Central Monitoring System Software is intended to conduct centralized monitoring of adult, pediatrics and neonatal patient's vital sign information from compatible bedside monitors. The software collects, stores, displays and alarms the information provided on the bedside monitor. Rate(HR), The monitoring parameters include Electrocardiogram(ECG), Heart Respiration(RESP), Pulse Oxygen Saturation(SpO2), Pulse Rate(PR), Non-invasive Blood (NIBP), Pressure Invasive -Blood Pressure(IBP), Impedance Cardiograph(ICG), TEMP(Temperature), Carbon dioxide (CO2), Anesthetic Gas (AG), Fetal Heart Rate (FHR), Uterine contraction (TOCO) and Fetal Movement (FM) etc. It is intended to be used in the hospital or medical institutions, and it is not intended for home use.
Device Story
M6000C Central Monitoring System is software-only device for centralized patient monitoring; collects, stores, displays, and alarms vital sign data received via network from compatible bedside monitors. Inputs include ECG/HR, RESP, SpO2, PR, NIBP, IBP, ICG, TEMP, CO2, AG, FHR, TOCO, and FM. System displays waveforms and parameters for multiple patients simultaneously; triggers visual/audible alarms when data exceeds set thresholds. Used in hospitals/medical institutions by clinicians (doctors/nurses) to facilitate patient management. Output allows clinicians to view real-time patient status, review historical records, and enter notes. Benefits include centralized oversight of multiple patients, enabling timely clinical intervention for out-of-limit vital signs.
Clinical Evidence
No clinical study is included in this submission. Evidence consists of bench testing, including performance testing for compatibility with various bedside monitors and data latency verification, and software verification and validation testing per FDA guidance.
Technological Characteristics
Software-only device; operates on networked hardware. Connectivity via wired and wireless networks. Complies with IEC 62304 (software lifecycle), EN ISO 14971 (risk management), and IEC 62366 (usability). Features include multi-patient display, alarm management, data archiving, and electronic patient record interfacing.
Indications for Use
Indicated for centralized monitoring of vital signs in adult, pediatric, and neonatal patients in hospital or medical institutional settings. Parameters include ECG, HR, RESP, SpO2, PR, NIBP, IBP, ICG, TEMP, CO2, AG, FHR, TOCO, and FM. Not for home use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing right, stacked on top of each other, representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 24, 2017
Guangdong Biolight Meditech Co., Ltd. % Diana Hong General Manager Mid-link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 CN
Re: K170514
Trade/Device Name: M6000C Central Monitoring System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX Dated: February 14, 2017 Received: February 21, 2017
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
M.A. Hillemann
for
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170514
Device Name M6000C Central Monitoring System
### Indications for Use (Describe)
Central Monitoring System Software is intended to conduct centralized monitoring of adult, pediatrics and neonatal patient's vital sign information from compatible bedside monitors. The software collects, stores, displays and alarms the information provided on the bedside monitoring parameters include Electrocardiogram(ECG), Heart Rate(HR), Respiration(RESP), Pulse Oxygen Saturation(SpO2), Pulse Rate(PR), Non-invasive Blood Pressure (NIBP), Invasive Blood Pressure(IBP), Impedance Cardiograph(ICG), TEMP(Temperature), Carbon dioxide (CO2), Anesthetic Gas (AG), Fetal Heart Rate (FHR), Uterine contraction (TOCO) and Fetal Movement (FM) etc. It is intended to be used in the hospital or medical institutions, and it is not intended for home use.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Tab #6 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K170514
- 1. Date of Preparation: 05/04/2017
- 2. Sponsor Identification Guangdong Biolight Meditech Co., Ltd. No.2 Innovation First Road, Technology Innovation Coast, Hi-Tech Zone, Zhuhai, Guangdong, 519085, P.R. China
Establishment Registration Number: 3007305624 Contact Person: Jin Liang Position: R&D Director Tel: +86-756-3399967 Fax: +86-756-3399989 Email: liang_j@blt.com.cn
- 3. Designated Submission Correspondent Ms. Diana Hong (Primary Contact Person) Ms. Cindy Wang (Alternative Contact Person) Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
- 4. Identification of Proposed Device Trade Name: Central Monitoring System Common Name: Central Monitoring System Model(s): M6000C Regulatory Information Classification Name: Perinatal monitoring system and accessories Classification: II; Product Code: MHX; Regulation Number: 21 CFR 870.1025; Review Panel: Cardiovascular;
{4}------------------------------------------------
## Intended Use Statement:
Central Monitoring System Software is intended to conduct centralized monitoring of adult, pediatrics and neonatal patient's vital sign information from compatible bedside monitors. The software collects, stores, displays and alarms the information provided on the bedside monitor. Rate(HR), The monitoring parameters include Electrocardiogram(ECG), Heart Respiration(RESP), Pulse Oxygen Saturation(SpO2), Pulse Rate(PR), Non-invasive Blood (NIBP), Pressure Invasive -Blood Pressure(IBP), Impedance Cardiograph(ICG), TEMP(Temperature), Carbon dioxide (CO2), Anesthetic Gas (AG), Fetal Heart Rate (FHR), Uterine contraction (TOCO) and Fetal Movement (FM) etc. It is intended to be used in the hospital or medical institutions, and it is not intended for home use.
