K170169 · Nuvasive Specialized Orthopedics, Inc. · HSB · Jun 9, 2017 · Orthopedic
Device Facts
Record ID
K170169
Device Name
PRECICE Intramedullary Limb Lengthening System
Applicant
Nuvasive Specialized Orthopedics, Inc.
Product Code
HSB · Orthopedic
Decision Date
Jun 9, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PRECICE Intramedullary Limb Lengthening System is indicated for limb lengthening of the tibia and femur.
Device Story
System consists of an intramedullary nail (implanted in tibia/femur), locking screws, end cap, and an External Remote Controller (ERC 3P). ERC 3P is a handheld, electrically powered unit containing two large rare-earth magnets. After surgical implantation of the nail, the patient or caregiver places the ERC 3P over the nail's actuator site. Rotating magnets in the ERC 3P drive rotation of the internal magnet within the nail, mechanically lengthening or shortening the device. Used in home environment for daily distraction sessions; distraction progress is monitored by clinicians via routine X-rays during weekly office visits. ERC 3P features an LCD touchscreen, camera, and alignment indicators to assist patient positioning and verify pairing with the implant. System enables controlled, non-invasive limb lengthening, reducing the need for additional invasive procedures.
Clinical Evidence
No clinical trials were performed. Evidence consists of bench testing and a usability study. Usability evaluation conducted on 17 participants to assess home-use safety and interface effectiveness. Bench testing included risk management (EN ISO 14971), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), shock/vibration, and ingress protection (IEC 60601-1-11).
Technological Characteristics
System includes titanium/metal intramedullary nail and screws. ERC 3P is an electromechanical handheld device using rare-earth magnets for remote actuation. Features LCD touchscreen, camera, and pairing detection. Connectivity is limited to internal magnetic coupling with the implant. Sterilization: Nail is gamma-irradiated; screws/instruments are non-sterile (autoclavable). Standards: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home use), EN ISO 14971 (risk management).
Indications for Use
Indicated for limb lengthening of the tibia and femur in patients requiring such procedures.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked one behind the other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Nuvasive Specialized Orthopedics, Inc. Cora Sim Associate Manager, Regulatory Affairs 101 Enterprise. Suite 100 Aliso Viejo, California 92656
June 9, 2017
Re: K170169
Trade/Device Name: PRECICE Intramedullary Limb Lengthening System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB Dated: May 23, 2017 Received: May 24, 2017
Dear Cora Sim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170169
#### Device Name
PRECICE Intramedullary Limb Lengthening System
#### Indications for Use (Describe)
The PRECICE Intramedullary Limb Lengthening System is indicated for limb lengthening of the tibia and femur.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized purple and silver graphic to the left of the company name. The company name is in a silver sans-serif font, with the words "SPECIALIZED ORTHOPEDICS, INC." in a smaller purple font below.
## PRECICE® Intramedullarv Limb Lengthening System 510(k) Summary - K170169 Mav 23, 2017
- 1. Company: NuVasive Specialized Orthopedics, Inc. 101 Enterprise, Suite 100 Aliso Viejo, CA 92656 Phone: (949) 837-3600 Fax: (949) 837-3664
- Cora Sim Contact: Associate Manager, Regulatory Affairs Phone: (949) 544-6478 Fax: (949) 837-3664
Date of Submission: May 23, 2017
- 2. Proprietary Trade Name: PRECICE® Intramedullary Limb Lengthening System
- 3. Classification Name: Intramedullary Fixation Rod (21 CFR 888.3020)
- Product Code: HSB (Rod, Fixation, Intramedullary and Accessories) 4.
- ട്. Product Description: The PRECICE Intramedullary Limb Lengthening System is composed of the PRECICE Nail (supplied sterile), locking screws, end cap, surgical instruments and an external remote controller (ERC, ERC2P, or ERC 3P). The nail is available in tibia or femur models with various diameters, lengths and screw hole configurations to accommodate a variety of patient anatomies. The locking screws are also available in a variety of diameters and lengths. The PRECICE Nail is supplied sterile by gamma radiation while the locking screws and PRECICE surgical instruments are supplied non-sterile and must be sterilized prior to use.
