The dynaMX™ Intramedulary Fixation Device is indicated to stabilize and aid in the fixation of fractures, fusions, and osteotomies of the phalanges.
Device Story
The dynaMX™ Intramedullary Implant is a single-use, cannulated bone fixation device made of Nitinol. It features a threaded distal end and a barbed proximal end. During surgery, the threaded end is inserted into a pre-drilled hole in the middle phalange, while the barbed end is inserted into a broached canal in the proximal phalange. The barbs deflect inward during insertion and superelastically expand to lock the implant in place. The device is used in an OR setting by a surgeon to stabilize phalangeal fractures or fusions. It is supplied in a kit with a drill bit, broach, and guide wire. The implant provides permanent fixation, aiding in bone healing and alignment.
Clinical Evidence
Bench testing only. Studies included static bending, bending fatigue, pull-out force, rotational stability, corrosion testing, transformation temperature determination, package seal strength, and shelf-life verification. Biocompatibility of Nitinol is well-established via reference literature. Endotoxin testing confirmed the device is non-pyrogenic.
Indicated for stabilization and fixation of fractures, fusions, and osteotomies of the phalanges in patients requiring small bone reconstruction.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Sterile Kirschner Wires and Steinmann Pins (K960385)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, composed of three human profiles facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus, indicating the department's name and national affiliation.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Arthrex, Inc. %Howard Schraver Mx Orthopedics. Corp. 1050 Waltham Street. Suite 510 Lexington, Massachusetts 02421 July 20, 2017
Re: K170326
Trade/Device Name: dynaMXTM Intramedullary Implant Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HTY Dated: June 20, 2017 Received: June 21, 2017
Dear Mr. Schrayer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### INDICATIONS FOR USE
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K170326
Device Name
dynaMXTM Intramedullary Fixation Device
Indications for Use (Describe)
The dynaMX™ Intramedulary Fixation Device is indicated to stabilize and aid in the fixation of fractures, fusions, and osteotomies of the phalanges.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (101) 443-6700 BF
{3}------------------------------------------------
#### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (Per 21 CFR 807.92)
### General Company Information
| Name: | Arthrex, Inc.<br>Originally, MÖ Orthopedics, Corp. |
|------------|----------------------------------------------------|
| Contact: | Howard Schrayer<br>Regulatory Affairs Consultant |
| Address: | 1050 Waltham St. Suite 510<br>Lexington, MA 02421 |
| Telephone: | (617) 270 - 6608 |
- April 24, 2017 Date Prepared
#### General Device Information
- Product Name: dynaMŸ™ Intramedullary Implant
- Classification: "Bone Fixation Fastners" Product code: HTY - Class II 21 CFR 888.3040
#### Predicate Device
| DePuy (Primary Predicate) | Sterile Kirschner Wires and Steinmann Pins |
|---------------------------|--------------------------------------------|
| | [510(k) Number K960385] |
| OrthoHelix | Intraosseous Fixation System |
| | [510(k) Number K120165] |
| Wright Medical | PRO-TOE® Hammertoe Fixation System |
| | [510(k) Number K140148] |
| BioMedical Enterprises | HammerLock® 2 |
| | [510(k) Number K133520] |
| Memometal Technologies | Smart Toe and X-Fuse |
| | [510(k) Number K070598] |
{4}------------------------------------------------
#### Description
The dynaMX™ Intramedullary Implant is intended to stabilize and aid in the fixation of fractures, fusions, and osteotomies of the phalanges,
- . The implant is fully cannulated with threads on the distal side and two barbs on the proximal side. The implant is offered in multiple combinations of diameter and length to accommodate various anatomies.
- . The dynaMÄ™ Intramedullary Implant is made of biocompatible Nitinol. The barbs of the implant are designed to exhibit superelastic properties at room and body temperature.
- . As manufactured, one end of the implant has a screw thread, and the other end a pair of barbs. The screw thread is inserted into a drilled hole in the middle phalange. The barbed end is implanted into a broached canal in the proximal phalange. While inserting the barbed end of the implant, the barbs deflect inward to allow the implant to pass through the broached canal. Once inside the canal, the barbs superelastically return to their flared outward configuration to securely hold the implant in the bone.
- . The dynaMÄ™ Intramedullary Implant is sold as part of a single use disposable kit. The kit includes an implant, pre-loaded onto a delivery device, a drill bit, a broach and a quide wire.
