K170236 · Guangdong Jinme Medical Technology Co., Ltd. · EFB · Nov 21, 2017 · Dental
Device Facts
Record ID
K170236
Device Name
Dental Low-speed Turbine Handpiece
Applicant
Guangdong Jinme Medical Technology Co., Ltd.
Product Code
EFB · Dental
Decision Date
Nov 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4200
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
Dental Low-speed Turbine Handpiece is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.
Device Story
Dental Low-speed Turbine Handpiece is an air-powered instrument used by dental professionals in clinics or hospitals for drilling, grinding, and repairing tooth disease. The device consists of a handpiece, connector, air motor, and geared angle handpiece. It operates via air pressure (245–392 KPa) to drive rotation at speeds up to 22,000 rpm. The device features a slide-proof handle and various gear ratios (1:1, 20:1, 1:5) to facilitate operational vision and efficiency. It requires routine maintenance with specified lubricant (PANA SPRAY Plus) and is compatible with steam autoclave sterilization at 135°C. The output is mechanical rotation for dental procedures, allowing the clinician to perform precise tooth preparation and restoration tasks, ultimately benefiting the patient through effective treatment of dental conditions.
Clinical Evidence
Bench testing only. Performance testing conducted per ISO 14457; biocompatibility testing per ISO 10993-1; reprocessing and sterilization validation per ISO 17665:2006 and ISO 11134:2003. No clinical data provided.
Technological Characteristics
Materials: Stainless steel, brass, titanium. Energy source: Air-powered. Dimensions/Form: Handpiece with air motor and geared angle attachments. Connectivity: None. Sterilization: Steam autoclave (135°C). Standards: ISO 14457, ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 17665, ISO 11134.
Indications for Use
Indicated for dental professionals for removal of carious material, cavity/crown preparation, finishing restorations, root canal preparation, and tooth polishing.
Regulatory Classification
Identification
A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
Guangdong JINME Medical Technology Co., Ltd. % Cecilia Ceng Vice President Guangzhou GLOMED Biological Technology Co., Ltd. Suite 306, Kecheng Mansion, No.121 Science Road, Guangzhou S Guangzhou, 510663 Cn
November 27, 2017
Re: K170236
Trade/Device Name: Dental Low-speed Turbine Handpiece Regulation Number: 21 CFR 872.4200 Regulation Name: Dental handpiece and accessories Regulatory Class: Class I Product Code: EFB. EGS Dated: October 24, 2017 Received: November 2, 2017
Dear Cecilia Ceng:
This letter corrects our substantially equivalent letter of Tuesday, November 21, 2017.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170236
# Device Name
Dental Low-speed Turbine Handpiece
#### Indications for Use (Describe)
Dental Low-speed Turbine Handpiece is intended for removing carious material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Chapter 6. 510(k) Summary K170236
#### 1. Submitter's Information
Company Name: Guangdong JINME Medical Technology Co., Ltd. Address: A15, New Light Source Industrial Base, Nanhai District, Foshan, Guangdong, China Name of contact person: Kristi Yang Title: Manager Tel: 86-757-88777997 Fax: 86-757-81267508 E-mail: kristi@jinmedental.com
## Application Correspondent:
## Contact Person: Ms. Cassie Lee
Guangzhou GLOMED Biological Technology Co., Ltd. Address: Suite 306, Kecheng Mansion, No.121 Science Road, Guangzhou Science Park, Guangzhou 510663, China Tel: +86-20-61099984 E-mail: regulatory@glomed-info.com
## 2. Subject Device Information:
- Type of 510(k) Submission: Traditional �
- � Common Name: Dental Handpiece
- � Trade Name: Dental Low-speed Turbine Handpiece
- ♦ Models: LN, L
- ♦ Classification Name: Dental Handpiece and accessories
- ♦ Product Code: EFB
- � Regulation Number: 21CFR 872.4200
- Regulation Class: 1 �
## 3. Predicate Device Information:
## Predicate Device 1:
{4}------------------------------------------------
| Sponsor: | Guangdong JINME Medical Technology Co., Ltd. |
|-----------------|------------------------------------------------------------|
| Subject Device: | Dental Low-speed Turbine Handpiece, Models: LN, L |
| File No.: | 510(k) Submission Report (V1.0), Chapter 6. 510(k) Summary |
- ◆ 510(K) Number: K150798
- ♦ Company Name: Codent Technical Industry Co., Ltd.
