The CARDIOVIT FT-1 is a 12-channel ECG unit used for the recording, analysis, viewing, storage and transmission of ECG waveforms. The CARDIOVIT FT-1 is designed for indoor use and can be used for all patient populations. The CARDIOVIT FT-1 is used to diagnose cardiac abnormalities, and detect acute myocardial ischemia and infarctions in chest pain patients. The CARDIOVIT FT-1 is intended for use in hospitals, cardiology units, outpatient clinical units and general physician's offices.
Device Story
12-channel ECG unit; records, analyzes, views, stores, and transmits ECG waveforms. Inputs: patient ECG signals via contact sensors. Operation: touch panel or barcode scanner input; internal strip chart recorder for output; color display for viewing. Context: indoor use in hospitals, cardiology units, outpatient clinics, and physician offices; operated by healthcare professionals. Connectivity: standalone or networked via Ethernet/WiFi to SCHILLER SEMA3 Data Management System for report storage and work order retrieval. Benefits: assists clinicians in diagnosing cardiac abnormalities and detecting acute myocardial ischemia/infarctions; provides digital workflow integration.
Clinical Evidence
Bench testing only. Device verified against electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), usability (IEC 60601-1-6, IEC 62366), and electrocardiograph performance (IEC 60601-2-25) standards. Software verification and validation conducted per FDA guidance for moderate level of concern.
Indicated for all patient populations for the recording, analysis, viewing, storage, and transmission of ECG waveforms to diagnose cardiac abnormalities and detect acute myocardial ischemia and infarctions in chest pain patients.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 19, 2017
Schiller AG % Jim Chickering Regulatory Affairs Manager Zoe Medical. Inc 460 Boston Street Topsfield, Massachusetts 01983
Re: K170182
Trade/Device Name: CARDIOVIT FT-1 Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: June 8, 2017 Received: June 12, 2017
Dear Jim Chickering:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
M.A. Willehemen
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K170182
Device Name CARDIOVIT FT-1
#### Indications for Use (Describe)
The CARDIOVIT F - 1 is a 12-channel ECG unit used for the recording, viewing, storage and transmission of ECG waveforms.
The CARDIOVIT FT-1 is designed for indoor use and can be used for all patient populations.
The CARDIOVIT FT-1 is used to diagnose cardiac abnormalities, and detect acute myocardial ischemia and infarctions in chest pain patients.
The CARDIOVIT FT-1 is intended for use in hospitals, cardiology units, out-patient clinical units and general physician's offices.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
SCHILLER AG Altgasse 68, Postfach CH-6341 Baar, Switzerland Tel: +41 41 766 42 42 Fax: +41 41 761 08 80
MWST Nr. 212 858 e-mail: sales@schiller.ch http://www.schiller.ch
Image /page/3/Picture/3 description: The image shows a logo for Schiller, a company that specializes in diagnostics. The logo features three colorful icons: a yellow heart on a red background, pink lungs on an orange background, and a green heart on a yellow background. To the right of the icons is the company name "SCHILLER" in red, with the tagline "The Art of Diagnostics" in black underneath.
# 510(k) Summary
## 1. General information
| Submitter | |
|-----------------|-----------------------------------------------------------------------------------|
| Address: | Schiller AG<br>Altgasse 68<br>CH-6341 Baar<br>Switzerland |
| Phone : | +41 41 766 42 52 |
| Fax: | +41 41 761 08 80 |
| Contact Person: | Zhenrong Yu, MD, Ph.D.<br>Vice President Regulatory Affairs and Quality Assurance |
| Date Prepared: | May 22, 2017 |
| Device | |
|-------------------|--------------------------------------------|
| Device Trade Name | CARDIOVIT FT-1 |
| Manufacturer | Schiller AG |
| | Altgasse 68 |
| | CH-6341 Baar |
| | Switzerland |
| Common Name: | Electrocardiograph |
| Product Code: | DPS - Electrocardiograph (21 CFR 870.2340) |
| Regulatory Class | II |
## 2. Predicate Device
The primary predicate device is the SCHILLER CARDIOVIT AT-10 Plus.
A secondary predicate device is cited for its use of the same electrocardiograph physiological parameter technology as the subject device.
No reference devices were used in this submission.
| Predicate Scope | Predicate Device | 510(k) | Classification |
|-------------------|--------------------|---------|----------------|
| Primary Predicate | SCHILLER CARDIOVIT | K050686 | DPS |
{4}------------------------------------------------
SCHILLER AG Altgasse 68, Postfach CH-6341 Baar, Switzerland Tel: +41 41 766 42 42 Fax: +41 41 761 08 80
MWST Nr. 212 858 e-mail: sales@schiller.ch http://www.schiller.ch
Image /page/4/Picture/3 description: The image shows a logo for Schiller, a company that specializes in diagnostics. The logo features three colorful squares, each containing a different organ: a yellow heart, pink lungs, and a green heart. To the right of the squares is the company name, "SCHILLER," in bold red letters, followed by the tagline "The Art of Diagnostics" in smaller black letters.
| Predicate Scope | Predicate Device | 510(k) | Classification |
|--------------------------------------------------------------------------------------|-------------------------------------|---------|----------------|
| | AT-10 Plus | | |
| Secondary Predicate<br>(electrocardiograph<br>physiological parameter<br>technology) | SCHILLER Diagnostic<br>Station DS20 | K152043 | DPS |
## 3. Device Description
The CARDIOVIT FT-1 is a 12-lead ECG (Electrocardiograph) device used in the recording, analysis, viewing, storage and transmission of ECG waveforms.
