K170260 · Oriental Inspiration Limited · OLP · Apr 25, 2017 · General, Plastic Surgery
Device Facts
Record ID
K170260
Device Name
BC-001+ Acne Purifier
Applicant
Oriental Inspiration Limited
Product Code
OLP · General, Plastic Surgery
Decision Date
Apr 25, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BC-001+ Acne Purifier is intended/indicated for over-the-counter use in the treatment of mild to moderate acne.
Device Story
Hand-held, battery-powered LED device for OTC treatment of mild to moderate acne; emits narrow-band red (660nm) and blue (415nm) light; light penetrates 3-5mm into skin; device features power switch, treatment button, battery indicator, and micro-USB charging; user-operated by placing device on skin; treatment duration up to 1.5 minutes per area; low-level light therapy intended to improve acne condition; benefits include non-invasive, home-based acne management.
Clinical Evidence
Bench testing only. Evidence includes biocompatibility (ISO 10993-1: cytotoxicity, sensitization, irritation), electrical safety (IEC 60601-1-11, IEC 60601-2-57, IEC 62471, IEC 62133), EMC (EN 60601-1-2), software verification/validation, and usability testing per FDA guidance.
Technological Characteristics
Hand-held LED device; 660nm (red) and 415nm (blue) wavelengths; lithium-ion rechargeable battery; dimensions 70mm x 70mm x 23mm; treatment area 9.1 cm²; timer up to 1.5 min/area; software level of concern: moderate; compliant with IEC 60601-1-11, IEC 60601-2-57, IEC 62471, EN 60601-1-2, and IEC 62133.
Indications for Use
Indicated for over-the-counter treatment of mild to moderate acne in patients.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
LightStim for Acne/ LightStim for Acne Mini (K142246)
Sonilase Blue-UV Clean Plus Model-SL113BUV+ (K152889)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 25, 2017
Oriental Inspiration Limited Andy Wu Director Unit 607, The Lakeside 2, West Wing, No. 10 Science Park West Avenue Hong Kong, CN
Re: K170260
Trade/Device Name: BC-001+ Acne Purifier Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OLP Dated: November 10, 2016 Received: January 27, 2017
Dear Andy Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170260
Device Name BC-001+ Acne Purifier
Indications for Use (Describe)
BC-001+ Acne Purifier is intended/indicated for over-the-counter use in the treatment of mild to moderate acne.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
001_510 (k) Summary Version: A/3
Department of Health and Human Services Centre of Device and Radiological Health Office of Device Evaluation Traditional 510(k) section
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
as required by section 21 CFR 807.92
#### Submitter of 510(K): 1.
| Company Name : | Oriental Inspiration Limited |
|------------------|--------------------------------------------------------------------------------------------------------------------------|
| Address : | Unit 607, The Lakeside 2, West Wing, No. 10 Science Park<br>West Avenue, Hong Kong Science Park, Shatin, N.T. Hong Kong. |
| Contact person : | Mr. Andy Wu |
| TEL : | +(852) 2654 0872 |
| FAX : | +(852) 2687 4148 |
| E-mail : | andy_wu@netop.com.hk |
| Date Prepared : | 25th April, 2017 |
#### Proposed Device and code: 2.
| Device Trade Name | BC-001+ Acne Purifier |
|------------------------------|---------------------------|
| Regulation Medical Specialty | General & Plastic Surgery |
| Product Code | OLP |
| Regulation number | 878.4810 |
| Device Class | 2 |
#### 3. Predicate Device:
| 510(K) | Trade or Proprietary or Model Name | Manufacturer |
|---------|---------------------------------------------|----------------------------------|
| K142246 | LightStim for Acne/ LightStim for Acne Mini | LED Intellectual Properties, LLC |
| K153081 | Clear Bi-Light | Michael Todd, LP |
| K152889 | Sonilase Blue-UV Clean Plus Model-SL113BUV+ | Biorenew Labs, LLC |
#### 4. Description of Proposed Device:
This devices adopts narrow band of low level Red 660nm and Blue light (415nm), the selected light
{4}------------------------------------------------
001_510 (k) Summary Version: A/3
can get about 3~5mm depth into skin. The BC-001+ Acne Purifier is a hand-held device consisting of a low intensity light emitting diode (LED) lamp that is intended/indicated for over-the-counter use in the treatment of mild to moderate acne. Finally, this product belongs to the low level light therapy, it is safe to use on acne treatment. The device has a power on/off switch, treatment button, battery status indicator, Micro- USB socket for battery charging.
#### 5. Intended Use
BC-001+ Acne Purifier is intended for over-the-counter use in the treatment of mild to moderate acne.
