenePOC GBS LB
Genepoc, Inc.
Residual de-identified LIM broth samples from routine clinical diagnostic procedures
Residual clinical samples from 839 pregnant women were used to evaluate the clinical sensitivity and specificity of the GenePOC GBS LB assay compared to a composite reference method (culture).
Pregnant women at 35-37 weeks of gestation for whom GBS diagnostic procedures were ordered; Sample Size: 771 evaluable specimens; Number of Sites: 4 (2 Canadian, 2 US)
Composite reference method (LIM Broth enrichment followed by subculture and biochemical identification)
Clinical sensitivity and specificity
Indications for Use
The GenePOC™ GBS LB assay performed on the revogene™ instrument is a qualitative in vitro diagnostic test designed to detect Group B Streptococcus (GBS) DNA from 18-24 hour LIM broth enrichments of vaginal/rectal specimen swabs obtained from pregnant women. The GenePOC GBS LB assay utilizes automated sample processing and real-time polymerase chain reaction (PCR) to detect a cfb gene sequence specific to the Streptococcus agalactiae genome. The GenePOC GBS LB assay is indicated for the identification of antepartum GBS colonization and does not provide susceptibility results. It is not intended to diagnose or monitor treatment of GBS infection. Culture isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women.
Device Story
The GenePOC GBS LB assay is a qualitative, automated in vitro diagnostic test for detecting GBS DNA. Input consists of 18-24 hour LIM broth-enriched vaginal/rectal swabs. The user transfers a sample aliquot into a provided Sample Buffer Tube (SBT) and then into a single-use microfluidic cartridge. The revogene instrument performs automated sample processing, including cell lysis, dilution, and real-time PCR amplification using TaqMan chemistry to target the cfb gene. The instrument monitors an internal process control (PrC) to validate extraction and amplification. Output results (positive, negative, indeterminate, unresolved) are generated by the instrument's embedded algorithms based on fluorescent signal analysis. The device is intended for use in clinical laboratory settings by trained personnel. It aids in identifying antepartum GBS colonization, allowing healthcare providers to determine appropriate prophylactic strategies for pregnant patients.
Clinical Evidence
Prospective multicenter trial (n=771 compliant specimens) compared the GenePOC GBS LB assay to a composite reference method (LIM broth enrichment followed by subculture and biochemical identification). Results showed 95.9% sensitivity (95% CI: 91.7-98.0%) and 95.5% specificity (95% CI: 93.5-96.9%).
Technological Characteristics
Single-use microfluidic cartridge; real-time PCR with fluorogenic TaqMan probes; automated sample processing (lysis, dilution, amplification); 4-channel optical detection; revogene instrument platform; internal process control (PrC) for monitoring inhibition and system failure; detects 190 bp cfb gene sequence.
Indications for Use
Indicated for the identification of antepartum Group B Streptococcus (GBS) colonization in pregnant women using 18-24 hour LIM broth enriched vaginal/rectal swabs. Not for diagnosing or monitoring GBS infection treatment; does not provide susceptibility results.
Regulatory Classification
Identification
Streptococcus spp. serological reagents are devices that consist of antigens and antisera (excluding streptococcal exoenzyme reagents made from enzymes secreted by streptococci) used in serological tests to identify Streptococcus spp. from cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genus Streptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K170557
B. Purpose for Submission:
Clearance of New Device
C. Measurand:
Target DNA sequence in *cfb* gene of *Streptococcus agalactiae* (Group B *Streptococcus*, GBS)
D. Type of Test:
An *in vitro* molecular diagnostic test for the qualitative detection of GBS DNA in enriched LIM Broth culture by real-time PCR.
E. Applicant:
GenePOC Inc.
F. Proprietary and Established Names:
GenePOC GBS LB Assay
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3740—*Streptococcal* spp. serological reagents
2. Classification:
Class I, non-exempt
3. Product code:
NJR—Nucleic Acid Amplification Assay System, Group B Streptococcus, Direct Specimen
4. Panel:
Microbiology (83)
{1}
H. Intended Use:
1. Intended use(s):
The GenePOC GBS LB assay performed on the revogene instrument is a qualitative *in vitro* diagnostic test designed to detect Group B *Streptococcus* (GBS) DNA from 18-24 hour LIM broth enrichments of vaginal/rectal specimen swabs obtained from pregnant women. The GenePOC GBS LB assay utilizes automated sample processing and real-time polymerase chain reaction (PCR) to detect a *cfb* gene sequence specific to the *Streptococcus agalactiae* genome.
The GenePOC GBS LB assay is indicated for the identification of antepartum GBS colonization and does not provide susceptibility results. It is not intended to diagnose or monitor treatment of GBS infection. Culture isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements:
The GenePOC GBS LB Assay is run on the revogene instrument, which is capable of automated nucleic acid extraction, amplification, real-time PCR detection of target nucleic acid sequences with data analysis using a GBS-specific assay definition file. The final software version reported for this 510(k) submission was V3.1.5 and 0.3.4 for the firmware.
*Note: Refer to K170558 for additional information regarding the revogene instrument and related software.*
I. Device Description:
The GenePOC GBS LB Assay is a single-use test for the qualitative detection of GBS DNA from enriched vaginal/rectal swab specimens using real-time PCR technology and fluorogenic oligonucleotide probes. The assay is automated and utilizes a microfluidic cartridge for the simultaneous detection of the target GBS DNA and the internal process control (PrC) DNA on the revogene instrument.
The GenePOC GBS LB Assay is composed of a GBS-specific disposable microfluidic cartridge (PIE), Sample Buffer Tube (SBT) and Disposable Transfer Tool (DTT). These components are used to lyse and dilute the sample, amplify, and detect GBS nucleic acid from vaginal/rectal swabs following LIM Broth enrichment. User intervention is required for sample preparation, adding LIM Broth enriched samples into the SBT, transferring the
{2}
sample into the cartridge, and loading/unloading the cartridge into the revogene instrument. Once the sample is added into the cartridge, the process is then fully automated. Each test kit contains 24 individual pouches.
