← Product Code [MMK](/productcode/MMK) · K170513

# MAUSER PG II Sharps Container (K170513)

_Mauzer USA, LLC · MMK · Nov 1, 2017 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170513

## Device Facts

- **Applicant:** Mauzer USA, LLC
- **Product Code:** [MMK](/productcode/MMK.md)
- **Decision Date:** Nov 1, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

MAUSER PG II Sharps Container is intended to be used in healthcare facilities including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, veterinarian offices and other waste generators for the safe collection and disposal of hazardous sharps. MAUSER PG II Sharps Container is designed for use on sites with no unsupervised patient access. The single-use, non-sterile device is intended for over-the-counter use. The manufacturer's pedal bin frame is required if the device is used on medication carts. The pedal bin frame cart is constructed of stainless steel and comes fully preassembled. The cart offers a hands free collection option and fully integrates with all sizes of container and lid styles. The full containers are collected and transported by a licensed health care waste contractor for disposal in accordance with federal and state regulations.

## Device Story

Single-use, non-sterile sharps container; available in 30L, 50L, and 60L sizes; consists of rectangular body and lid (grip or click-with-porthole styles). Used in healthcare facilities (nursing stations, labs, ORs, etc.) for safe collection/disposal of hazardous sharps (needles, blades, pipets). Optional stainless steel pedal frame cart enables hands-free collection. Device provides puncture-resistant, leak-resistant containment for sharps; prevents puncture wounds, cuts, or tears in skin/mucous membranes. Healthcare providers manually insert sharps through container top; fill line visualized through aperture. Full containers transported by licensed waste contractors. Benefits include safe containment of biohazardous sharps, reducing risk of accidental injury to staff.

## Clinical Evidence

No clinical data submitted. Bench testing only: Puncture resistance (ASTM F2132-01), drop test (49 CFR 178.603), leak resistance (49 CFR 173.197), stack test (49 CFR 178.606), vibration test (49 CFR 178.608), stability (ISO 23907), and handle strength (ISO 23907). All tests passed.

## Technological Characteristics

Materials: Polypropylene. Construction: Injection molded. Energy source: None (mechanical). Form factor: Rectangular containers (30L, 50L, 60L) with grip or click-with-porthole lids. Accessories: Stainless steel pedal frame cart. Standards: ASTM F2132-01 (puncture resistance), ISO 23907 (stability/handle strength), 49 CFR 178/173 (DOT transport standards).

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- Oak Ridge Products LLC. Sharps Containers ([K130281](/device/K130281.md))
- Oak Ridge Products LLC. Multi-purpose Sharps Containers ([K141759](/device/K141759.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Please wait...

If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.

You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.

For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.

Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image shows the word "MAUSER" in large, bold, blue letters. Above the "M" is a small, red, downward-pointing triangle. The font is sans-serif and the letters are spaced closely together. The overall impression is one of strength and precision.

## 510K SUMMARY

#### 5.1 SUBMITTER

MAUSER USA, LLC. 2 Tower Center Drive, 20th floor East Brunswick, NJ 08816 Telephone: (732) 353-7015 Fax: (732) 353-7030

Contact person: Tatiana Smoleeva Date prepared: November 1, 2017

#### 5.2 DEVICE NAME AND CLASSIFICATION

Common name: Classification Name: Classification Regulations: Product Code: Trade/Proprietary Name:

Sharps container Hyperdermic single lumen needle 21 CFR 880.5570, Class II MMK MAUSER PG II Sharps Container

#### 5.3 PREDICATE DEVICE

MAUSER PG II Sharps Container is substantially equivalent in indications and design principle to the following predicate devices

- Oak Ridge Products LLC. Sharps Containers (K130281) O
- Oak Ridge Products LLC. Multi-purpose Sharps Containers (K141759) O

#### DEVICE DESCRIPTION 5.4

MAUSER PG II Sharps Container is a single-use device, which consist of rectangular body and available in three sizes 30 liter, 50 liter and 60 liter. Each container has two parts, the body and the lid, which assemble together to form a unit. Two lid styles are available: grip, and click with port hole.

