The Kobold Prostate HDR Template and Stepper Template are indicated for use as an accessory for high dose rate brachytherapy treatment of the prostate. The Stainless Steel Interstitial Needles and corresponding stylets are indicated for use as an accessory for high dose rate brachytherapy treatment of the prostate.
Device Story
Kobold Prostate HDR Templates and Stepper Templates are single-use, disposable accessories for ultrasound-guided interstitial temporary HDR prostate brachytherapy; used with biplanar transrectal ultrasound and fluoroscopy. Templates feature suture holes for perineal fixation and are customizable for prostate size/shape. Stepper Template fits Elekta Martinez Template Holder. Stainless steel interstitial needles (17 gauge, pencil or bevel tip) with stylets provide stiffening during implantation and between radiation fractions. Devices are used in clinical settings by physicians; contact duration is <24 hours. System facilitates precise needle placement for radiation delivery, aiding in localized prostate cancer treatment.
Clinical Evidence
No clinical data. Evidence consists of bench testing and biocompatibility evaluation (cytotoxicity, irritation, sensitization, systemic toxicity) per ISO 10993. Real-time aging testing established a 5-month shelf life. Performance validation confirmed all predetermined acceptance criteria were met.
Technological Characteristics
Templates are single-use, disposable, non-sterile. Needles are 17-gauge stainless steel with pencil or bevel tips. Includes stylets for stiffening. Principle of operation is mechanical guidance for needle insertion in HDR brachytherapy. No software or electronic components.
Indications for Use
Indicated for patients undergoing high dose rate (HDR) brachytherapy treatment of the prostate.
Regulatory Classification
Identification
A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 1, 2017
Kobold, LLC % Ms. Deanna Hughes Director of Regulatory and Quality Systems 23403 E. Mission Avenue, Suite 220E LIBERTY LAKE WA 99019
Re: K170203
Trade/Device Name: Kobold® Prostate HDR Template(s) and Kobold® Prostate HDR Stepper Template(s); Kobold® Interstitial Pencil Point Stainless Steel Needle Set(s) and Kobold® Interstitial Bevel Tip Stainless Steel Needle Set(s) Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: II Product Code: JAQ Dated: April 26, 2017 Received: May 1, 2017
Dear Ms. Hughes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K170203
### Device Name
Kobold® Prostate HDR Template(s) and Kobold® Prostate HDR Stepper Template(s) Kobold® Interstitial Pencil Point Stainless Steel Needle Interstitial Bevel Tip Stainless Steel Needle Set(s)
### Indications for Use (Describe)
The Kobold Prostate HDR Template and Stepper Template are indicated for use as an accessory for high dose rate brachytherapy treatment of the prostate.
The Stainless Steel Interstitial Needles and corresponding stylets are indicated for use as an accessory for high dose rate brachytherapy treatment of the prostate.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image features the word "Kobold" in a stylized, sans-serif font. The word is rendered in a bright, cyan color, giving it a modern and clean appearance. A registered trademark symbol is present to the right of the word.
# 510(k) Summary
#### 1. 510(k) Owner:
Kobold, LLC
#### 2. Address:
23403 E. Mission Ave., Suite 220E Liberty Lake, WA 99019
#### 3. Contact Person:
Deanna Hughes, MS, RAC Director of Regulatory and Quality Systems Kobold, LLC Email: deanna.hughes@koboldmedical.com Direct: 818.321.0116
#### Date 510(k) Summary Prepared: 4.
January 20, 2017
#### 5. Trade Name:
Kobold Prostate HDR Template(s) Kobold Prostate HDR Stepper Template(s) Kobold Interstitial Pencil Point Stainless Steel Needle Set(s) Kobold Interstitial Bevel Tip Stainless Steel Needle Set(s)
#### Common Name: 6.
System, Applicator, Radionuclide, Remote Controlled
#### 7. Classification Name:
21 CFR § 892.5700, Remote controlled radionuclide applicator system (JAQ)
#### 8. Classification Panel:
Radiology
#### Regulatory Classification: 9.
Class II
## 10. Predicate Device(s):
HDR Prostrate Template and Accessories cleared in K091230 on August 12, 2009
## 11. Device Description:
## Kobold Prostate HDR Templates and HDR Stepper Templates:
The Kobold Prostate HDR Template and Stepper Template are accessories to be used where ultrasound-guided interstitial temporary high dose rate (HDR)
{4}------------------------------------------------
brachytherapy treatment of the prostate is accepted by up-to-date clinical guidelines. The templates are ergonomically designed for freehand HDR technique. The templates are best used with biplanar transrectal ultrasound and confirmatory fluoroscopy.
