K170412 · Toshibamedical Systems Corporation · LNH · Jun 7, 2017 · Radiology
Device Facts
Record ID
K170412
Device Name
Vantage Titan 1.5T, MRT-1510, M-Power GX
Applicant
Toshibamedical Systems Corporation
Product Code
LNH · Radiology
Decision Date
Jun 7, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
Vantage Titan systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are: - Proton density (PD) (also called hydrogen density) - Spin-lattice relaxation time (T1) - Spin-spin relaxation time (T2) - Flow dynamics - Chemical Shift Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
Device Story
Vantage Titan 1.5T is a magnetic resonance imaging (MRI) system; acquires cross-sectional transaxial, coronal, sagittal, and oblique images of head/body. Operates by detecting spatial distribution of protons (hydrogen nuclei) via NMR properties (T1, T2, PD, flow, chemical shift). System uses various pulse sequences (FFE3D, mUTE, MP2RAGE, FSE2D, FASE DWI, PSIR, MOLLI, V-TISP) to generate images. Operated by trained clinicians in clinical settings. Output is digital images interpreted by physicians for diagnostic purposes. New software functionalities include motion correction (2D-RMC), R-wave detection (surfECG), and enhanced post-processing (SPEEDER). Benefits include reduced acoustic noise, improved motion handling, and advanced tissue mapping (T1/T2/T2*).
Clinical Evidence
Bench testing, phantom imaging, and volunteer clinical imaging were performed. No formal clinical trial with patient outcomes was required. Testing demonstrated that new software functionalities (pulse sequences and post-processing) perform as intended and maintain safety/effectiveness comparable to predicate devices.
Technological Characteristics
1.5T superconducting magnet MRI system. Software-controlled pulse sequences and image reconstruction. Connectivity via standard DICOM. Complies with IEC 60601-1, IEC 60601-2-33, IEC 62304, and IEC 62366. Software is Moderate Level of Concern. Features include motion correction, R-wave detection, and multi-phase ASL.
Indications for Use
Indicated for patients requiring diagnostic cross-sectional imaging of head or body structures via MRI, including non-contrast MRA. No specific age or gender restrictions; contraindications include presence of ferromagnetic objects or other standard MRI contraindications.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The eagle is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 7, 2017
TOSHIBA MEDICAL SYSTEMS CORPORATION PAUL BIGGINS DIRECTOR REGULATORY AFFAIRS 2441 MICHELLE DRIVE TUSTIN CA 92780
Re: K170412
Trade/Device Name: Vantage Titan 1.5T. MRT-1510. M-Power GX Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: May 16, 2017 Received: May 17, 2017
Dear Mr. Paul Biggins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. O'Hara
For
Robert A. Ochs. PhD Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
### K170412
Device Name Vantage Titan 1.5T, MRT-1510, M-Power GX
Indications for Use (Describe)
Vantage Titan systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.
MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
- · Proton density (PD) (also called hydrogen density)
- · Spin-lattice relaxation time (T1)
- · Spin-spin relaxation time (T2)
- · Flow dynamics
- · Chemical Shift
Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Toshiba Medical. The word "TOSHIBA" is written in large, bold, red letters on the top line. Below it, the word "MEDICAL" is written in the same font and color, but slightly smaller. The logo is simple and clean, with a focus on the company name.
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92
### 1. CLASSIFICATION and DEVICE NAME
| Classification Name: | Magnetic Resonance Diagnostic Device |
|-------------------------|--------------------------------------|
| Regulation Number: | 90-LNH (Per 21 CFR § 892.1000) |
| Trade Proprietary Name: | Vantage Titan 1.5T |
| Model Number: | MRT-1510 |
### 2. SUBMITTER'S NAME
Toshiba Medical Systems Corporation (TMSC) 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
### 3. OFFICIAL CORRESPONDENT
Akinori Hatanaka Senior Manager, Regulatory Affairs and Vigilance
### 4. CONTACT PERSON, U.S. AGENT and ADDRESS
Paul Biggins Director, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive, Tustin, CA 92780 (714) 730-5000 Fax: (714) 730-1310 pbiggins@tams.com
### 5. MANUFACTURING SITE
Toshiba Medical Systems Corporation (TMSC) 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan
- 6. ESTABLISHMENT REGISTRATION 9614698
- 7. DATE PREPARED February 6, 2017
- 8. DEVICE NAME Vantage Titan 1.5T, MRT-1510, V4.0 Software
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the words "TOSHIBA MEDICAL" stacked on top of each other. The words are in a bold, sans-serif font and are colored red. The word "TOSHIBA" is on the top line, and the word "MEDICAL" is on the second line.
