The MiroCam® Capsule Endoscope System is intended for visualization of the small bowel mucosa. - It may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy. - It may be used in the visualization and monitoring of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy. - It may be used in the visualization and monitoring of lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy. It may be used as a tool in the detection of abnormalities of the small bowel and this device is indicated for adults only. The Suspected GI Bleeding Indicator (SGIB) is intended to mark frames of the video suspected of containing blood or red areas.
Device Story
MiroCam Capsule Endoscope System captures small bowel images via ingestible capsule; utilizes CMOS imaging sensor and 12 white LEDs. Capsule transmits image data via Human Body Communication (HBC) to sensor pads affixed to patient abdomen; pads connect to wearable receiver (MR2000) via data cables. Receiver stores data for up to 12 hours; features LCD for real-time viewing and status monitoring. Post-procedure, receiver connects to PC via USB to upload data to MiroView U 4.0 software. Gastroenterologists review video data to identify abnormalities. SGIB algorithm marks frames suspected of containing blood. System aids in diagnosis of Crohn's disease, obscure bleeding, and iron deficiency anemia. Bi-directional capsule design eliminates blind spots compared to single-direction predecessors.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included MTF measurement, minimum distinguishable contrast, depth of field, geometric distortion, field of view, color reproduction, optical spectra, intensity change over shelf life, bite testing, waterproof testing, pH resistance, and comparative performance testing between MR1100 and MR2000 receivers.
Technological Characteristics
Capsule: 30.1mm x 10.8mm, biocompatible plastic, COP dome. CMOS sensor, 12 white LEDs, 170° FOV. HBC transmission. Receiver: 145mm x 82.5mm x 29.8mm, 330g, Li-ion battery. Software: MiroView U 4.0 (Server/Operator/Client). Connectivity: USB, LAN. Sterilization: Not specified. Algorithm: Auto Exposure (P-control) and SGIB (Suspected GI Bleeding) marker.
Indications for Use
Indicated for adults for visualization of small bowel mucosa to monitor lesions associated with Crohn's disease, obscure GI bleeding, or iron deficiency anemia not detected by upper/lower endoscopy. Includes Suspected GI Bleeding Indicator (SGIB) to mark frames with suspected blood/red areas.
Regulatory Classification
Identification
An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”
Given PillCam® SB3 Capsule endoscopy system (K123864)
Reference Devices
Given PillCam® COLON 2 capsule endoscopy system (K123666)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized human figure. To the right of that is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 30, 2018
Intromedic Co., Ltd. Jinyoung Lee General Manager Suite 1105, 1106, E&C Venture Dream Tower 6-Cha Guro-dong, Guro-gu Seoul. 08375 Korea
Re: K170438 Trade/Device Name: MiroCam® Capsule Endoscope System Regulation Number: 21 CFR§ 876.1300 Regulation Name: Ingestible Telemetric Gastrointestinal Capsule Imaging System Regulatory Class: II Product Code: NEZ Dated: December 7, 2017 Received: December 14, 2017
Dear Jinyoung Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold, sans-serif font. The text is black and appears to be the main subject of the image. The background is plain and does not distract from the text.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
#### 510(k) Number (if known) K170438
Device Name
MiroCam® Capsule Endoscope System
Indications for Use (Describe)
The MiroCam® Capsule Endoscope System is intended for visualization of the small bowel mucosa.
- It may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy.
- It may be used in the visualization and monitoring of lesions that may be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy.
- It may be used in the visualization and monitoring of lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy.
It may be used as a tool in the detection of abnormalities of the small bowel and this device is indicated for adults only.