The revogene instrument can process from one up to a maximum of 8 samples simultaneously in the same run. On completion of a run, the results are reported by the revogene instrument from measured fluorescent signals and embedded calculation algorithms. For the GBS application, two signals are processed: GBS target and PrC. The output results can be reported as positive, negative, indeterminate, or unresolved. On completion of a run, the user removes the used cartridges and disposes of them in normal biological waste. Results may be viewed, printed, transferred, and/or stored by the user.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
BD MAX GBS Assay
2. Predicate 510(k) number(s):
K111860
3. Comparison with predicate:
| Item | GenePOC GBS LB Assay (Subject Device) K170557 | BD MAX GBS Assay (Predicate Device) K111860 |
| --- | --- | --- |
| SIMILARITIES | | |
| Intended Use | The GenePOC GBS LB assay performed on the revogene instrument is a qualitative in vitro diagnostic test designed to detect Group B Streptococcus (GBS) DNA from 18-24 hour LIM broth enrichments of vaginal/rectal specimen swabs obtained from pregnant women. The GenePOC GBS LB assay utilizes automated sample processing and real-time polymerase chain reaction (PCR) to detect a cfb gene sequence specific to the Streptococcus agalactiae genome. The GenePOC GBS LB assay is indicated for the identification of antepartum GBS colonization and does not provide susceptibility results. It is not intended to diagnose | The BD MAX GBS Assay as implemented on the BD MAX System is a qualitative in vitro diagnostic test designed to detect Group B Streptococcus (GBS) DNA in Lim Broth cultures after incubation for greater than or equal to (>)18 hours, obtained from vaginal-rectal swab specimens from antepartum pregnant women. The test incorporates automated DNA extraction to isolate the target nucleic acid from the specimen and real-time polymerase chain reaction (PCR) to detect a 124 bp region of the cfb gene sequence of the Streptococcus agalactiae chromosome. Results from the BD MAX GBS Assay can be used as an aid in determining colonization |
{3}
| Item | GenePOC GBS LB Assay (Subject Device) K170557 | BD MAX GBS Assay (Predicate Device) K111860 |
| --- | --- | --- |
| | or monitor treatment of GBS infection. Culture isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women. | status in antepartum women.
The BD MAX GBS Assay does not provide susceptibility results.
Cultured isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women. Subculture to solid media for additional testing when indicated.
The BD MAX System is intended for in vitro diagnostic (IVD) use in performing FDA cleared or approved nucleic acid testing in clinical laboratories. The BD MAX System is capable of automated extraction and purification of nucleic acids from multiple specimen types as well as the automated amplification and detection of target nucleic acid sequences by fluorescence-based PCR. |
| Specimen type | Vaginal-Rectal Swab (Enriched LIM Broth) | Same |
| Sample Preparation Method | Manual; vaginal/rectal swab enriched overnight in LIM Broth, then an aliquot of culture added to sample buffer | Same |
| DNA Extraction | Automated by instrument | Same |
| Assay Format | • Amplification: Real-Time PCR
• Detection: Fluorogenic | Same |
| Result Interpretation | Automated using software | Same |
| Internal Process Control | To help monitor presence of potential inhibitory substances as well as any system or reagent failures | Same |
| External Control | Materials available commercially | Same |
{4}
| Item | GenePOC GBS LB Assay (Subject Device) K170557 | BD MAX GBS Assay (Predicate Device) K111860 |
| --- | --- | --- |
| DIFFERENCES | | |
| DNA Target | 190 bp region of the cfb gene | 124 bp region of cfb gene |
| Probe Design | TaqMan | Scorpion |
| GBS Assay Format | Fully integrated sample processing and PCR reaction/detection in a cartridge | Requires combination of a reagent strip and a PCR cartridge |
| Single Use | Cartridge (PIE) can be used once | Cartridge can be used twice (contains 24 test channels) |
| Instrument Optical Channels | revogene contains 4 channels | BD MAX Instrument contains 6 optical channels |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI Guideline EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline.
# L. Test Principle:
Vaginal-rectal swab specimens are collected and used to inoculate LIM Broth. After incubation at $35 - 37^{\circ}\mathrm{C}$ for 18-24 hours in ambient air, $15\mu \mathrm{l}$ of the enriched broth culture is transferred to the SBT. After vortexing the SBT for 15 seconds, approximately $150~\mu \mathrm{l}$ of the inoculated sample buffer is transferred into the GenePOC microfluidic cartridge (PIE) using the DTT. The loaded GBS LB cartridge is placed into the revogene for further sample processing. No operator intervention is necessary once the clinical sample is loaded onto the revogene.
Each GBS LB microfluidic cartridge is a completely integrated and self-contained cartridge. Each sample is transferred by centrifugation from one microfluidic chamber to the next in sequence, where all reagents specific for the PCR reaction are incorporated and dried within the PCR wells. The step-wise process includes sample homogenization, lysis of cells, and specimen dilution followed by subsequent real-time PCR steps within 1 PCR well in the cartridge. An internal Process Control (PrC) is contained in the homogenization chamber and is present in every test to monitor the analytical process (including sample lysis, dilution and nucleic acid amplification, and detection) for the presence of potential inhibitory substances as well as system or reagent failures. The amplified products are detected in real time using target-specific TaqMan chemistry-based probes. The GBS LB-specific designed primers and probe detect a target region of 190 base pairs of the cfb gene of the Streptococcus agalactiae genome. The results are reported by the revogene from measured fluorescent signals and embedded calculation algorithms.
