K170262 · K-Jump Health Co., Ltd. · FLL · Sep 28, 2017 · General Hospital
Device Facts
Record ID
K170262
Device Name
Digital Thermometer
Applicant
K-Jump Health Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Sep 28, 2017
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Digital Thermometer is intended to measure oral temperature of human body. It can be used in home environment or at a doctor's clinic. It is indicated for use by people of all ages.
Device Story
Digital Thermometer (Model 2180) is a battery-powered, handheld probe device using a thermistor sensor to detect human body thermal energy. The probe connects to a mobile device via a headphone jack. A mobile application (iOS/Android) processes the electronic signal from the probe, converts it to a temperature reading, and displays the result on the mobile screen. The application provides measurement activity, historical data, trend analysis, and settings. Used by patients or clinicians in home or clinical environments. The device aids in temperature monitoring, allowing users to track health trends via the mobile interface.
Clinical Evidence
Bench testing only. Compliance verified against AAMI/ANSI ES60601-1 (electrical safety), AAMI/ANSI/IEC 60601-1-2 (EMC), and ASTM E1112-00 (performance/accuracy). Biocompatibility testing (ISO 10993-1, -5, -10) confirmed no cytotoxicity, sensitization, or irritation.
Technological Characteristics
Thermistor-based sensor; internal SR41 battery power; dimensions 87.5 x 14.2 x 8.7 mm; weight 6g. Connectivity via mobile device headphone jack. Software-based display on iOS/Android platforms. Complies with ASTM E1112-00, IEC 60601-1, and IEC 60601-1-2. Biocompatible materials.
Indications for Use
Indicated for oral temperature measurement in humans of all ages in home or clinical settings.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
K-jump Health Co., Ltd. Basal Digital Thermometer (K141505)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for K-Jump. The logo is written in a stylized, sans-serif font, and the color is blue. The "K" and "jump" are connected, and there is a registered trademark symbol to the right of the word "jump".
### 510(k) Summary
### 1. Submitter
K-jump Health Co., Ltd.
No. 56, Wu Kung 5th Rd. New Taipei Industrial Park New Taipei City 24890, Taiwan
Contact person: JM Lin, Regulatory Affairs Representative +886-2-2299-1378 ext. 237 Phone: Fax: +886-2-2299-1385 Email: jm@kjump.com.tw
Date Prepared: January 25, 2017
# 2. Name of Device
Trade Name: Digital Thermometer, Model 2180 Common/Usual Name: Digital Thermometer Regulation Number: 21 CFR 880.2910 Classification Name: Clinical electronic thermometer Regulatory Class: II Product Code: FLL
# 3. Predicate Device
| Device Name | 510(k) Number | Decision Date |
|------------------------------------------------------|---------------|---------------|
| K-jump Health Co., Ltd.<br>Basal Digital Thermometer | K141505 | 8/26/2014 |
This predicate has not been subject to a design-related recall.
{1}------------------------------------------------
#### 4. Device Description
The Digital Thermometer consists two parts. The thermometer probe part is a compact, small, light-weight, and battery-powered unit with thermometer sensor for oral temperature measurements. The thermometer probe must be connected to a compatible mobile phone device through its headphone jack for displaying measured temperatures.
The mobile application could be conducted for operation and display temperature measurement when the Digital Thermometer connects with mobile device. The mobile application can be conducted on iOS operating system and Android operating system of mobile device.
The mobile application could perform the functionality on mobile device screen including measurement activity, historical data activity, trend activity, basic setting and initial setting activity.
#### 5. Indications for Use
The Digital Thermometer is intended to measure oral temperature of human body. It can be used in home environment or at a doctor's clinic. It is indicated for use by people of all ages.
#### 6. Technological Characteristics
The Digital Thermometer uses a thermistor sensor to detect the thermal energy from human body. The sensor in the thermometer detects temperature change and transfers to the proper electronic signal. The electronic signal could be received and converted to the temperature reading. The measured data will be displayed on the screen of a mobile device. The technological characteristics of Digital Thermometer have no significant difference with predicate device.
#### 7. Comparison of technological characteristics with Predicate Device
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for K-Jump. The logo is written in a stylized blue font. There is a registered trademark symbol to the right of the word jump.
The Digital Thermometer is substantially equivalent to the predicate devices, K141505, the Basal Digital Thermometer. The comparison of their technological characteristics is summarized in the table below.
