K170171 · Carl Zeiss Meditec, AG · HJO · Apr 6, 2017 · Ophthalmic
Device Facts
Record ID
K170171
Device Name
IOLMaster 700
Applicant
Carl Zeiss Meditec, AG
Product Code
HJO · Ophthalmic
Decision Date
Apr 6, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1850
Device Class
Class 2
Indications for Use
The IOLMaster 700 is intended for biometric measurements and visualization of ocular structures. The measurements and visualization assist in the determination of the appropriate power and type of intraocular lens. The IOLMaster 700 measures: - Lens thickness - Corneal curvature and thickness - Axial length - Anterior chamber depth - Pupil diameter - White-to-white distance (WTW) For visualization, the IOLMaster 700 employs optical coherence tomography (OCT) to obtain two-dimensional images of ocular structures of the anterior and posterior segments of the eye. The Reference Image functionality is intended for use as a preoperative and postoperative image capture tool.
Device Story
Non-invasive optical biometry instrument; uses swept-source laser and spectral domain interferometry (OCT) to measure ocular structures (axial length, lens thickness, corneal curvature/thickness, anterior chamber depth, pupil diameter, WTW distance); captures 2D images of anterior/posterior segments; includes Reference Image functionality for pre/postoperative scleral vessel imaging. Used in clinical settings by eye care professionals. Outputs biometric data and ocular images to assist clinicians in selecting intraocular lens power/type. Benefits include precise preoperative planning for cataract surgery.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-5:2009 (cytotoxicity) and ISO 10993-10:2014 (skin irritation/sensitization). Electrical safety and EMC evaluated against IEC 60601-1 (Ed 3.1), IEC 60601-1-6 (Ed 3.1), IEC 62366 (Ed 1.1), and IEC 60601-1-2 (Ed 4.0).
Indicated for biometric measurements and visualization of ocular structures to assist in determining intraocular lens power and type in patients undergoing cataract surgery or similar procedures requiring ocular biometry.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 6, 2017
Carl Zeiss Meditec Ag % Mr. Rahul Ram Consultant Biologics Consulting Group, Inc. 400 N. Washington Street, Suite 100 Alexandria, VA 22314
Re: K170171
Trade/Device Name: IOLMaster 700 Regulation Number: 21 CFR 886.1850 Regulation Name: AC-Powered Slitlamp Biomicroscope Regulatory Class: Class II Product Code: HJO Dated: January 18, 2017 Received: January 19, 2017
Dear Mr. Ram:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170171
Device Name IOLMaster 700
#### Indications for Use (Describe)
The IOLMaster 700 is intended for biometric measurements and visualization of ocular structures. The measurements and visualization assist in the determination of the appropriate power and type of intraocular lens. The IOLMaster 700 measures:
- · Lens thickness
- · Corneal curvature and thickness
- · Axial length
- · Anterior chamber depth
- · Pupil diameter
- · White-to-white distance (WTW)
For visualization, the IOLMaster 700 employs optical coherence tomography (OCT) to obtain two-dimensional images of ocular structures of the anterior and posterior segments of the eye.
The Reference Image functionality is intended for use as a preoperative image capture tool.
| Type of Use (Select one or both, as applicable) | | | | |
|-------------------------------------------------|---------------------------------------------|--|--|--|
| 区 Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the IOLMASTER 700 is provided below.
| Device Common Name: | Biometer |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Device Trade Name: | IOLMASTER 700 |
| Applicant: | Carl Zeiss Meditec AG<br>Goeschwitzer Strasse 51-52<br>D-07745 Jena<br>Germany |
| Contact: | Dr. Christian Muenster<br>Director Regulatory and Clinical Affairs<br>+49 7364 206985<br>christian.muenster@zeiss.com |
| Prepared by: | Rahul Ram<br>Biologics Consulting Group<br>rram@biologicsconsulting.com<br>(571) 777-9518 |
| Date Prepared: | April 5, 2017 |
| Classification Regulation: | 21 CFR 886.1850, Class II, AC-powered slitlamp<br>biomicroscope |
| Panel: | Ophthalmology |
| Product Code: | HJO - AC-powered slitlamp biomicroscope. |
| Predicate Device: | Carl Zeiss Meditec AG IOLMASTER 700<br>K143275<br>Product Code: HJO - AC-powered slitlamp biomicroscope. |
### Indications for Use:
The indications for use are identical between the subject and predicate IOLMaster 700 devices.
The IOLMaster 700 is intended for biometric measurements and visualization of ocular structures. The measurements and visualization assist in the determination of the appropriate power and type of intraocular lens. The IOLMaster 700 measures:
{4}------------------------------------------------
- . Lens thickness
- Corneal curvature and thickness
- Axial length
- Anterior chamber depth ●
- Pupil diameter ●
- White-to-white distance (WTW)
For visualization, the IOLMaster 700 employs optical coherence tomography (OCT) to obtain two-dimensional images of ocular structures of the anterior and posterior segments of the eye.
The Reference Image functionality is intended for use as a preoperative and postoperative image capture tool.
### Device Description:
The IOLMaster 700 is a non-invasive optical biometry instrument for visualization and measurement of ocular structures. The IOLMaster 700 is the latest generation device in the IOLMaster series. The version of the IOLMaster 700 that is the subject of this submission is a modified version of the IOLMaster 700 cleared under K143275.
The differences between the subject IOLMaster 700 and the predicate IOLMaster 700 that are the subject of this 510(k) submission are:
- . Labeling changes, including inclusion of additional clinical data and minor updates;
- Materials changes to the forehead rest and chin rest ●
The changes described in this submission do not affect how the hardware is used to acquire images, nor do these changes affect the principle of operation of the device.
