← Product Code [FOZ](/productcode/FOZ) · K170336

# BD Nexiva Closed IV Catheter System (Single Port) with MaxZero Needleless Connector (K170336)

_Becton, Dickinson and Company · FOZ · Mar 10, 2017 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170336

## Device Facts

- **Applicant:** Becton, Dickinson and Company
- **Product Code:** [FOZ](/productcode/FOZ.md)
- **Decision Date:** Mar 10, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5200
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Indications for Use

BD Nexiva Closed IV Catheter System - Single Port with MaxZero Needleless Connector devices are intended to be inserted into a patient's vascular system for short term use to sample blood pressure, or administer fluids. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed.

## Device Story

Over-the-needle intravascular catheter system; includes radiopaque polyurethane catheter, needle, needle shield, septum, stabilization platform, extension tubing, clamp, Luer adapter, vent plug, and MaxZero needleless connector. Closed system design contains blood during insertion; septum wipes blood from needle surface upon withdrawal; passive needle tip protection reduces needlestick risk. Features BD Instaflash technology for immediate blood return visualization. Used by clinicians in clinical settings for vascular access. MaxZero connector provides zero-reflux valve mechanism to prevent fluid displacement upon disconnection. Device configuration facilitates fluid administration or blood sampling; benefits include reduced blood exposure and accidental injury risk.

## Clinical Evidence

Bench testing only. Performance testing included insertion force, droplet size, high back pressure, droplet separation, flow rate, plug retention, packaging integrity, and device damage. Biocompatibility testing for the EtO-sterilized MaxZero connector included cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, haemocompatibility, and pyrogenicity (LAL and material-mediated). All tests met predetermined acceptance criteria.

## Technological Characteristics

Materials: Stainless steel needle, polyurethane catheter/tubing, TPE wings, copolyester adapter, polycarbonate hub/shield, acetal clamp, silicone valve. Sensing: Passive needle protection, zero-reflux valve. Dimensions: 18-24 GA; 0.75-1.25 inch lengths. Sterilization: Ethylene oxide (EtO). Connectivity: None. Standalone mechanical device.

## Regulatory Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

## Predicate Devices

- BD Nexiva Closed IV Catheter System - Single Port ([K161777](/device/K161777.md))
- MaxZero Needleless Connector (MZ1000) ([K132413](/device/K132413.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with three heads, representing the department's commitment to health, human services, and science. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 10, 2017

Becton Dickinson Infusion Therapy Systems, Inc. % Mark Job Regulatory Technology Services, LLC 1394 25th Street North West Buffalo, Minnesota 55313

Re: K170336

Trade/Device Name: BD Nexiva™ Closed IV Catheter System-Single Port with MaxZero™ Needleless Connector Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ, FPA Dated: March 7, 2017 Received: March 8, 2017

Dear Mr. Mark Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Mark Job

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Image /page/1/Picture/8 description: The image shows a signature of Tina Kiang. The signature is in black ink and is on a white background. There is a large looping design to the left of the name. There is also a small "-s" below the name.

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### Device Name

BD Nexiva Closed IV Catheter System - Single Port with MaxZero Needleless Connector

#### Indications for Use (Describe)

BD Nexiva Closed IV Catheter System - Single Port with MaxZero Needleless Connector devices are intended to be inserted into a patient's vascular system for short term use to sample blood pressure, or administer fluids. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18 GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed.

| Type of Use (Select one or both, as applicable)                                      |                                                                                     |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Becton Dickinson Infusion Therapy Systems Inc. 9450 South State Street Sandy, UT 84070 USA

bd.com

Image /page/3/Picture/2 description: The image shows the logo for BD, a medical technology company. The logo consists of two parts: a circular orange symbol on the left and the letters "BD" in blue on the right. The orange symbol appears to be a stylized sun or starburst design. The letters "BD" are in a bold, sans-serif font.