MAUSER PG II Sharps Container is designed for use on sites with no unsupervised patient access.

Manufacturer offers a stainless steel pedal frame cart that offers a hands free collection process and integrates with all sizes of container and lid type.

The container is designed for safe collection and disposal of sharps, i.e. items that can cause puncture wounds, cuts or tears in skin or mucous membranes, including, but not limited to

MAUSER USA. LLC. 2 Tower Center drive, 20th floor East Brunswick, NJ 08816 USA

Managing Director
Hans-Peter Schaefer, CEO
Bjorn Kreiter, CFO

al Court : Cologne

info.us@mausergroup.com www.mausergroup.com

CONFIDENTIAL

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the word "MAUSER" in blue, with a red triangle above the "M". The font is bold and sans-serif. The word is likely a logo or brand name. The overall impression is clean and professional.

hypodermic, surgical, suture, and IV needles; Pasteur pipets, lancets, razors, scalpels, and other blades.

The following MAUSER PG II Sharps Container models and accessories are described in this submission:

| Model or part                                                      | Product Description and Closure Type                                                                       | Overall size<br>(mm)           | Aperture (mm)                                       |
|--------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|--------------------------------|-----------------------------------------------------|
| 060V4080RO<br>060V4080GE<br>060V4080Z0<br>060V4080WI               | MAUSER PG II Sharps Container, 60<br>liter, Grip lid (Red, Yellow, Black,<br>White w/blue lid)             | (L) 376 x (W)<br>309 x (H) 640 | (L) 375 x (W) 305                                   |
| 050V4080RO<br>050V4080GE<br>050V4080ZW<br>050V4080WI               | MAUSER PG II Sharps Container, 50<br>liter, Grip lid,<br>(Red, Yellow, Black, White w/blue lid)            | (L) 376 x (W)<br>309 x (H) 534 | (L) 375 x (W) 305                                   |
| 030V4080RO<br>030V4080GE<br>030V4080ZW<br>030V4080WI               | MAUSER PG II Sharps Container, 30<br>liter, Grip lid<br>(Red, Yellow, Black, White w/blue lid)             | (L) 376 x (W)<br>309 x (H) 312 | (L) 375 x (W) 305                                   |
| 060V4080RO<br>060V4080GE<br>060V4080Z0<br>060V4080WI               | MAUSER PG II Sharps Container, 60<br>liter, Click lid w/porthole<br>(Red, Yellow, Black, White w/blue lid) | (L) 376 x (W)<br>309 x (H) 640 | (L) 375 x (W) 305(Major)<br>152 (porthole diameter) |
| 050V4080RO<br>050V4080GE<br>050V4080ZW<br>050V4080WI               | MAUSER PG II Sharps Container, 50<br>liter, Click lid w/porthole<br>(Red, Yellow, Black, White w/blue lid) | (L) 376 x (W)<br>309 x (H) 534 | (L) 376 x (W) 309(Major)<br>152 (porthole diameter) |
| 030V4080RO<br>030V4080GE<br>030V4080ZW<br>030V4080WI               | MAUSER PG II Sharps Container, 30<br>liter, Click lid w/porthole<br>(Red, Yellow, Black, White w/blue lid) | (L) 376 x (W)<br>309 x (H) 312 | (L) 376 x (W) 309(Major)<br>152 (porthole diameter) |
| 710374 (Red)<br>710371(Black)<br>710258 (Yellow)<br>710303 (Blue)  | Grip lid                                                                                                   | (L) 395.5 x (W)<br>329         | NA                                                  |
| 700260 (Red)<br>700156 (Black)<br>700256 (Yellow)<br>700901 (Blue) | Click lid w/porthole                                                                                       | (L) 395.5 x (W)<br>329         | 152 (porthole diameter)                             |
| N/A                                                                | Steel pedal bin frame                                                                                      | N/A                            | N/A                                                 |

### INDICATIONS FOR USE 5.5

 MAUSER USA, LLC. 2 Tower Center drive, 20th floor East Brunswick, NJ 08816 USA Managing Director Hans-Peter Schaefer, CEO Bjorn Kreiter, CFO

 Registered Seat 50321 Brüehl
 Local Court : Cologne Registration HBB 58469

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the word "MAUSER" in a bold, sans-serif font. The word is blue, and there is a red triangle above the "M". The triangle is pointing downwards. The overall design is simple and modern.

