K170405 · Care Fusion · KPM · Nov 2, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K170405
Device Name
BD Curve Ascites Shunt
Applicant
Care Fusion
Product Code
KPM · Gastroenterology, Urology
Decision Date
Nov 2, 2017
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.5955
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Peritoneovenous shunting is indicated for patients with malignant or non-malignant intractable ascites (including patients with chylous ascites, hepatorenal syndrome, and idiopathic ascites), not responding to standard medical management, and not considered candidates for portal-venous shunting. The 17Fr valved peel-away introducer is intended for use in the percutaneous insertion of catheters in the peritoneal space. The 15.5Fr Pigtail Revision Catheter is indicated for revising the peritoneal catheter of a BD Curve Ascites Shunt.
Device Story
Fully implantable peritoneovenous shunt; transfers ascitic fluid from peritoneal space to circulatory system for symptom relief. System includes pump chamber (single or double valved), venous catheter, and peritoneal catheter. Placed percutaneously via Seldinger technique. 17Fr valved peel-away introducer facilitates catheter insertion while minimizing air/fluid leakage. 15.5Fr Pigtail Revision Catheter used to replace occluded peritoneal limbs. Used in clinical settings by physicians. Provides palliative management for intractable ascites.
Clinical Evidence
No clinical tests were conducted for this submission. Substantial equivalence is supported by bench-level performance testing, including biocompatibility, sterilization, and mechanical integrity evaluations.
Technological Characteristics
Materials: Silicone tubing with barium sulfate stripe and silique surface treatment. Pump chamber: Silicone with silique surface treatment. Design: Fully implantable shunt with single/double valved pump chamber. Introducer: Peel-away sheath with integral valve seal and dilator. Standards: ISO 10993 (biocompatibility), ISO 11135 (EO sterilization), ISO 11607 (packaging), EN ISO 14630 (surgical implants), BS EN 1617/1618 (catheters).
Indications for Use
Indicated for patients with malignant or non-malignant intractable ascites, including chylous ascites, hepatorenal syndrome, and idiopathic ascites, who are unresponsive to standard medical management and are not candidates for portal-venous shunting.
Regulatory Classification
Identification
A peritoneo-venous shunt is an implanted device that consists of a catheter and a pressure activated one-way valve. The catheter is implanted with one end in the peritoneal cavity and the other in a large vein. This device enables ascitic fluid in the peritoneal cavity to flow into the venous system for the treatment of intractable ascites.
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(2) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” and
(3) Backflow specification and testing to prevent reflux of blood into the shunt.
Predicate Devices
Denver Ascites Shunt And Percutaneous Access Kit With Ascites Shunt (K011862)
Medcomp, Inc. - Valved Tearaway Introducer Generation II (K132498)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 2, 2017
CareFusion Tamara Brev Regulatory Affairs Advisor 75 North Fairway Drive Vernon Hills, IL 60061
Re: K170405
> Trade/Device Name: BD Curve Ascites Shunt Regulation Number: 21 CFR§ 876.5955 Regulation Name: Peritoneo-Venous Shunt Regulatory Class: II Product Code: KPM, DYB Dated: September 29, 2017 Received: October 2, 2017
Dear Tamara Brey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170405
Device Name
BD Curve Ascites Shunt
#### Indications for Use (Describe)
Peritoneovenous shunting is indicated for patients with malignant intractable ascites (including patients with chylous ascites, hepatorenal syndrome, and idiopathic ascites), not responding to standard medical management, and not considered candidates for portal-venous shunting
The 17Fr valved peel-away introducer is intended for use in the percutaneous insertion of catheters in the peritoneal space.
