# Coveris Cervical Cage System (K170550)

_Camber Spine Technologies · ODP · Jul 13, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170550

## Device Facts

- **Applicant:** Camber Spine Technologies
- **Product Code:** [ODP](/productcode/ODP.md)
- **Decision Date:** Jul 13, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K170550 · Jul 13, 2017 | Coveris Cervical Cage System | Camber Spine Technologies | Clinical literature review | A clinical literature review was conducted to evaluate the risks and benefits of using allogenic bone graft and multi-level (two contiguous levels) application of the cervical cage system, supporting the expansion of the device's indications for use. | Literature review; Allogenic bone graft; Multi-level fusion |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Clinical literature review; Systematic literature review | Patients undergoing cervical intervertebral body fusion | Not applicable for this study | Risks and benefits associated with allogenic bone graft and multi-level use |

## Indications for Use

When used as a cervical intervertebral fusion device, the Coveris devices are indicated for use at one or two contiguous levels in the cervical spine, from C2-T1, in skeletally mature patients who have had six weeks of non-operative treatment for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with autogenous bone graft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft and with supplemental fixation (such as anterior cervical plating systems, or posterior systems) systems cleared for use in the cervical spine.

## Device Story

Intervertebral body fusion device; maintains disc space distraction; used in cervical spine (C2-T1). Device features hollow center for bone graft material; superior/inferior surface ridges for endplate grip/expulsion prevention. Used in conjunction with supplemental spinal fixation (e.g., anterior cervical plating). Implanted by surgeons in clinical settings. Benefits include stabilization of spinal segments to facilitate fusion in patients with DDD.

## Clinical Evidence

No clinical trials performed. Evidence consists of a comprehensive clinical literature review investigating risks and benefits of allograft use and two-level application, supplemented by bench testing (static compression and static torsion per ASTM F2077).

## Technological Characteristics

Materials: PEEK (ASTM F2026) with embedded tantalum x-ray markers (ASTM F560). Design: Intervertebral cage with hollow center and surface ridges. Connectivity: None. Energy source: None. Sterilization: Not specified.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Coveris Cervical Cage System ([K133529](/device/K133529.md))
- Medtronic Anatomic PEEK PTC Cervical Fusion System ([K160528](/device/K160528.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Camber Spine Technologies % Margeaux Rogers Senior Associate, Regulatory Affairs MCRA, LLC 1050 K Street NW, Suite 1000 Washington, District of Columbia 20001 July 13, 2017

Re: K170550

Trade/Device Name: Coveris Cervical Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP Dated: June 22, 2017 Received: June 23, 2017

Dear Ms. Rogers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

Image /page/0/Picture/11 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K170550

Device Name Coveris Cervical Cage System

#### Indications for Use (Describe)

When used as a cervical intervertebral fusion devices are indicated for use at one or two contiguous levels in the cervical spine, from C2-T1, in skeletally mature patients who have had six weeks of non-operative treatment for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with autogenous bone graft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft and with supplemental fixation (such as anterior cervical plating systems) systems) systems cleared for use in the cervical spine.

| Type of Use (Select one or both, as applicable)                                                                 |                                                                                                                |
|-----------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CER 801 Subpart D)             </span> | <span> <span style="font-size: 16px;">□</span> Over-The-Counter Use (21 CER 801 Subpart C)             </span> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

| Device Trade Name: | Coveris Cervical Cage System                                                                                                                                                  |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:      | Camber Spine Technologies<br>418 E. Lancaster Ave.<br>Wayne, PA 19087                                                                                                         |
| Contact:           | Mr. Joseph Nyahay<br>Product Development Engineer<br>Camber Spine Technologies<br>418 E. Lancaster Ave.<br>Wayne, PA 19087                                                    |
| Prepared by:       | Ms. Margeaux Rogers<br>Musculoskeletal Clinical Regulatory Advisers, LLC<br>1050 K Street NW, Suite 1000<br>Washington, DC 20001<br>Phone: (202) 552-5800<br>mrogers@mcra.com |
| Date Prepared:     | February 23, 2017                                                                                                                                                             |
| Classification:    | 21 CFR §888.3080, Intervertebral body fusion device                                                                                                                           |
| Class:             | II                                                                                                                                                                            |
| Product Codes:     | ODP                                                                                                                                                                           |

## Indications For Use:

When used as a cervical intervertebral fusion device, the Coveris devices are indicated for use at one or two contiguous levels in the cervical spine, from C2-T1, in skeletally mature patients who have had six weeks of non-operative treatment for the treatment of degenerative disc disease (DDD) with up to Grade 1 spondylolisthesis. DDD is defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with autogenous bone graft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft and with supplemental fixation (such as anterior cervical plating systems, or posterior systems) systems cleared for use in the cervical spine.

{4}------------------------------------------------

#### Device Description:

The Camber Spine Coveris Cervical Cage series of intervertebral body fusion devices are used to maintain disc space distraction in skeletally mature adults requiring intervertebral body fusion. They are designed to be used in conjunction with supplemental spinal fixation instrumentation. The series is comprised of cages of various fixed heights and shapes for placement in the cervical spine. There are different cages designed for specific regions of the spine and approaches to the spine. Each cage has a hollow center to allow placement of graft material inside of the cage. Ridges on the superior and inferior surfaces of the device help to grip the endplates and prevent expulsion.

The Coveris Cervical Cage of intervertebral body fusion devices are made from the PEEK radiolucent material with embedded tantalum x-ray markers as specified in ASTM F2026 and ASTM F560, respectively.

#### Predicate Device:

The primary predicate is the Coveris Cervical Cage System (K133529). The Medtronic Anatomic PEEK PTC Cervical Fusion System (K160528) is an additional predicate. The subject Coveris Cervical Cage System is substantially equivalent to the previously cleared Coveris Cervical Cage System (K133529) with respect to design, function and materials and predicate Medtronic Anatomic PEEK PTC Cervical Fusion System (K160528) with respect to indications, design, function, and materials.

#### Substantial Equivalence:

The Coveris Cervical Cage System and predicate Medtronic Anatomic PEEK PTC Cervical Fusion System (K160528) are similar in design, material, and indicated use, and are both cleared devices. They do differ in that the Medtronic Anatomic PEEK PTC Cervical Fusion System is indicated for use with allogenic bone graft at one or two contiguous levels. A comprehensive, clinical literature review has been conducted to investigate the risks and benefits associated with using allogenic bone graft with the Coveris Cervical Cage System and mutli-level use. The clinical literature suggests that there is a positive benefit associated with allograft and use at two contiguous levels with minimal risk.

## Nonclinical Performance Testing:

Comprehensive, clinical literature review data have been provided to investigate the risks and benefits associated with using allogenic bone graft with the Coveris Cervical Cage System and use at two contiguous levels. Static compression and static torsion testing were performed per ASTM F2077.

## Conclusion:

The Coveris Cervical Cage System has been modified to expand the indications to permit use with allograft and use at two contiguous levels, C2-T1. The 510(k) demonstrates substantial equivalence to predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K170550](https://fda-staging.innolitics.com/device/K170550)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).