- ♦ Trade Name: Low Speed Handpiece and Accessories
- ♦ Common Name: Dental Handpiece and Accessories
- ♦ Product Code: EFB, EGS
## Reference Device K141886
- ◆ Company Name: MODERN KOREA Co., Ltd.
- ♦ Trade Name: MDK handpieces Low-speed handpieces
- ♦ Common Name: Dental Handpiece and Accessories
- ♦ Product Code: EFB
#### Reference Device K163483
- ◆ Company Name: NAKANISHI, INC.
- ♦ Trade Name: Pana Spray Plus
- ♦ Common Name: Dental Handpiece and Accessories
- ♦ Product Code: EFB
## 4. Device Description
The Dental Low-speed Handpiece is the dental clinic, hospital treatment for patients with tooth disease tools, which is an effective instrument for drilling, grinding, repairing. It composed of handpiece and a connector, including low speed air motor, straight handpiece and geared angle handpiece, for the model LN and L. The device description of the Dental Low-speed Handpiece is as following:
The gear ratios of handpieces have various gear ratios (for different geared angle handpiece) (1:1 (constant), 20:1 (speed reduction) and 1:5 (speed increase));
The handpieces have maximum Forward rotation speed 19000 rpm and maximum Reverse rotation speed of 18000 rpm.
The air motors are capable of running up to a speed of 22000 rpm, but different pressure of air supply; refer to section 2.3 for details.
The handle is slide-proof, comfortable and easy to clean. Handpiece and adaptors can bear steam disinfection at 135°C. The head angle can provide better operational vision and angle so that work efficiency can be improved. Cartridges have high precision when rotating; cartridges have low noise and high efficiency. The scope of application: for dental professional use only.
{5}------------------------------------------------
| Sponsor: | Guangdong JINME Medical Technology Co., Ltd. |
|-----------------|------------------------------------------------------------|
| Subject Device: | Dental Low-speed Turbine Handpiece, Models: LN, L |
| File No.: | 510(k) Submission Report (V1.0), Chapter 6. 510(k) Summary |
Lubricant should be used during routine maintenance (e.g. after each patient use and prior to sterilization). In order to avoid the risk, user must buy and use specified lubricant type "PANA SPRAY Plus" manufactured by NAKANISHI INC (cleared in K163483).
## 5. Intended Use / Indications for Use
Dental Low-speed Turbine Handpiece is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations, root canal preparations and polishing teeth.
## 6. Test Summary
Dental Low-speed Turbine Handpiece is designed, tested and will be manufactured in accordance with both mandatory and voluntary standards, including:
Performance test according to ISO 14457 standard
Biocompatibility test according to ISO 10993-1
Reprocessing and Sterilization test according to ISO 17665:2006 and ISO 11134:2003 standards
Reprocessing validation according to FDA guidance document Dental Handpieces - Premarket
Notification [510(k)] Submissions
The result of tests indicates that the Dental Low-speed Turbine Handpiece is substantially equivalent to the legally marketed predicate device.
| Elements of<br>Comparison | | Predicate Device<br>(Primary) | Reference<br>Device | Reference<br>Device | Remark |
|-----------------------------------------|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|--------------------------------------------|--------|
| Manufacturer | Guangdong<br>JINME Medical<br>Technology Co.,<br>Ltd. | Codent Technical<br>Industry Co., Ltd. | MODERN<br>KOREA Co., Ltd. | NAKANISHI,<br>INC. | -- |
| Device Name | Dental Low-speed<br>Turbine<br>Handpiece | Codent Low<br>Speed Dental<br>Handpieces and<br>Accessories | MDK handpieces<br>- Low-speed<br>handpieces | Pana Spray Plus | -- |
| Model | LN, L | LEIPB1, LEIPA1,<br>A61L, A61, A65L,<br>A65, A11, E-<br>2000, E-4010, E-<br>4000, AI6C, AI2C,<br>E-1110, E-<br>1100 | -- | -- | -- |
| 510(k)<br>Number | Applying | K150798 | K141886 | K163483 | -- |
| Product Code | EFB, EGS | EFB, EGS | EFB | EFB | SE |