MAUSER PG II Sharps Container is intended to be used in healthcare facilities including nursing stations, medication carts, laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, veterinarian offices and other waste generators for the safe collection and disposal of hazardous sharps.

MAUSER PG II Sharps Container is designed for use on sites with no unsupervised patient access.

The single-use, non-sterile device is intended for over-the-counter use. The manufacturer's pedal bin frame is required if the device is used on medication carts. The pedal bin frame cart is constructed of stainless steel and comes fully preassembled. The cart offers a hands free collection option and fully integrates with all sizes of container and lid styles.

The full containers are collected and transported by a licensed health care waste contractor for disposal in accordance with federal and state regulations.

#### 5.6 EQUIVALENCE TO MARKETED DEVICE

MAUSER USA, LLC. Submits the following information in this premarket notification to demonstrate, for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indication and design principle to the following legally marketed predicate device:

|                       | Subject Device                                                                                                                          | Predicate Device                                                                                                                                                                  | Predicate Device                                                                                       |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
|                       | MAUSER PG II Sharps<br>Container                                                                                                        | Oak Ridge Products Sharps<br>Containers (K130281)                                                                                                                                 | Oak Ridge Products Multi-<br>purpose Sharps Containers<br>(K141759)                                    |
|                       | MAUSER USA, LLC.<br>2 Tower Center Drive, 20th<br>floor<br>East Brunswick, NJ 08816<br>USA                                              | Oak Ridge Products L.L.C<br>4612 Century Court<br>McHenry, Illinois, 60050<br>USA                                                                                                 | Oak Ridge Products L.L.C<br>4612 Century Court<br>McHenry, Illinois, 60050<br>USA                      |
| Target<br>population  | Healthcare professionals                                                                                                                | Healthcare professionals                                                                                                                                                          | Healthcare professionals                                                                               |
| Indication for<br>use | MAUSER PG II Sharps<br>Container is intended to be<br>used in healthcare facilities<br>including nursing stations,<br>medication carts, | Oak Ridge Products Sharps<br>containers are single-use,<br>disposable, non-sterile<br>containers intended to be used<br>for healthcare purposes for<br>safe disposal of hazardous | Oak Ridge sharps containers<br>are intended to be used for<br>the safe disposal of<br>hazardous sharps |

- Oak Ridge Products LLC. Sharps Containers (K130281) O
- Oak Ridge Products LLC. Multi-purpose Sharps Containers (K141759) O

MAUSER USA. LLC 2 Tower Center drive, 20th floo East Brunswick, NJ 08816 US/

Managing Director Hans-Peter Schaefer, CEO Registered Seat 50321 Brüe Local Court : Cologne

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the word "MAUSER" in large, bold, blue letters. Above the "M" is a red triangle pointing downwards. The font is sans-serif and the overall design is clean and modern.

