This device is intended for monitoring PEF (Peak Expired Flow Rate) and FEV 1 (Forced Expiratory Volume in one second) for patient home use. The device is designed for pediatric to adult patients. The device is intended for monitoring respiratory conditions such as asthma.
Device Story
Hand-held pulmonary function meter; measures peak expiratory flow (PEF) and forced expiratory volume in 1 second (FEV1) via turbine sensor; flow integration used for volume calculation. Designed for home use by pediatric to adult patients. Device stores up to 250 recordings; transfers data via Bluetooth (BLE 4.0) to mobile app for tracking. Healthcare providers use tracked data to monitor respiratory conditions like asthma. Benefits include longitudinal tracking of lung function to assist in disease management.
Clinical Evidence
No clinical data. Bench testing only. Performance testing evaluated accuracy, intra-instrument repeatability, inter-instrument repeatability, and flow resistance per ATS 2005 standards. Biocompatibility testing (cytotoxicity, irritation, sensitization) performed per ISO 10993. Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Hand-held device; turbine sensor; PP mouthpiece. Measures PEF (50-900 L/min) and FEV1 (0.01-9.99 L). Power: 2x AAA 1.5V alkaline batteries. Connectivity: BLE 4.0. Software: Moderate level of concern. Standards: ATS 2005, IEC 60601-1, IEC 60601-1-2, ISO 10993.
Indications for Use
Indicated for pediatric to adult patients for home monitoring of respiratory conditions such as asthma, specifically measuring Peak Expired Flow Rate (PEF) and Forced Expiratory Volume in one second (FEV1).
Regulatory Classification
Identification
A peak-flow meter for spirometry is a device used to measure a patient's maximum ventilatory flow rate.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is a symbol of medicine and healing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 6, 2017
Beijing M&B Electronic Instruments Co., Ltd. % Mike Gu Regulatory Affairs Manager Guangzhou Osmunda Medical Device Technical Services Co.,Ltd. 8-9Th Floor, R&D Building, No.26 Qinglan St, Panyu District Guangzhou. 510006 Cn
Re: K170281
Trade/Device Name: MSA100BT Peak Flow Meter Regulation Number: 21 CFR 868.1860 Regulation Name: Peak-Flow Meter for Spirometry Regulatory Class: Class II Product Code: BZH Dated: September 1, 2017 Received: September 5, 2017
Dear Mike Gu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Tara A. Ryan -S
for
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170281
Device Name MSA100BT Peak Flow Meter
Indications for Use (Describe)
This device is intended for monitoring PEF (Peak Expired Flow Rate) and FEV 1 (Forced Expiratory Volume in one second) for patient home use. The device is designed for pediatric to adult patients. The device is intended for monitoring respiratory conditions such as asthma.
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a blue logo. The logo appears to be a stylized flower or plant with two petals or leaves extending upwards and outwards. There is a registered trademark symbol (®) located to the upper right of the logo.
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
# 1. SUBMITTER
Beijing M&B Electronic Instruments Co., Ltd.
Room 6319, Building 1, No. 27, Yongwang Road, Beijing Bioengineering and Medicine Industry Base, Huangcun Town, Daxing district, Beijing, P.R. China
Phone: +86-10-61253716 Fax: +86-10-61253794
| Contact Person: | Yang Shisheng |
|-----------------|----------------------------------------------|
| | Project Manager |
| | Beijing M&B Electronic Instruments Co., Ltd. |
| | Tel: (+86)-10-61253716-8037 |
| | Fax: (+86)-10-61253794 |
Date prepared Oct. 03, 2017
- 2. DEVICE
| Device Name: | MSA100BT Peak Flow Meter |
|--------------------|-----------------------------------------|
| Common/Usual Name: | Peak Flow Meter |
| Regulation number | 868.1860 Peak-flow meter for spirometry |
| Regulation Class: | II |
| Product Code: | BZH Meter, Peak Flow, Spirometry |
- 3. PREDICATE DEVICE
K133975, MSA100 Peak Flow Meter
This predicate legally marketed device has not been subject to a design-related recall.
No reference device was used in this submission.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with a stylized blue symbol that resembles a flame or a stylized flower. The symbol has a flowing, upward motion. To the upper right of the symbol is a circled "R", indicating that the logo is a registered trademark.
#### 4. DEVICE DESCRIPTION
MSA100BT Peak Flow Meter is a hand-held pulmonary function measuring medical device that measures patient's maximum possible exhalation which is called peak expiratory flow (PEF) and forced expiratory volume in 1 second (FEV1). The accuracy meets American Thoracic Society (ATS) Standard 2005 Revision.
The MSA100BT is designed for pediatric to adult patients for home use and for single user, and is intended for monitoring respiratory conditions such as asthma.
The MSA100BT has an automatic memory of 250 recordings and can connect to cell phone by Bluetooth. Patients can transfer the records to phone via BLE4.0, then using app to help save and track long term.
The MSA100BT is capable of measuring PEF from 50 to 900 L/min and FEV1 from 0.01 to 9.99 L. Its measuring accuracy is ± 20 L/min or ± 10% of the reading for PEF, and ± 0.050 L or ± 3% of the reading for FEV1.
The MSA100BT is composed of main unit, turbine sensor, and mouth piece. After each use, the mouthpiece and turbine of the device should be cleaned by soap solution and water within 30 minutes.
