HyperVit™ Vitrectomy Probe (23, 25 gauge) The HyperVit™ Vitrectomy Probe is designed for use with the Constellation® Vision System and is intended to be used to remove vitreous and dissect tissue in the eye. HyperVit™ Vitrectomy Probe (27 gauge) The HyperVit™ Vitrectomy Probe is designed for use with the Constellation® Vision System and is intended for use as follows: - Vitreous aspiration & cutting - Membrane cutting & aspiration - Lens removal - Dissect tissue in the eye
Device Story
HyperVit Vitrectomy Probe is a sterile, single-use, pneumatically actuated accessory for the Constellation Vision System. It features a modified cutter tip with two ports, enabling two cuts per operating cycle. Input consists of pneumatic pulses from the console; output is mechanical guillotine-style cutting and aspiration of vitreous and ocular tissue. Operated by surgeons in an OR setting during vitreoretinal or cataract procedures. The device interfaces with the console via RFID tags. The dual-port design allows for 20,000 cuts per minute at 10,000 cycles per minute, improving cutting efficiency compared to single-port predecessors. The healthcare provider uses the probe to perform precise tissue removal, facilitating surgical access and reducing potential for retinal traction or damage. Benefits include enhanced cutting performance and tissue dissection capabilities.
Clinical Evidence
Bench testing only. Performance evaluations included cut quality verification, aspiration flow characterization, heat generation testing (max temperature increase 0.8°C at 150% max speed), and metal particulate analysis (no particulates detected after 20 minutes of continuous operation).
Technological Characteristics
Single-use, pneumatically actuated vitrectomy probe. Materials: stainless steel. Dimensions: 23, 25, and 27 gauge. Connectivity: RFID-enabled interface with Constellation Vision System. Sterilization: EtO. Features dual-port cutter tip design for guillotine-style cutting. Operating speed: 10,000 cycles/min (20,000 cuts/min).
Indications for Use
Indicated for patients undergoing vitreoretinal or cataract surgery requiring vitreous removal, tissue dissection, membrane cutting/aspiration, or lens removal.
Regulatory Classification
Identification
A vitreous aspiration and cutting instrument is an electrically powered device, which may use ultrasound, intended to remove vitreous matter from the vitreous cavity or remove a crystalline lens.
Special Controls
*Classification.* Class II (special controls). The device, when it is phacofragmentation unit replacement tubing, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of the human profile in triplicate.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 22, 2017
Alcon Research, Ltd. Karen Mudd. Ph.D. Director. Regulatory Affairs 20511 Lake Forest Drive Lake Forest, CA 92630
Re: K170520
Trade/Device Name: Hypervit Vitrectomy Probe, 23 Ga, Hypervit Vitrectomy Probe, 25+, Hypervit Vitrectomy Probe, 27+ Regulation Number: 21 CFR 886.4150 Regulation Name: Vitreous Aspiration and Cutting Instrument Regulatory Class: Class II Product Code: MLZ Dated: February 16, 2017 Received: February 22, 2017
Dear Dr. Mudd:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170520
Device Name HyperVit™ Vitrectomy Probe
Indications for Use (Describe)
HyperVit™ Vitrectomy Probe (23, 25 gauge)
The HyperVit™ Vitrectomy Probe is designed for use with the Constellation® Vision System and is intended to be used to remove vitreous and dissect tissue in the eye.
HyperVit™ Vitrectomy Probe (27 gauge) The HyperVit™ Vitrectomy Probe is designed for use with the Constellation® Vision System and is intended for use as follows:
- · Vitreous aspiration & cutting
- Membrane cutting & aspiration
- · Lens removal
- · Dissect tissue in the eye
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> <path d="M3,3 L13,13 M3,13 L13,3" stroke="black" stroke-width="2"></path> </svg> |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> </svg> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# (510)k SUMMARY
This 510(k) summary document has been prepared in accordance with Section 21 CFR 807.92.