# Interpretation of Sample Results
The analysis software associated with the revogene instrument determines results for GBS and PrC based on the amplification cycle (Ct) values provided in the assay protocol file. The
{5}
GenePOC GBS LB Assay performed on the revogene requires the use of a Ct cut-off and a threshold on RFU to report results. All assay outcomes are described below in Table 1.
Table 1. Summary of all Potential Results for the GenePOC GBS LB Assay and Interpretation
| Symbol Displayed on User Screen | Overall Result Reported | Interpretation of PrC Result | Interpretation of Result |
| --- | --- | --- | --- |
| + | Positive | Positive or Negative | Sample contains GBS target DNA. The process control may not amplify in the presence of high load of GBS target DNA. |
| - | Negative | Positive | No GBS target DNA detected. Requires positive PrC result. |
| ? | Unresolved | Negative | Amplification/detection failure for the process control as well as for the GBS target. Repeat testing must be performed using the original corresponding enriched LIM Broth specimen or inoculated SBT within the defined timeframe as described in the package insert. |
| ! | Indeterminate | N/A | No reportable result due to possible instrument errors. Repeat testing must be performed using the original corresponding enriched LIM Broth specimen or inoculated SBT within the defined timeframe as described in the package insert. |
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
The reproducibility of the GenePOC GBS LB Assay was evaluated by testing three reagent lots of the assay on three revogene instruments (one per site) with two operators at each of three sites for five days. A blinded and randomized reproducibility panel containing three GBS strains was tested in triplicate at two concentrations in negative matrix: low positive (735 CFU/ml) and moderate positive (1,125 CFU/ml). A Precision Study was based on the within-site variability over five days. The GBS strains used in the panel included two hemolytic strains (ATCC 12400 and ATCC 12403) and one non-hemolytic strain (ATCC 13813). Two true negative samples (LIM Broth matrix previously inoculated with vaginal/rectal swabs that was determined to be GBS negative) were also included as panel members. Across three sites, a total of 720 replicates were tested with the GenePOC GBS LB Assay. For site-to-site reproducibility, the overall percent agreement with expected result (i.e., negative for true negative and positive for moderate positive and low positive samples) was $>95\%$ for all GBS strains tested at low positive and high positive concentration. The true negatives yielded expected results $100\%$ of the time. During the Reproducibility and Precision Study, one unresolved result and one indeterminate
{6}
result were reported for a rate of $0.1\%$ for the indeterminate rate and the unresolved rate. The results from the site-to site Reproducibility Study for the GenePOC GBS LB Assay are presented in Table 2 below. Bacterial cell numbers in the suspensions were verified by colony counts.
Table 2. Summary of the Percent Agreement Analysis across all Sites
| Panel Member | Observed/Expected | | | All Sites | % Agreement |
| --- | --- | --- | --- | --- | --- |
| | Site 1 | Site 2 | Site 3 | | |
| ATCC 13813, Low Positive | 29/30 | 30/30 | 30/30 | 89/90 | 98.9% |
| ATCC 13813, High Positive | 28/30 | 29/30 | 30/30 | 87/90 | 96.7% |
| ATCC 12403, Low Positive | 28/30 | 30/30 | 30/30 | 88/90 | 97.8% |
| ATCC 12403, High Positive | 29/30 | 30/30 | 30/30 | 89/90 | 98.9% |
| ATCC 12400, Low Positive | 30/30 | 30/30 | 30/30 | 90/90 | 100% |
| ATCC 12400, High Positive | 30/30 | 30/30 | 30/30 | 90/90 | 100% |
| True Negative | 60/60 | 60/60 | 60/60 | 180/180 | 100% |
The quantitative results per panel member for between-site, between-operator and between-day reproducibility for Ct values are summarized in Table 3 below. Table 4 shows the $\% \mathrm{CV}$ from the Precision Study. The Reproducibility/Precision Studies are acceptable.
Table 3. Summary of the Overall SD and % CV for the Ct values in the Reproducibility Study
| Target Load | Strain | Ct GBS | | Inter-site | | Inter-operator | | Inter-day | | Overall | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | N | Mean Ct | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Low Positive | ATCC 13813 | 89 | 36.4 | 2.6 | 7.1 | 2.4 | 6.6 | 3.9 | 10.6 | 5.2 | 14.4 |
| | ATCC 12403 | 88 | 36.9 | 3.2 | 8.5 | 2.7 | 7.3 | 4.3 | 11.5 | 5.9 | 16.1 |
| | ATCC 12400 | 90 | 35.8 | 2.6 | 7.4 | 2.2 | 6.2 | 3.8 | 10.5 | 5.1 | 14.3 |
| Moderate Positive | ATCC 13813 | 87 | 35.9 | 2.5 | 7.1 | 2.5 | 7.0 | 4.0 | 11.2 | 5.4 | 15.0 |
| | ATCC 12403 | 89 | 35.9 | 2.8 | 7.8 | 2.4 | 6.7 | 3.9 | 10.7 | 5.3 | 14.9 |
| | ATCC 12400 | 90 | 35.3 | 2.7 | 7.6 | 2.2 | 6.4 | 3.7 | 10.4 | 5.1 | 14.4 |
| Target Load | Strain | Ct PrC | | Inter-site | | Inter-operator | | Inter-day | | Overall | |
| | | N | Mean | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| True Negative | None | 180 | 31.9 | 1.7 | 5.5 | 1.8 | 5.7 | 2.9 | 9.2 | 3.9 | 12.1 |
{7}
Table 4. Summary of the SD and % CV for the Ct values for the Within-Site Precision Study