## K170336 510(k) Summary (21 CFR §807.92) BD Nexiva™ Closed IV Catheter System – Single Port with MaxZero™ Needleless Connector

| Submitter<br>Information                            | Submitter Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Becton Dickinson Infusion Therapy Systems Inc.                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                              |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                     | Submitter Address:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 9450 South State Street<br>Sandy, UT 84070                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Contact Person:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Kimberly Geisler<br>Regulatory Affairs Manager                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Email Address:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | kimberly.geisler@bd.com                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Phone Number:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | (801) 565-2422                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Fax Number                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | (801) 304-3963                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Date of Preparation:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | March 3, 2017                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                              |
| Subject Device                                      | Trade Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | BD Nexiva™ Closed IV Catheter System – Single Port<br>with MaxZero™ Needleless Connector                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Common Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Peripheral Intravascular or IV Catheter                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Regulation Number:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 21 CFR §880.5200                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Regulation Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Intravascular Catheter                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Regulatory Class:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | II                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Product Code:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | FOZ (primary); FPA (secondary)                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Classification Panel:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | General Hospital                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                              |
| Predicate<br>Device<br>(BD Nexiva)                  | Trade Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | BD Nexiva™ Closed IV Catheter System – Single Port                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | 510(k) Reference:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K161777                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Common Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Peripheral Intravascular or IV Catheter                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Regulation Number:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 21 CFR §880.5200                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Regulation Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Intravascular Catheter                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Regulatory Class:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | II                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Product Code:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | FOZ                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Classification Panel:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | General Hospital                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                              |
| Predicate<br>Device<br>(MaxZero)                    | Trade Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | MaxZero™ Needleless Connector (MZ1000)                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | 510(k) Reference:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | K132413                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Common Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Intravascular Administration Set or Needleless Connector                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Regulation Number:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 21 CFR §880.5440                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Regulation Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Intravascular Administration Set                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Regulatory Class:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | II                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Product Code:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | FPA                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | Classification Panel:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | General Hospital                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                              |
| Reason for<br>Submission                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | The reason for this submission is the addition of a MaxZero Needleless Connector<br>to the BD Nexiva Closed IV Catheter System - Single Port unit package.                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                              |
| Device<br>Description                               | BD Nexiva Closed IV Catheter System - Single Port with MaxZero Needleless<br>Connector devices are over-the-needle, intravascular catheters. These devices<br>have a radiopaque BD Vialon catheter, needle, needle shield, septum, stabilization<br>platform, integrated extension tubing, clamp, Luer adapter, and vent plug. The<br>needle and catheter are protected by a needle cover. A MaxZero device with<br>protective cover is provided in the unit package.                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | The closed system is designed to keep blood contained within the device<br>throughout the insertion process. The septum is designed to wipe visible blood<br>from the needle surface as the needle is withdrawn from the catheter, further<br>reducing the risk of blood exposure. The needle tip is passively protected when<br>the needle is removed, reducing the risk of accidental needlestick injury.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | These devices have BD Instaflash needle technology, allowing for immediate<br>visualization of blood return along the catheter. Continuous blood return is seen in<br>the extension tubing. The vent plug prevents blood leakage from the extension<br>tubing during insertion. Both the stabilization platform and Luer connector are color<br>coded to indicate catheter gauge size (24 GA (0.7 mm)=Yellow, 22 GA (0.9<br>mm)=Blue, 20 GA (1.1 mm)=Pink, 18 GA (1.3 mm)=Green).                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                              |
| Indications for<br>Use<br>(21 CFR<br>§807.92(a)(5)) | The subject device Indications for Use is the same as the predicate BD Nexiva<br>Closed IV Catheter System - Single Port device (K16177), with the exception of<br>the device trade name. The subject device Indications for Use does not include the<br>predicate MaxZero Needleless Connector Indications for Use (K132413) since the<br>MaxZero Needleless Connector is a component of the primary BD Nexiva Closed<br>IV Catheter System - Single Port device. The addition of the MaxZero Needleless<br>Connector to the BD Nexiva Closed IV Catheter System unit package does not<br>impact the intended use of the MaxZero Needleless Connector compared to the<br>predicate device (K132413). As such, the differences between the subject and<br>predicate device Indications for Use do not change the intended use of the<br>predicate devices or raise different questions of safety and effectiveness. |                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                              |
|                                                     | SUBJECT<br>BD Nexiva Closed IV Catheter System -<br>Single Port with MaxZero Needleless<br>Connector<br>Indications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | PREDICATE 1<br>BD Nexiva Closed IV Catheter<br>System - Single Port (K161777)<br>Indications for Use                                                                                                                                                                                 | PREDICATE 2<br>MaxZero Needleless<br>Connector (K132413)<br>Indications for Use                                                                                                                                                                                                                                                                              |
| Attribute                                           | SUBJECT<br>BD Nexiva Closed IV Catheter<br>System - Single Port with<br>MaxZero Needleless<br>Connector                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | PREDICATE 1<br>BD Nexiva Closed IV Catheter<br>System - Single Port<br>(K161777)                                                                                                                                                                                                     | PREDICATE 2<br>MaxZero Needleless<br>Connector (K132413)                                                                                                                                                                                                                                                                                                     |
| Owner                                               | Becton Dickinson Infusion<br>Therapy Systems Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Becton Dickinson Infusion<br>Therapy Systems Inc.                                                                                                                                                                                                                                    | Becton, Dickinson and<br>Company (formerly<br>CareFusion)                                                                                                                                                                                                                                                                                                    |
| Classification                                      | 21 CFR §880.5200<br>Class II<br>FOZ - Intravascular Catheter<br>(primary)<br>FPA - IV Administration Set<br>(secondary)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 21 CFR §880.5200<br>Class II<br>FOZ - Intravascular Catheter                                                                                                                                                                                                                         | 21 CFR §880.5440<br>Class II<br>FPA - IV Administration Set                                                                                                                                                                                                                                                                                                  |
| 510(k) Status                                       | Subject of this premarket<br>notification                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | K161777 - Clearance date<br>August 29, 2016                                                                                                                                                                                                                                          | K132413 - Clearance date<br>August 29, 2013                                                                                                                                                                                                                                                                                                                  |
| Fundamental<br>Scientific<br>Technology             | BD Nexiva Closed IV Catheter<br>System - Single Port device<br>Same as Predicate 1<br>(BD Nexiva SP)<br>MaxZero Needleless<br>Connector device<br>Same as Predicate 2<br>(MaxZero)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Closed peripheral intravascular<br>catheter system designed with<br>an integrated extension set<br>incorporating a single port<br>injection site. Incorporates BD<br>Instaflash™ technology to assist<br>with flashback visualization.                                               | Closed, Luer activated<br>needleless connector<br>designed to prevent fluid<br>displacement into the<br>catheter upon disconnection<br>of the male user (i.e., zero<br>reflux).  During activation, a<br>male Luer lock connector<br>causes the valve to deflect<br>and enables fluid to flow<br>freely around the valve and<br>out the male Luer connector. |
| Primary<br>Device<br>Components<br>/ Materials      | BD Nexiva Closed IV Catheter<br>System - Single Port device<br>Same as Predicate 1<br>(BD Nexiva Closed IV Catheter<br>System - Single Port)<br>MaxZero Needleless<br>Connector device<br>Same as Predicate 2<br>(MaxZero Needleless Connector)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Needle:<br>Stainless Steel<br>Catheter:<br>Polyurethane<br>Catheter Wings:<br>TPE<br>Catheter Adapter:<br>Copolyester<br>Needle Hub:<br>Polycarbonate<br>Tip Shield:<br>Polycarbonate<br>Extension Tubing:<br>Polyurethane<br>Luer Adapter:<br>Copolyester<br>Pinch Clamp:<br>Acetal | Top:<br>Polycarbonate<br>Polycarbonate<br>Base:<br>Silicone<br>Valve:                                                                                                                                                                                                                                                                                        |

