K170291 · Soterix Medical, Inc. · EGJ · May 1, 2017 · Physical Medicine
Device Facts
Record ID
K170291
Device Name
IontoDC
Applicant
Soterix Medical, Inc.
Product Code
EGJ · Physical Medicine
Decision Date
May 1, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5525
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
IontoDC is intended to use a direct current to introduce ions of soluble salts or other drugs into the body.
Device Story
IontoDC is a bench-top, battery-powered, microcontroller-controlled iontophoresis device. It delivers direct current to a drug delivery electrode placed on the patient's skin to expedite the migration of ionic drug solutions into localized tissue. The device features two output jacks, a digital display, and shielded banana cable connectors. It provides current intensity between 1.0-2.0 mA DC with 1% accuracy. Safety features include low battery warnings and overcurrent protection. The device is intended for prescription use. By facilitating transdermal drug delivery, it serves as an alternative to hypodermic injection, potentially avoiding needle-related pain or tissue trauma.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical output parameter verification, safety testing per IEC 60601-1, and EMC testing per IEC 60601-1-2.
Technological Characteristics
Bench-top, battery-powered (two 9V alkaline batteries). Microcontroller-controlled. Delivers 1.0-2.0 mA DC. Voltage range 0-40V. Shielded banana electrode cable jacks. Digital display. Complies with IEC 60601-1 and IEC 60601-1-2. Software is classified as Moderate level of concern.
Indications for Use
Indicated for the administration of soluble salts or other drugs into the body via iontophoresis using direct current.
Regulatory Classification
Identification
An iontophoresis device is a device that is intended to use a direct current to introduce ions of soluble salts or other drugs into the body and induce sweating for use in the diagnosis of cystic fibrosis or for other uses if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug. When used in the diagnosis of cystic fibrosis, the sweat is collected and its composition and weight are determined. Additionally, an iontophoresis device intended for any other purposes is a prescription device that is intended to use a current to introduce ions of drugs or non-drug solutions into the body for medical purposes other than those specified in paragraph (a) of this section, meaning that the device is not intended for use in diagnosis of cystic fibrosis, or a specific drug is not specified in the labeling of the iontophoresis device.
Special Controls
*Classification.* Class II (special controls). The device is classified as class II. The special controls for this device are:(i) The following performance testing must be conducted:
(A) Testing using a drug approved for iontophoretic delivery, or a solution if identified in the labeling, to demonstrate safe use of the device as intended;
(B) Testing of the ability of the device to maintain a safe pH level; and
(C) If used in the ear, testing of the device to demonstrate mechanical safety.
(ii) Labeling must include adequate instructions for use, including sufficient information for the health care provider to determine the device characteristics that affect delivery of the drug or solution and to select appropriate drug or solution dosing information for administration by iontophoresis. This includes the following:
(A) A description and/or graphical representation of the electrical output;
(B) A description of the electrode materials and pH buffer;
(C) When intended for general drug delivery, language referring the user to drug labeling approved for iontophoretic delivery to determine if the drug they intend to deliver is specifically approved for use with that type of device and to obtain relevant dosing information; and
(D) A detailed summary of the device-related and procedure-related complications pertinent to use of the device, and appropriate warnings and contraindications, including the following warning:
*Warning:* Potential systemic adverse effects may result from use of this device. Drugs or solutions delivered with this device have the potential to reach the blood stream and cause systemic effects. Carefully read all labeling of the drug or solution used with this device to understand all potential adverse effects and to ensure appropriate dosing information. If systemic manifestations occur, refer to the drug or solution labeling for appropriate action.(iii) Appropriate analysis/testing must demonstrate electromagnetic compatibility, electrical safety, thermal safety, and mechanical safety.
(iv) Appropriate software verification, validation, and hazard analysis must be performed.
(v) The elements of the device that may contact the patient must be demonstrated to be biocompatible.
(vi) The elements of the device that may contact the patient must be assessed for sterility, for devices labeled as sterile.
(vii) Performance data must support the shelf life of the elements of the device that may be affected by aging by demonstrating continued package integrity and device functionality over the stated shelf life.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of what appears to be an eagle.