The Suspected GI Bleeding Indicator (SGIB) is intended to mark frames of the video suspected of containing blood or red areas.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data nepoled annound the and review the collection of information. Send comments regarding this burden estimate or any other onspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
### 1. Company and Correspondent making the submission
| Name | IntroMedic Co., Ltd. |
|-----------|---------------------------------------------------------------------------------------------------|
| Address | Suite 1105, 1106, E&C Venture Dream Tower 6-Cha<br>197-28 Guro-dong, Guro-gu, Seoul, 08375, Korea |
| Telephone | +82-2-801-9300 |
| Fax | +82-2-801-9330 |
| Contact | Jinyoung Lee |
| Internet | http://www.intromedic.com |
### 2. Device
| Proprietary Name | MiroCam® Capsule Endoscope System |
|---------------------|-----------------------------------------------------------------------------------------------------|
| Common Name | Capsule Imaging System |
| Classification Name | 21 CFR 876.1300 (Product Code NEZ)<br>Ingestible telemetric gastrointestinal capsule imaging system |
### 3. Predicate Device
| Manufacturer | IntroMedic Co., Ltd. |
|---------------|------------------------------------------------|
| Device | MiroCam® Capsule Endoscope System |
| 510(K) Number | K143663 |
| Manufacturer | Given Imaging Ltd. |
| Device | Given PillCam SB3 Capsule endoscopy system |
| 510(K) Number | K123864 |
| Manufacturer | Given Imaging Ltd. |
| Device | Given PillCam COLON 2 Capsule endoscopy system |
| 510(K) Number | K123666 (Reference device) |
{4}------------------------------------------------
### 4. Description
The MiroCam® Capsule Endoscope System is comprised of the following core components:
- -MiroCam® Capsule: MC2000-B
- MiroCam® Receiver: MR2000 -
- Data Cable / Data Belt (Optional): MR1000-D / MR1000-D(S) or MR1000-D(M) or MR1000--D(L)
- MiroView™ Software: MiroView™ U 4.0 -
The general usage workflow of the MiroCam® system is as follows, the MiroCam® capsule captures images of the GI tract, which are sent via Human Body Communication to sensor pads which are affixed to the patient body. The sensor pads are connected to the receiver by the data cables. The image data is stored on the receiver for the duration of the patient procedure. After removing the receiver set from the patient body, the receiver is connected via USB to the Commercial PC and the image data is uploaded. Following upload, the physician (Gastroenterologist) reviews the patient image data for suspected abnormalities of the small bowel. The key system components are explained in detail below.
### MiroCam® Capsule
Functionally, the MiroCam® captures imaging via a CMOS imaging sensor for at least 12 hours at the rate of 6 images per second. Twelve white LEDs flash in concert with the imaging sensor. A sensor PCB links the imaging sensor to the two gold electrodes of the capsule, from which a weak current is emitted containing the image data via Human Body Communication technology. The weak current is then passed through the bodily tissue and fluid to be picked up by the sensor pads. Human Body Communications is uni-directional from the capsule to the sensor pads.
Physically, the capsule is 30.1mm in length, and 10.8mm in diameter. The exterior of the capsule is composed of biocompatible materials, capable of withstanding potential bite forces and exposure to fluids in the GI tract.
### MiroCam® Receiver
The image data from the capsule is saved to the MiroCam® Receiver via the sensor pads and data cables. The sensor pads are standard ECG type sensor pads (3M Red Dot), which are affixed to the patient's abdomen according to the location guide. The 9 sensor pads are attached to the 9 leads of the data cable, and the data cable is subsequently attached to the Receiver sits in a shoulder pouch, which is put on the patient for the duration of the procedure. A waist pouch is used to organize the data cables.
{5}------------------------------------------------
The Receiver can store images for up to 12 hours. Using LCD display it shows icons that indicate the state of receiver, real time image view and preview. User can capture image for the real time images are displayed. Three LEDs also represent the receiver state by switching its power. And MiroCam® receiver play some sounds for notifying to user. To reduce the battery consumption of MiroCam® receiver backlight brightness is controlled. User can control MiroCam® receiver by using three buttons.
The dimensions of the receiver are 145mm in height, 82.5mm in width, and 29.8mm in length. The units weight 330 grams, which is data cable. The battery is lithium ion, and is unique to the Receiver. The battery is rechargeable, and can be charged by placing in the battery charger provided with the Receiver. The receiver uses a storage flash for storing the images.
#### MiroView™ U 4.0
The MiroView™ U 4.0 helps the user to upload the image data from the Receiver and review the image data with displaying.
MiroView™ U 4.0 will be deployed as three types of software: MiroView™M Server, MiroView™M Operator and MiroView™ Client.
Server PC includes two types of S/W: MiroView™ Server and Operator.
MiroView™ Server and MiroView™ Operator will be deployed on a Server PC. MiroView™ Client S/W will be deployed on reviewer's PCs. 10 Client S/Ws will be able to connect to one Server PC via Network (LAN) at a time.
