The Dental Implants OKTAGON® Bone Level Tapered Design are surgically placed in the maxillary and/or mandibular arches to provide support for prosthetic restorations in edentulous patients. The implants are intended to be used with OKTAGON® abutments and prosthetic parts. The implants are intended for delayed loading in a two-stage surgery or for immediate loading when good primary stability is achieved and with appropriate occlusal load.
Device Story
Endosseous dental implants; surgically placed in maxillary/mandibular arches; provide support for prosthetic restorations (crowns, bridges, dentures). Device features conical coronal region; cylindrical apical region; micro-structured surface (grit-blasted/acid-etched); polished shoulder. Prosthetic connection via inner cone with octagonal anti-rotation device. Includes sterile cover screw for internal thread occlusion. Used by dental clinicians in clinical settings. Benefits include restoration of chewing function and oral rehabilitation for edentulous/partially edentulous patients.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated through bench testing, including fatigue testing (ISO 14801), biocompatibility (ISO 10993-1, ISO 7405), cytotoxicity (ISO 10993-5), and sterility/packaging validation (ISO 11137-1/2, ISO 11607-1, ASTM F88/F1929).
Technological Characteristics
Material: Commercially pure Titanium Grade 4 (ASTM F67). Design: Root-form, conical coronal/cylindrical apical shape. Surface: Micro-structured (grit-blasted/acid-etched). Connection: Inner cone with octagonal anti-rotation. Dimensions: Diameters 3.3-4.1 mm, lengths 8-12 mm. Sterilization: Validated per ISO 11137-1/2.
Indications for Use
Indicated for edentulous or partially edentulous patients requiring prosthetic restoration support via surgical placement in maxillary and/or mandibular arches. Suitable for delayed loading (two-stage surgery) or immediate loading given sufficient primary stability and appropriate occlusal load.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
Dental Implant System OKTAGON® Bone Level (K143539)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is a symbol of medicine and healing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 24, 2017
Hager & Meisinger GmbH Melanie May Regulatory Affairs Hansemannstrasse 10 Neuss. 41468 GERMANY
Re: K170287
Trade/Device Name: Dental Implants OKTAGON® Bone Level Tapered Design Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: May 4, 2017 Received: May 8, 2017
Dear Melanie May:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Lori A. Wiggins -S
for
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K170287
#### Device Name
Dental Implants OKTAGON® Bone level Tapered Design
Indications for Use (Describe)
The Dental Implants OKTAGON® Bone Level Tapered Design are surgically placed in the maxillary and/or mandbular arches to provide support for prosthetic restorations in edentulous patients. The implants are intended to be used with OKTAGON® abutments and prosthetic parts. The implants are intended for delayed loading in a two-stage surgery or for immediate loading when good primary stability is achieved and with appropriate occlusal load.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Dental Implants OKTAGON® Bone Level Tapered Design
### 510(k) Summary
#### 1. Applicant's Name and Address
| Hager & Meisinger GmbH |
|-------------------------------------------------------|
| Hansemannstrasse 10 |
| 41468 Neuss |
| Germany |
| Phone: 0049 2131 2012 292 |
| Fax: 0049 2131 2012 223 |
| Contact Person: Dr. Melanie May<br>Regulatory Affairs |
#### 2. Date prepared
Date prepared:
05/23/2017
#### 3. Name of the device
| Trade Name: | Dental Implants OKTAGON® Bone level Tapered Design |
|----------------------|----------------------------------------------------|
| Common Name: | Endosseous dental implants |
| Classification Name: | Endosseous dental implants |
| Product Code: | DZE |
| Regulation No: | 872.3640 |
| Class: | II |
| Panel: | Dental |
#### 4. Predicate Devices:
| 510(k) No. | Manufacturer | Trade Name |
|------------|------------------------|-------------------------------------------|
| K143539 | Hager & Meisinger GmbH | Dental Implant System OKTAGON® Bone Level |
| K140878 | Straumann USA,LLC | Straumann® Bone Level Tapered Implants |
#### 5. Device Description:
The Dental Implants OKTAGON® Bone Level Tapered Design are an addition to the currently distributed Dental Implant System® OKTAGON Bone Level. The implants can be used for immediate or early implantation after loss or extraction of natural teeth.
The endosteal implant body of the subject device has a conical shape in the coronal region and is cylindrical in the apical region. The apical cylindrical part of the implant is relevant for the given endosteal diameter of 3.3 to 4.1 mm. The implants are made of commercially pure Titanium Grade 4 conforming to ASTM- F67. The surface is micro-structured in the endosted part and has been blasted with high-grade corundum and afterwards acid-etched. The implant shoulder is polished.