| MAUSER USA LLC, 2 Tower Center Drive 20th floor, East Brunswick, New Jersey 08816 USA |                                                                                                                                                                                           |                                                                          |                                                              |
|---------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------|
|                                                                                       | laboratories, dental offices, emergency rooms, surgical rooms, treatment rooms, veterinarian offices and other waste generators for the safe collection and disposal of hazardous sharps. | sharps such as hypodermic needles, syringes, lancets, and blood needles. |                                                              |
| Performance                                                                           | Single use                                                                                                                                                                                | Single use                                                               | Single use                                                   |
| Manufacturing                                                                         | Injection molded                                                                                                                                                                          | Injection molded                                                         | Injection molded                                             |
| Materials of construction                                                             | Polypropylene                                                                                                                                                                             | Polypropylene                                                            | Polypropylene                                                |
| Color                                                                                 | Red, yellow, black, white                                                                                                                                                                 | Red, natural                                                             | Red                                                          |
| Sharps elimination                                                                    | Sharps inserted through the top                                                                                                                                                           | Sharps inserted through the top                                          | Sharps inserted through the top                              |
| Method of visualizing the fill line                                                   | Large apertures opening provides clear and unrestricted view of fill line                                                                                                                 | A translucent container component provides view of fill line             | A translucent container component provides view of fill line |
| Mounting Accessory                                                                    | Available for use on medication carts                                                                                                                                                     | Available and necessary for 1 quart and 5.4 quart models                 | Unknown                                                      |
| Impact resistance (Drop test)                                                         | Yes                                                                                                                                                                                       | Yes                                                                      | Yes                                                          |
| Vibration test                                                                        | Yes                                                                                                                                                                                       | Unknown                                                                  | Unknown                                                      |
| Stack test                                                                            | Yes                                                                                                                                                                                       | Unknown                                                                  | Unknown                                                      |
| Puncture resistance                                                                   | Yes                                                                                                                                                                                       | Yes                                                                      | Yes                                                          |
| Leak Resistance                                                                       | Yes                                                                                                                                                                                       | Yes                                                                      | Yes                                                          |

#### 5.7 PERFORMANCE DATA

No FDA performance standard for MAUSER PG II Sharps Container have been established. The following tests were performed to ensure that the performance of the subject device meets MAUSER USA, LLC. requirements.

|   | Test                             | Test Method                       | Results |
|---|----------------------------------|-----------------------------------|---------|
| 1 | Puncture resistance              | ASTM F2132-01² (Reapproved 2008)  | Pass    |
| 2 | Drop test (Impact<br>resistance) | §178.603¹                         | Pass    |
| 3 | Leak resistance test             | §173.197¹ Regulated medical waste | Pass    |
| 4 | Stack Test                       | §178.606¹ (c ) (1)                | Pass    |

MAUSER USA, LLC.
2 Tower Center drive, 20th floor
East Brunswick, NJ 08816 USA

Managing Director Hans-Peter Schaefer, CEO Bjorn Kreiter, CFO

 Registered Seat 50321 Brüehl
 Local Court : Cologne Registration HBB 58469

{5}------------------------------------------------

| 5 | Vibration Test      | §178.608¹  | Pass |
|---|---------------------|------------|------|
| 6 | Stability           | ISO 23907³ | Pass |
| 7 | Strength of handles | ISO 23907³ | Pass |

1United States Department of Transportation Code of Federal Regulations (CFR) Title 49, Transportation, Parts 100-199

2 Standard Specification for Puncture Resistance of Materials Used in Containers for Discarded Medical Needles and Other Sharps

3 Sharps Injury Protection - Requirements and test methods - Sharps Containers (First Edition 2012-09-01)

Puncture resistance – ASTM F2132-01 (Reapproved 2008) Standard Specification for Puncture Resistance of Materials used in Containers for Discarded Medical Needles and Other Sharps.

Puncture resistance was measured by taking various sections of container and subjecting them to penetration with a 21 gage by 1 inch needle. The force required to puncture the container was measured. The puncture resistance of each section of the container was greater than 15 N and the passing criterion was met.

Drop test - \$178.603 (United States Department of Transportation Code of Federal Regulations (CFR) Title 49, Transportation, Parts 100-199)

Drop test was conducted by filling six containers to maximum authorized gross mass (60 liter -30 kg; 50 liter -28 kg; 30 liter -15 kg) with simulated medical waste. The containers were then conditioned at -18° C for 24 hours and dropped in specific orientations from a height of no less than 1.2 meters. No release of contents was observed and the passing criterion was met.

