K170515 · Primus Gloves Pvt Limited · KGO · Jun 9, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170515
Device Name
Sterile Nitrile Surgical Gloves, Powder-free
Applicant
Primus Gloves Pvt Limited
Product Code
KGO · General, Plastic Surgery
Decision Date
Jun 9, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
The powder-free Nitrile Surgical Gloves , is a sterile and single use devise made of synthetic rubber intended to be worn by operation room personnel to protect a surgical wound from contamination. The gloves do not contain lubricating or dusting powder.
Device Story
Sterile, powder-free, anatomic surgical glove made of synthetic nitrile latex compound. Used by operating room personnel to protect surgical wounds from contamination. Device acts as a physical barrier. Performance verified against ASTM D3577-09 standards for dimensions, tensile strength, and elongation. Biocompatibility confirmed via ISO 10993-10 testing (non-irritant, non-sensitizer). Sterility achieved to SAL 10^-6 per ISO 11737-2.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via ASTM D5151-06 (pinholes), ASTM D3577-09 (dimensions, physical properties), ASTM D6124-06 (powder residue), and ISO 10993-10 (biocompatibility). All tests passed.
Technological Characteristics
Material: Synthetic nitrile latex compound. Form factor: Anatomic, textured, sterile, powder-free. Standards: ASTM D3577-09, ASTM D5151-06, ASTM D412-06, ASTM D6124-06, ISO 10993-10, ISO 11737-2. Sterilization: SAL 10^-6.
Indications for Use
Indicated for use by operating room personnel to protect surgical wounds from contamination. Single-use, sterile, powder-free, synthetic rubber gloves.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
POWDER-FREE NITRILE SURGICAL GLOVES, WHITE COLOUR (K140989)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a triple helix-like structure extending from the heads.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 9, 2017
Primus Gloves Pvt. Limited Jose Paul M Manager Oa & RA Plot No 14a. Cochin Special Economic Zone. Kakkanad Cochin, 682037 IN
Re: K170515
Trade/Device Name: Sterile Nitrile Surgical Gloves, Powder-free Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: Class I Product Code: KGO Dated: May 4, 2017 Received: May 9, 2017
Dear Jose Paul M:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Lori A. Wiggins -S6
Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K170515
Device Name
Sterile Nitrile Surgical Gloves Powder-free
Indications for Use (Describe)
The powder-free Nitrile Surgical Gloves , is a sterile and single use devise made of synthetic rubber intended to be worn by operation room personnel to protect a surgical wound from contamination. The gloves do not contain lubricating or dusting powder.
Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K170515
#### 1.0 SUBMITTER
3.0
| 1.1 | Company Name | : | PRIMUS GLOVES PRIVATE LIMITED |
|-----|----------------|---|------------------------------------------------------------------------------------------|
| 1.2 | Address | : | Plot No: 14-A, Cochin Special Economic Zone,<br>Kakkanad, Cochin, Kerala, India - 682037 |
| 1.3 | Telephone | : | + 91 484 2413063 |
| 1.4 | Fax | : | +91 484 2413089 |
| 1.5 | Email | : | josepaul@primusgloves.com |
| 1.6 | Contact Person | : | Mr. JOSE PAUL M<br>MANAGER – QA & RA |
#### 2.0 OFFICIAL CORRESPONDENT
| 2.1 | Company Name | : | PRIMUS GLOVES PRIVATE LIMITED |
|------------------|----------------|---|------------------------------------------------------------------------------------------|
| 2.2 | Address | : | Plot No: 14-A, Cochin Special Economic Zone.<br>Kakkanad, Cochin, Kerala, India - 682037 |
| 2.3 | Telephone | : | + 91 484 2413063 |
| 2.4 | Fax | : | +91 484 2413089 |
| 2.5 | Email | : | josepaul@primusgloves.com |
| 2.6 | Contact Person | : | Mr. JOSE PAUL M<br>MANAGER – QA & RA |
| Preparation date | | : | June 2, 2017 |
{4}------------------------------------------------
| 4.0 | Name of the device | | : |
|-----|--------------------|-------------------|----------------------------|
| | 4.1 | Device Name | : STERILE NITRILE SURGICAL |
| | | | GLOVES, POWDER-FREE, |
| | | Trade Name | : PRIMUS NITRILE GLOVES |
| | 4.2 | Common Name | : SURGICAL GLOVES |
| | 4.3 | Classification | : SURGEON"S GLOVES |
| | 4.4 | Class | : CLASS I |
| | 4.5 | Product code | : KGO |
| | 4.6 | Regulation Number | : 21 CFR 878.4460 |
#### 5.0 Identification of the legally marketed predicate device
| 5.1 | Device Name | : | POWDER-FREE NITRILE SURGICAL<br>GLOVES, WHITE COLOUR |
|-----|----------------------|---|--------------------------------------------------------------------------------------------------|
| 5.2 | 510(k) Number | : | k140989 |
| 5.3 | Company | : | Hebei Hongsen Plastics Technology Co Ltd<br>853 Dorchester Ln, Unitb, New Millford,<br>NJ 07646. |
| 5.4 | Device Description | : | Sterile Nitrile Surgical Gloves |
| 5.5 | Classification | : | SURGEON'S GLOVES |
| 5.6 | Class | : | CLASS I |
| 5.7 | Product code | : | KGO |
| 5.8 | Classification Panel | : | General Hospital |
{5}------------------------------------------------
#### 6.0 Description of the Device
The subject device is a surgical nitrile qlove made of synthetic nitrile latex compound. It is sterile, powder-free. The device is anatomic. The device meets ASTM D 3577-09: Standard specification for Rubber Surgical gloves for Medical Application.
