← Product Code [JDQ](/productcode/JDQ) · K170206

# Arthrex FiberTape Cerclage (K170206)

_Arthrex, Inc. · JDQ · Nov 15, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K170206

## Device Facts

- **Applicant:** Arthrex, Inc.
- **Product Code:** [JDQ](/productcode/JDQ.md)
- **Decision Date:** Nov 15, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3010
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Arthrex FiberTape Cerclage suture is intended for use in soft tissue approximation and or ligation. These sutures may be incorporated, as components, into surgeries where constructs including those with allograft or autograft tissues are used for repair. When used as a bone fixation cerclage the suture is intended for: • Trochanteric reattachment after trochanteric osteotomy following total hip arthroplasty. • Sternotomy indications including the "rewiring" of osteotomized sternums. • Trauma surgery indications including olecranon, ankle, patella and some shoulder fracture rewiring. • Repair of long bone fractures due to trauma or reconstruction.

## Device Story

Arthrex FiberTape Cerclage is a flat braided suture or loop configuration used for soft tissue approximation, ligation, and bone fixation. Constructed from a polyblend of UHMWPE and polyester; loop assembly features a hitch over a sheath. Used by surgeons in orthopedic and trauma procedures (e.g., hip arthroplasty, sternotomy, fracture repair). Device provides mechanical fixation for bone segments or soft tissue; aids in stabilization during healing. Operates as a manual surgical implant; no electronic or automated components.

## Clinical Evidence

Bench testing only. Testing included fatigue strength, tensile force, creep, and knot strength. Wear testing with particle analysis compared the device to the DSM Biomedical DPR Cable, showing no statistically significant differences. Bacterial Endotoxins Test performed per USP <85> to confirm compliance with pyrogen limits.

## Technological Characteristics

Materials: Polyblend of UHMWPE and polyester. Form factor: Flat braided suture or loop configuration with sheath. Principle: Mechanical fixation/cerclage. Sterilization: Not specified. Software: None.

## Regulatory Identification

A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.

## Predicate Devices

- DSM Biomedical DPR Cable ([K143716](/device/K143716.md))

## Reference Devices

- ISO-Elastic Cerclage System ([K102834](/device/K102834.md))
- Arthrex FiberTape Family ([K032245](/device/K032245.md))
- Arthrex Suture Grafting Kit ([K041553](/device/K041553.md))
- Arthrex Suture ([K122374](/device/K122374.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized depiction of human figures. The FDA logo is on the right and includes the agency's acronym in a blue square, followed by the full name "U.S. Food & Drug Administration" in blue text.

November 15, 2017

Arthrex, Inc. Ms. Ivette Galmez Senior Regulatory Affairs Specialist 1370 Creekside Boulevard Naples, Florida 34108-1945

Re: K170206

Trade/Device Name: Arthrex FiberTape Cerclage Regulation Number: 21 CFR 888.3010 Regulation Name: Bone Fixation Cerclage Regulatory Class: Class II Product Code: JDQ, GAT Dated: October 18, 2017 Received: October 19, 2017

Dear Ms. Galmez:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Katherine D. Kavlock -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K170206

Device Name

Arthrex FiberTape Cerclage

Indications for Use (Describe)

The Arthrex FiberTape Cerclage suture is intended for use in soft tissue approximation and or ligation. These sutures may be incorporated, as components, into surgeries where constructs including those with allograft or autograft tissues are used for repair. When used as a bone fixation cerclage the suture is intended for:

- · Trochanteric reattachment after trochanteric osteotomy following total hip arthroplasty.
- · Sternotomy indications including the "rewiring" of osteotomized sternums.
- · Trauma surgery indications including olecranon, ankle, patella and some shoulder fracture rewiring.
- · Repair of long bone fractures due to trauma or reconstruction.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary 1.