| Target Load | Strain | Mean Ct, SD, and % CV for GBS | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Site 1 | | | | Site 2 | | | | Site 3 | | | |
| | | N | Mean | SD | %CV | N | Mean | SD | %CV | N | Mean | SD | %CV |
| True Negative | None | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA | NA |
| Low Positive | ATCC 13813 | 29 | 37.6 | 1.8 | 4.9 | 30 | 36.3 | 1.0 | 2.8 | 30 | 35.4 | 1.5 | 4.3 |
| | ATCC 12403 | 28 | 37.7 | 2.4 | 6.4 | 30 | 37.0 | 1.6 | 4.3 | 30 | 36.1 | 1.3 | 3.5 |
| | ATCC 12400 | 30 | 36.7 | 1.6 | 4.4 | 30 | 35.4 | 1.5 | 4.1 | 30 | 35.2 | 1.5 | 4.2 |
| Moderate Positive | ATCC 13813 | 28 | 37.2 | 1.9 | 5.2 | 29 | 36.0 | 1.1 | 3.2 | 30 | 34.7 | 1.2 | 3.3 |
| | ATCC 12403 | 29 | 36.6 | 2.1 | 5.7 | 30 | 35.9 | 1.5 | 4.1 | 30 | 35.2 | 1.2 | 3.4 |
| | ATCC 12400 | 30 | 36.0 | 1.2 | 3.5 | 30 | 35.1 | 1.3 | 3.6 | 30 | 34.7 | 2.0 | 5.8 |
| Target Load | Strain | Mean Ct, SD, and % CV for PrC | | | | | | | | | | | |
| | | Site 1 | | | | Site 2 | | | | Site 3 | | | |
| | | N | Mean | SD | %CV | N | Mean | SD | %CV | N | Mean | SD | %CV |
| True Negative | None | 60 | 32.1 | 1.2 | 3.7 | 60 | 32.7 | 0.7 | 3.2 | 60 | 30.8 | 1.1 | 2.4 |
| Low Positive | ATCC 13813 | 30 | 32.5 | 1.0 | 3.0 | 30 | 32.4 | 0.9 | 2.6 | 30 | 31.2 | 1.0 | 3.1 |
| | ATCC 12403 | 30 | 32.5 | 2.2 | 6.9 | 30 | 32.2 | 0.9 | 2.7 | 30 | 31.2 | 1.6 | 5.0 |
| | ATCC 12400 | 30 | 32.4 | 1.5 | 4.7 | 30 | 31.9 | 0.9 | 2.8 | 30 | 31.1 | 0.7 | 2.4 |
| Moderate Positive | ATCC 13813 | 30 | 32.7 | 1.2 | 3.8 | 30 | 32.9 | 1.2 | 3.5 | 30 | 31.0 | 0.9 | 3.1 |
| | ATCC 12403 | 30 | 31.6 | 0.9 | 2.9 | 30 | 32.5 | 1.1 | 3.5 | 30 | 30.8 | 0.8 | 2.6 |
| | ATCC 12400 | 29 | 32.1 | 0.8 | 2.5 | 30 | 32.1 | 0.8 | 2.4 | 29 | 31.3 | 1.0 | 3.3 |
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Controls
There are three types of controls for the GenePOC GBS LB Assay including:
Internal Processing Control: Each GenePOC GBS LB Assay cartridge contains an Internal Process Control (PrC). The PrC is lysed, amplified, and detected along
{8}
with each specimen tested and monitors the efficacy of the DNA extraction and PCR amplification processes.
**External Controls:** External controls should be tested according to guidelines or requirements of local, provincial and/or federal regulations or accreditation organizations. Commercial control material (e.g., *Streptococcus agalactiae* ATCC 13813) can be used as a positive external control. It is recommended that bacterial strains be freshly prepared in LIM Broth. Transfer a 15 μl aliquot of an 18-24 hour culture in LIM broth into a SBT. Pure LIM Broth is recommended for use as a negative external control. The GenePOC GBS LB Assay does not include external positive and negative controls.
During the clinical study, positive and negative external controls were tested with the GenePOC GBS LB Assay daily. Overall, it was reported that 99.5% of the external controls yielded the expected results (189/190). The submitted external control data are acceptable.
## Reagent Stability
A Reagent Stability Study was conducted with three GenePOC GBS LB reagent lots and three revogene instruments. Stability of the kits stored at both 4°C and 25°C was evaluated at the following time points: Day 0, Day 14, Month 1, Month 2, Month 4, and Month 5. The three reagent lots were not produced on the same date and, therefore, testing was staggered. For every stability point, six cartridges per condition were processed. All SBT were inoculated with 15 μl of LIM Broth containing GBS gDNA.
Results of the study indicated that the GenePOC GBS LB Assay reagents were stable up to 4 months with conditioning at 4°C and 25°C. Testing of all lots is ongoing at GenePOC until the 15 month time point is reached.
## LIM Broth Stability and SBT Stability
For both the LIM Broth Culture and inoculated SBT Stability Studies, performance of the GenePOC GBS LB Assay was assessed when broth cultures and SBT were stored at room temperature and 2 – 8°C across several time points: 0, 1, 2, 3, 5, 7, and 12 days. All testing was performed on three revogene instruments. One strain of GBS was tested at 1,125 CFU/ml and prepared in negative matrix for the positive samples. An aliquot of negative matrix was used to prepare negative samples. Four replicates were tested at each time point and storage condition per lot across three lots. Results from the nested stability study at room temperature (25 ± 2°C) showed that enriched LIM Broth culture could be stored at 25 ± 2°C for up to 48 hrs before inoculating SBT (then held at room temperature for an extra 3 days) and still detect 4/4 GBS samples per condition. Two unexpected results were obtained in the room temperature stability test. One sample during the room temperature study yielded an unresolved assay result; it was re-tested and yielded a positive result. Also, one replicate yielded a false negative GBS result with LIM Broth stored for 48 hrs at room temperature only, which was considered a worst-case scenario. For the LIM Broth stored at 2-8°C up to 3 days, all tested samples produced the expected results.