Leak resistance test - §173.197 Regulated medical waste (United States Department of Transportation Code of Federal Regulations (CFR) Title 49, Transportation, Parts 100-199)

Leak resistance was conducted by filling container to rated liquid capacity with water. The containers were then closed as per closing instructions and positioned upside down on cardboard paper to show any evidence of leaking. Contained upside down for 60 minutes while being observed for leakage. No leakage was observed and the passing criterion was met.

Stack test - §178.606 (c) (United States Department of Transportation Code of Federal Regulations (CFR) Title 49, Transportation, Parts 100-199)

Containers were filled to the required test capacity and closed as for shipment. Each container was subjected to a force applied to the top surface equivalent to the total weight of identical packages which might be stacked on it during transport. The minimum height of the stack, including the test sample, was 3 meters. Upon completion of the stack test, containers were observed for deformation. The containers did not deteriorate or release contents, and the passing criteria were met.

MAUSER USA. LLC. 2 Tower Center drive, 20th floor East Brunswick, NJ 08816 USA

Managing Director Hans-Peter Schaefer, CEO Registered Seat 50321 Brileh Local Court : Cologne Registration HBB 58469

info.us@mausergroup.com www.mausergroup.com

CONFIDENTIAL

{6}------------------------------------------------

Image /page/6/Picture/2 description: The image shows the word "MAUSER" in large, bold, blue letters. Above the "M" is a red triangle pointing downwards. The font is sans-serif and the overall design is simple and modern. It is likely a logo for the Mauser company.

MAUSER USA LLC, 2 Tower Center Drive 20th floor, East Brunswick, New Jersey 08816 USA Vibration test - §178.608 (United States Department of Transportation Code of Federal Regulations (CFR) Title 49, Transportation, Parts 100-199)

Containers were filled with simulated medical waste, closed as for shipment and placed on vibration table. The containers were then vibrated sing a vibration with a 1 in. vertical peak-to-peak displacement, for 60 minutes at a frequency that causes them to be raised enough such that a piece of material 1.6-mm thick can be passed between the container and the vibration platform. The containers did not rupture, leak or deteriorate, and the passing criteria were met.

Container Stability - ISO 23907 (Sharps Injury Protection - Requirements and test methods - Sharps Containers, First Edition 2012-09-01)

Container stability was evaluated by filling containers with a material of density 0.20+/-0.01 kg/l and placed on an incline with a minimum angle of 15°. Testing was conducted for each side of the container. The container did not slide or topple and the passing criteria were met.

Strength of Handles - ISO 23907 (Sharps Injury Protection - Requirements and test methods - Sharps Containers, First Edition 2012-09-01)

Handle strength was measured by filling container to 150% of its authorized gross mass and suspending it from handles for an hour at 23+/-5°C. The containers remained intact and handles did not rupture, tear. crack, or separate from the container and the passing criteria were met.

Clinical data were not submitted in this premarket notification

#### 5.8 CONCLUSION

MAUSER PG II Sharps Container is substantially equivalent to Oak Ridge Products Sharps Container and Oak Ridge Products Multi-purpose Sharps Container (K130281 and K141759).

MAUSER USA. LLC 2 Tower Center drive, 20th floo East Brunswick, NJ 08816 USA Managing Director

Registered Seat 50321 Brüe Local Court : Cologne

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 1, 2017

Mauzer Usa LLC Tatiana Smoleeva Sr. Associate, Technology & Regulatory Affairs 2 Tower Center Drive, 20th Floor East Brunswick, New Jersey 08816

Re: K170513

Trade/Device Name: Mauser PG II Sharps Container Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: MMK Dated: September 25, 2017 Received: September 28, 2017

Dear Tatiana Smoleeva:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.qov

{8}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

---

**Source:** [https://fda-staging.innolitics.com/device/K170513](https://fda-staging.innolitics.com/device/K170513)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).