| Indications for<br>Use&<br>Intended Use | Dental Low-speed<br>Turbine<br>Handpiece is | Codent Low<br>Speed Dental<br>Handpieces and | MDK low-speed<br>handpieces used<br>for teeth cutting. | PANA SPRAY<br>Plus is a<br>lubricant to be | SE |
## 7. Comparison to predicate device and conclusion
{6}------------------------------------------------
| Sponsor: | Guangdong JINME Medical Technology Co., Ltd. |
|-----------------|------------------------------------------------------------|
| Subject Device: | Dental Low-speed Turbine Handpiece, Models: LN, L |
| File No.: | 510(k) Submission Report (V1.0), Chapter 6. 510(k) Summary |
| | intended for<br>removing carious<br>material, excess<br>filling material,<br>cavity and crown<br>preparation,<br>finishing tooth<br>preparations and<br>restorations, root<br>canal preparations<br>and polishing<br>teeth. | Accessories are<br>intended for<br>removing carious<br>material, excess<br>filling material,<br>cavity and crown<br>preparation,<br>finishing tooth<br>preparations and<br>restorations, root<br>canal<br>preparations and<br>polishing teeth. | cavity and crown<br>preparation,<br>restorations and<br>polishing teeth.<br>All the devices<br>are designed for<br>use by a trained<br>professional in<br>the field of<br>general<br>dentistry. | used during<br>routine<br>maintenance of<br>dental and<br>medical surgical<br>handpieces after<br>each patient use<br>and prior<br>to sterilization<br>and is intended<br>for use to clean<br>and lubricate the<br>dental<br>and medical<br>surgical<br>handpieces. | |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Operational<br>Modes | Air-powered | Air-powered | Air-powered | -- | SE |
| Type of<br>Chunk | Push button<br>Latch-type chuck | Push button | Push button,<br>latch, screw,<br>snap-on<br>or tip-lock chuck<br>options | -- | SE |
| Composition<br>of Main<br>Materials | Stainless Steel,<br>Brass, Titanium | Stainless Steel,<br>Brass, Titanium | Stainless steel<br>and titanium | -- | SE |
| Operating<br>Pressure | $245 ">– 392 KPa$ | -- | 36psi to 43 psi | -- | SE<br>Note 1 |
| Motor Speed | 18,000 ~ 22,000<br>rpm | 25,000rpm | Up to 20,000<br>rpm | -- | SE<br>Note 1 |
| Sterilization | Steam autoclave<br>method | Steam autoclave<br>method | -- | -- | SE |
| Compliance<br>Standards | Complied with ISO<br>10993-5, ISO<br>10993-10,<br>ISO14457 | -- | Complied with<br>ISO 10993-5,<br>ISO 10993-10,<br>ISO14457 | -- | SE<br>Note 1 |
| Lubricant | The specified<br>lubricant, type<br>"PANA SPRAY<br>Plus"<br>manufactured by<br>NAKANISHI INC<br>(cleared in<br>K163483), must<br>be used during<br>routine<br>maintenance. | -- | Pana-Spray<br>made by<br>NSK(K052700) | Lubricant "PANA<br>SPRAY Plus"<br>manufactured<br>by NAKANISHI<br>INC | SE<br>Note 3 |
# Comparison in Detail
{7}------------------------------------------------
| Sponsor: | Guangdong JINME Medical Technology Co., Ltd. |
|-----------------|------------------------------------------------------------|
| Subject Device: | Dental Low-speed Turbine Handpiece, Models: LN, L |
| File No.: | 510(k) Submission Report (V1.0), Chapter 6. 510(k) Summary |
## Note 1:
Although the subject devices are a little difference from predicate devices in Operating Pressure, Motor Speed, Compliance Standards; the subject devices are compliance with "ISO 14457:2012 Dentistry -Handpieces and Motors". So the difference will not raise any safety or effectiveness issue.
## Note 2:
We add lubricant "PANA SPRAY Plus" (cleared in K163483) as additional predicate device. And we need user to buy and use the specified lubricant, type "PANA SPRAY Plus" manufactured by NAKANISHI INC, during routine maintenance.
## Note 3:
The user need to buy and use the specified lubricant, type "PANA SPRAY Plus" (cleared in K163483) manufactured by NAKANISHI INC, during routine maintenance.
## Finial conclusion:
The subject device Dental Low-speed Turbine Handpiece (Models: LN, L) has all features of predicate devices. The differences between them do not raise new question of safety and effectiveness. Thus, the subject device is substantially equivalent to the predicate device.
## 8. Date of the summary prepared: November 21, 2017
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.