Total and hemi-hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Total or hemi-hip arthroplasty may be considered for younger patients if, in the opinion of the surgeon, an unequivocal indication for total or hemi-hip replacement outweighs the risks associated with the age of the patient and if limited demands regarding activity and hip joint loading can be assured. This includes severely crippled patients with multiple joint involvement for whom a gain in hip mobility may lead to an expectation of significant improvement in the quality of their lives.
Device Story
Summit Hip System consists of tapered femoral stems for total or hemi-hip arthroplasty. Stems are manufactured from titanium alloy (Ti-6Al-4V) or Cobalt-Chromium-Molybdenum. Porous-coated versions feature sintered commercially pure titanium beads (Porocoat), some with hydroxyapatite (HA) coating. Cemented versions use PMMA centralizers. Device is used by orthopedic surgeons in clinical settings to replace damaged hip joint articulation. System provides increased patient mobility and pain reduction. Output is a mechanical reconstruction of the hip joint. Clinical decision-making relies on surgeon assessment of bone quality and patient activity demands.
Clinical Evidence
No clinical data. Bench testing only, including hip head center analysis, corrosion analysis, range of motion analysis, and pyrogenicity testing.
Technological Characteristics
Materials: Titanium alloy (Ti-6Al-4V), Cobalt-Chromium-Molybdenum, commercially pure titanium bead porous coating (Porocoat), hydroxyapatite (HA) coating. Design: Tapered femoral stems, collarless or flanged/collared, various body sizes and offsets. Fixation: Cemented (with PMMA centralizers) or uncemented (press-fit).
Indications for Use
Indicated for patients requiring total or hemi-hip replacement due to osteoarthritis, rheumatoid arthritis, congenital hip dysplasia, avascular necrosis of the femoral head, acute traumatic fracture of the femoral head/neck, failed previous hip surgery, ankylosis, fracture dislocation, or non-union of femoral neck fractures. Hemi-hip replacement is also indicated for degenerative arthritis involving only the femoral head or pathology of the proximal femur treatable by hemi-arthroplasty. Specific stems are indicated for either cemented or uncemented use.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Depuy Orthopaedics, Inc. Daniel J. Williman Project Manager, Regulatory Affairs 700 Orthopaedic Drive Warsaw, Indiana 46582
April 26, 2018
### Re: K170339
Trade/Device Name: Summit Hip System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, JDI, LWJ, LPH, MEH, KWY, KWL, LZY Dated: March 28, 2018 Received: March 29, 2018
Dear Daniel J. Williman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Katherine D. Kavlock -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170339
Device Name Summit Hip System
Indications for Use (Describe)
Total hip replacement is indicated in the following conditions:
1. A severely painful and/or disabled joint from osteoarthritis, rheumatoid arthritis, or congenital hip dysplasia.
2. Avascular necrosis of the femoral head.
3. Acute traumatic fracture of the femoral head or neck.
4. Failed previous hip surgery including joint reconstruction, arthrodesis, hemiarthroplasty, surface
replacement arthroplasty, or total hip replacement.
5. Certain cases of ankylosis.
Hemi hip replacement is indicated in the following conditions:
1. Acute fracture of the femoral head or neck that cannot be appropriately reduced and treated fixation.
2. Fracture dislocation of the hip that cannot be appropriately reduced and treated with internal fixation.
3. Avascular necrosis of the femoral head.
4. Non-union of femoral neck fractures.
5. Certain high subcapital and femoral neck fractures in the elderly.
6. Degenerative arthritis involving only the femoral head in which the acetabulum does not require replacement.
7. Pathology involving only the femoral head/neck and/or proximal femur that can be adequately treated by hemi-hip arthroplasty.
The SUMMIT POROCOAT Porous, SUMMIT DuoFix Porous, and SUMMIT Basic Press Fit stems are indicated for uncemented use only.
The SUMMIT Cemented and SUMMIT Basic Cemented stems are indicated for cemented use only.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| <span style="display:inline-block; margin-right: 5px;"> <input checked="true" type="checkbox"/> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="display:inline-block; margin-right: 5px;"> <input type="checkbox"/> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Submitter Information | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | DePuy Orthopaedics |
| Address | 700 Orthopedic Drive<br>Warsaw, IN 46582 |
| Phone number | 574-372-7745 |
| Fax number | 574-371-4987 |
| Establishment Registration<br>Number | 1818910 |
| Name of contact person | Daniel J. Williman |
| Date prepared | March 28, 2017 |
| Name of device | |
| Trade or proprietary name | Summit Hip System |
| Common or usual name | Hip Joint Replacement Prosthesis |
| Classification name | Hip joint metal/ceramic/polymer semi-constrained cemented or non-<br>porous uncemented prosthesis; |
| Class | II |
| Classification panel | 87 Orthopedic and Rehabilitation Devices |
| Regulation | 21 CFR 888.3353, 888.3350, 888.3358, 888.3360, and 888.3390 |
| Product Codes | LZO, JDI, LWJ, LPH, MEH, KWY, LZY, KWL |
| Predicate Devices | K001991 Titan Porocoat Hip Prosthesis |
| | K011489 Summit Duofix Hip Prosthesis |
| | K013352 Summit Cemented Hip Prosthesis |
| | K023453 DePuy Cemented FX Cemented Hip Prosthesis |
| | K030122 DePuy Summit Basic Press Fit Hip Prosthesis |
| | K150862 DePuy Actis DuoFix Hip Prosthesis |
| | K123991 DePuy Corail AMT Hip Prosthesis |
| Reason for 510(k) Submission | The purpose of this submission is to append to the existing<br>indication for use statement to include hemi-hip replacement for the<br>Summit Hip System. |