9
{9}
LIM Broth culture stored for 3 days at $2 - 8^{\circ}\mathrm{C}$ was then used to inoculate SBT and held up to 12 days at $2 - 8^{\circ}\mathrm{C}$ . All SBT samples produced the expected results when stored for 12 days at $2 - 8^{\circ}\mathrm{C}$ . In addition, all positive and negative external controls yielded the expected results during the study. The initial rate of unresolved results was $0.3\%$ (1/336). No indeterminate results were obtained during the study. Table 5 shows results of the stability studies with LIM Broth culture and inoculated SBT. The results support claims for the following in the package insert:
- LIM Broth Culture stored at $25^{\circ}\mathrm{C}$ for up to 2 days or at $2 - 8^{\circ}\mathrm{C}$ for up to 3 days.
- Inoculated SBT stored at $25^{\circ}\mathrm{C}$ for up to 3 days or at $2 - 8^{\circ}\mathrm{C}$ for up to 5 days.
Table 5. LIM Broth and Inoculated SBT Stability Results
| Parameter | Storage Temperature | Results |
| --- | --- | --- |
| LIM Broth culture storage and conservation without refrigeration | 25 ± °2C | Stable up to 2 days (48 hrs) |
| Inoculated SBT conservation and storage without refrigeration | | Stable up to 3 days (72 hrs) |
| LIM Broth culture conservation and storage with refrigeration | 4 ± 3°C | Stable up to 3 days |
| Inoculated SBT conservation and storage with refrigeration | | Stable up to 12 days |
# PIE Stability
To establish the sample stability after loading the sample into a cartridge, a true negative and one strain of GBS (ATCC 12403) spiked at 1,125 CFU/ml in SB were tested at room temperature in duplicate across three time points: 0, 60 minutes, and 120 minutes. Assays were performed using three reagent lots and three revogene instruments. All samples yielded the expected results. A sample stability of 60 minutes in the PIE will be claimed. No unresolved or indeterminate results were obtained in the study.
# d. Detection limit:
# LoD
A Limit of Detection (LoD) study was performed to evaluate the analytical sensitivity of the GenePOC GBS LB Assay using two representative strains of GBS (ATCC 12403 and ATCC 13813). Preliminary LoD concentrations were determined by serially diluting strains in negative clinical LIM Broth matrix from 5,000 CFU/ml down to 150 CFU/ml per strain. For each strain, eight replicates were tested at each dilution across three GenePOC GBS LB assay lots. The observed LoD of each GBS strain was determined as the lowest concentration that had a positivity rate of $\geq$ $95\%$ . While the preliminary LoD for ATCC 13813 GBS strain was estimated to be
{10}
at 200 CFU/ml, the LOD for ATCC 12403 GBS strain was determined to be 375 CFU/ml. The analytical studies were performed using these two strains based on the final assay LoD claims. Only one sample was reported as unresolved for an initial unresolved rate of 0.3% (1/384). Repeat testing yielded a valid result.
The LoD concentrations determined in the preliminary study were confirmed with the same GBS reference strains (diluted to the preliminary LoD concentrations) and tested with twenty-four (24) replicates. A third GBS strain was added (ATCC 12400) to the confirmation panel, where the LoD was estimated to be 375 CFU/ml. The final LoDs values for these strains are presented in Table 6 below.
Table 6. Limit of Detection of the GenePOC GBS LB Assay
| GBS Strain | LoD Concentration (CFU/ml) | Observed/Expected (% Agreement) |
| --- | --- | --- |
| ATCC 12403 | 375 | 23/24 (96%) |
| ATCC 13813 | 200 | 23/24 (96%) |
| ATCC 12400 | 375 | 23/24 (96%) |
## Analytical Reactivity
The analytical reactivity (inclusivity) of the GenePOC GBS LB Assay was evaluated with a panel of twelve *Streptococcus agalactiae* strains representing 11 serotypes (Ia, Ib, Ic, II, III, IV, V, VI, VII, VIII, IX) and one non-hemolytic strain. Each isolate was tested with twenty-four (24) replicates in negative matrix at a concentration near the LoD based on the average CFU/ml from the three isolates tested in the LoD study. If the positivity rate for GBS was not 100%, sample GBS strain concentrations were increased until all replicates were positive. Table 7 below provides a list of GBS strains tested and the test concentration where 100% of the replicates were positive for GBS.
Table 7. Analytical Reactivity Study
| GBS Strain | Serotype | Concentration at which 100% positivity observed^a |
| --- | --- | --- |
| ATCC 12400 | Serotype Ia | 735 CFU/ml |
| ATCC 51487 | Serotype Ib | 5,625 CFU/ml |
| ATCC 27591 | Serotype Ic | 1,875 CFU/ml |
| ATCC 12973 | Serotype II | 735 CFU/ml |
| ATCC 12403 | Serotype III | 375 CFU/ml |
| ATCC 49446 | Serotype IV | 2,625 CFU/ml |
| ATCC BAA-611 | Serotype V | 1,875 CFU/ml |
| ATCC BAA-2671 | Serotype VI | 1,125 CFU/ml |
| ATCC BAA-2670 | Serotype VII | 2,625 CFU/ml |
| ATCC BAA-2669 | Serotype VIII | 3,750 CFU/ml |
| ATCC BAA-2668 | Serotype IX | 1,125 CFU/ml |
| ATCC 13813 | Non-Hemolytic | 500 CFU/ml |
<a>^a</a> The higher concentrations needed to obtain 100% positivity are not a concern as the enriched LIM Broth culture is expected to have GBS organism concentrations well-above the assay LoD.
{11}
# e. Analytical specificity:
A study was conducted to determine the cross-reactivity of 75 microorganisms (or gDNA/RNA) that represented various non-GBS groups of Streptococci, along with other bacteria, parasites, and viruses normally found in vaginal and anal flora, with the GenePOC GBS LB Assay. Bacteria and yeasts were tested at $\geq 10^{6}$ CFU/ml in SB, whereas viruses, parasites and human DNA were tested at $\geq 10^{5}$ DNA or RNA cp/ml in SB. Samples were tested in triplicate and spiked into negative enriched LIM Broth.