#### I. Submitter of the 510(k)
| Company: | Alcon Laboratories, Inc. |
|-------------------------|--------------------------------|
| | 6201 South Freeway |
| | Fort Worth, TX 76134-2099, USA |
| Primary Contact Person: | Karen Mudd, Ph.D. |
| Phone: | 949-505-7269 |
| Fax: | 949-505-6237 |
| Email: | karen.mudd@alcon.com |
| Back up Contact Person: | Amy Tezel, Ph.D. |
| Phone: | 817-363-1513 |
| Fax: | 817-551-4630 |
| Email: | amy.tezel@alcon.com |
| Date Prepared: | 16 February 2017 |
#### II. Devices Subject to this 510(k)
| Trade Names: | HyperVit™ Vitrectomy Probe |
|----------------------|---------------------------------------------------------------|
| Common Name: | Vitrectomy Device |
| Classification Name: | Vitreous Aspiration & Cutting Instrument (21 CFR<br>886.4150) |
| Product Code: | MLZ |
#### III. Predicate Devices
- Enhanced UltraVit® Probe K093305 .
- 27+ UltraVit® Probe, K110951 .
{4}------------------------------------------------
#### Device Description IV.
The HyperVit" Vitrectomy Probe is an accessory to the Constellation® Vision System (K141065, K101285 and K063583). The HyperVit™ Vitrectomy Probe is a sterile, single-use pneumatically actuated device which supports aspiration and guillotine style cutting functions for the purpose of vitreous cutting and aspiration, as well as tissue dissection, in the eye during vitreoretinal or cataract surgery. The HyperVit™ Vitrectomy Probe is comprised of a pre-assembled hand piece and tubing. The tubing assembly interfaces with a Constellation® Vision System console by means of RFID tags at the proximal end of the probe. The HyperVit" Vitrectomy Probe is offered in three different sizes: 23, 25 and 27 gauge.
The HyperVit" Vitrectomy Probe is a modification of the existing UltraVit® Vitrectomy Probes currently used with the Constellation® Vision System. A port opening has been added to the cutter tip to create a second cutting edge. Thus, two cuts are made during a single operating cycle, compared to one cut for the existing UltraVit® Vitrectomy Probes. With an operating actuation speed of 10,000 cycles per minutes, the HyperVit™ Vitrectomy Probe can achieve 20,000 cuts per minute.
#### Indications for Use V.
# HyperVit™ Vitrectomy Probe (23, 25 gauge)
The HyperVit" Vitrectomy Probe is designed for use with the Constellation® Vision System and is intended to be used to remove vitreous and dissect tissue in the eye.
## HyperVit™ Vitrectomy Probe (27 gauge)
The HyperVit™ Vitrectomy Probe is designed for use with the Constellation® Vision System and is intended for use as follows:
- Vitreous aspiration & cutting
- Membrane cutting & aspiration
- Lens removal
- Dissect tissue in the eye
#### Comparison of Technological Characteristics with the Predicate Device VI.
The technological characteristics of the HyperVit™ Vitrectomy Probe are substantially equivalent to those of the predicate device, as shown in the table below.