MiroView™ Server has functionalities: Data Storage, PACS connectivity.
MiroView™ Operator has functionalities: Receiver Management and Account Management.
MiroView™ Client has functionalities: List Mode (Case Data Management), Review Mode (Video Display, Image Capture, Annotation Box, etc.), Report Mode (User Comment, Auto Disease Name Check, etc), Export Mode (JPG, AVI, MiroVlew™ file format)
MiroView™ U 4.0 provides some kinds of Modes and Views to help the user review patient video data quickly and conveniently; however, IntroMedic, Co., Ltd. recommends that the doctor review the patient video in the Normal Mode / Single View to prevent from a wrong diagnosis.
{6}------------------------------------------------
MiroView™ U 4.0 is a new version of MiroView™ U 3.0, which has new updated features from the MiroView™M U 3.0. The features are described below:
- Compatibility with New Receiver MR2000: MiroView™ U Operator 4.0 -
- UI Function for Reviewing Patient Data of Dual Tip Capsule: MiroView™ U Client 4.0 -
- Support of H.264 Codec: MiroView™ U Common 4.0 -
- Express Play ver 3.0: MiroView™ U Client 4.0 -
- Supplementary Play: MiroView™ U Client 4.0 -
### 5. Indications for Use
Capsule Endoscope System is intended for visualization of the small The MiroCam® bowel mucosa.
- It may be used in the visualization and monitoring of lesions that may indicate Crohn's disease not detected by upper and lower endoscopy.
- -It may be used in the visualization and monitoring of lesions that mav be a source of obscure bleeding (either overt or occult) not detected by upper and lower endoscopy.
- -It may be used in the visualization and monitoring of lesions that may be potential causes of iron deficiency anemia (IDA) not detected by upper and lower endoscopy.
It may be used as a tool in the detection of abnormalities of the small bowel and this device is indicated for adults only.
The Suspected GI Bleeding Indicator (SGIB) is intended to mark frames of the video suspected of containing blood or red areas.
#### 6. Technological Characteristics and Substantial Equivalence
The characteristics of the IntroMedic Co., Ltd. MiroCam® Capsule Endoscope System are substantially equivalent to the following current legally marketed predicate devices based on indications for use, typical clinical use, and operational and fundamental technological characteristics
- MiroCam® Capsule Endoscope System by IntroMedic Co., Ltd. K143663
- Given® PillCam® SB3 Capsule endoscopy system by Given Imaging Ltd. K123864
- Given PillCam® COLON 2 capsule endoscopy system by Given Imaging Ltd. K123666 (Reference device)
{7}------------------------------------------------
### Table 5-1. Side-by-Side Comparison of the MiroCam® Capsule Endoscope System
## with the Predicate Device
| Type | Characteristics | MiroCam®<br>Endoscope System<br>Subject device | MiroCam®<br>Endoscope System<br>K143663 | Given® PillCam® SB3<br>Capsule system<br>K123864 | Given PillCam®<br>COLON 2 capsule<br>endoscopy system<br>K123666<br>(Reference device) |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Indications for Use | The MiroCam® Capsule<br>Endoscope System is<br>intended for visualization<br>of the small bowel<br>mucosa.<br>- It may be used in the<br>visualization and<br>monitoring of<br>lesions that may<br>indicate Crohn's<br>disease not detected<br>by upper and lower<br>endoscopy.<br>- It may be used in the<br>visualization and<br>monitoring of<br>lesions that may be a<br>source of obscure<br>bleeding (either<br>overt or occult) not<br>detected by upper<br>and lower<br>endoscopy.<br>- It may be used in the<br>visualization and<br>monitoring of<br>lesions that may be<br>potential causes of<br>iron deficiency<br>anemia (IDA) not<br>detected by upper<br>and lower<br>endoscopy.