The results for all replicates in the cross-reactivity study were negative, demonstrating that the organisms or gDNA/RNA tested at these concentrations did not cross react with the GenePOC GBS LB Assay. During the study, no unresolved or indeterminate results were obtained with any of the organisms tested. Table 8 shows the cross-reactivity panel tested for this study.
Table 8. Cross-Reactivity Panel
| Bacteria (strain or gDNA) | |
| --- | --- |
| Acinetobacter baumannii | Mycoplasma genitalium gDNA |
| Aerococcus viridans | Mycoplasma hominis gDNA |
| Aeromonas hydrophila | Neisseria gonorrhoeae |
| Bacillus cereus | Peptostreptococcus anaerobius |
| Bacillus subtilis | Porphyromonas asaccharolytica |
| Bacteroides fragilis | Prevotella melaninogenica |
| Bifidobacterium adolescentis | Propionibacterium acnes |
| Bifidobacterium breve | Proteus mirabilis |
| Brevibacterium linens | Pseudomonas aeruginosa |
| Campylobacter jejuni | Salmonella enterica subsp. enterica serovar Dublin |
| Chlamydia trachomatis gDNA | Salmonella enterica subsp. enterica serovar Minneapolis |
| Citrobacter freundii | Salmonella enterica subsp. enterica serovar typhimurium |
| Clostridium difficile | Salmonella enterica subsp. enterica serovar Newport |
| Clostridium perfringens | Serratia marcescens |
| Corynebacterium genitalium | Shigella sonnei |
| Enterobacter aerogenes | Staphylococcus aureus |
| Enterobacter cloacae | Staphylococcus aureus (Cowan) |
| Enterococcus avium | Staphylococcus epidermidis |
| Enterococcus dispar | Staphylococcus saprophyticus |
| Enterococcus durans | Streptococcus anginosus |
| Enterococcus faecalis | Streptococcus bovis |
| Enterococcus faecium | Streptococcus dysgalactiae subsp. disgalactiae |
| Escherichia coli | Streptococcus dysgalactiae equisimilis |
| Escherichia fergusonii | Streptococcus intermedius |
| Gardnerella vaginalis | Streptococcus oralis |
| Klebsiella oxytoca | Streptococcus pneumoniae |
| Klebsiella pneumoniae | Streptococcus pyogenes |
| Lactobacillus acidophilus | Streptococcus salivarius |
| Lactobacillus brevis | Streptococcus sanguinis |
{12}
| Lactobacillus casei | Streptococcus uberis |
| --- | --- |
| Lactobacillus delbreuckii | Ureaplasma urealyticum gDNA |
| Lactobacillus jensenii | Yersinia enterocolitica |
| Yeasts | |
| Candida albicans | Candida parapsilosis |
| Candida glabrata | Candida tropicalis |
| Viruses (gDNA or RNA) | |
| HerpesSimplexVirus-1 gDNA | Norovirus GII RNA |
| HerpesSimplexVirus-2 gDNA | Human Papillomavirus (HPV) gDNA |
| Parasites (gDNA) | |
| Blastocystis hominis gDNA | Trichomonas vaginalis gDNA |
f. Interfering Studies
## Microbial Interference
This study was performed to evaluate potential interference from common microorganisms (associated with vaginal/rectal flora) with the detection of GBS by the GenePOC GBS LB Assay. All testing was completed with three reagent lots and three revogene instruments. For testing, non-target organisms were grouped into 7 pools according to Table 9 shown below. Pools were diluted to obtain a concentration of $10^{5}$ or $10^{6}$ CFU/ml (or cp/ml) of SB and tested in triplicate. Each of the seven pools was a tested in the presence of negative LIM Broth culture with two GBS strains. Both ATCC 13813 and ATCC 12403 were tested at 735 CFU/ml. No interference from non-target organisms was observed, except for GBS strains in the presence with Enterococcus faecalis (ATCC 19433), Enterococcus faecium (ATCC 19434), and Lactobacillus acidophilus (ATCC 4356). Interference with Enterococcus faecalis (ATCC 19433) was observed with both GBS strains at $10^{6}$ CFU/ml. No interference was observed with both GBS strains when Enterococcus faecalis was lowered to $10^{5}$ CFU/ml. For Enterococcus faecium, all three replicates of GBS strain ATCC 12403 were detected at $10^{4}$ CFU/ml. At $10^{5}$ CFU/ml of Lactobacillus acidophilus, GBS strain ATCC 12403 was inhibited. When the concentration of Lactobacillus acidophilus was lowered to $10^{4}$ CFU/ml, all three replicates of GBS strain ATCC 12403 could be detected. All pools, when tested in the absence of GBS, yielded the expected negative result with the assay. No interference with the internal process control was observed when microorganisms were present in high loads ($10^{5}$ or $10^{6}$ CFU/ml of SB). The initial rate of unresolved results obtained during the study was $0.8\%$ (1/133). This unresolved sample replicate was successfully repeated. No indeterminate results were obtained. The potential for high concentrations of $E.$ faecalis, $E.$ faecium, and $L.$ acidophilus to inhibit the detection of GBS is noted as a Limitation in the device labeling. The results of the Microbial Interference Study are acceptable.