#### Indications for Use
The dynaMX™ Intramedullary Implant is intended to stabilize and aid in the fixation of fractures, fusions, and osteotomies of the phalanges,
#### Substantial Equivalence
A series of laboratory studies (bench tests and cadaver tests) have been conducted to verify the suitability of the dynaMX™ Intramedullary Implant for its intended use, establish Substantial Equivalence with the predicate devices and confirm reproducibility of the packaging.
These tests include:
Elastic Static Bending Testing Bending Fatigue Testing Implant Pull-Out Force Rotational Stability Corrosion Testina Transformation Temperature Determination Package Seal Strength Verification Shelf Life
{5}------------------------------------------------
The biocompatibility of Nitinol has been well-established. A reference publication that describes the biocompatibility was appended, together with a copy of the shelf-life / stability protocol that has been designed to support expiry dating of up to 36 months.
Endotoxin testing of the dynaMX™ Intramedullary Implant has been conducted to confirm that the implant is non-pyrogenic.
This submission supports the position that the MX Orthopedics dynaMX™ Intramedullary Implant is substantially equivalent to previously cleared devices, including the devices listed above. A number of predicate devices list the same range of clinical uses.
#### Conclusions
MX Orthopedics, Corp. believes that the information provided establishes that similar legally marketed devices have been used for the same clinical applications as the MX Orthopedics dynaMX™ Intramedullary Implant. The materials from which the MX Orthopedics device is fabricated have an established history of use, and the devices have been tested in accordance with applicable FDA guidelines.
{6}------------------------------------------------
## SUBSTANTIAL EQUIVALENCE INFORMATION
| M <b>X</b> Orthopedics,<br>Corp. | DePuy Inc. | OrthoHelix | Wright Medical | BioMedical<br>Enterprises | Memometal<br>Technologies | |
|----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|--|
| dynaMX™<br>Intramedullary Implant | Sterile Kirschner Wires<br>and Steinmann Pins | Intraosseous Fixation<br>System | PRO-TOE®<br>Hammertoe Fixation<br>System | HammerLock® 2 | Smart Toe and X-Fuse® | |
| 510(k) K170326 | 510(k) K960385 | 510(k) K120165 | 510(k) K140148 | 510(k) K133520 | 510(k) K070598 | |
| Similarities | | | | | | |
| Device is indicated for<br>stabilizing and aiding in<br>the fixation of fractures,<br>fusions, and<br>osteotomies of the<br>phalanges. | Device is indicated for<br>fixation of bone<br>fractures, for bone<br>reconstruction, as guide<br>pins for insertion of<br>other implants. | Device is intended to<br>stabilize and aid in the<br>fixation of fractures,<br>fusions, and<br>osteotomies of the<br>phalanges. | Device is indicated for<br>the fixation of<br>osteotomies and<br>reconstruction of the<br>lesser toes following<br>correction procedures. | Device is indicated for<br>small bone<br>reconstruction and<br>fusion such as inter-<br>digital fusion of fingers<br>and toes. | Device is indicated for<br>small bone<br>reconstruction limited to<br>inter-digital fusion of<br>fingers and toes and<br>small bones fusions. | |
| Product code: HTY -<br>Class II | Product code: HTY and<br>JDW - Class II 21 CFR | Product code: HTY -<br>Class II 21 CFR | Product code: HWC<br>and JDW - Class II 21 | Product code: HTY -<br>Class II 21 CFR | Product code: HTY -<br>Class II 21 CFR | |
| 21 CFR 888.3040 | 888.3040 | 888.3040 | CFR 888.3040 | 888.3040 | 888.3040 | |
| Prescription use in OR<br>setting | Prescription use in OR<br>setting | Prescription use in OR<br>setting | Prescription use in OR<br>setting | Prescription use in OR<br>setting | Prescription use in OR<br>setting | |
| Implant is barbed on<br>one side, threaded on<br>the other | Device may be<br>threaded or smooth | Implant is barbed on<br>one side, threaded on<br>the other | Implant is barbed on<br>one side, threaded on<br>the other | Implant is barbed on<br>both sides | Implant is barbed on<br>both sides | |