### 9. TRADE NAME
Vantage Titan 1.5T
### 10. CLASSIFICATION NAME
Magnetic Resonance Diagnostic Device (MRDD)
### 11. CLASSIFICATION PANEL
Radiology
### 12. DEVICE CLASSIFICATION
Class II (per 21 CFR 892.1000, Magnetic Resonance Diagnostic Device)
### 13. PRODUCT CODE
90-LNH
### 14. PREDICATE DEVICE
Primary Predicate Device (system): Vantage Titan 1.5T, MRT-1510, V3.6
Reference Predicate Device (added software functionalities): Vantage Galan 3T, MRT-3020, V4.0
| | Subject | Primary Predicate | Reference Predicate |
|----------------|---------------------------------------|---------------------------------------|-------------------------------------|
| System | Vantage Titan 1.5T,<br>MRT-1510, V4.0 | Vantage Titan 1.5T,<br>MRT-1510, V3.6 | Vantage Galan 3T,<br>MRT-3020, V4.0 |
| Marketed By | Toshiba America<br>Medical Systems | Toshiba America Medical<br>Systems | Toshiba America<br>Medical Systems |
| 510(k) Number | This Submission | K160632 | K162183 |
| Clearance Date | | August 23, 2016 | November 25, 2016 |
### 15. REASON FOR SUBMISSION
Modification of a cleared device
### 16. SUBMISSION TYPE
Traditional 510(k) Premarket Notification
### 17. DEVICE DESCRIPTION
The Vantage Titan (Model MRT-1510) is a 1.5 Tesla Magnetic Resonance Imaging (MRI) System, previously cleared under K160632. This system is based upon the technology and materials of previously marketed Toshiba MRI systems and is intended to acquire and display cross-sectional transaxial, coronal, sagittal, and oblique images of anatomic structures of the head or body.
### 18. SUMMARY OF CHANGE(S)
This submission is to report the following software functionalities have been added:
- . FFE3D:
- o mUTE: Ultra short TE T1 weighted sequence with reduced acoustic noise o mUTE 4D-MRA: Ultra short TE time resolved MR angiography with reduced acoustic noise
- o 2D-RMC: Two dimensional real-time motion correction
- MP2RAGE: Images with multiple TI (Inversion time) to create T1 map ●
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Toshiba Medical. The word "TOSHIBA" is written in large, bold, red letters on top of the word "MEDICAL", which is also written in large, bold, red letters. The logo is simple and clean, with a focus on the company name.
**Made For life**
- FSE2D mEcho: Images with multiple TE (Echo time) to create T2 map
- o FFE2D mEcho: Images with multiple TE (Echo time) to create T2* map
- mASTAR: Multi Phase ASL (Arterial Spin Labeling)
- FASE DWI: Spin Echo based Diffusion Weighted Imaging with reduced acoustic noise
- PSIR: Phase Sensitive Inversion Recovery
- MOLLI: Modified Look Locker Inversion recovery
- V-TISP: Variable Ti*1 and TS*2 Prep (*1 inversion recovery time; *2 saturation recovery time)
- o Enhancement of SPEEDER post-processing: Improvement of SPEEDER algorithm
- Improvement of surfECG R-wave detection: Morphology-matching based R-wave detection.