{13}
Table 9. List of Seven Pools for the Microbial Interference Study
| Poola | Strains |
| --- | --- |
| 1 | Enterococcus faecalis, ATCC 19433 |
| 2 | Enterococcus faecium, ATCC 19434 |
| 3 | Acinetobacter baumannii, ATCC 19606 |
| | Bacillus subtilis, ATCC 27370 |
| | Bacteroides fragilis, ATCC 25285 |
| | Campylobacter jejuni, ATCC 33560 |
| | Clostridium difficile, ATCC 9689 |
| 4 | Corynebacterium genitalium, ATCC 33031 |
| | Escherichia coli, ATCC 11775 |
| | Gardnerella vaginalis, ATCC 49145 |
| | Klebsiella oxytoca, ATCC 8724 |
| | Lactobacillus acidophilus, ATCC 4356 |
| 5 | Neisseria gonorrhoeae, ATCC 43069 |
| | Peptostreptococcus anaerobius, ATCC 27337 |
| | Propionibacterium acnes, ATCC 11827 |
| | Pseudomonas aeruginosa, ATCC 35554 |
| | Salmonella enterica subsp. enterica serovar Dublin, ATCC 39184 |
| | Serratia marcescens, ATCC 13880 |
| 6 | Shigella sonnei, ATCC 29930 |
| | Staphylococcus aureus, ATCC 33592 |
| | Staphylococcus epidermidis, ATCC 14990 |
| | Streptococcus pyogenes, ATCC 19615 |
| | Yersinia enterocolitica, ATCC 27729 |
| | Candida albicans, ATCC 20735 |
| Poola | gDNA or RNA |
| 7 | Chlamydia trachomatis, VR-878D |
| | Herpes Simplex Virus-1, ATCC VR-539D |
| | Human Papillomavirus (HPV), ATCC VR-3240SD |
| | Trichomonas vaginalis, ATCC 30001 |
| | Mycoplasma genitalium, ATCC BAA-2641S |
${}^{a}$ Even if a pool only contained one microorganism,the term "pool" was kept all along the list.
# Interfering Substances
This study was conducted to evaluate the potential interference of a panel of 31 endogenous and exogenous substances with the GenePOC GBS LB Assay. Potentially interfering substances were distributed into 11 pools for testing. Samples were prepared in triplicate with one of two strains of GBS (ATCC 13813 and ATCC 12403) in negative LIM Broth and diluted to $735\mathrm{CFU / ml}$ in SB in the presence of the potential interferent. LIM Broth served as a negative sample. Interfering substances were added at concentrations that would be expected to be observed among patient specimens. Interference was observed against GBS strain ATCC 12403 in the
{14}
presence of Pool E (1.5 μg/ml loperamide hydrochloride, 45 μg/ml bismuth subsalicylate, and 0.735 μg/ml sennosides). When tested individually, no interference on the detection of GBS strain ATCC 12403 was observed with the substances from Pool E. Interference pools used for the study are highlighted in Table 10 and Table 11 below. Fecal fat was not tested due to lack of supply. One indeterminate result was obtained for a positive external control to yield an initial indeterminate rate of 0.75% (1/133). After re-testing, the control yielded a valid positive result. No unresolved results were obtained during the study. Eight samples were excluded due to: errors in the placement of the PIE cartridge, wrong sample tested, or wrong PIE lot tested. All samples were re-tested according to the study protocol and the re-test results were used to replace the previous results. Results of the Interference Study are acceptable.
Table 10. List of Exogenous Interferents Tested and Grouping into the Six Pools
| Pool | ID | Substances | Commercial Name | Active Ingredient | Absorption, Substances | Concentration in SBT |
| --- | --- | --- | --- | --- | --- | --- |
| A | SI1 | Fungicide | Micatin | Miconazole nitrate cream | Topical, cream | 0.023% w/v |
| | SI2 | Hemorrhoid cooling gel | Preparation H | Phenylephrine HCl | Topical, gel | 0.023% w/v |
| | SI3 | Lubricating gel | K-Y Personal Lubricant | NA | Topical, gel | 0.023% w/v |
| | SI4 | Body powder | Vagisil deodorant powder | NA | Topical, powder | 0.023% w/v |
| | SI5 | Moisturizing lotion | Aveeno moisturizing lotion | NA | Topical, cream | 0.023% w/v |
| B | SI6 | Body oil | Neutrogena Body oil | NA | Topical, Oil | 0.023% v/v |
| | SI7 | Deodorant Spray | Summer's Eve Spray | NA | Topical, spray | 0.023% v/v |
| | SI8 | Enemas | Life BRAND Heavy Mineral Oil USP | Mineral oil | Topical, oil | 0.023% v/v |
| | SI9 | Antimicrobials | Canesten | Clotrimazole | Topical, cream | 0.023% w/v |
| C | SI10 | Enemas | Pentasa | Mesalazine | Oral, powder | 0.495 μg/ml |
| | SI11 | Radiology Oral Compounds | Barium Sulfate | Barium Sulfate | Oral, powder | 0.113 μg/ml |
| | SI12 | Gastritis Medications | Nexium | Esomeprazole Magnesium | Oral, powder | 0.011 μg/ml |
| | SI13 | Antimicrobials | Flagyl | Metronidazole | Oral, powder | 0.016 μg/ml |
{15}
Table 11. List of Endogenous Interferents Tested and Grouping into the Five Pools
| Pool | ID | Substances | Concentration in SBT |
| --- | --- | --- | --- |
| G | SI23 | Whole blood | 0.023% v/v |
| | SI24 | Leukocytes | 15000 cells/ml |
| H | SI25 | Amniotic Fluid | 0.023% v/v |
| | SI26 | Mucous | 0.023% v/v |
| I | SI27 | Seminal Fluid | 0.023% v/v |
| | SI28 | Urine | 0.023% v/v |
| J | SI29 | Feces | 0.023% v/v |
| | SI30 | Meconium | 0.023% v/v |
| K | SI31 | Human DNA | 4.650 ng/ml |
{16}
g. Carry-over/Cross Contamination
Carry-over and cross contamination for the GenePOC GBS LB Assay was assessed with two operators conducting a total of 20 runs of 8 samples in each run. Samples were tested in an alternating pattern with high positive samples and negative samples (within-run and between-run). No carry-over and cross contamination was observed. The overall percent agreement was 100% for positive and negative samples. No unexpected results, including unresolved or indeterminate results, were obtained during the study.