| Device description | The Summit Hip System is series of tapered femoral hip stems<br>which may be used in either total or hemi-hip arthroplasty<br>procedures in conjunction with DePuy hip arthroplasty devices.<br>The Summit Porous Hip Prosthesis is a series of collarless, titanium,<br>tapered, press fit femoral stems The hip stem is manufactured from |
| | titanium alloy (Ti-6Al-4V) and has a sintered commercially pure<br>titanium bead porous coating (Porocoat) applied to the stem. There<br>are 10 body sizes ranging in diameter from 7mm (Size 1) to 18mm<br>(Size 10) with each body size having two offset options. These<br>stems are intended for uncemented use only. |
| | The Summit DuoFix Porous Hip Prosthesis is a series of collarless,<br>titanium alloy, tapered, press-fit femoral stems. The hip stem is<br>manufactured from titanium alloy (Ti-6Al-4V) and has a sintered<br>commercially pure titanium bead porous coating (Porocoat) applied<br>to the stem with a thin layer of hydroxyapatite (HA) coating. There<br>are 10 body sizes ranging in diameter from 7mm (Size 1) to 18mm<br>(Size 10) with each body size having two offset options. These<br>stems are intended for uncemented use only. |
| | The Summit Cemented Hip Prosthesis is series a flanged, collared<br>tapered Cobalt-Chromium femoral stems with a smooth finish.<br>There are seven proportional body sizes with a standard offset, and<br>six proportional body sizes with a high offset. Distal and proximal<br>PMMA centralizers help assure the stem is centered in the femoral<br>canal. These stems are intended for cemented use only. |
| | The Summit Basic Cemented Hip Prosthesis is a series of flanged,<br>collared, tapered Cobalt-Chromium-Molybdenum femoral stems<br>with a smooth surface finish. There are seven total sizes with a<br>constant offset. Only six of the seven sizes are available in the high<br>offset option. A distal PMMA centralizer helps assure that the stem<br>is centered in the femoral canal. These stems are intended for<br>cemented use only. |
| | The Summit Basic Press Fit Hip Prosthesis is a series of collared,<br>tapered Titanium femoral stem with a grit-blasted finish. There are<br>seven sizes with a constant offset and a distal centralizer to assure<br>that the stem is centered in the femoral canal. These stems are<br>intended for uncemented use only. |
| Intended use of the device | Total and hemi-hip arthroplasty is intended to provide increased<br>patient mobility and reduce pain by replacing the damaged hip joint<br>articulation in patients where there is evidence of sufficient sound<br>bone to seat and support the components. |
| | Total or hemi-hip arthroplasty may be considered for younger<br>patients if, in the opinion of the surgeon, an unequivocal indication<br>for total or hemi-hip replacement outweighs the risks associated<br>with the age of the patient and if limited demands regarding activity<br>and hip joint loading can be assured. This includes severely crippled<br>patients with multiple joint involvement for whom a gain in hip<br>mobility may lead to an expectation of significant improvement in<br>the quality of their lives. |
| Indications for use | Total hip replacement is indicated in the following conditions:<br>1. A severely painful and/or disabled joint from osteoarthritis,<br>traumatic arthritis, rheumatoid arthritis, or congenital hip<br>dislocation. |
| dysplasia.<br>2. Avascular necrosis of the femoral head.<br>3. Acute traumatic fracture of the femoral head or neck.<br>4. Failed previous hip surgery including joint reconstruction, internal<br>fixation, arthrodesis, hemiarthroplasty, surface replacement<br>arthroplasty, or total hip replacement.<br>5. Certain cases of ankylosis.<br><br>Hemi hip replacement is indicated in the following conditions:<br>1. Acute fracture of the femoral head or neck that cannot be<br>appropriately reduced and treated with internal fixation.<br>2. Fracture dislocation of the hip that cannot be appropriately reduced<br>and treated with internal fixation.<br>3. Avascular necrosis of the femoral head.<br>4. Non-union of femoral neck fractures.<br>5. Certain high subcapital and femoral neck fractures in the elderly.<br>6. Degenerative arthritis involving only the femoral head in which<br>the acetabulum does not require replacement.<br>7. Pathology involving only the femoral head/neck and/or proximal<br>femur that can be adequately treated by hemi-hip arthroplasty.<br><br>The SUMMIT POROCOAT Porous, SUMMIT DuoFix<br>Porous, and SUMMIT Basic Press Fit stems are indicated for<br>uncemented use only. The SUMMIT Cemented and SUMMIT<br>Basic Cemented stems are indicated for cemented use only. | |
| <b>SUBSTANTIAL EQUIVALENCE INFORMATION</b> | |
| The subject and predicate Summit stem devices are identical in design, materials, and coatings. Non-<br>clinical testing demonstrated performance equivalence between the subject and predicate devices. | |
| <b>SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF<br/>SUBSTANTIAL EQUIVALENCE</b> | |
| To support a determination of substantial equivalence, non-clinical testing/evaluations were conducted. | |
| The following studies were provided:<br>● Hip Head Center Analysis – Demonstrated that Summit stem hemi-arthroplasty constructs do not<br>introduce new fatigue strength risks when compared to Summit stem total hip arthroplasty<br>constructs.<br>● Corrosion Analyses - Demonstrated that Summit stem hemi-arthroplasty constructs pose no<br>greater risk for corrosion when compared to the predicate device Summit and/or Actis stems.<br>● Range of Motion (ROM) Analysis – Concluded that the Summit stem hemi-arthroplasty<br>constructs have ROM equivalent to that of the predicate device Actis stem hemi-arthroplasty<br>constructs.<br>● Pyrogenicity Testing – Concluded that the subject devices meet the required endotoxin limits |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.