{4}------------------------------------------------
#### Dental Implants OKTAGON® Bone Level Tapered Design
#### 510(k) Summary
The prosthetic connection is achieved with the help of an inner cone with an additional octagonal antirotation device. A sterile cover screw is enclosed with the implant so that an immediate occlusion of the internal thread is possible after successful insertion.
The submission contains the following tapered design implant variations:
BLT NC Ø 3.3 mm, L 8.0 mm BLT NC Ø 3.3 mm, L 10.0 mm BLT NC Ø 3.3 mm, L 12.0 mm
BLT NC Ø 3.75 mm, L 8.0 mm BLT NC Ø 3.75 mm, L 10.0 mm BLT NC Ø 3.75 mm, L 12.0 mm
BLT NC Ø 4.1 mm, L 8.0 mm BLT NC Ø 4.1 mm, L 10.0 mm BLT NC Ø 4.1 mm, L 12.0 mm
#### 6. Indications for Use:
The Dental Implants OKTAGON® Bone Level Tapered Design are surgically placed in the maxillary and/or mandibular arches to provide support for prosthetic restorations or partially edentulous patients. The implants are intended to be used with OKTAGON® abutments and prosthetic parts. The implants are intended for delayed loading in a two-stage surgery or for immediate loading when good primary stability is achieved and with appropriate occlusal load.
#### 7. Performance tests and used standards
Information regarding biocompatibility testing, sterility validation and shelf life was leveraged from the predicate device KI43539.
The following standards have been followed for the development, production, performance and safety testing of the subject devices.
- Fatigue Testing according to ISO 14801 ●
- Evaluation of Biocompatibility according to ISO 10993-1, ISO 7405
- Cytotoxicity Testing according to ISO 10993-5
- Sterilization Validation for implants and accessories according to ISO I I I 37-1, ISO I I I I 37-2,
{5}------------------------------------------------
## Dental Implants OKTAGON® Bone Level Tapered Design
#### 510(k) Summary
- Sterile Barrier Testing according to ASTM F88/F88M-09 and ASTM F1929-98 ●
- Packaging of sterilized implants according to ISO I 1607-1
- . Material Specification according to ASTM F67
- Risk Management according to ISO I 497 I
In order to demonstrate substantial equivalence performance tests (breakage and fatigue tests) have been conducted fulfilling the requirements of ISO 14801 and the FDA's "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments".
#### 8. Basis for substantial equivalence
The intended use for Dental Implant OKTAGON® Bone Level Tapered Design is identical to the named predicated devices. The implants have the same material composition and the same surface treatments. In addition, the implants are of the same, root-form design
Since the Dental Implants OKTAGON® Bone Level Tapered Design and the predicate implants have the same design, are made of the same materials (see table), exhibit the same performance characteristics (e.g. strength and function), and are furthermore intended for the same use, no new performance or safety issues are raised.
| Manufacturer | Hager & Meisinger GmbH | Hager & Meisinger GmbH | Straumann USA, LLC | Manufacturer | Hager & Meisinger GmbH | Hager & Meisinger GmbH | Straumann USA, LLC |
|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|----------------------------------------------------|-------------------------------------------------------|--------------------------------------------------------------------|
| 510(k) Number | unknown | K143539 (Primary Predicate Device) | K140878 (Reference Predicate Device) | Material | Titanium Grade 4 | Titanium Grade 4 | Titanium<br>Grade<br>4<br>Titanium-13Zirconium<br>alloy (Roxolid®) |
| Intended Use | The Dental Implants<br>OKTAGON® Bone Level<br>Tapered Design are surgically<br>placed in the maxillary and/or<br>mandibular arches to provide<br>support for prosthetic<br>restorations in edentulous or<br>partially edentulous patients.<br>The implants are intended to<br>be used with OKTAGON®<br>abutments and prosthetic<br>parts. The implants are<br>intended for delayed loading in<br>a two-stage surgery or for<br>immediate loading when good<br>primary stability is achieved<br>and with appropriate occlusal<br>load. | The implants are surgically<br>placed in the maxillary and/<br>or mandibular arches to<br>provide support for<br>prosthetic restorations in<br>edentulous or partially<br>edentulous patients. The<br>parts are intended to be<br>used with OKTAGON®<br>Bone Level abutments and<br>prosthetic parts. The<br>Dental Implant system<br>OKTAGON® Bone Level is<br>intended for delayed<br>loading, or for immediate<br>loading when good primary<br>stability is achieved and<br>with appropriate occlusal<br>load. | Straumann® dental<br>implants are indicated for<br>oral endosteal implantation<br>in the upper and lower jaw<br>and for the functional and<br>esthetic oral rehabilitation<br>of edentulous and partially<br>dentate patients.