<br>- It may be used as a tool in<br>the detection of<br>abnormalities of the small<br>bowel and this device is<br>indicated for adults only.<br>- The Suspected GI<br>Bleeding Indicator<br>(SGIB) is intended to<br>mark frames of the video<br>suspected of containing<br>blood or red areas. | MiroCam® Capsule<br>Endoscope System is<br>intended for visualization<br>of the small bowel<br>mucosa. It may be used as<br>a tool in the detection of<br>abnormalities of the small<br>bowel in adults and<br>children from two years<br>of age.<br>- The Suspected GI<br>Bleeding Indicator<br>(SGIB) is intended to<br>mark frames of the video<br>of suspected of containing<br>blood or red areas. | The PillCam SB capsule<br>is intended for<br>visualization of the small<br>bowel mucosa.<br>- It may be used in the<br>visualization and<br>monitoring of<br>lesions that may<br>indicate Crohn's<br>disease not detected<br>by upper and lower<br>endoscopy.<br>- It may be used in the<br>visualization and<br>monitoring of<br>lesions that may be<br>a source of obscure<br>bleeding (either<br>overt or occult) not<br>detected by upper<br>and lower<br>endoscopy.<br>- It may be used in the<br>visualization and<br>monitoring of<br>lesions that may be<br>potential causes of<br>iron deficiency<br>anemia (IDA) not<br>detected by upper<br>and lower<br>endoscopy.<br>- The suspected Blood<br>Indicator (SBI) feature is<br>intended to mark frames<br>of the video suspected of<br>containing blood or red<br>areas.<br>- The PillCam SB capsule<br>may be used as a tool in<br>the detection of<br>abnormalities of the small<br>bowel and is intended for<br>use in adults and children<br>from two years of age. | The PillCam COLON 2<br>System is indicated to<br>provide visualization of<br>the colon. It is intended to<br>be used for detection of<br>colon polyps in patients<br>that may after an incomplete<br>optical colonoscopy with<br>adequate preparation, and<br>a complete evaluation of<br>the colon was not<br>technically possible.<br>(※The Indications for<br>Use of the device do not<br>apply to the subject<br>device.) | |
| Manufacturer | IntroMedic Co., Ltd.<br>MC2000-B | IntroMedic Co., Ltd.<br>MC1200-B | Given Imaging Ltd.<br>PillCam SB3 | Given Imaging Ltd.<br>PillCam COLON 2 | |
| | | | | | |
| ule | Size | Length: 30.1mm<br>Diameter: 10.8mm | Length: 24.5mm<br>Diameter: 10.8mm | Length: 26.2mm<br>Diameter: 11.4mm | Length: 31.5mm<br>Diameter: 11.6mm |
| | Weight | $3.5g\pm0.1g$ | 3.25g | $3.0\pm0.1g$ | $2.9\pm0.03g$ |
| | Material | Human Compliance<br>Plastic | Human Compliance<br>Plastic | Biocompatible plastic | Biocompatible plastic |
| | Dome Material | COP (Cyclo Olefin<br>Polymers) | COP (Cyclo Olefin<br>Polymers) | Polycarbonate | Polycarbonate |
| | Light | 12 White LEDs (each<br>White LEDs) | 6 White LEDs | 4 White LEDs | 8 White LEDs (each<br>White LEDs) |
| | LED Size | 2.0mm(L)*1.25mm(W)*<br>0.8mm(H) | 2.0mm(L)*1.25mm(W)*<br>0.8mm(H) | | |
| | LED Viewing<br>Angle | 130° | 130° | | |
| | Exposure Time | 0.5 ~ 24 ms | 0.5~24 ms | | |
| | Field of View | 170° | 170° | 156° | |
| | Image Sensor | CMOS | CMOS | CMOS | CMOS |
| | Depth of Field | Length: 3cm | Length: 3cm | Length: 3cm | Length: 3cm |
| | Enlargement<br>Ratio | 1:8 | 1:8 | | |
| | Detectable<br>Range | Under 0.1mm | Under 0.1mm | At least 0.07mm | |
| | Sampling Ratio | 6 frames per second (each<br>3 frames per head) | 3 frames per second | 2 or 2~6 frames per<br>second | Up to 35 Frames per<br>Second per head |
| | Pixel size of<br>Sensor | $6.0 \mu m*6.0 \mu m$ | $6.0 \mu m*6.0 \mu m$ | | |
| | Working Time | 12 Hours | 12 Hours | Over 8 Hours | 10 Hours |
| | Chemical Safety | Safe in pH=2 ~ pH=8 | Safe in pH=2 ~ pH=8 | Safe in pH=2 ~ pH=8 | Safe in pH=2 ~ pH=8 |
| | Battery Type | Silver Oxide Cell…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.