| Characteristics | Digital Thermometer | Basal Digital Thermometer |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K170262 | K141505 |
| Intended Use | Similar | Similar |
| Technology | | |
| Measurement method | Thermistor detection | Thermistor detection |
| Feature | Compatible with iOS / Android<br>Mobile Device via headphone jack<br>connection<br>Measured data display on the<br>screen of mobile device | Compatible with iOS/Android<br>Mobile Device via<br>headphone/NFC connection<br>Measured data transfer to<br>mobile device |
| Function | | |
| Measuring Range | 32.0°C~43.9°C | 32.00°C~43.90°C |
| C/F Switchable | Yes | Yes |
| Performance | | |
| Display | 3 1/2 digits<br>on the screen of a mobile device | 4 1/2 digits<br>on the LCD of medical device |
| Display Resolution | 0.1°C (°F) | 0.01°C (°F) |
| Measuring accuracy | ±0.1°C (35.5°C to 42.0°C);<br>±0.2°C (other range) | ±0.1°C (34.00°C to 42.00°C);<br>±0.2°C (other range) |
| Low battery detection | Yes | Yes |
| Memory | 1 set memory | 40 set memory |
| Environmental | | |
| Operating condition | 10°C~40°C (50 ~ 104°F)<br>15%-95% RH | 5°C~40°C (41.0 ~ 104°F)<br>15%-95% RH |
| Storage condition | -20°C ~50°C (-4.0 ~ 122°F)<br>15%-95% RH | -25°C ~55°C (-13.0 ~ 131°F)<br>15%-95% RH |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for K-Jump. The logo is blue and features a stylized "K" with a dot above it, resembling a person jumping. The word "Jump" is written in a cursive font next to the "K", and there is a registered trademark symbol next to the word "Jump".
| Characteristics | Digital Thermometer | Basal Digital Thermometer | |
|------------------|----------------------------------------------------------------|----------------------------------------------------------------|-----------------------------------------------------|
| Material | Component materials | Stainless Steel,<br>Acrylonitrile butadiene styrene | Stainless Steel,<br>Acrylonitrile butadiene styrene |
| Biocompatibility | No Cytotoxicity<br>No Sensitization &<br>Irritation-negligible | No Cytotoxicity<br>No Sensitization &<br>Irritation-negligible | |
| Power | Internal power<br>(one SR41 battery) | Internal power<br>(one CR2032 battery) | |
| Physical | Dimensions | 87.5 x 14.2 x 8.7 mm | 125.5 x 34 x 9.6 mm |
| Weight | 6g (with battery) | 24g (with battery) | |
#### > Discussion of Main Differences:
The Digital Thermometer connects to the mobile device through headphone jack. The measured data will be displayed on the screen of a mobile device. The predicate device could display measured data on the device and transfer measured data to a mobile device.
While connected with mobile device, the usage of Digital Thermometer is identical to the predicate device. Risk analysis, software validation and associated verification have been performed and demonstrated that the display feature of the subject device is acceptable;
#### 8. Performance Summary
The performance of the Digital Thermometer is verified and validated to comply with following recognized standards and does not raise any safety and effectiveness issues.
- 1. AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012,, C1:2009/(R)2012 and A2:2010/(R)2012 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for K-Jump. The logo is written in a stylized blue font. There is a registered trademark symbol to the right of the word jump.
The Digital Thermometer complies with to applicable ANSI/AAMI ES60601-1 requirements including general requirements, protection against electrical hazards, protection against mechanical hazards, protection against excessive temperatures, hazardous situations and fault conditions, and constructions.
- 2. AAMI / ANSI/ IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
The Digital Thermometer complies with applicable ANSI/AAMI/IEC 60601-1-2 requirements including radiated emission test, electrostatic discharge immunity test, radiated RF electromagnetic field immunity test, and power frequency magnetic field immunity test.
- 3. ASTM E1112-00 (Reapproved 2011), Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature
The Digital Thermometer complies applicable ASTM E1112 requirements including displayed temperature range, maximum permissible laboratory error, ambient conditions for operation and storage, human interface and mechanical test.
- 4. Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
The Digital Thermometer complicable the guidance requirements including level of concern, software description, software requirements specification, architecture design chart, software design specification, verification and validation documentation.
- 5. Biocompatibility:
AAMI / ANSI / ISO 10993-1:2009(R)2013 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process;
AAMI / ANSI / ISO 10993-5:2009(R)2014 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity and
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for K-Jump. The logo is written in a stylized blue font. To the right of the word "K-Jump" is the registered trademark symbol.
AAMI / ANSI / ISO 10993-10:2010(R)2014 Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization
The Digital Thermometer complies applicable biocompatibility requirements and the device met the endpoints for cytotoxicity, sensitization & irritation-negligible
The Digital Thermometer is verified and validated by evaluating safety, EMC, device-specific standard ASTM E1112-00, software validation and biocompatibility testing without raising any safety and effectiveness issues.
## 9. Conclusions
The Digital Thermometer has the similar intended use, materials, technological characteristics and the same fundamental scientific technology with the predicate device. Moreover, both devices comply with same performance standards. All information described above can demonstrate the subject device is substantial equivalent to the predicate device.
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a flowing, abstract design connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 28, 2017
K-jump Health Co., Ltd. JM Lin Regulatory Affairs Representative No. 56. Wu Kung 5th Rd., New Taipei Industrial Park New Taipei City, 24890 TAIWAN
Re: K170262
Trade/Device Name: Digital Thermometer, Model 2180 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: August 30, 2017 Received: August 31, 2017
Dear JM Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.