The device is for over-the counter single use.
#### Indications for use 7.0
The powder-free nitrile surgical gloves, is a sterile and single use device made of synthetic rubber intended to be worn by operation room personnel to protect a surgical wound from contamination. The gloves do not contain lubricating or dusting powder.
#### 8.0 Summary of performance data
There is no difference in technological characteristics compared to the predicate device. Gloves are made from Nitrile latex compound, sterile, Powder-free. The gloves have the same technological characteristics compared to ASTM or equivalent standards as given below.
{6}------------------------------------------------
| Characteristics | Test | Test standard | Sampling<br>plan/Inspe<br>ction<br>level/AQL | Sterile Nitrile<br>Surgical<br>gloves powder-<br>free, - PRIMUS | RESULT |
|---------------------------|-----------------------------------|----------------------------------------|------------------------------------------------------------------------|---------------------------------------------------------------------------------|--------|
| Freedom from<br>Pin holes | FDA 1000 ml<br>water leak<br>test | ASTM D5151-06<br>(reap 2011) | ISO 2859-1<br>/ G1/AQL<br>2.5 | PASS | PASS |
| Dimensions | Length | ASTM D 3577-09 | ISO 2859-1<br>/ S2/AQL<br>4.0 | > 265 mm | PASS |
| | Width | ASTM D 3577-09 | ISO 2859-1<br>/ S2/AQL<br>4.0 | 76±6 mm to<br>114±6 mm (<br>sizes 6 to 9) | PASS |
| | Thickness | ASTM D 3577-09 | ISO 2859-1<br>/ S2/AQL<br>4.0 | > 0.10 mm<br>(Cuff, palm &<br>finger) | PASS |
| Physical | Before aging | | ISO 2859-1 | Tensile strength | PASS |
| properties | | ASTM D 3577-09<br>and ASTM D412-06 | / S2/AQL<br>4.0 | > 17 Mpa<br>Ultimate<br>Elongation : ><br>650 %<br>Stress at 500%<br>Elongation | |
| | After<br>Accelerated<br>aging | ASTM D 3577-09<br>and ASTM D412-06 | ISO 2859-1<br>/ S2/AQL<br>4.0 | Tensile strength :<br>> 12 Mpa<br>Ultimate<br>Elongation : ><br>490 % | PASS |
| Powder-free<br>residue | Powder-free<br>residue | ASTM D 3577-09<br>and ASTM<br>D6124-06 | N=5 | Less than 2 mg<br>per glove | PASS |
| Biocompatibility | Primary skin<br>irritation | ISO 10993-10 | Under the conditions of the<br>study the device is not an<br>irritant | | PASS |
| | Skin/Dermal<br>Sensitization | ISO 10993-10 | Under the conditions of the<br>study the device is not a<br>sensitizer | | PASS |
| Sterility | | ISO 11737-02 | Sterile | | Pass |
{7}------------------------------------------------
{8}------------------------------------------------
### 9.0 Summary of the technological characteristics of device compared to the legally marketed predicate device
| Characteristics | PREDICATE – 510(K) :<br>K140989 | SUBJECT DEVICE : | Comparison |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | Hebei Hongsen<br>Plastics Technology Co.<br>Ltd853 Dorchester Ln,<br>Unitb, New Millford,NJ<br>07646. | PRIMUS GLOVES PRIVATE<br>LIMITED, Plot No: 14-A, CSEZ,<br>Kakkanad, Cochin, Kerala,<br>India -682037 | |
| Product Name | POWDER- FREE<br>NITRILE SURGICAL<br>GLOVES, WHITE<br>COLOUR | STERILE NITRILE SURGICAL<br>GLOVES, POWDER-FREE, | Similar |
| Intended Use | The powder-free nitrile<br>surgical glove is a sterile<br>and single use device<br>made of synthetic rubber<br>intended to be worn by<br>operation room personnel<br>to protect a surgical<br>wound from | The powder-free nitrile surgical<br>glove is a sterile and single use<br>device made of synthetic<br>rubber intended to be worn by<br>operation room personnel to<br>protect a surgical wound from<br>contamination. The gloves do<br>not contain lubricating or | Similar |