{4}------------------------------------------------

bd.com

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{5}------------------------------------------------

#### bd.com

#### Technological Technological characteristics of the subject and predicate devices are substantially Characteristics equivalent with respect to the device design and materials. The subject BD Nexiva Closed IV Catheter System - Single Port with MaxZero Needleless Connector achieves its intended use based on the same technology and principles of operation as the predicate BD Nexiva Closed IV Cather System - Single Port and MaxZero Needleless Connector, respectively.

The BD Nexiva Closed IV Catheter System - Single Port is being offered with a MaxZero Needleless Connector in the unit package as a convenience to the clinician. This configuration represents a change the MaxZero sterilization method from irradiation (e-beam) to ethylene oxide (EtO).

A comparison of the subject and predicate device technological characteristics is provided in the table below.

{6}------------------------------------------------

Becton Dickinson Infusion Therapy Systems Inc. 9450 South State Street Sandy, UT 84070 USA

#### bd.com

| Attribute              | SUBJECT<br>BD Nexiva Closed IV Catheter<br>System - Single Port with<br>MaxZero Needleless<br>Connector | PREDICATE 1<br>BD Nexiva Closed IV Catheter<br>System - Single Port<br>(K161777)                 | PREDICATE 2<br>MaxZero Needleless<br>Connector (K132413) |
|-------------…

---

**Source:** [https://fda-staging.innolitics.com/device/K170336](https://fda-staging.innolitics.com/device/K170336)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
