K170369 · Cathay Manfacturing Corp. · GEI · Jul 19, 2018 · General, Plastic Surgery
Device Facts
Record ID
K170369
Device Name
Disposable Electrosurgical Pencils
Applicant
Cathay Manfacturing Corp.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 19, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Disposable Electrosurgical Pencil is used for cutting and coagulation to remove tissue and control bleeding by using high frequency during electrosurgical surgery with a specified Electrosurgery Unit (ESU) generator. The device is disposable for single use and supplied as sterile with an electrode tip.
Device Story
Disposable electrosurgical pencil; monopolar device; used for cutting and coagulation of soft tissue. Input: high-frequency alternating current (HFAC) from electrosurgical unit (ESU). Operation: surgeon activates via hand-piece buttons (cut/coag) or external monopolar footswitch; HFAC travels from ESU through pencil to active electrode tip; current passes through patient tissue to dispersive return pad. Output: thermal tissue effect (cutting/coagulation). Used in clinical settings by surgeons/physicians. Benefits: enables precise tissue removal and hemostasis during surgery.
Clinical Evidence
Bench testing only. No clinical data. Performance evaluated via thermal effect testing on porcine tissue (liver, kidney, muscle) comparing thermal damage zone sizes (length, width, depth) against the predicate. Safety testing conducted per IEC 60601-1 and IEC 60601-2-2. Biocompatibility confirmed per ISO 10993. Package integrity and shelf-life stability testing performed.
Technological Characteristics
Monopolar electrosurgical pencil; stainless steel hex blade electrode (2.36mm diameter); ABS handle and insulation; PVC cable (3m or 5m). Energy source: RF (high-frequency alternating current). Sterilization: Ethylene Oxide (EO). Standards: IEC 60601-1, IEC 60601-2-2, ISO 11135, ISO 11607, ISO 10993. Connectivity: Wired connection to ESU generator.
Indications for Use
Indicated for cutting and coagulation of soft tissue and control of bleeding during electrosurgical procedures using a compatible ESU generator. For single-use only.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 19, 2018
Cathay Manfacturing Corp. % Ren Ren Manager No. 328, Xishe Road Maogang Town, Songjiang Area Shanghai, China 201607
Re: K170369
Trade/Device Name: Disposable Electrosurgical Pencils, CP1001 Series Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: July 3, 2018 Received: July 3, 2018
Dear Ren Ren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Image /page/1/Picture/5 description: The image shows a signature block. It starts with the word "Sincerely,", followed by the name "Jennifer R. Stevenson -S3". The signature block ends with "For Binita S. Ashar, M.D., M.B.A., F.A.C.S."
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170369
Device Name
Disposable Electrosurgical Pencils (CP1001 Series)
Indications for Use (Describe)
The Disposable Electrosurgical Pencil is used for cutting and coagulation to remove tissue and control bleeding by using high frequency during electrosurgical with a specified Electrosurgery Unit (ESU) generator. The device is disposable for single use and supplied as sterile with an electrode tip.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| <div> <span> <b> </b> </span> </div> | |
| <div> <input checked="true" type="checkbox"/>Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <input type="checkbox"/>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Submitter: | CATHAY MANUFACTURING CORP.<br>No. 328, Xishe Road, Maogang Town, Songjiang Area<br>Shanghai, CN 201607 |
|-------------------------------------|--------------------------------------------------------------------------------------------------------|
| Contact: | Ms. Ren, Ren<br>Phone: (86)021-57878328<br>Email: rubyfantasy@qq.com |
| Date Prepared: | July. 6, 2018 |
| Device Trade Name: | Disposable Electrosurgical Pencils |
| Model: | CP1001 Series |
| Device Common Name: | Disposable Electrosurgical Pencils |
| Classification Name:<br>accessories | Electrosurgical cutting and coagulation device and |
| Class: | II |
| Classification Name: | 21 CFR 878.4400 |
| Product Code: | GEI |
### Predicate Device:
| Device | Company | Product Code | 510(k) Number |
|-------------------------------------------|----------------------------------------------------------------|--------------|---------------|
| OBS Disposable<br>Electrosurgical Pencils | Jiangmen City Xinhui<br>Baisheng Medical<br>Equipment Co., Ltd | GEI | K092634 |
# 1. Indication for Use:
The Disposable Electrosurgical Pencil is used for cutting and coagulation to remove tissue and control bleeding by using high frequency during electrosurgical surgery with a specified Electrosurgery Unit (ESU) generator. The device is disposable for single use and supplied as sterile with an electrode tip.
{4}------------------------------------------------
# 2. Product Description:
The Disposable Electrosurgical pencils are used for the cutting and coagulation of soft tissue and have a conductive cable which is designed for use with high frequency surgical generator (Electrosurgical unit).
The connector or plug fits into electrosurgical unit. The hand-pieces are made of plastic with two buttons (cutting and coagulation buttons) in the case of the handcontrolled pencils, while the foot controlled pencils is activated by a monopolar footswitch connected to the generator. One button or switch is to control the cutting mode, CUT, of the ESU while the other controls the coagulation mode, COAG. The applied electrosurgery, monopolar, is the emittance of the high frequency alternating current, HFAC, from the diathermy via an active electrode through the patient's body tissues and then returned back to the diathermy machine via a dispersive electrode (patient return pad).