| | contamination. The<br>gloves do not contain<br>lubricating or dusting<br>powder. | dusting powder. | |
| Indication for use | The powder-free nitrile<br>surgical glove is a sterile<br>and single use device<br>made of synthetic rubber<br>intended to be worn by<br>operation room personnel<br>to protect a surgical<br>wound from<br>contamination. The<br>gloves do not contain<br>lubricating or dusting<br>powder. | The powder-free nitrile surgical<br>glove is a sterile and single use<br>device made of synthetic<br>rubber intended to be worn by<br>operation room personnel to<br>protect a surgical wound from<br>contamination. The gloves do<br>not contain lubricating or<br>dusting powder. | Similar |
| Description | Sterile Nitrile surgical<br>gloves is made of nitrile<br>The textured gloves are<br>provided in sizes<br>6.0,6.5,7.0,7.5,8.0,8.5&9.<br>0. | Sterile Nitrile surgical gloves is<br>made of nitrile The textured<br>gloves are provided in sizes<br>6.0,6.5,7.0,7.5,8.0,8.5&9.0. | Similar |
| Presentation | Sterile gloves are | Sterile gloves are provided in | Similar |
| | provided in pouches | pouches | |
| Material | Nitrile synthetic latex | Nitrile synthetic latex | Similar |
| Single Use | Yes | Yes | Similar |
| Hand Specific | Yes | Yes | Similar |
| Dimensions | Meets ASTM D 3577-09 | Overall length min 265 mm,<br>width varies from 76 mm for<br>6.0 size to 114 mm for 9.0 size,<br>thickness in finger ,Palm & cuff<br>has a minimum 0.10 mm | Similar |
| Tensile Strength | Meets ASTM D 3577-09 | Tensile strength 17 Mpa min for<br>before aging and 12 Mpa min<br>for after aging<br>Aging done at 70 ±2 deg C for<br>166±2 hrs or 100±2deg C for<br>22±0.3 hrs | Similar |
| Ultimate Elongation | Meets ASTM D 3577-09 | Ultimate elongation 650 % min<br>for before aging and 450 % min<br>for after aging. Aging done at<br>70 ±2 deg C for 166±2 hrs or<br>100±2deg C for 22±0.3 hrs | Similar |
| Stress at 500%<br>Elongation | Meets ASTM D 3577-09 | Stress at 500% elongation for<br>before agening is 7.0% min | Similar |
| | Meets ASTM D 5151 -06 | Meets ASTM D 5151 -06 | Similar |
| Freedom from<br>pinholes | and ASTM D 3577-09 | (2011) and ASTM D 3577-09 | |
| Residual Powder | Meets ASTM D 6124-06 | Less than 2 mg per glove | Similar |
| Biocompatibility<br>Tests<br>ISO 10993-10 | Non-irritant -Primary Skin<br>Irritation In Rabbits | Under the conditions of the<br>study the device is not an<br>irritant | Similar |
| | Non-sensitizer - skin<br>Sensitization in Guinea<br>pigs | Under the conditions of the<br>study the device is not a<br>sensitizer | Similar |
| | | | |
| Labeling | Powder-free,<br>Nitrile Surgical glove<br>sterile<br>Single use only<br>Anatomic<br>Manufactured for<br>Lot No<br>Intended use<br>Quantity<br>Country of origin | *Powder-free,<br>* Nitrile Surgical glove<br>glove<br>* sterile<br>*Single use only<br>*Anatomic<br>*Manufactured for<br>*Lot No<br>*Intended use<br>*Quantity<br>*Country of origin | Similar |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| Sterility | SAL of 10^6 achieved<br>according to ISO 11737-2 | SAL of 10^6 achieved<br>according to ISO 11737-2 | Similar |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
### 10.0 CONCLUSION
The physical performance of the subject device is substantially equivalent to predicate K140989 and will perform according to the glove performance and biocompatibility standards referenced. Based on the intended use, physical properties and technological characteristics, the subject device is as safe, as effective and performs as well as the legally marketed predicate device
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.