May 1, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Soterix Medical,inc. Abhishek Datta Chief Technology Officer 237 W 35 St. 1401 New York, New York 10001
Re: K170291
Trade/Device Name: IontoDC Regulation Number: 21 CFR 890.5525 Regulation Name: Iontophoresis Device Regulatory Class: Class II Product Code: EGJ Dated: January 31, 2017 Received: January 31, 2017
Dear Abhishek Datta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Our substantially equivalent decision does not apply to the drugs that you will label or promote for use with your device. Therefore, you may neither label nor promote your device for use with specific drugs, nor package drugs with your device prior to FDA having approved the drugs for iontophoretic administration. For information on the requirements for marketing new drugs, you may contact:
Director Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857
As you are aware, there are concerns relating to the fact that no drug is currently labeled for administration via an iontophoresis device. The Agency currently is evaluating this public health concern regarding the safety and effectiveness of this route of administration of drugs, and in the near future will inform manufacturers of certain additional steps the Agency believes are necessary to address this concern.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
{2}------------------------------------------------
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K170291
Device Name IontoDC
Indications for Use (Describe)
IontoDC is intended to use a direct current to introduce ions of soluble salts or other drugs into the body.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) SUMMARY
Date Prepared: December 26, 2016 Submitter Information: Company Name: Soterix Medical, Inc. 237 W 35 ST. 1401 Company Address: New York, NY 10001 Contact Person: Abhishek Datta Phone: 888-990-8327 Fax: 212-315-3232 Device Information: IontoDC Trade Name: Common Name: IontoDC Classification Name: Iontophoresis Device (21 CFR 890.5525, Product Code EGI) Device Class: Class II Predicate Device: 1. Dynatron® ibox Iontophoresis device (K043047) Dynatronics Corporation Class III (primary predicate ) 2. ActivaDose II (K020476) Selective Med Components, Inc. acquired by Activatek Inc. Class III (reference device) Device Description: IontoDC is a bench-top, battery powered, microcontrollercontrolled iontophoresis device that delivers direct current to a drug delivery electrode placed on the patient's skin. This current expedites the migration of ionic drug solutions through the skin into localized tissue. Intended Use: IontoDC is intended to use a direct current to introduce ions of soluble salts or other drugs into the body. IontoDC uses the same technological principle as the Technological Comparison: predicate device to accomplish its intended use, namely the introduction of ions to the body through direct current. The technology is based on the principle that an electric potential will cause ions in solution to migrate according to their
{5}------------------------------------------------
electrical charges. A technical comparison of the electrical output parameters of the device with the predicate device has been completed and supports a substantial equivalence determination.
### Basis for Equivalence:
| -Performance testing: | Bench testing demonstrated that the electrical output parameters of the IontoDC device are substantially equivalent to those of the predicate device. The subject device complies with the IEC 60601-1 standard for safety and the IEC 60601-1-2 standard for EMC. Software verification and validation testing were conducted and documented per FDA's Guidance for Industry and Staff. The software for this device was considered as a "Moderate" level of concern. |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| -Labeling: | The labeling of IontoDC device is substantially equivalent to that of the predicate device. |
| -Discussion of Clinical Tests Performed: | Not Applicable. |
| Conclusions from Testing: | The intended use and the basic technological characteristic of the IontoDC device are equivalent with those of the predicate device. The IontoDC device complies with the requirement of IEC 60601-1 and IEC 60601-1-2. Bench testing and safety report documentation supports the conclusion that electrical output is substantially equivalent to the predicate device, and any differences between the devices do not pose new questions of safety and effectiveness. Thus IontoDC device is substantially equivalent to the predicate device. |
| Parameter | IontoDC | ActivaDose II<br>(Reference Device) | Dynatron® IBox<br>(Primary<br>Predicate) | Comparison | |
|-------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|-------------------------------------|
| 510(k) | To Be Determined | K020476 | K043047 | -- | |
| Device Name | IontoDC | ActivaDose II | Dynatron® ibox | -- | |
| Manufacturer | Soterix Medical,<br>Inc. | ActivaTek. Inc. | Dynatronics<br>Corporation | -- | |
| Parameter | IontoDC | ActivaDose II<br>(Reference Device) | Dynatron® IBox<br>(Primary<br>Predicate) | Comparison | |
| Indications For<br>Use | IontoDC is<br>intended to use a<br>direct current to<br>introduce ions of<br>soluble salts or<br>other drugs into<br>the body. | The ActivaDose II<br>Iontophoresis<br>Delivery Unit is<br>indicated for the<br>administration of<br>soluble salts or other<br>drugs into the body<br>for medical purposes<br>as an alternative to<br>hypodermic<br>injection in<br>situations when it is<br>advisable to avoid<br>the pain that may<br>accompany needle<br>insertion and drug<br>injection, when it is<br>advisable to<br>minimize the<br>infiltration of carrier<br>fluids, or to avoid<br>the damage caused<br>by needle insertion<br>when tissue is<br>traumatized. | Dynatron® ibox<br>iontophoresis<br>System is intended<br>to use a direct<br>current to<br>introduce ions of<br>soluble salts or<br>other drugs into the<br>body. | Identical to<br>Primary<br>Predicate | |
