Model#'s 3543A, 4343A, 2929A and A843B; Solid State X-ray Imager (Flat Panel/Digital Imager) are indicated for use in generating radiographic images of human anatomy. They are intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding dental, fluoroscopic, angiographic, and mammographic applications. As prescribed by a licensed physician.
Device Story
Solid state X-ray imagers (flat panel detectors) used in medical X-ray imaging systems; consist of GOS scintillator screen and thin-film transistors (TFTs). Scintillator converts X-rays to visible light; TFTs convert light to electrical charge; system performs analog-to-digital conversion to produce digital images. Operated by trained personnel in medical facilities for single-image diagnosis. Output is a DICOM image viewed by healthcare providers to assist in clinical decision-making for general radiographic procedures. Benefits include replacement of traditional film/screen systems with digital imaging capabilities.
Clinical Evidence
No clinical trials performed. Substantial equivalence supported by bench testing and comparative image quality analysis using phantom/clinical images against the Atlaim Atal 9 (K132151) predicate.
Technological Characteristics
Flat panel digital X-ray detector; GOS scintillator; thin-film transistor (TFT) array; 16-bit A/D conversion; pixel pitch 140 µm. Connectivity: Gigabit Ethernet (wired) and IEEE 802.11ac (wireless). Power: AC line or rechargeable lithium battery. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for use in generating radiographic images of human anatomy for general-purpose diagnostic procedures. Excludes dental, fluoroscopic, angiographic, and mammographic applications. For use as prescribed by a licensed physician.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a symbol above them that resembles a caduceus without the snake.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Shanghai PZ Medical Technology Co., Ltd. % Jeff Moeller Vice President OrganizeUS. Inc. 8317 Lakeside Drive Downers Grove, Illinois 60516
July 7, 2017
Re: K170480
Trade/Device Name: Models 3543A, 4343A, 2929A and A843B Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: MOB Dated: April 30, 2017 Received: May 5, 2017
Dear Jeff Moeller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Michael D. O'Hara
For
Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170480
Device Name Models 3543A, 4343A, 2929A and A843B
Indications for Use (Describe)
Model#'s 3543A, 4343A, 2929A and A843B; Solid State X-ray Imager (Flat Panel/Digital Imager) are indicated for use in generating radiographic images of human anatomy. They are intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding dental, fluoroscopic, and mammographic applications. As prescribed by a licensed physician
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
�rganizeUS, Inc.
8317 Lakeside Drive Downers Grove, IL 60516
# 510(k) Summary
(Revised 06/30/17)
| Owners name: | Shanghai PZ Medical Technology Co., Ltd.<br>5F, Building 13 No.2 Suide Road<br>200331 Shanghai |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person / firm: OrganizeUS, Inc. | 8317 Lakeside Drive<br>Downers Grove, IL 60516<br>847-337-3004<br>Jeff Moeller, Vice President |
| Date: | March 17, 2017 |
| Trade name: | Models 3543A, 4343A, 2929A and A843B |
| Common name: | Solid State X-Ray Imager (Flat Panel/Digital Imager) |
| Classification name: | Stationary X-Ray system, 892.1680, code MQB |
| Predicate devices: | Atlaim's model ATAL 9 - K152151<br>Stationary X-Ray system, 892.1680, code MQB |
| | IRAY model 1417 - K161730<br>Stationary X-Ray system, 892.1680, code MQB |
| | CareRay model Care view 1500C & 1500L - K153058<br>Stationary X-Ray system, 892.1680, code MQB |
| Description: | Our device is used in medical x-ray imaging systems. The product can only be<br>used by trained personnel of medical facilities. The product is only used for<br>single image diagnosis applications. The flat panel detector consists of GOS<br>scintillator screen and thin-film transistors. The scintillator screen converts the<br>x-rays into visible light. Thin-film transistors convert the visible light to an<br>electrical charge. The flat panel detector can then obtain a digital image by<br>analog to digital conversion and associated circuits. |
{4}------------------------------------------------
Intended use: Model#'s 3543A, 4343A, 2929A and A843B; Solid State X-ray Imager (Flat Panel/Digital Imager) are indicated for use in generating radiographic images of human anatomy. They are intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding dental, fluoroscopic, angiographic, and mammographic applications. As prescribed by a licensed physician.