The 15.5Fr Pigtail Revision Catheter is indicated for revising the peritoneal catheter of an BD Curve Ascites Shunt.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY- K170405
# A summary of 510(k) safety and effectiveness information in accordance with 21 CFR 807.92.
| SUBMITTER INFORMATION | | |
|----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | CareFusion | |
| Address | 75 North Fairway Drive, Vernon Hills, IL 60061 USA | |
| Phone number | (847) 362-9485 | |
| Fax number | (312) 949-9245 | |
| Establishment Registration<br>Number | 1423507 | |
| Name of contact person | Tamara Brey | |
| Date prepared | November 1, 2017 | |
| DESCRIPTION OF DEVICE | | |
| Trade or proprietary name | BD Curve Ascites Shunt | |
| Device name | Ascites Shunt | |
| Classification name | Peritoneo-Venous Shunt | |
| Classification panel | Gastroenterology/Urology | |
| Regulation | Class II per 21CFR §876.5955 | |
| Product Code(s) | KPM, DYB | |
| Legally marketed device(s) to<br>which equivalence is claimed | Denver Ascites Shunt And Percutaneous Access Kit With Ascites Shunt:<br>K011862<br>Medcomp, Inc. - Valved Tearaway Introducer Generation II: K132498<br>CareFusion – Valved Peelable Introducers: K122422 | |
| Reason for 510(k) submission | This 510(k) submission is for minor design modifications to the Ascites<br>Shunt pump chamber and the peritoneal catheter. A minor design<br>modifications to the peritoneal revision catheter design and addition of a<br>new placement accessory. Updates to labeling include a revision of<br>placement techniques and updated patient selection considerations. | |
| Device description | The Ascites Shunt is a fully implantable device that transfers fluid from the<br>peritoneal space to the circulatory system to provide symptom relief for<br>patients with malignant or non-malignant intractable ascites. | |
| Intended use of the device | Peritoneovenous shunting is indicated for patients with malignant or non-<br>malignant intractable ascites (including patients with chylous ascites,<br>hepatorenal syndrome, and idiopathic ascites), not responding to standard<br>medical management, and not considered candidates for portal-venous<br>shunting.<br>The 17Fr valved peel-away introducer is intended for use in the<br>percutaneous insertion of catheters in the peritoneal space.<br>The 15.5Fr Pigtail Revision Catheter is indicated for revising the peritoneal<br>catheter of a BD Curve Ascites Shunt. | |
| Characteristic | New Device | Predicates: |
| | | CareFusion Denver Ascites Shunt PAK:<br>K011862<br>Medcomp, Inc. - Valved Tearaway<br>Introducer Generation II: K132498<br>CareFusion - Valved Peelable Introducers:<br>K122422 |
| Ascites Shunt Overall<br>Design | Shunt with pump chamber<br>(Single Valved and Double<br>Valved) with venous and<br>peritoneal catheters | Same as predicate: CareFusion Denver<br>Ascites Shunt PAK: K011862 |
| Ascites Shunt Materials<br>of Construction | Tubing: Silicone with a barium<br>sulfate stripe and silique surface<br>treatment.<br>Pump Chamber Material:<br>Silicone with silique surface<br>treatment | Same as predicate: CareFusion Denver<br>Ascites Shunt PAK: K011862 |
| Ascites Shunt Placement<br>Technique | Minimally invasive (Seldinger<br>technique) for all shunts | Same as predicate: CareFusion Denver<br>Ascites Shunt PAK: K011862 |
| Ascites Shunt<br>Design Intent | Placed percutaneously to<br>manage ascites by recirculating<br>ascitic fluid from the peritoneal<br>space to the circulatory system | Same as predicate: CareFusion Denver<br>Ascites Shunt PAK: K011862 |
| Valved Peelable<br>Introducer Overall Design | Peel-away introducer sheath<br>which contains an integral valve<br>seal and dilator comprised of a<br>cylindrical tube with a locking<br>hub | Same as predicate: Medcomp, Inc. -<br>Valved Tearaway Introducer Generation II:<br>K132498 |
| Valved Peelable<br>Introducer Design Intent | To obtain peritoneal access and<br>facilitate catheter insertion while<br>minimizing air/fluid leakage. | Same as predicate: CareFusion - Valved<br>Peelable Introducers: K12242 |
| Pigtail Revision Catheter<br>Design Intent | Placed percutaneously to<br>replace occluded peritoneal limb<br>of Ascites shunt. | Same as predicate: CareFusion Denver<br>Ascites Shunt PAK: K011862 |
| CONCLUSION OF DEVICE COMPARISON | | |
| The technological characteristics of the proposed devices are substantially equivalent to the predicates | | |
{4}------------------------------------------------
### SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
## PERFORMANCE DATA
### SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE
Bench-level testing was carried out on Ascites Shunt to demonstrate substantial equivalence. The performance testing requirements were determined by the predicate devices and further defined by the standards listed below. The testing and evaluation conducted includes displacement volume, resealing capacity, interface testing, declogging, aging, biocompatibility, sterilization, and EO residuals.