The third generation External Remote Controller (ERC 3P) which is the subject of this premarket notification, is a non-invasive adjustment component of the system. The ERC 3P is an electrically powered handheld unit. The ERC 3P contains two large rare-earth magnets that are rotated using gears. After the PRECICE Nail has been implanted into the patient, the external device can be placed over the actuator portion of the nail and activated, the magnets within the ERC 3P rotate, which causes the magnet in the actuator portion of the PRECICE Nail to rotate,
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized purple and silver graphic on the left, followed by the word "Nuvasive" in silver lettering. Below "Nuvasive" is the phrase "Specialized Orthopedics, Inc." in purple lettering. The logo is clean and modern, with a focus on the company's name and specialization.
lengthening or shortening the nail. Periodic lengthening (typically daily) of the nail is performed after the implantation surgery to lengthen the limb. The ERC 3P also encompasses on-unit prescription programming. The distraction is confirmed in office using standard, routine x-ray of the limb. These office visits usually occur on a weekly basis. The principles of operation of the ERC 3P are the same as those of the ERC and ERC2P previously cleared.
- Indications: The PRECICE Intramedullary Limb Lengthening System is indicated for limb 6. lengthening of the tibia and femur.
- Substantial Equivalence: Documentation demonstrates substantial equivalence to the PRECICE 7. Intramedullary Limb Lengthening System cleared under K101997 (cleared on July 12, 2011), K 1 13219 (cleared on October 19, 2012), K 13 1490 (cleared on January 23, 2014), K 133289 (cleared on December 9, 2013), K141023 (cleared on May 20, 2014), K150656 (cleared on April 23, 2015), K151131 (cleared on June 19, 2015), and most recently K160325 (cleared on April 4, 2016) . The purpose of this premarket notification is to include a third generation External Remote Controller (ERC 3P) into the PRECICE® Intramedullary Limb Lengthening System. Data provided in this submission includes information relevant to the third generation ERC (ERC 3P). Substantial equivalence is based on similar indications for use, designs, in vitro testing, software validation, and Usability and labeling readability evaluations performed. The in vitro evaluations included specific tests performed on the ERC 3P to demonstrate the suitability of the device for its intended use, and electrical safety and electromagnetic compatibility tests. Usability evaluation of the ERC 3P in a representative population was performed to demonstrate its suitability for use by the proposed patient population in the home environment and in accordance with the indications.
The ERC 3P was designed to improve ergonomic handling and user interfaces, and has the same methods and principles of operation as the first and second generation ERCs. The ERC 3P incorporates an updated user interface and combines the controller and hand piece into one unit, to be more ergonomic and user friendly. The ERC 3P incorporates an updated user interface that utilizes an LCD touchscreen, camera, external magnet line to assist the patient in proper alignment with the implant location, as well as features to detect the pairing status of the ERC 3P with the implanted PRECICE Nail during lengthening sessions. The third generation ERC for the PRECICE Intramedullary Limb Lengthening System was developed and evaluated in accordance with recognized standards and with in-house developed test methodologies. This testing includes risk assessment of the device, testing to applicable IEC standards, and a usability study undertaken on 17 participants to evaluate the usability of the ERC 3P in an equivalent patient population. Risk analysis, draft labeling, and test results are included in this premarket notification. The results of
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized purple and gray leaf-like shape on the left. To the right of the shape is the company name, "NUVASIVE," in gray, with the words "SPECIALIZED ORTHOPEDICS, INC." in purple underneath.
testing demonstrate that the ERC 3P for the PRECICE Intramedullary Limb Lengthening System that is the subject of this premarket notification is substantially equivalent to the predicate External Remote Controller (ERC 2P) device.
The following documentation and testing have been included in order to establish equivalence to the predicate device. Testing includes a usability evaluation for the Home Use of the ERC 3P by the patient, minimum rated voltage testing, shock and vibration testing, and ingress protection testing performed in accordance with IEC 60601-1-11;2015. The following tests have been performed in order to establish equivalence to the predicate device:
| Test/Document Description | Applicable test standard |
|------------------------------------------------|---------------------------------|
| Risk Management Report | EN ISO 14971: 2012 |
| Electrical Safety | IEC 60601-1 (3rd edition): 2005 |
| Electromagnetic Compatibility and Interference | IEC 60601-1-2: 2014 |
| Minimum rated voltage testing | |
| Shock and Vibration Testing | IEC 60601-1-11: 2015 |
| Ingress protection | |
| Usability evaluation | N/A |
| Labeling Readability | N/A |
Conclusions can be drawn from the testing performed that the PRECICE Intramedullary Limb Lengthening System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.