h. Assay cut-off:
The assay cut-off was established and validated by evaluating the performance of the GenePOC GBS LB Assay on the revogene instrument in two separate studies. A total of 258 LIM specimens were tested with the GenePOC GBS LB Assay and with the Reference Culture Method. In the first study, the assay cut-off and thresholds for the GenePOC GBS LB Assay were established by testing 133 fresh specimens previously enriched in LIM Broth. Testing was conducted on one revogene instrument with three unique GBS LB PIE lots. In the second study, testing was performed to validate the cut-offs from an additional 125 frozen LIM specimens. Testing was conducted on one revogene with the same three GBS LB PIE lots utilized for Study 1. The cut-off values for a positive result for both the internal process control and GBS were set at Ct values <45 and an EP Threshold at 125. Using these cut-offs, a sensitivity of 98.3% (58/59) (95% CI: 91.0-99.7%) and a specificity of 99.0% (197/199)(95% CI: 96.4-99.7%) were reported after testing the 258 specimens.
2. Comparison studies:
a. Method comparison with predicate device:
Not Applicable. Please refer to 3a below.
b. Matrix comparison:
Not Applicable
3. Clinical studies:
a. Clinical Sensitivity:
GenePOC conducted a prospective multi-center trial at 4 geographically diverse clinical trial sites (two Canadian and two US clinical sites). Vaginal/rectal swab specimens were collected from pregnant women at 35-37 weeks of gestation for whom GBS diagnostic procedures were indicated and ordered. Vaginal/rectal specimen swabs were collected and cultured according to established guidelines; swabs were placed in LIM Broth and incubated for 18-24 hrs at 35-37°C in ambient
{17}
air or 5% CO₂. Residual de-identified LIM Broth cultures were used for testing with the GenePOC GBS LB Assay. A total of 839 vaginal/rectal swab specimens initially met the study criteria. Sixty-eight (68) specimens were shown to be non-compliant for the following reasons:
- Transport and storage times exceeded study protocol requirements (1)
- Samples did not have all test results as required by the study protocol (4)
- Reference Method and/or GenePOC GBS LB Assay not performed according to the clinical trial protocol (63)
For the Reference Culture Method, LIM Broth cultures were subcultured to a non-selective blood agar, and all colonies with a characteristic appearance suggestive of GBS were screened to confirm the presence of GBS using established laboratory methods: gram stain, catalase, latex agglutination, and CAMP test. The CAMP test was only used at one site in addition to the other biochemical tests listed. If GBS was not presumptively identified after incubation for 18-24 hrs on sheep blood agar plates, the plates were further incubated overnight and then re-inspected to identify suspect colonies. Originally, the Reference Culture Method revealed 156 positive specimens. To improve the recovery of GBS by the Reference Culture Method, a supplemental study was conducted with frozen LIM Broth culture aliquots (in glycerol) from all negative samples by the Reference Culture Method (n=615). Aliquots were subcultured onto non-selective agar and again examined for the presence of GBS colonies. GBS identification was confirmed by biochemical testing according to the clinical reference method protocol. Any positive GBS samples identified by this method were combined with the results of the first culture. No GenePOC GBS LB Assay results were changed for the supplemental study. For six isolates in the supplemental study, results were excluded due to inconclusive culture, the frozen LIM Broth tube was not found, or the frozen LIM Broth tube was not stored with glycerol; therefore, the original Reference Culture Method results for the six samples were included in the analysis instead of excluding these samples from the final analysis. With the addition of the secondary study, the Reference Culture Method was considered to be a Composite Reference Method. A total of 14 additional reference positive samples were identified by the supplemental study and added to previous results for a total of 170 reference positive results [156 (original) + 14 (supplemental study)].
The GenePOC GBS LB Assay was performed according to the package insert. A 15 μl aliquot of enriched LIM Broth culture was transferred to a sample buffer tube (SBT), and a 150 μl was loaded into each PIE. The assay was run on the revogene instrument. Residual SBT with sample was stored at 2-8°C for repeat testing, if needed. One hundred thirty-seven (137) GenePOC GBS LB runs out of one hundred thirty-nine (139) were valid (98.6% success rate). The overall initial unresolved and indeterminate rates were 0.74% (6/812) and 1.48% (12/812), respectively. The final non-reportable result rate was 0.25% for the GenePOC GBS LB Assay. Results of the GenePOC GBS LB Assay were compared to a Composite Reference Culture Method as shown in Table 12.
18
{18}
Table 12. Clinical Performance Data for the GenePOC GBS LB Assays. Composite Reference Culture Method (all sites)
| GBS LB Assay | Composite Reference Method | | |
| --- | --- | --- | --- |
| | Positive | Negative | Total |
| Positive | 163 | 27b | 190 |
| Negative | 7a | 574 | 581 |
| Total | 170 | 601 | 771 |
| Sensitivity: 95.9% (95%CI: 91.7 - 98.0 %) Specificity: 95.5% (95%CI: 93.5 – 96.9 %) | | | |
${}^{a}$ 5 out of 7 false negative GenePOC GBS LB results were tested on a FDA-cleared molecular device and yielded negative results. An FDA-cleared GBS NAAT was used to test all samples at only one site.
${}^{b}$ 10 out 27 false positive GenePOC GBS LB results were tested on a FDA-cleared molecular device and yielded positive results. An FDA-cleared GBS NAAT was used to test all samples at only one site.
b. Clinical Specificity:
See above
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
In the investigational study with the GenePOC GBS LB Assay on the revogene (K170558), the overall GBS prevalence rate as determined by the Composite Reference Method (defined above) was $22.0\%$ (170/771).
# N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
# O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.