| Power source | Battery Powered | Battery Powered | Battery Powered | Identical | |
| Number, Size, and<br>Type of Batteries | Two 9 Volt<br>Alkaline batteries | One 9V DC<br>Alkaline Battery | Two 1.5 Volt AA<br>Alkaline batteries | SE<br>Note 1 | |
| Output Jacks | 2 | 1 | 2 | Identical to<br>Primary<br>Predicate | |
| Current Intensity | 1.0 - 2.0 mA DC | 0.0 - 4.0 mA DC | 0.5 - 4.0 mA DC | SE<br>Note 2 | |
| Current Accuracy | +/- 1% | -- | +/- 10% | SE | |
| Parameter | IontoDC | ActivaDose II<br>(Reference Device) | Dynatron® IBox<br>(Primary<br>Predicate) | Comparison | |
| | | | | Note 3 | |
| Current Precision | 0.1 mA | 0.1 mA | 0.1 mA | Identical | |
| Voltage at<br>Electrodes | +/- 0-40 Volts | 80 Volts | +/- 0-46 Volts | SE<br>Note 4 | |
| Electrode Cable<br>Jacks | Shielded banana | -- | Custom 2-pin<br>polarized | SE<br>Note 5 | |
| Maximum Total<br>Deliverable Dose | 80 mA-min | 80 mA-min | 3360 mA-min | Identical to<br>Reference<br>Device | |
| Single Treatment<br>Deliverable Dose | 10mA - 80 mA-<br>min | 0 - 80 mA-min | 1 - 160 mA-min | SE<br>Note 6 | |
| Number of 40 mA-<br>min. Treatments | 2 | 2 | 89 | Identical to<br>Reference<br>Device | |
| Minimum<br>Treatment Time | 10 min | 1.5 seconds | 15 seconds | SE<br>Note 7 | |
| Maximum<br>Treatment Time | 40 min | 20 minutes | 100 minutes | SE<br>Note 8 | |
| Timing Accuracy | +/- 3 second | +/- 1 second | +/- 1 second | SE<br>Note 9 | |
| Polarity<br>Options: | Positi<br>ve (+) | YES | YES | YES | Identical |
| | Negat<br>ive (-) | YES | YES | YES | Identical |
| | Dual<br>Polar<br>ity | NO | NO | YES | Identical to<br>Reference<br>Device |
| Parameter | | IontoDC | ActivaDose II<br>(Reference Device) | Dynatron® Ibox<br>(Primary Predicate) | Comparison |
| | (+/-) | | | | |
| Low<br>Battery<br>Warning: | Durin<br>g<br>Treat<br>ment | YES | YES | YES, at 2.0 V<br>remaining | Identical |
| | At<br>Powe<br>r-On | YES | YES | YES, if <2.2 V<br>remaining | Identical |
| Overcurrent<br>Warning | | YES, if >2.2 mA | YES, at 4 mA | Yes, if >6 mA on<br>any channel | SE<br>Note 10 |
| Certificate<br>s: | IEC<br>60601<br>-1 | YES | | YES | Identical |
| | CSA/<br>NRT<br>L | YES | | YES | Identical |
| Dimensions (in.) [L<br>x W x H] | | 7.91" x 5.9" x<br>2.83" | 6.1" x 3.5" x 1.9" | 4.5" x 2.66" x<br>1.25" | SE<br>Note 11 |
| Weight (lbs., oz.)<br>[Without battery<br>or electrode<br>cables] | | 1 lb 3oz | .4 lbs | 3.6 oz. | SE<br>Note 12 |
| Display | | Digital Display | Digital Display | Organic Light<br>Emitting Diodes<br>(OLED) | Identical to<br>Reference<br>Device |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
#### Note 1:
The number of batteries used in subject device is different than the predicate device but all devices comply to IEC 60601-1 requirements. So the difference will not raise any safety issue.
#### Note 2:
The subject device can deliver a maximum of 2 mA of current as opposed to a maximum 4mA delivered by each of the predicates but all devices comply to IEC 60601-1 requirements. The differences in functional specifications will not raise any new safety and effectiveness issues.
{9}------------------------------------------------
#### Note 3:
The subject device has a current accuracy of 1% which is different than the 10% accuracy provided by the predicate. While both devices comply to IEC 60601-1 requirements, increased current accuracy in subject device ensures more accurate measurements. The differences in functional specifications will not raise any new safety and effectiveness issues.
#### Note 4:
The voltage provided by the subject device is slightly different than the predicate devices and is the lowest of all. All devices comply with IEC 60601-1 requirements so the differences in functional specifications will not raise any new safety and effectiveness issues.
#### Note 5:
The electrode cable jacks of the subject device is shielded banana while it is 2-pin polarized jack for the predicate device. Both jacks ensure a reliable connection so the difference does not raise new questions of safety and effectiveness.
#### Note 6:
The single treatment deliverable dose range for the subject device is slightly different than the predicate device. The upper limit of the total deliverable dose however does not exceed any of the predicate devices. This difference in functional specification will not raise any safety and effectiveness issue.
#### Note 7:
The minimum treatment time of the subject device is more than the predicate devices but this can be manually adjusted by pressing abort for any desired treatment time. This difference in functional specification does not raise any new safety and effectiveness issue.
#### Note 8:
The maximum treatment time of the subject device is more than one predicate device but less than the other. The treatment time can however be manually adjusted by pressing abort for any desired treatment time. This difference in functional specification does not raise any new safety and effectiveness issue.
#### Note 9:
The timing accuracy of the subject device is different than the predicate devices. This difference does not raise new safety and effectiveness issue for iontophoretic use.
#### Note 10:
The subject device can deliver a maximum of 2 mA and as a result the overcurrent warning threshold is is different than the predicate devices. This difference in functional specification will not raise a safety and effectiveness issue.
#### Note 11:
The subject device has different dimensions than the predicate devices. This difference in functional specification will not raise a safety and effectiveness issue.
#### Note 12:
The subject device has a different weight than the predicate devices. This difference in functional specification will not raise a safety and effectiveness issue.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.