### 5. INDICATIONS FOR USE
This device is intended for monitoring PEF (Peak Expired Flow Rate) and FEV1 (Forced Expiratory Volume in one second) for patient home use. The device is designed for pediatric to adult patients. The device is intended for monitoring respiratory conditions such as asthma.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a blue logo. The logo appears to be a stylized representation of a plant or flower with three petals or leaves. There is a registered trademark symbol to the right of the logo.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE 6. PREDICATE DEVICE
| Specification | Predicate MSA100 Peak<br>Flow Meter | Proposed<br>MSA100BT Peak Flow Meter | Discussion of<br>Differences |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|-------------------------------------------------------------------------------------------------------|
| K number | K133975 | | |
| Device name | MSA100 Peak Flow Meter | MSA100BT Peak Flow Meter | |
| Manufacturer | Beijing M&B Electronic Instruments Co., Ltd | | Identical |
| Indications for use | The device is intended for monitoring Peak Expired Flow<br>(PEF) and Forced Expiratory Volume in one second (FEV1)<br>for patient home use. The device is intended for monitoring<br>respiratory conditions such as asthma. | | Identical |
| Patient Population | The device is designed for pediatric to adult patients | | Identical |
| Environment of Use | OTC | | Identical |
| Materials | PP (mouthpiece) | | Identical |
| Measuring range | PEF from 50 to 900 L/min<br>FEV1 from 0.01 to 9.99 L | | Identical |
| Measuring method | Flow: Turbine sensor<br>Volume: Flow Integration | | Identical |
| Accuracy | PEF ± 20 L/min or PEF ± 10% of the reading;<br>FEV1 ± 0.05 L or FEV1 ± 3% of the reading | | Identical |
| Measuring solution | PEF 1 L/min; FEV1 0.01 L | | Identical |
| Data safety | Data rememorized by flash memory | | Identical |
| Data transmission | USB wire transmission | BLE wireless transmission | Different<br>This difference does not<br>raise different questions<br>of safety and<br>effectiveness. |
| Memory volume | 300 recordings | 250 recordings | Different<br>This difference does not<br>raise different questions<br>of safety and<br>effectiveness. |
| Power source | AAA 1.5×2 alkaline batteries | | Identical |
| Electrical Safety and<br>EMC | IEC 60601-1 and IEC 60601-1-2 | | Identical |
### Table 6.1 Comparison with the Predicate Device
According to the above comparison, the proposed device is identical to the predicate device in indications for use, materials, measuring method, technical specification, power source and electrical safety.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a stylized, abstract blue flower. The flower has three petals that are curved and appear to be blowing in the wind. The flower is set against a white background. In the upper right corner of the image, there is a small, circular icon with a symbol inside.
The differences are the way of data transmission and memory, which are USB wire transmission and 300 recordings for the predicate device and BLE (Blue Smart) wireless transmission and 250 recordings for the proposed device. The differences will not raise different questions of safety and effectiveness, as both devices were demonstrated to be electromagnetic compatibility and electrical safety by testing.
In conclusion, Beijing M&B Electronic Instruments Co., Ltd. believes that the MSA100BT Peak Flow Meter is as safe and effective, and performs in a substantially equivalent manner to the predicate device.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a blue logo on a white background. The logo appears to be a stylized flower or plant with three distinct petals or leaves at the top and a stem extending downwards. There is a small circle with the registered trademark symbol (®) in the upper right corner of the logo.
#### 7. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
The mouthpiece of the MSA100BT Peak Flow Meter is considered to contact directly with the human body for a duration of less than 24 hours.
The mouthpiece of the MSA100BT Peak Flow Meter in its final finished form is identical to the mouthpiece of the MSA100 Peak Flow Meter in formulation and processing, and no other chemicals have been added.
The biocompatibility evaluation for the mouthpiece was conducted in accordance with the International Standard ISO 10993-1:2009/(R)2013, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as recognized by FDA. The biocompatible testing included the following tests:
- Cytotoxicity
- Irritation
- Skin sensitization
Testing for cytotoxicity, skin sensitization and oral mucosa irritation complied with ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity and ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization. Testing demonstrates that the test article is non-cytotoxic, non-sensitizing, and non-oral-mucosal-irritating.
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the MSA100BT Peak Flow Meter. The device complies with the IEC 60601-1:2012, standard for electrical safety and the IEC 60601-1-2:2014 standard for EMC. It demonstrates substantial equivalences to the predicate device.
Wireless testing was conducted on the MSA100BT Peak Flow Meter. The device complies with FDA's guidance: Guidance Radio Frequency Wireless Technology in Medical Devices Guidance for Industry and Food and Drug Administration Staff.
# Performance testing
Performance testing was conducted on the MSA100BT Peak Flow Meter. Technical parameters of accuracy, intra instrument repeatability, PEF inter instrument repeatability, performance of flow resistance were evaluated in the performance
{8}------------------------------------------------
testing, according to Standards of Spirometry (2005 Revision). All of the tested parameters meet the requirements in the standards. The performance of the proposed device is demonstrated to be comparable with the predicate device in the comparison testing, so it is concluded that the proposed device is substantially equivalent to the predicate device.
# Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern. Software validation demonstrated that the software functions as specified in the software requirement specifications. It demonstrates substantial equivalences to the predicate device.
# Animal Study
The subject of this premarket submission, MSA100BT Peak Flow Meter, did not require animal studies to support substantial equivalence.
### Clinical Study
The subject of this premarket submission, MSA100BT Peak Flow Meter, did not require clinical studies to support substantial equivalence.
#### 8. CONCLUSION
The differences between the MSA100BT Peak Flow Meter and its predicate device do not raise different questions of safety and effectiveness. The non-clinical data including the performance testing supports that the MSA100BT Peak Flow Meter is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.