The CARDIOVIT FT-1 does not provide a patient monitoring capability with alarm annunciation.
The CARDIOVIT FT-1 has a color display. It accepts user input via a touch panel or barcode scanner. It can generate a variety of reports that can be viewed on the display or printed on a strip chart recorder that is built into the device.
The CARDIOVIT FT-1 is mains- or battery- powered and uses sensors that come in contact with the patient.
The CARDIOVIT FT-1 is intended to function in the patient vicinity alongside other medical devices. It can operate as a stand-alone device or can be connected to the SCHILLER SEMA3 Data Management System via Ethernet (land-line or WiFi) in order to store reports and retrieve work orders for a given patient.
## 4. Indication For Use
The CARDIOVIT FT-1 is a 12-channel ECG unit used for the recording, analysis, viewing, storage and transmission of ECG waveforms.
The CARDIOVIT FT-1 is designed for indoor use and can be used for all patient populations.
The CARDIOVIT FT-1 is used to diagnose cardiac abnormalities, and detect acute myocardial ischemia and infarctions in chest pain patients.
{5}------------------------------------------------
SCHILLER AG Altgasse 68, Postfach CH-6341 Baar, Switzerland Tel: +41 41 766 42 42 Fax: +41 41 761 08 80
MWST Nr. 212 858
e-mail: sales@schiller.ch
http://www.schiller.ch
Image /page/5/Picture/3 description: The image shows a logo for Schiller. The logo has three different colored squares with a different organ in each square. The first square is red and has a yellow heart, the second square is orange and has pink lungs, and the third square is yellow and has a green heart. To the right of the squares is the word "SCHILLER" in red, with the words "The Art of Diagnostics" in black below it.
The CARDIOVIT FT-1 is intended for use in hospitals, cardiology units, outpatient clinical units and general physician's offices.
# 5. Comparison of Technological Characteristics with Predicate
The subject device has the same device characteristics and intended use as the primary predicate device. The subject device has exactly the same electrocardiograph physiological parameter technology as the secondary predicate device.
At a high level, the subject and predicate devise are based on the following same technological elements:
- 12 lead ECG electrocardiographs -
- with data analysis
- without alarm annunciation
The following technological differences, corresponding to technological upgrades, exist between the subject and the predicate device:
- -Slimmer and lighter design
- Virtual keyboard and keys instead of an integrated keyboard -
- Higher display resolution -
- -Increased Database Storage
- Modern network technology (including WLAN) -
- Wider range of operating and storage conditions -
- -Use of a more readily available electrical components (e.q., microcontrollers, memory chips)
- -Use of an external (rather than internal) medical-grade power supply
- Use of a newer battery technology -
These differences do not raise new questions of safety and effectiveness.
### 6. Performance Data
#### 6.1. Electrical safety, essential performance and electromagnetic compatibility (EMC) testing
The CARDIOVIT FT-1 was successfully tested to the following requlatory standards:
K170182
{6}------------------------------------------------
SCHILLER AG Altgasse 68, Postfach CH-6341 Baar, Switzerland Tel: +41 41 766 42 42 Fax: +41 41 761 08 80
MWST Nr. 212 858 e-mail: sales@schiller.ch http://www.schiller.ch
Image /page/6/Picture/3 description: The image shows a logo for Schiller. The logo has a yellow heart on a red background, pink lungs on an orange background, and a green heart on a yellow background. To the right of the images is the word "SCHILLER" in red, with the words "The Art of Diagnostics" in a smaller font below.
| Standard | Standard Title |
|-------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1: 2005 +<br>CORR. 1:2006 + CORR.<br>2:2007 + AM1:2012 (or<br>IEC 60601-1: 2012<br>reprint) | 3rd Edition, Medical Electrical Equipment, Part 1: General<br>requirements for basic safety and essential performance |
| IEC 60601-1-2:2014 | 4th Edition. Medical Electrical Equipment, Part 1-2: General<br>requirements for basic safety and essential performance -<br>Collateral Standard: Electromagnetic Compatibility |
| IEC 60601-1-6 : 2010<br>(Third Edition) + A1:2013 | Medical electrical equipment - Part 1-6: General<br>requirements for basic safety and essential performance -<br>Collateral standard: Usability |
| IEC 60601-2-25:2011 | 2nd Edition. Medical Electrical Equipment, Part 2-25:<br>Particular requirements for the basic safety and essential<br>performance of electrocardiographs |
| IEC 62366:2007 (First<br>Edition) + A1:2014 | Medical Devices - Application of usability engineering to<br>medical devices |
| IEC 62304:2006 | Medical Device Software - Software life cycle processes |
#### 6.2. Software Verification and Validation Testing
Software verification and validation testing was conducted as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device is considered as a "moderate" level of concern, since a failure or latent flaw in the software could indirectly result in minor injury to the patient through incorrect or delayed information or through the action of a care provider.
### 7. Conclusions
Based upon a comparison of devices and performance testing results, the SCHILLER CARDIOVIT FT-1 is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.