Refer to the table below (2 pages) Technological:
| Characteristic | Atlaim ATAL 9 K152151 | PZMedical 4343A,3543A,2929A,A843B |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Indicated for use in general<br>radiographic images of human<br>anatomy. It is intended to replace<br>radiographic film/screen systems in all<br>general-purpose diagnostic procedures,<br>excluding fluoroscopic, angiographic,<br>and mammographic applications. | Model#'s 3543A, 4343A, 2929A and A843B;<br>Solid State X-ray Imager (Flat Panel/Digital<br>Imager) are indicated for use in generating<br>radiographic images of human anatomy. They are<br>intended to replace radiographic film/screen<br>systems in all general-purpose diagnostic<br>procedures, excluding dental, fluoroscopic,<br>angiographic, and mammographic applications.<br>As prescribed by a licensed physician. |
| Configuration | This submission is for the Digital<br>Panel and Software only, no generator<br>or stand provided. | This submission is for the Digital Panel and<br>Software only, no generator or stand provided. |
| Digital Panel | Atlaim ATAL 9 | PZMedical 4343A,3543A,2929A,A843B |
| Scintillator | GOS/CsI | GOS/CsI |
| Pixel Pitch | 139 µm | 140 µm |
| Limiting<br>Resolution | Over 3 lp/mm | 3.6 lp/mm |
| DOE | 26% at 2 lp/mm (CsI) Not specified for<br>GOS | 32% at 2 lp/mm (CsI) 15% at 2 lp/mm (GOS) |
| MTF | 42% at 2 lp/mm (CsI) Not specified for<br>GOS | 33% at 2 lp/mm (CsI) 24% at 2 lp/mm (GOS) |
| A/D Conversion | 16 bit | 16 bit |
| Active Area | 17 x 17 inch | 4343A: 430.08 x 430.08 mm or 16.9 x 16.9 inch<br>3543A: 350.00 x 427.28 mm or 13.8 x 16.8 inch<br>2929A: 286.72 x 286.72 mm or 11.3 x 11.3 inch<br>A843B: 1075.2 x 430.08 mm or 42.3 x 16.9 inch |
| Dimensions | 460(W)×461(L)×15(H) | 4343A: 460 x 460 x 15 mm<br>3543A: 383 x 460 x 15 mm<br>2929A: 316 x 316 x 15 mm<br>A843B: 1120 x 465 x 20 mm |
| Weights | 2.9kg (2.4kg w/o Battery) (up to 10<br>hours of battery life) | 4343A: 4.3 kg (wireless, w/ batteries) 4.3 kg<br>(wired, w/o batteries)<br>3543A: 3.3 kg (wireless, w/ batteries) 2.9 kg<br>(wired, w/o batteries)<br>2929A: 1.8 kg (wireless, w/ batteries) 1.5 kg<br>(wired, w/o batteries)<br>A843B: 11.6 kg (wireless, w/ batteries) 10.9 kg<br>(wired, w/o batteries) |
| Pixels | 3072 x 3072 (9.4 Million) | 3543A: 2500 x 3052 (7.6 Million)<br>4343A: 3072 x 3072 (9.4 Million)<br>2929A: 2048 x 2048 (4.2 Million)<br>A843B: 7680 x 3072 (23.6 Million) |
| Interface | Wired: Gigabit Ethernet (1000Base-T)<br>Wireless: IEEE802.11ac, backward<br>compatible | Wired: Gigabit Ethernet (1000Base-T)<br>Wireless: IEEE802.11ac, backward compatible |
| Power Source | AC Line and/or Rechargeable Lithium<br>Battery (10 hr run time) | AC Line and/or Rechargeable Lithium Battery |
| Electrical safety<br>and EMC | Electrical Safety per IEC 60601-1 and<br>EMC per<br>IEC 60601-1-2 as well as IEEE<br>802.11ac.<br>Meets FCC requirements. | Electrical Safety per IEC 60601-1 and EMC per<br>IEC 60601-1-2 as well as IEEE 802.11ac.<br>Meets FCC requirements. |
{5}------------------------------------------------
## SOFTWARE RELATED
| Characteristic | Atlaim ATAL 9 K152151 | PZMedical 4343A,3543A,2929A,A843B |
|-----------------------------------------------------------------------------------------------|-----------------------|-----------------------------------|
| Software | Outputs a DICOM image | Outputs a DICOM image |
| SW revision | K152151 | PZDR V2.0.1 |
| DICOM | Yes | Yes |
| DICOM Image Transmission | Yes | Yes |
| Image Acquisition Interface | Yes | Yes |
| Patient Body Part Selection | Yes | Yes |
| Image Processing | Yes | Yes |
| Basic Image Editing and<br>Marking (pan, zooming,<br>window/level adjusting, text<br>marking) | Yes | Yes |
| Image Browsing | Yes | Yes |
| Patient Registration and<br>Management | Yes | Yes |
Non-Clinical data:
Testing results have shown that our devices Models 3543A, 4343A, 2929A and A843B, Solid State X-Ray Imager (Flat Panel/Digital Imager) are substantially equivalent to cleared predicate devices.
Atlaim Corporation, Atal 9, K132151 Careray, Care View 1500C / Care View 1500L, K153058 IRay, Wireless Digital Flat Panel Detertor, K161730
Specifically in relation too; DQE, MTF, Quantum limited performance, Effects of aliasing, Sensitivity linearity, Lag, Change in detection sensitivity, Dose requirement and reciprocity changes, Stability of the device characteristics with time, Uniformity of device characteristics, NPS, spatial resolution, Image Acquisition time and Black level.
{6}------------------------------------------------
### Clinical data:
We have included comparison images using our devices Models 3543A, 4343A, 2929A and A843B and a Atlaim Atal 9 cleared device (K132151) as direct comparison of the image quality of our devices in order to support substantial equivalence. We believe our image quality is substantially equivalent to cleared predicate devices.
### Conclusions:
In accordance with 21 CFR, Part 807 and supported by the information provided in this premarket notification (K170480). Shanghai PZ Medical Technology Co., Ltd. 5F, Building 13 No.2 Suide Road 200331 Shanghai concludes that our devices Models 3543A, 2929A and A843B Solid State X-Ray Imager (Flat Panel/Digital Imager) is substantially equivalent to predicate devices in regards to safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.