Device Description:
M6000C central monitoring system software, the risk management object, can central monitor significant vital sign parameters of multi-patients, including ECG/HR, RESP, SpO2, PULSE, NIBP, IBP, ICG, TEMP, CO2, AG, FHR, TOCO and FM. It is connected by network with bedside units and receives data from bedside units. Then, the data are displayed on the screen or recorded or printed as per needs. When the monitored data exceed a set value, the central monitoring system software will start alarm system and gives out alarm to remind doctors and nurses for attention.
- ર. Identification of Predicate Device and Reference Device
Predicate Device 510(k) Number: K120727 Product Name: Central monitoring system Model Name: MFM-CMS
Reference Device 510(k) Number: K153580 Product Name: Central monitoring system Model Name: M6000C
#### 6. Non-Clinical Test Conclusion
Non-clinical tests were conducted to verify that the proposed device met all design specifications. The Central Monitoring System also complies with voluntary standards for medical devices.
The test results demonstrated that the proposed device complies with the following standards:
{5}------------------------------------------------
- EN ISO 14971: 2012 Medical Device Risk Management A
- A IEC 62366: 2007 Medical Devices - Application of Usability Engineering to Medical Devices
The following data were provided to support the substantial equivalence determination:
- ≫ Performance testing was conducted to verify that the Central Monitoring System Software is compatible with various bedside monitors.
- > Performance testing was conducted to verify that the data latency rates were acceptable for the clinical environment.
- ≫ Software verification and validation testing as recommended in FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (May 11, 2005)
Please note that the subject is a software only product, EMC and Electrical Safety Evaluation are not required.
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
| ITEM | Proposed Device<br>Central Monitoring System<br>Truscope CNS | Predicate Device |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product<br>Code | MHX | MHX |
| Regulation<br>No. | 870.1025 | 870.1025 |
| Class | II | II |
| Intended<br>Use | Central Monitoring System<br>Software is intended to conduct<br>centralized monitoring of adult,<br>pediatrics and neonatal patient's<br>vital sign information from<br>compatible bedside monitors. The<br>software collects, stores, displays<br>and alarms the information<br>provided on the bedside monitor.<br>The monitoring parameters include<br>Electrocardiogram(ECG), Heart<br>Rate(HR), Respiration(RESP), | MFM-CMS provided centralized<br>monitoring and critical care<br>management for patients<br>monitored by EDAN bedside<br>monitors, From MFM-CMS,<br>clinicians can gain access to<br>patient information from patients<br>on the Network. MFM-CMS<br>display waveforms , parameters<br>and alarm status of EDAN bedside<br>monitors for up to 32 patients on a<br>single screen or up to 64 patients |
| | | |
| | Pulse Oxygen Saturation(SpO2),<br>Pulse Rate(PR), Non-invasive<br>Blood Pressure (NIBP), Invasive<br>Blood Pressure(IBP), Impedance<br>Cardiograph(ICG),<br>TEMP(Temperature), Carbon<br>dioxide (CO2), Anesthetic Gas<br>(AG), Fetal Heart Rate (FHR),<br>Uterine contraction (TOCO) and<br>Fetal Movement (FM) etc. It is<br>intended to be used in the hospital<br>or medical institutions, and it is not<br>intended for home use. | using two screens |
| Network | Wired and Wireless | Wired and wireless |
| Display | Patient demography data, and<br>notes.<br>Providing the means to display<br>multiple beds simultaneously. | Waveforms and notes<br>Providing the means to display<br>multiple beds simultaneously |
| Print | Print Patient's monitoring records | Print Patient's monitoring records |
| Archive | Providing the ability to archive<br>files to a secondary or tertiary<br>storage medium (i.e. optical disk).<br>Providing automatic archiving of<br>the data. | Providing the ability to archive<br>files to a secondary or tertiary<br>storage medium (i.e. optical disk).<br>Providing automatic archiving of<br>the data. |
| Alarm | Visual alerts of out-of-limit data | Visual alerts of out-of-limit data |
| Electronic<br>patient<br>record | Easy interfacing with any IT<br>patient record system for data<br>acquisition, viewing and storage of<br>electronic patient record. | Easy interfacing with any IT<br>patient record system for data<br>acquisition, viewing and storage of<br>electronic patient record. |
| Notes | Providing the user the ability to<br>enter comments and specific data. | Providing the user the ability to<br>enter comments and specific data. |
| Complied<br>Standard | IEC 62304 | IEC 62304 |
### Substantially Equivalent (SE) Comparison 8.
Table 1 Comparison of Technology Characteristics
{6}------------------------------------------------
### 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.