{5}------------------------------------------------
| Characteristic | Predicate Devices<br>K093305, Enhanced UltraVit Probe<br>(23 and 25 gauge)<br>K110951, 27+ UltraVit Probe (27 gauge) | HyperVit Vitrectomy Probe<br>(23, 25 and 27 gauge) |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation | 886.4150 | Same |
| Intended use /<br>Indications for use | K093305, Enhanced UltraVit<br>Probe (23 and 25 gauge):<br>The ALCON® Enhanced UltraVit®<br>Probe is designed for use with the<br>Constellation® Vision System and is<br>intended to be used to remove<br>vitreous and dissect tissue in the eye.<br>K110951, 27+ UltraVit Probe<br>(27 gauge):<br>• Vitreous aspiration & cutting<br>• Membrane cutting & aspiration<br>• Remove a crystalline lens<br>• Dissect tissue in the eye<br>(The 27 UltraVit Probe is intended to<br>be used with the Alcon Constellation<br>Vision System) | 23 and 25 gauge:<br>The HyperVit™ Vitrectomy<br>Probe is designed for use with<br>the Constellation® Vision System<br>and is intended to be used to<br>remove vitreous and dissect<br>tissue in the eye.<br>27 gauge:<br>The HyperVit™ Vitrectomy<br>Probe is designed for use with<br>the Constellation® Vision System<br>and is intended for use as<br>follows:<br>• Vitreous aspiration & cutting<br>• Membrane cutting & aspiration<br>• Lens removal<br>• Dissect tissue in the eye |
| Power source for cutter<br>activation | Forward and return pneumatic<br>pressure pulses | Same |
| Probe needle gauge | 23, 25, 27 | Same |
| Cutting action format | Guillotine | Same |
| Frequency of cutter<br>activation | Up to 15,000 pulses per minute | Up to 10,000 pulses per minute |
| Cutting port format | Side port, one | Side port, two |
| Cutting rate | Up to 15,000 cuts per minute | Up to 20,000 cuts per minute |
| Aspiration channel | Through the cutter tubing cannula | Same |
| Patient contact material<br>(direct or through fluid<br>path) | Stainless steel | Same |
| Overall length of probe<br>(nominal) | 3.3" | Same |
| Single-use / Reusable | Single use only | Same |
| Shelf-life | Two years | Same |
| Sterile product<br>packaging | Tyvek pouch | Same |
| Sterility method | EtO-sterilized | Same |
| Gauge identifier | Color coded and molded in the<br>product | Same |
| Connector | RFID (connectors for pneumatic lines<br>and luer connectors for aspiration<br>line) | Same |
{6}------------------------------------------------
#### VII. Performance Data
Testing and validations have demonstrated that the functional requirements and specifications have been met. Therefore, the HyperVit" Vitrectomy Probe is as safe and effective as the predicate device. An overview of key performance tests conducted is provided below:
## Mechanical Testing
- . Cut Quality: Testing was conducted on the vitreous cutting surfaces in order to verify that the forward and return cutting surfaces met the same acceptance criteria for cut quality and cut reliability as the predicate device. Results show that the probe met the acceptance criteria and, therefore, have the same cut quality as the predicate device.
- Aspiration Flow: The aspiration flow characteristics of the HyperVit™ Vitrectomy Probe . at the default vacuum settings were characterized and compared with the predicate device. The flow characteristics of the HyperVit™ Vitrectomy Probe at the default vacuum setting were similar to those of the predicate device as a result of the identical dimensions of the fluid path (e.g., cutter and aspiration tubing). Therefore, the aspiration flow is equivalent to the predicate device.
- Heat Generation: Testing was conducted to measure the temperature increase at the . vitreous cutter tip operated at 150% of HyperVit™ Vitrectomy Probe maximum speed. Testing showed that the maximum temperature increase was 0.8℃ throughout 30 minutes of operation; this is less than the maximum allowed temperature increase at the probe tip of no more than 10°C, based on literature review (Ref: Wu et al., Photochemical damage of the retina. Survey of Ophthalmology, 51(5), 461–481, 2006).
- Metal Particulates: Testing was performed to evaluate whether metal particulates were . generated by the HyperVit™ Vitrectomy Probe during operation. The probe was actuated at the maximum 10.000 cycles per minute in a water container. After running the test continuously for 20 minutes, the filter in the container was microscopy-examined for metal particulate and no particulates were detected, thus meeting the acceptance criteria.
#### Safety and Effectiveness Information VIII.
A review of the indications for use and technical characteristics demonstrates that the HyperVit™ Vitrectomy Probe is substantially equivalent to the predicate devices and is safe and effective for use for the stated indications for use.
{7}------------------------------------------------
#### Conclusion IX.
The HyperVit™ Vitrectomy Probe was found to be substantially equivalent to the predicate device.
The HyperVit™ Vitrectomy Probe shares identical indications for use, similar design features and functional features with the predicate device, and thus is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.