The proposed device mainly consists of electrode tip, hand piece handle, cable, and plug. For all applied device models, only one kind of dimension of electrode tip, standard hex blade electrode is available. However, two kind of cable length, 3m or 5m, and three kind of available hand-piece color, white/blue/green color, are available
| ELEMENT OF<br>COMPARISON | SUBJECT DEVICE | Predicate Device |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | K092634<br>OBS Disposable Electrosurgical Pencils |
| Indications for Use | used for cutting and coagulation to remove tissue and control bleeding by using high frequency during electrosurgical surgery with a specified Electrosurgery Unit (ESU) generator. The device is disposable for single use and supplied as sterile with an electrode tip. | used for cutting and coagulation to remove tissue and control bleeding by using high frequency during electrosurgical surgery with a specified Electrosurgery Unit (ESU) generator. The device is disposable for single use and supplied as sterile with an electrode tip. |
| Classification | 878.4400 | 878.4400 |
| Product Code | GEI | GEI |
| OTC or<br>Prescription | For Prescription Use | For Prescription use |
| Energy source<br>type | RF Energy | RF Energy |
| Sterility and for<br>single use | Yes, Sterile and for single use | Yes, Sterile and for single use |
| Sterilization<br>method | EO Sterilization | EO Sterilization |
| Monopolar or<br>Bipolar electrode | Monopolar electrode | Monopolar electrode |
| Electrode<br>materials | Stainless steel | Stainless steel |
| Electrode<br>insulation material | ABS | ABS |
| Electrode coating | N/A | N/A |
| Electrode shape<br>and dimensions | standard hex blade electrode, 2.36 x 70mm | standard hex blade electrode, 2.36 x 70mm;<br>other shape and dimensions |
| Shaft materials | ABS | ABS |
| Handle length | 95 mm | 90 mm |
# 4. Summary of Technology Characteristics and Performance Testing (Comparison to the Predicate Device)
{5}------------------------------------------------
#### K170369
#### CATHAY MANUFACTURING CORP.
| Shaft length | 50mm | 50 mm | |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Total length | 145mm | 140mm | |
| Shaft diameter | 2.36 mm | 2.36 mm | |
| Cable materials | Blue PVC cable | Blue PVC cable | |
| Cable length | 3m or 5m | 3m or 5m | |
| Foot control model | Yes, has foot control mode | Yes, has foot control mode | |
| Standards applied | IEC 60601-1; IEC 60601-2-2;<br>ISO 11135; ISO 11607 | IEC 60601-1; IEC 60601-2-2;<br>ISO 11135; ISO 11607 | |
| Biocompatibility | Conforms to ISO10993 | Conforms to ISO10993 | |
| Labeling | confirm to 21 CFR Part 801 | confirm to 21 CFR Part 801 | |
| Performance | Conforms to IEC 60601-1 and<br>IEC 60601-2-10:2012,<br>Part 2 particular requirements for the basic<br>safety and essential performance of high<br>frequency surgical equipment and high<br>frequency surgical accessories.<br>Performance conforms the requirements<br>specified in 'Premarket Notification (510(k))<br>Submissions for Electrosurgical Devices for<br>General Surgery' | Conforms to IEC 60601-1 and<br>IEC 60601-2-10:2012,<br>Part 2 particular requirements for the basic<br>safety and essential performance of high<br>frequency surgical equipment and high<br>frequency surgical accessories Performance<br>conforms the requirements specified in<br>'Premarket Notification (510(k)) Submissions<br>for Electrosurgical Devices for General<br>Surgery' | |
| | Stability and shelf<br>life | 3 year | 3 year |
### 5. Bench Testing (non-clinical):
Compare to predicate product specified in K092634, our device and the predicate device are same in Essential Components, raw materials, physical features, and same manufacturing processes. The biocompatibility performance equivalence evidence of proposed electrode can be demonstrated.
And also following the 'Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery, the proposed Disposable Electrosurgical Pencils had evaluated the Performance of Thermal effects on tissues, comparing with reference device specified in K092634. Three tissue types, liver, kidney and muscle tissue, were used for this evaluation by measuring the thermal damage zone sizes (length, width, and depth).
The comparison tests conclusively prove that the proposed electrode is similar / equivalence in the performance of thermal effects on tissues as the predicate device device.
The safety performances of subject device are demonstrated by the Third party, TUV Rheiland, through testing following IEC 60601-1:2005+CORR. 1(2006) + CORR.2(2007), Part 1General requirements for basic safety and essential performance, and IEC 60601-2-2:2009, Medical electrical equipment -Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories.
{6}------------------------------------------------
### CATHAY MANUFACTURING CORP.
The results of bench testing & type testing provide reasonable assurance that the proposed device has been designed and validated to assure conformance to the requirements for its indication for use.
Package integrity and functional performance testing were completed on the subject device following aging and real time stability tests to support the proposed shelf life.
## 6. Conclusion:
Based on the indication for use, technology characteristics, and performance testing, the proposed product, Disposable Electrosurgical Pencils, has been shown to be appropriate for its indication for use and is considered to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.