| Item | Subject Device: Vantage Titan<br>1.5T, V4.0 | Predicate Device: Vantage Titan<br>1.5T, V3.6 (K160632) | Notes |
|------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-------|
| Static field strength | 1.5T | 1.5T | Same |
| Operational Modes | Normal and 1st Operating Mode | Normal and 1st Operating Mode | Same |
| i. Safety parameter<br>display | SAR dB/dt | SAR dB/dt | Same |
| ii. Operating mode<br>access requirements | Allows screen access to 1st level<br>operating mode | Allows screen access to 1st level<br>operating mode | Same |
| Maximum SAR | 4W/kg for whole body (1st<br>operating mode specified in<br>IEC 60601-2-33:2010<br>+Amd.1:2013) | 4W/kg for whole body (1st<br>operating mode specified in<br>IEC 60601-2-33(2010)) | Same |
| Maximum dB/dt | <1st operating mode specified in<br>IEC60601-2-33(2010)<br>+Amd.1:2013 | <1st operating mode specified in<br>IEC 60601-2-33 (2010) | Same |
| Potential emergency condition<br>and means provided for shutdown | Shut down by Emergency Ramp<br>Down Unit for collision hazard<br>for ferromagnetic objects | Shut down by Emergency Ramp<br>Down Unit for collision hazard<br>for ferromagnetic objects | Same |
#### 19. SAFETY PARAMETERS
#### 20. IMAGING PERFORMANCE PARAMETERS
No change from the previous predicate submission, K160632.
#### 21. INDICATIONS FOR USE
Vantage Titan systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Toshiba Medical. The word "TOSHIBA" is stacked on top of the word "MEDICAL". Both words are written in a bold, sans-serif font. The color of the text is red.
MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
- . Proton density (PD) (also called hydrogen density)
- Spin-lattice relaxation time (T1)
- o Spin-spin relaxation time (T2)
- . Flow dynamics
- Chemical Shift
Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images vield information that can be useful in diagnosis.
No changes to the previously cleared indication, K160632.
#### 22. SUMMARY OF DESIGN CONTROL ACTIVITIES
PS Risk List for new software functionalities and pulse sequences are included in this submission. The test methods used are the same as those submitted in the previously cleared submissions of the primary predicate device, Vantage Titan 1.5T, MRT-1510, V3.6 (K160632), and the reference predicate device, Vantage Galan 3T, MRT-3020, V4.0 (K162183). A declaration of conformity with design controls is included in this submission.
#### 23. SAFETY
This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. The changes to this device are limited to the application software that is being presented for review. As the hardware and related safety considerations remain unchanged from the previous submission no information related to these aspects are presented for the determination of safety. Performance and/or clinical testing are being provided to demonstrate the safety and effectiveness of the application software.
This device is in conformance with the applicable parts of the following consensus standards published by the International Electrotechnical Commission (IEC) for medical devices and the National Electrical Manufacturers Association (NEMA).
#### LIST OF APPLICABLE STANDARDS
- IEC 60601-1:2005 +Amd.1:2012 ●
- IEC 60601-2-33:2010 +Amd.1:2013 .
- IEC 60601-1-2:2007 o
- IEC 60601-1-8:2006 +Amd.1:2012
- . IEC 60825-1:2007
- IEC 62304:2006 ●
- IEC 62366:2007 +Amd.1:2014 ●
#### 24. TESTING
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the words "TOSHIBA MEDICAL" stacked on top of each other. The words are in a bold, sans-serif font and are colored red. The background is white.
### 25. SUBSTANTIAL EQUIVALENCE
Toshiba Medical Systems Corporation believes that the Vantage Titan 1.5T, MRT-1510, V4.0 Magnetic Resonance Imaging (MRI) System is substantially equivalent to the previously cleared predicate devices referenced in this submission. Toshiba Medical Systems Corporation believes that the changes incorporated into the Vantage Titan 1.5T, MRT-1510, V4.0 software are substantially equivalent to the previously cleared predicate devices.
#### 26. CONCLUSION
The modifications incorporated into the Vantage Titan 1.5T, MRT-1510, V4.0 software do not change the indications for use or the intended use of the device. Based upon bench testing, phantom imaging, volunteer clinical imaging, successful completion of software validation and application of risk management and design controls, it is concluded that the subject device is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.