#### 6. Technical and Performance
The following table compares the device to the predicate devices with basic technological characteristics.
| Parameter | New Device | Predicate Device | Predicate Device | Predicate Device | Remark |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | (to be assigned) | K142246 | K153081 | K152889 | -- |
| Device Name<br>and Model | BC-001+ Acne Purifier | LightStim for Acne/<br>LightStim for Acne Min | Clear Bi-Light | Sonilase Blue-UV Clean Plus Model-SL113BUV+ | -- |
| Manufacturer | Oriental Inspiration<br>Limited | LED Intellectual<br>Properties, LLC | Michael Todd, LP | Biorenew Labs, LLC | -- |
| Regulation<br>Number | 878.4810 | 878.4810 | 878.4810 | 878.4810 | -- |
| Classification | 2 | 2 | 2 | 2 | -- |
| Intend of use | BC-001+ Acne<br>Purifier is<br>intended/indicated<br>for over-the-counter<br>use in the treatment<br>of mild to moderate<br>acne. | LightStim for Acne/<br>LightStim for Acne<br>Mini is an<br>over-the-counter<br>hand-held device<br>intended for the use<br>in the treatment of<br>mild to moderate<br>acne. | The Clear Bi-Light is<br>indicated for the<br>treatment of mild to<br>moderate inflammatory<br>acne. | The Sonilase Blue-UV<br>Clean Plus is a<br>handheld OTC device<br>intended to emit<br>energy in the blue<br>region of the light<br>spectrum for use in the<br>treatment of mild to<br>moderate acne vulgaris | SE |
| Over-the-coun<br>ter use? | Yes | Yes | Yes | Yes | SE |
| Overall design | Lithium-ion<br>rechargeable battery<br>AC charger: 100-<br>240V at 50-60Hz, 500 | Mains powered<br>handheld device<br>applied to the face<br>providing LED light | Lithium-ion<br>rechargeable battery<br>AC charger: 100-<br>240V at 50-60Hz, 500 | One 7.4V<br>rechargeable<br>Lithium-ion battery | Same as<br>K153081 and<br>K152889,but<br>different with |
| | mA | output. | mA | | |
| IEC 60601<br>Compliant | Yes | Yes | Yes | Yes | K142246,<br>Note 1<br>SE |
| Handheld | Yes | Yes | Yes | Yes | SE |
| Dimensions | 70mm*70mm*23mm | unknown | 64 mm *35 mm *14.6 mm | Not stated | Different,<br>Note 1 |
| Performance<br>Data | Complies with applicable performance specifications and usability requirements | | | | SE |
| Performance Features | | | | | |
| Treatment<br>Area<br>(cm^2) | 9.1 [cm²] | 16.6 [cm²] | 20 [cm²] | Over 9.6cm² | Different with<br>K153081,<br>K152889 and<br>K142246, but<br>within the<br>scope of<br>treatment<br>area of<br>predicate<br>device. Note<br>1 |
| Wavelength(s)<br>(nm) | | | | | Blue light is<br>same as the<br>K153081,<br>K152889 and<br>K142246, And<br>red light is<br>same as the<br>k153081 |
| Blue | 415 ± 5 | 411 ± 4 | 405-420 | 415 ± 5 | |
| Red | 660 ± 5 | 640 ± 4 | 630-660 | | |
| Treatment<br>Timer | Up to 1.5 minutes per<br>area | Up to 3 minutes per<br>area | Up to 3 minutes per<br>area | 4 minutes per area,<br>daily | Different with<br>K141242, and<br>K140381,<br>Note 1,<br>Blue dose |
| Dose rate<br>(mW/cm2) | | | | | rate is same<br>as K152889.<br>Red dose rate<br>is within the<br>scope of<br>K153081,<br>Note 1 |
| Blue | 50 ( blue) | 12.4 (blue)<br>6.3 (red) | 31.1 (blue)<br>54.6 (red) | 50 (blue) | |
| Red | 23 (red) | | | | |
{5}------------------------------------------------
### Note 1
Although it is a little bit different from the predicate devices, it will not affect the main function and the intended use of the device. They all also comply with safety and performance requirements. So the differences will not raise any
{6}------------------------------------------------
001 510 (k) Summary Version: A/3
safety or effectiveness issues.
#### 7. PERFORMANCE DATA
The testing for BC-001+ Acne Purifier included performance, software, electrical safety, EMC, biocompatibility and bench. BC-001+ Acne Purifier passed all testing in support of the substantial equivalence determination:
### 7.1. Biocompatibility testing
The biocompatibility evaluation for the BC-001+ Acne Purifier was conducted in accordance with International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process." As dictated by the application and duration of contact with the intact skin, the enclosure of testing included the following tests:
- Cytotoxicity
- Sensitization
- Irritation
### 7.2. Electrical safety and electromagnetic compatibility
Electrical safety, performance and EMC testing were conducted on the BC-001+ Acne Purifier. The device with compliance with the IEC 60601-1-11, IEC 60601-2-57, and IEC 62471 standards for safety and the EN 60601-1-2 standard for EMC and the IEC 62133 standard for battery safety.
### 7.3. Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern.
### 7.4. Usability Testing
Usability testing according to following FDA Guidance 1757, Applying Human Factors and Usability Engineering to Optimize Medical Device Design, was conducted.
#### 8. Conclusions
The proposed device has the same intended use and similar characteristics as the predicate devices, the LightStim for Acne/ LightStim for Acne Mini device (K142246), Clear Bi-Light device (K153081) and Sonilase Blue-UV Clean Plus (K152889). Meanwhile, performance testing, bench testing, and safety report documentation supplied in this submission demonstrates that any differences in their
{7}------------------------------------------------
001_510 (k) Summary Version: A/3
technological characteristics do not raise any new questions of safety or effectiveness. Based on performance testing, the proposed device is Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.