<br>Straumann dental implants<br>can also be used for<br>immediate or early<br>implantation following<br>extraction or loss of natural<br>teeth. Implants can be<br>placed with immediate<br>function on single-tooth<br>and/or multiple tooth<br>applications when good<br>primary stability is achieved<br>and with appropriate | Surface<br>endosseous | Micro-structured (grit blasted<br>and acid etched) | Micro-structured<br>(grit<br>blasted and acid etched) | SLA (grit blasted then acid<br>etched), SLA active |
| Manufacturer | Hager & Meisinger GmbH | Hager & Meisinger GmbH | Straumann USA, LLC | Connection | Inner cone with octagonal<br>anti-rotation device | Inner cone with octagonal<br>anti-rotation device | Inner cone with octagonal<br>anti-rotation device |
| | | | occlusal loading, to restore<br>chewing function. The<br>prosthetic restorations used<br>are single crowns, bridges<br>and partial or full dentures,<br>which are connected to the<br>implants by the<br>corresponding elements<br>(abutments). In cases of<br>fully edentulous patients, 4<br>or more implants must be<br>used in immediately loaded<br>cases. | Length | 8 to 12 mm | 8 to 14mm | 8 to 16 mm |
| Description | The endosteal implant body of<br>the subject device has a<br>conical shape in the coronal<br>region and is cylindrical in the<br>apical region. The apical<br>cylindrical part of the implant<br>is relevant for the given<br>endosteal diameter of 4.1<br>mm. The implants are made<br>of commercially pure Titanium<br>Grade 4 conforming to ASTM-<br>F67. The surface is micro-<br>structured in the endosteal<br>part and has been blasted<br>with high-grade corundum and<br>afterwards acid-etched. The<br>implant shoulder is polished.<br>The prosthetic connection is<br>achieved with the help of an<br>inner cone with an additional<br>octagonal anti-rotation device.<br>A sterile cover screw is<br>enclosed with the implant so<br>that an immediate occlusion<br>of the internal thread is<br>possible after successful<br>insertion.<br>Subject to this submission is a<br>tapered design implant<br>variation of 4.1 mm, which is<br>available in a range of<br>endosseous lengths. | The root-form designed<br>implant is made of<br>commercially pure Titanium<br>Grade 4 conforming to<br>ASTM Standard<br>Specification F67. The<br>surface is micro-structured<br>in the endosteal section and<br>the implant shoulder is<br>polished. The implant<br>surface has been blasted<br>with high-grade corundum<br>and afterwards acid-etched.<br>The prosthetic connection is<br>achieved with the help of an<br>inner cone with an<br>additional octagonal anti-<br>rotation device.<br>A sterile locking screw is<br>enclosed with the implant<br>so that an immediate<br>occlusion of the internal<br>thread is allowed after<br>successful insertion.<br>The endosseous dental<br>implants are available in a<br>range of endosseous<br>diameters and lengths. | The devices represent a line<br>extension of the previously<br>cleared Bone Level implants<br>of the Straumann Dental<br>Implant System (K062129,<br>K083550 and K121131).<br>The apical part of the<br>implants has a tapering<br>thread form and three<br>cutting flutes.<br>The implants are made of<br>Titanium Grade 4<br>respectively Titanium-<br>13Zirconium alloy<br>(Roxolid®). The surface<br>finishes are either SLA (grit<br>blasted then acid etched)<br>or SLAactive<br>The implant-to-abutment<br>connection (NC, RC) is<br>achieved by an inner cone<br>with octagonal anti-<br>rotation device.<br>The devices are available in<br>the endosseous diameters<br>3.3, 4.1 and 4.8 mm. | Endosseous<br>Diameter | 3.3 to 4.1 mm | 3.3 to 4.8 mm | 3.3 to 4.8 mm |
{6}------------------------------------------------
## Dental Implants OKTAGON® Bone Level Tapered Design
## 510(k) Summary
{7}------------------------------------------------
## Dental Implants OKTAGON® Bone Level Tapered Design
## 510(k) Summary
Based on these observations, we conclude that the Dental Implant OKTAGON® Bone Level Tapered Design are substantially equivalent to the legally marketed predicate devices described.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.