### Electrical Safety & Electromagnetic Compatibility:
The IOLMaster 700 was evaluated against the requirements of Edition 3.1 of IEC 60601-1 (IEC 60601-1:2005 + Amendment 1 (2012)), Edition 3.1 of IEC 60601-1-6 (IEC 60601-1-6:2010 + Amendment 1 (2013)) and Edition 1.1 of IEC 62366 (IEC 62366:2007 + Amendment 1 (2014)), and found to comply.
The IOLMaster 700 was evaluated against the requirements of Edition 4.0 of IEC 60601-1-2 (IEC 60601-1-2:2014) and found to comply.
### Biocompatibility:
Testing was conducted as per the following biocompatibility standards:
- ISO 10993-10:2014 (skin irritation and sensitization) ●
- . ISO 10993-5:2009 (cytotoxicity)
Testing demonstrated that the new materials are biocompatible for the proposed use.
{5}------------------------------------------------
# Device Comparison Table:
| DEVICE CHARACTERISTICS | PROPOSED IOLMASTER 700<br>(CARL ZEISS MEDITEC AG)<br>K170171 | PREDICATE IOLMASTER 700<br>(CARL ZEISS MEDITEC AG)<br>K143275 |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| Indications for Use | The IOLMaster 700 is intended for<br>biometric measurements and<br>visualization of ocular structures. The<br>measurements and visualization assist in<br>the determination of the appropriate<br>power and type of intraocular lens. The<br>IOLMaster 700 measures:<br>Lens thickness<br>●<br>Corneal curvature and thickness<br>●<br>Axial length<br>●<br>Anterior chamber depth<br>●<br>Pupil diameter<br>●<br>White-to-white distance (WTW)<br>●<br>For visualization, the IOLMaster 700<br>employs optical coherence tomography<br>(OCT) to obtain two-dimensional images<br>of ocular structures of the anterior and<br>posterior segments of the eye.<br>The Reference Image functionality is<br>intended for use as a preoperative and<br>postoperative image capture tool. | Identical |
| Principles of Operation | Spectral domain interferometry (OCT<br>principle),<br>Light spot projection (infrared LEDs),<br>Image capturing | Identical |
| Feature - Corneal Curvature Measurement (KER): | | |
| Technology for obtaining<br>measurements/images | Telecentric keratometry = distance<br>independent,<br>Light spot projection (infrared LEDs) | Identical |
| Keratometry algorithm | | Similar |
| Feature - Lens Thickness Measurement (LT): | | |
| Technology for obtaining<br>measurements/images | Swept source laser<br>Spectral domain interferometry (OCT<br>principle),<br>Multiple A-scans provide a B-scan | Identical |
{6}------------------------------------------------
| DEVICE CHARACTERISTICS | PROPOSED IOLMASTER 700<br>(CARL ZEISS MEDITEC AG) | PREDICATE IOLMASTER 700<br>(CARL ZEISS MEDITEC AG)<br>K143275 | |
|--------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|--|
| Feature - Central Corneal Thickness Measurement (CCT): | | | |
| Technology for obtaining<br>measurement | Swept source laser<br>Spectral domain interferometry (OCT<br>principle),<br>Multiple A-scans provide a B-scan | Identical | |
| Feature - Anterior Chamber Depth Measurement (ACD): | | | |
| Technology for obtaining<br>measurement | Swept source laser<br>Spectral domain interferometry (OCT<br>principle),<br>Multiple A-scans provide a B-scan | Identical | |
| Feature - Axial Length Measurement (AL): | | | |
| Technology for obtaining<br>measurement | Swept source laser 1055 nm.<br>Spectral domain interferometry (OCT<br>principle),<br>Multiple A-scans provide a B-scan | Identical | |
| Feature - Pupil Diameter Measurement (P): | | | |
| Technology for obtaining<br>measurement | Image capturing of the iris with internal<br>digital camera. | Identical | |
| Feature - White-to-White Measurement (WTW): | | | |
| Technology for obtaining<br>measurement | Image capturing of the iris with internal<br>digital camera. | Identical | |
| Feature - Reference Image Functionality: | | | |
| Technology for obtaining<br>measurement | Green LEDs for green light illumination<br>for image capturing of scleral vessels<br>with internal digital camera. | Identical | |
| Feature - Computational<br>formulas | Haigis suite (includes Haigis, Haigis-L<br>and Haigis-T);<br>Hoffer Q;<br>Holladay 2;<br>SRK®/T | Identical | |
| Electrical Data: | | | |
| Rated voltage / frequency | 100 V to 240 V AC (± 10 %) / 50/60 Hz | Identical | |
| Power consumption:<br>Basic unit<br>In standby mode | 150 W<br>1 W | Identical | |
{7}------------------------------------------------
| Ambient conditions: | | |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| for intended use | Temperature: 10°C to +35°C,<br>Relative humidity: 30% to 80%<br>(noncondensing) | Temperature: 10°C to +35°C,<br>Relative humidity: 30% to 90%<br>(noncondensing) |
| for storage and transport | Temperature: -20°C to +60°C,<br>Relative humidity: 10% to 90%<br>(noncondensing) | for storage:<br>Temperature: -10°C to +55°C,<br>Relative humidity: 10% to 95%<br>(noncondensing)<br><br>for storage and transport:<br>Temperature: -40°C to +70°C,<br>Relative humidity: 10% to 95%<br>(noncondensing) |
| Material composition of patient<br>interface with forehead rest<br>and chin rest cap | ABS | PC/ABS Blend |
# Substantial Equivalence Summary:
Given the acceptable results of performance testing for the significant device modifications, we believe that the subject IOLMaster 700 is substantially equivalent to the predicate IOLMaster 700, cleared under K143275.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.