| Middle phalange<br>engaged with 4.0 to<br>4.5mm thread diameter. | Range in diameter from<br>0.7mm to 4.7mm | Middle phalange<br>engaged with 4.2 to<br>4.5mm thread diameter | Middle phalange<br>engaged with 4.0mm<br>thread diameter. | Middle phalange<br>engaged with 5.0 to<br>7.0mm superelastic<br>barbs | Middle phalange<br>engaged with 6.6mm<br>superelastic barbs | |
| Proximal phalange<br>engaged with 6.3mm<br>superelastic barbs. | Range in diameter from<br>0.7mm to 4.7mm | Proximal phalange<br>engaged with 2.2 to<br>2.9mm barbs. | Proximal phalange<br>engaged with 2.0 to<br>3.5mm screw threads. | Proximal phalange<br>engaged with 5.0 to<br>6.0mm barbs. | Proximal phalange<br>engaged with 6.0 to<br>7.0mm shape memory<br>base. | |
| Polished and<br>chemically passivated | Polished and<br>chemically passivated | Polished and<br>chemically passivated | Polished and<br>chemically passivated | Polished and<br>chemically passivated | Polished and<br>chemically passivated | |
| Device is one-<br>component construction | Device is one-<br>component construction | Device is one-<br>component construction | Device is one-<br>component construction | Device is one-<br>component construction | Device is one-<br>component construction | |
| Device is delivered into<br>pre-drilled and pre-<br>broached channels | No pre-drilling or<br>broaching required | Device is delivered into<br>pre-drilled and pre-<br>broached channels | Device is delivered into<br>pre-drilled and pre-<br>broached channels | Device is delivered into<br>pre-drilled and pre-<br>broached channels | Device is delivered into<br>pre-drilled and pre-<br>broached channels | |
| Repaired toe is straight | Repaired toe is straight | Repaired toe is straight | Repaired toe is straight<br>or 10° | Repaired toe is straight<br>or 10° | Repaired toe is straight<br>or 10° | |
| Device is intended as<br>permanent implant | Device is intended as<br>temporary implant | Device is intended as<br>permanent implant | Device is intended as<br>permanent implant | Device is intended as<br>permanent implant | Device is intended as<br>permanent implant | |
| Device is radiopaque | Device is radiopaque | Device is radiopaque | Device is radiopaque | Device is radiopaque | Device is radiopaque | |
{7}------------------------------------------------
#### Differences
| Device is fabricated<br>from Nitinol per ASTM<br>F2063-12 | Device is fabricated<br>from 316 LVM Stainless<br>Steel | Device is fabricated<br>from stainless steel | Device is fabricated<br>from stainless steel or<br>titanium | Device is fabricated<br>from Nitinol per ASTM<br>F2063-12 | Device is fabricated<br>from Nitinol per ASTM<br>F2063-12 |
|-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|------------------------------------------------------------|-------------------------------------------------------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Nitinol material<br>provides fixation to<br>repair | Nitinol not used | Nitinol not used | Nitinol not used | Nitinol material<br>provides fixation to<br>repair | Nitinol material<br>provides fixation to<br>repair |
| Device is provided with<br>single-use, sterile<br>accessory delivery tools | Device provided without<br>accessories | Part of a reusable kit | Part of a reusable kit | Device is provided with<br>single-use, sterile<br>accessory delivery tools | Part of a reusable kit |
| Implant is available in<br>13mm, 14mm, and<br>15mm lengths | Device is available in<br>lengths ranging from 4"<br>to 9" | Implant is available in<br>12mm, 13mm, and<br>14mm lengths | Implant is available in<br>13mm and 16mm<br>lengths. | Implant is available in<br>17mm to 22mm<br>lengths. | Implant is available in<br>15mm to 22mm<br>lengths. |
| Initial proposed shelf-<br>life 6 months following<br>completion of initial<br>shelf-life study | Shelf-life 3 years | Shelf-life 3 years | Shelf-life 3 years | Shelf-life 5 years | Shelf-life 3 years |
| Device is provided<br>sterile by gamma<br>radiation and stored at<br>room temperature for<br>single-patient-use | Device is provided<br>sterile or non-sterile. | Device is provided non-<br>sterile. | Device is provided non-<br>sterile. | Device is provided<br>sterile and stored at<br>room temperature for<br>single patient use. | Device is provided<br>sterile and stored cold<br>for single patient use. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.