The templates are single-use, disposable devices shipped nonsterile. It is customizable for almost any prostate size or shape. The templates include four suture holes at the corners to secure the template to the perineum. The templates are available in various gauge sizes to match the most common needles sizes. These devices are for continuous use of up to 24 hours of contact with the patient.
For workflows that require a stepper, the Kobold Prostate HDR Stepper Template fits into the Elekta Martinez Template Holder.
## Kobold Stainless Steel Interstitial Needle Sets:
The stainless steel needles are 17 gauge in either a pencil point tip or a bevel tip. The associated stylets are inserted into needles to stiffen the needles during implantation of the needles into the patient and to stiffen the needles between radiation therapy fractions.
The Kobold Stainless Steel Interstitial Needles are intended for single use and are disposable. The maximum implantation time for these needles is less than 24 hours.
## 12. Indications for Use:
## Kobold Prostate HDR Templates and HDR Stepper Templates:
The Kobold Prostate HDR Template and Stepper Template are indicated for use as an accessory for high dose rate brachytherapy treatment of the prostate.
## Kobold Stainless Steel Interstitial Needle Sets:
The Stainless Steel Interstitial Needles and corresponding stylets are indicated for use as an accessory for high dose rate brachytherapy treatment of the prostate.
## 13. Technological Characteristics
The Kobold Prostate HDR Template(s), Kobold Prostate HDR Stepper Template(s), Kobold Interstitial Pencil Point Stainless Steel Needle Set(s), and Kobold Interstitial Bevel Tip Stainless Steel Needle Set(s) have substantially equivalent indications for use, principle of operation, and technological characteristics as the HDR Prostate Template and Accessories, K091230.
The Kobold Prostate HDR Template(s), Kobold Prostate HDR Stepper Template(s), Kobold Interstitial Pencil Point Stainless Steel Needle Set(s), and Kobold Interstitial Bevel Tip Stainless Steel Needle Set(s) have substantially equivalent fundamental scientific principles as the HDR Prostate Template and Accessories. The subject and predicate devices have substantially equivalent methods of template and needle use (principle of operation). The devices have substantially equivalent technological characteristics.
{5}------------------------------------------------
Kobold. – Traditional 510(k) – Prostate HDR Templates & Needle Sets Section 7: 510(k) Summary
The use environment, target user, and patient population of the subject devices are the same as the predicate device.
## 14. Performance Data (Nonclinical)
The Kobold Prostate HDR Template(s), Kobold Prostate HDR Stepper Template(s), Kobold Interstitial Pencil Point Stainless Steel Needle Set(s), and Kobold Interstitial Bevel Tip Stainless Steel Needle Set(s) was evaluated using biocompatibility, and bench testing to confirm the performance characteristics.
- . Shelf Life - Real Time aging testing conducted shows that the templates included in this submission has an initial shelf life of 5 months.
- Biocompatibility - Cytotoxicity, Irritation, Sensitization, and Systemic Toxicity testing of the templates was conducted per ISO 10993. All testing passed.
- . Performance Evaluation -The Non-Clinical Performance Bench Validation Testing demonstrated that the subject device met all predetermined acceptance criteria.
The data presented in this submission show that the subject devices will perform in a substantially equivalent manner to the predicate device.
Clinical trial data was not needed for these devices to show substantial equivalence.
## 15. Conclusion
The Kobold Prostate HDR Template(s), Kobold Prostate HDR Stepper Template(s). Kobold Interstitial Pencil Point Stainless Steel Needle Set(s), and Kobold Interstitial Bevel Tip Stainless Steel Needle Set(s) successfully met all predetermined acceptance criteria for the design validation.
The Kobold Prostate HDR Template(s), Kobold Prostate HDR Stepper Template(s), Kobold Interstitial Pencil Point Stainless Steel Needle Set(s), and Kobold Interstitial Bevel Tip Stainless Steel Needle Set(s) is substantially equivalent to the legally marketed predicate device and does not raise any new safety or effectiveness questions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.