| Date Prepared                      | November 13, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter                          | Arthrex Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Contact Person                     | Ivette Galmez<br>Senior Regulatory Affairs Specialist<br>1-239-643-5553, ext. 71263<br>lvette.galmez@arthrex.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Name of Device                     | Arthrex FiberTape Cerclage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Common Name                        | Bone Fixation Cerclage, Suture                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Product Code                       | JDQ, GAT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification Name                | 21 CFR 888.3010: Bone Fixation Cerclage<br>21 CFR 878.5000: Nonabsorbable Poly(ethylene) Terephthalate Surgical Suture                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Regulatory Class                   | ll                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Predicate Device                   | K143716: DSM Biomedical DPR Cable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Reference Predicate                | K102834: ISO-Elastic Cerclage System<br>K032245: Arthrex FiberTape Family<br>K041553: Arthrex Suture Grafting Kit<br>K122374: Arthrex Suture                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Purpose of Submission              | This traditional 510(k) premarket notification is submitted to obtain clearance for the<br>Arthrex FiberTape Cerclage sutures for use as bone fixation cerclage. This Arthrex<br>FiberTape Cerclage submission is comprised of four suture devices, two of which<br>(FiberTape® and TigerTape™) have been previously cleared in K032245, K041553 and<br>K122374.                                                                                                                                                                                                                                                                                                                                      |
| Device Description                 | The Arthrex FiberTape Cerclage suture is available as a flat braided suture or<br>assembled in a loop configuration. The Arthrex FiberTape Cerclage suture is braided<br>from a polyblend of Ultra High Molecular Weight Polyethylene (UHMWPE) and<br>polyester materials identical to those used in other cleared Arthrex sutures (K032245,<br>K041553 and K122374). For the loop assembly, the looped end of the suture is tied as<br>a hitch over a sheath that secures a double looped suture.                                                                                                                                                                                                    |
| Intended Use                       | The Arthrex FiberTape Cerclage suture is intended for use in soft tissue<br>approximation and or ligation. These sutures may be incorporated, as components,<br>into surgeries where constructs including those with allograft or autograft tissues are<br>used for repair. When used as bone fixation cerclage the suture is intended for:<br>• Trochanteric reattachment after trochanteric osteotomy following total hip<br>arthroplasty<br>• Sternotomy indications including the "rewiring" of osteomized sternums<br>· Trauma surgery indications including olecranon, ankle, patella and some shoulder<br>fracture rewiring<br>• Repair of long bone fractures due to trauma or reconstruction |
| Substantial Equivalence<br>Summary | The Arthrex FiberTape Cerclage suture is substantially equivalent to the predicate<br>devices, in which the basic design features and intended uses are the same. Any<br>differences between the Arthrex FiberTape Cerclage and the predicates are considered<br>minor and do not raise new questions concerning safety and effectiveness.                                                                                                                                                                                                                                                                                                                                                            |
|                                    | The submitted testing data, fatigue strength, tensile force, creep and knot strength<br>demonstrates that the Arthrex FiberTape Cerclage is substantially equivalent to the<br>cleared DSM Biomedical DPR Cable and the Kinamed ISO-Elastic Cerclage System.<br>Wear testing with particle analysis and comparative testing demonstrated no<br>statistical significant differences between the proposed Arthrex FiberTape Cerclage<br>and the cleared DSM Biomedical DPR Cable.                                                                                                                                                                                                                       |
|                                    | Bacterial Endotoxins Test was performed in accordance to USP <85> to demonstrate<br>that the device meets pyrogen limit specifications.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                    | Based on the indication for use, the technological characteristics, and the summary of<br>data submitted, Arthrex, Inc. has determined that the Arthrex FiberTape Cerclage<br>suture is substantially equivalent to currently marketed predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                             |

{4}------------------------------------------------

## K170206 - Page 2 of 2

---

**Source:** [https://fda-staging.innolitics.com/device/K170206](https://fda-staging.innolitics.com/device/K170206)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com