Performance Test Summary
{5}------------------------------------------------
| Biocompatibility | ISO 10993-1: Part 1, Biological Evaluation of Medical Devices: Evaluation and<br>Testing within a Risk Management Process. |
|--------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | ISO 10993-5:2009/(R) 2014: Part 5, Biological Evaluation of Medical Devices --<br>Tests For In Vitro Cytotoxicity |
| Biocompatibility | ISO 10993-7 Second Edition 2008-10-15: Part 7, Biological Evaluation Of Medical<br>Devices - Ethylene Oxide Sterilization Residuals [Including: Technical Corrigendum<br>1 (2009)] |
| Biocompatibility | ISO 10993-12, 2012 Biological Evaluation Of Medical Devices - Part 12: Sample<br>Preparation and Reference Materials. |
| Biocompatibility | ISO 10993-17:2002: Part 17, Biological evaluation of medical devices -<br>Establishment of allowable limits for leachable substances |
| Biocompatibility | ISO 10993-18:2005: Part 18, Biological evaluation of medical devices - Chemical<br>characterization of materials |
| Sterilization | ISO 11135 2nd Ed: Sterilization of Health-Care Products -Ethylene Oxide -<br>Requirements for the Development, Validation and Routine Control of a Sterilization<br>Process for Medical Devices |
| Sterilization | ISO 11138-1 2nd Ed: Sterilization of Health Care Products - Biological Indicators -<br>Part 1: General Requirements |
| Sterilization | ISO 11737-1 2nd Ed: Sterilization of Medical Devices – Microbiological Methods<br>Part 1 |
| Sterilization | AAMI TIR28:2009(R) 2013: Product Adoption and Process Equivalency for Ethylene<br>Oxide Sterilization |
| Sterilization | ANSI/AAMI ST72:2011: Bacterial endotoxins - Test methods, routine monitoring,<br>and alternatives to batch testing |
| Packaging | AAMI / ANSI / ISO 11607-1:2006/(R)2010: Packaging For Terminally Sterilized<br>Medical Devices - Part 1:Requirements For Materials, Sterile Barrier Systems And<br>Packaging Systems [Including: Amendment 1 (2014)] |
| Packaging | AAMI/ANSI/ISO 11607-2:2006/(R)2010: Packaging For Terminally Sterilized Medical<br>Devices - Part 2: Validation Requirements For Forming, Sealing And Assembly<br>Processes [Including: Amendment 1 (2014)] |
| Risk Management | BS EN ISO 14971: 2012 Medical Devices. Application of Risk Management to<br>Medical Devices. |
| Performance | EN ISO 14630: Non-active Surgical Implants - General Requirements |
| Performance | BS EN 1617: 1997 Sterile Drainage Catheters and Accessory Devices for Single Use |
| Performance | BS EN 1618: 1997 Catheters other than Intravascular Catheters - Test Methods for<br>Common Properties |
| SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL<br>EQUIVALENCE AND/OR OF CLINICAL INFORMATION | |
| | N/A - No clinical tests were conducted for this submission. |
## CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA
The results of the non-clinical tests show that the CareFusion Ascites Shunt is as safe, as effective, and performs as well as the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.